S5 Mast Roller Pump 150, Item Number: 10-88-00 recallZ-0274-2013 · Sorin Group USA, Inc. · initiated Sep 19, 2012
Class II Terminated All recalls posted in November 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Sorin Group USA, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin
FDA root cause category: Mixed-up of materials/components
What the company is doing
The firm, Sorin Group, sent an "URGENT FIELD SAFETY NOTICE" dated September 19, 2012 to all affected customers by certified mail. The notice described the product, problem and actions to be taken. The customers were instructed to follow the provided instructions for the continued safe use of the Perfusion systems until they have been serviced or replaced by Sorin Group. The Sorin Group Service Team will contact the customer to schedule servicing and replacement of the affected product. The customers were also instructed to complete and return the Customer Response Form via fax to 303-467-6502 or by email to [email protected]; assure this notice is distributed to all personnel who need to be aware of this notice; and if they have transferred the affected products to a third party, pass this information with them as well as with Sorin Group Customer Service at 1-800-650-2623. For questions regarding this notice, contact Sorin Group USA Customer Service at 1-800-650-2623.
Recall details
- Recall number
- Z-0274-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 19, 2012
- Posted by the FDA
- Nov 8, 2012
- Recalling firm
- Sorin Group USA, Inc.
- Quantity
- 2 units
- Product code
- DTQ
- FDA submissions
- K071318, K093882
- Distribution
- Worldwide distribution: USA (nationwide) including states of: AZ, GA, MI, MS, NC, NY, PA, SC, and TX; and countries of: Austria, Australia, Belgium, Canada, People's Republic of China, Columbia, Denmark, Finland, Germany, India, Iran-Islamic Republic of, Ireland, Japan, Lebanon, Morocco, Pakistan, Poland, Russian Federation, Saudia Arabia, Spain, South Africa, Thailand, Turkey and United Kingdom.
Product description
S5 Mast Roller Pump 150, Item Number: 10-88-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
Code information (models, lots, serial numbers)
Serial Numbers: 50E00680--50E00681, 50E00688, 50E00696-50E00706, 50E00708-50E00718.
Other devices under the same FDA submission
23 devices cleared under K071318, K093882 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- S5 Roller Pump 150 RP 150LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 death120 reports
- S5 Console 4-positionLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 death65 reports
- S5 Double Roller Pump 85 DRP 85LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS54 reports
- Electrical Remote-Controlled Tubing Clamp 620 mmLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS52 reports
- S5 Console 3-positionLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS37 reports
- Electrical Remote-Controlled Tubing Clamp 500 mmLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS31 reports
- Bubble Detector Sensor 3/8"LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 death26 reports
- S5 Console 5-positionLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS2 deaths16 reports
- Bubble Detector Sensor 1/4"LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 reports
- Display & Control Module (S5)LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 reports
- SCP Pump Control Panel for S5/C5LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 reports
- Bubble Detector Sensor 3/16"LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 63383).
- Z-0271-2013 · TerminatedS5 Single Roller Pump 150, Item Number: 10-80-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
- Z-0272-2013 · TerminatedS5 Double Roller Pump 85, Item Number: 10-85-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
- Z-0273-2013 · TerminatedS5 Mast Roller Pump 85, Item Number: 10-88-60 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
- Z-0275-2013 · TerminatedSorin C5 Perfusion System, Item number 58-00-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page