LATITUDE Paceart Integration Software, version 1.00 Model… recallZ-0037-2012 · Boston Scientific CRM Corp · initiated Aug 12, 2011
Terminated All recalls posted in October 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific CRM Corp or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Boston Scientific has determined that under certain conditions, LATITUDE Paceart Integration (LPI) software may not properly transfer data resulting in the incorrect summary of programmed parameter values in the ICD Summary tab display and Current Programming Tachy summary section of the Paceart report.
FDA root cause category: Software design
What the company is doing
Consignees were sent on 8/12/11 a Boston Scientific letter dated August 2011. The letter was addressed to "Dear Doctor". The letter described the product and problem. It also gave Background Information, Description of Potential Data Translation Error, and Additional Information and Action information. A letter, including an HRS form, was also sent to all US LATITUDE Clinics that have active LPI software to notify them of the need to install the corrected LPI Software.
Recall details
- Recall number
- Z-0037-2012
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Aug 12, 2011
- Posted by the FDA
- Oct 12, 2011
- Recalling firm
- Boston Scientific CRM Corp
- Quantity
- 398
- Product code
- LWS
- FDA submissions
- P910077
- Distribution
- Nationwide Distribution -- AL, AK, AZ, AR, CA, CO, CT DE, DC, FL, GA, IL,IN, IA, KS, KY, LA,ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OJ, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA,WV, WI, and WY.
Product description
LATITUDE Paceart Integration Software, version 1.00 Model 6472. LATITUDE Paceart Integration is software and therefore does not have any serial numbers. However, the software version affected is version 1.00 and is distributed on USB pen drives. These pen drives are lot number controlled and denoted as Model 6493. When used in conjunction with the LATITUDE EMR Integration software transfers data files from the Boston Scientific LATITUDE Patient Management system to a clinic's Paceart system.
Code information (models, lots, serial numbers)
software verson 1.00
Other devices under the same FDA submission
34 devices cleared under P910077 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls1 death65 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls9 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls4 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls1 report
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls1 report
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls0 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls0 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls0 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls0 reports
- LATITUDE® Communicator 6290Boston Scientific Corporation · Pacemaker/icd/crt non-implanted components142 reports
- LATITUDE™ Programming System 3300Boston Scientific Corporation · Pacemaker/icd/crt non-implanted components141 reports
- LATITUDE™ Programming System 3300Boston Scientific Corporation · Pacemaker/icd/crt non-implanted components72 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page