Persona The Personalized Knee System, Natural Tibia… recallZ-0544-2013 · Zimmer, Inc. · initiated Nov 14, 2012
Class II Terminated All recalls posted in December 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening
FDA root cause category: Device Design
What the company is doing
Zimmer sent distributors beginning the week of November 14, 2012, E-mail information requesting all equipment be collected in distributors office before shipment back to Zimmer and replacement. A telephone conference discussing the recall and amended surgical technique was held the week of November 14, 2012, and Urgent Recall notices were sent to Distributors describing the devices; risks; updated surgical technique with directions to hand deliver surgeon/risk manager letters to all affected surgeons and risk managers in hospitals receiving the product. The notices included directions to remove all affected product and ensure risk manager letter is provided. The letter stated a Zimmer Inc. employee will conduct onsite reconciliation and asked return of distribution Spreadsheet to [email protected] within 24 hours of receiving notification. Product is to be returned with certificate of sterilization to Zimmer Product Service Department 1777 West Center St. Warsaw, IN 46580 or Zimmer International Logistics, Eschbach Germany for International Accounts. For questions regarding this recall call 574-372-4807.
Recall details
- Recall number
- Z-0544-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 14, 2012
- Posted by the FDA
- Dec 18, 2012
- Recalling firm
- Zimmer, Inc.
- Quantity
- 195
- Product code
- JWH
- FDA submissions
- K113369
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Germany and Europe.
Product description
Persona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile TIBIA SIZE H, LEFT TIBIA SIZE H RIGHT This device is indicated for patients with severe knee pain and disability due to: o Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o Moderate valgus, varus, or flexion deformities.
Code information (models, lots, serial numbers)
Part Number 42-5320-083-01 TIBIA SIZE H, LEFT, Lot: 11006513, 11006722, 11007119, 62018828, 62018829, 62033403, 62060748, 62068423, 62076595, 62079771, 62091371, 62110207, 62117667 Part Number 42-5320-083-02 TIBIA SIZE H, RIGHT, Lot: 11007062, 11007123, 62018830, 62018831, 62033404, 62051804, 62060749, 62068424, 62076597, 62079772, 62091372, 62117668, 77001707
Other devices under the same FDA submission
249 devices cleared under K113369 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- Persona® The Personalized Knee® 42-5270-005-05Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER234 reports
- Persona® The Personalized Knee® 42-5170-005-05Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER185 reports
- Persona® The Personalized Knee® 42-5170-003-03Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER135 reports
- Persona® The Personalized Knee® 42-5270-007-07Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER129 reports
- Persona® The Personalized Knee® 42-5170-007-07Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER116 reports
- Persona® The Personalized Knee® 42-5270-003-03Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER112 reports
- PERSONA™ 42-5099-093-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER72 reports
- PERSONA® Natural Tibia® 42-5320-071-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER64 reports
- PERSONA™ Natural Tibia™ 42-5321-083-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- PERSONA™ Natural Tibia™ 42-5321-083-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- PERSONA® 42-5399-083-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 63690).
- Z-0539-2013 · TerminatedPersona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile SIZE C, LEFT; SIZE C RIGHT This device is indicated for patients with severe …
- Z-0540-2013 · TerminatedPersona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile This device is indicated for patients with severe knee pain and disability due …
- Z-0541-2013 · TerminatedPersona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile TIBIA SIZE E, LEFT TIBIA SIZE E RIGHT This device is indicated for patients with…
- Z-0542-2013 · TerminatedPersona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile TIBIA SIZE F, LEFT TIBIA SIZE F RIGHT This device is indicated for patients with…
- Z-0543-2013 · TerminatedPersona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile TIBIA SIZE G, LEFT TIBIA SIZE G RIGHT This device is indicated for patients with…
- Z-0545-2013 · TerminatedPersona, The Personalized Knee System,Tibial Drill, Cemented, 15.7 mm Diameter, Nonsterile This device is indicated for patients with severe knee pain and disability due to: o R…
- Z-0546-2013 · TerminatedPersona, The Personalized Knee System, Tibial Drill Guide, Cemented, 15.7 mm Diameter, Nonsterile, Nonsterile This drill guide is used with the Persona Knee System for patients …
- Z-0547-2013 · TerminatedPersona, The Personalized Knee System, Tibial Broach,, Cemented, Size C-D, Nonsterile, This device is used with the Persona Knee System for patients with severe knee pain and …
- Z-0548-2013 · TerminatedPersona, The Personalized Knee System, Tibial Broach, Cemented, Size E-F, Nonsterile, This device is used with the Persona Knee System for patients with severe knee pain and d…
- Z-0549-2013 · TerminatedPersona, The Personalized Knee System, Tibial Broach, Cemented, Size G-H, Nonsterile, This device is used with the Persona Knee System for patients with severe knee pain and d…
- Z-0550-2013 · TerminatedPersona, The Personalized Knee System, Tibial Sizing Plate, Cemented, Nonsterile, Size C Left; Size C Right This device is used with the Persona Knee System for patients with se…
- Z-0551-2013 · TerminatedPersona, The Personalized Knee System, Tibial Sizing Plate, Cemented, Nonsterile, Size D Left; Size D Right This device is used with the Persona Knee System for patients with se…
- Z-0552-2013 · TerminatedPersona, The Personalized Knee System, Tibial Sizing Plate, Cemented, Nonsterile, Size E Left; Size E Right This device is used with the Persona Knee System for patients with se…
- Z-0553-2013 · TerminatedPersona, The Personalized Knee System, Tibial Sizing Plate, Cemented, Nonsterile, Size F Left; Size F Right This device is used with the Persona Knee System for patients with se…
- Z-0554-2013 · TerminatedPersona, The Personalized Knee System, Tibial Sizing Plate, Cemented, Nonsterile, Size G Left; Size G Right This device is used with the Persona Knee System for patients with se…
- Z-0555-2013 · TerminatedPersona, The Personalized Knee System, Tibial Sizing Plate, Cemented, Nonsterile, Size H Left; Size H Right This device is used with the Persona Knee System for patients with se…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page