Skip to content

Persona The Personalized Knee System, Natural Tibia… recallZ-0539-2013 · Zimmer, Inc. · initiated Nov 14, 2012

Class II Terminated All recalls posted in December 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening

FDA root cause category: Device Design

What the company is doing

Zimmer sent distributors beginning the week of November 14, 2012, E-mail information requesting all equipment be collected in distributors office before shipment back to Zimmer and replacement. A telephone conference discussing the recall and amended surgical technique was held the week of November 14, 2012, and Urgent Recall notices were sent to Distributors describing the devices; risks; updated surgical technique with directions to hand deliver surgeon/risk manager letters to all affected surgeons and risk managers in hospitals receiving the product. The notices included directions to remove all affected product and ensure risk manager letter is provided. The letter stated a Zimmer Inc. employee will conduct onsite reconciliation and asked return of distribution Spreadsheet to [email protected] within 24 hours of receiving notification. Product is to be returned with certificate of sterilization to Zimmer Product Service Department 1777 West Center St. Warsaw, IN 46580 or Zimmer International Logistics, Eschbach Germany for International Accounts. For questions regarding this recall call 574-372-4807.

Recall details

Recall number
Z-0539-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Nov 14, 2012
Posted by the FDA
Dec 18, 2012
Recalling firm
Zimmer, Inc.
Quantity
318
Product code
JWH
FDA submissions
K113369
Distribution
Worldwide Distribution - USA (nationwide) and internationally to Germany and Europe.

Product description

Persona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile SIZE C, LEFT; SIZE C RIGHT This device is indicated for patients with severe knee pain and disability due to: o Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o Moderate valgus, varus, or flexion deformities.

Code information (models, lots, serial numbers)

Part Number 42-5320-064-01 TIBIA SIZE C, LEFT Lots: 62018336, 62018337, 62033380, 62051020, 62060739, 62068413, 62076438, 62079740, 62091360, 62117665, and 77001708,  Part Number 42-5320-064-02 TIBIA SIZE C RIGHT Lot: 62018338, 62018339, 62033383, 62051021, 62060740, 62060742, 62068415, 62076439, 62079741, 62091361,and 62117666

Other devices under the same FDA submission

249 devices cleared under K113369 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 63690).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page