Smiths Medical HOTLINE 3 Blood and Fluid Warmer recallZ-1979-2012 · Smiths Medical ASD, Inc. · initiated Jun 27, 2012

Class II Terminated All recalls posted in July 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

HOTLINE 3 Blood and Fluid Warmer( Reorder No. HL-390 ) disposable is not correctly installed , the microswitch may become damaged and unusable

FDA root cause category: Device Design

What the company is doing

Smiths Medical sent an Urgent Medical Device Correction -Field Safety Notice letter dated June 21, 2012 via U.S. Mail to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter included an Urgent Medical Device Field Safety Notice Acknowledgement Form ("Confirmation Form") that customers were instructed to complete and return to Smiths Medical for carrying out the correction and affectivity tracking purposes. The letter stated that a Smiths Medical Representative will visit the affected sites to correct the units with a shelf upgrade kit, which includes instructions and all the parts necessary to make the correction to their affected devices. Customers were provided with a form to record the serial numbers of the devices corrected so that this can be returned to Smiths Medical. For questions regarding this information call Smiths Medical Quality Department at 1-781-763-9335. .

Recall details

Recall number
Z-1979-2012
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 27, 2012
Posted by the FDA
Jul 9, 2012
Recalling firm
Smiths Medical ASD, Inc.
Quantity
2247 units
Product code
LGZ
FDA submissions
K911383
Distribution
US Nationwide Distribution

Product description

Smiths Medical HOTLINE 3 Blood and Fluid Warmer, Reorder Numbers: HL-390 Product Usage: The Micro-switch is used on the HOTLINE 3 Blood and Fluid Warmer. This product, for use in conjunction with HOTLINE 3 L-370 Fluid Warming Set are intended for the delivery of warmed blood and intravenous fluids at normothermic temperatures. These products are intended for use by appropriately trained healthcare professionals in clinical environments

Code information (models, lots, serial numbers)

All serial numbers

Other devices under the same FDA submission

10 devices cleared under K911383 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.