Synthes(R) Spine Vectra, Vectra-T and Vectra-One Anterior… recallZ-0782-2013 · Synthes USA HQ, Inc. · initiated Aug 6, 2012
Class II Terminated All recalls posted in February 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes USA HQ, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Although the reported incidence is low, there is the potential for the drill stop to slip on the drill bit resulting in an adverse event.
FDA root cause category: Component design/selection
What the company is doing
Synthes sent a "NOTICE: MEDICAL DEVICE RECALL" letter dated August 6, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 610-719-5450 for questions related to this notice.
Recall details
- Recall number
- Z-0782-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Aug 6, 2012
- Posted by the FDA
- Feb 4, 2013
- Recalling firm
- Synthes USA HQ, Inc.
- Quantity
- 48
- Product code
- KWQ
- FDA submissions
- K050451, K051665, K071667
- Distribution
- Nationwide Distribution-including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY.
Product description
Synthes(R) Spine Vectra, Vectra-T and Vectra-One Anterior Cervical Plating for Spinal Fusion, Adjustable Drill Stop, Part Number 03.613.012. Synthes Vectra System, Synthes Vectra-One System and Synthes Vectra-T System is intended for anterior plate and screw fixation of the cervical spine (C2-C7).
Code information (models, lots, serial numbers)
Adjustable Drill Stop, Part Number 03.613.012 - Lots 5456424, 5494147, 5557938, and 5557939
Other devices under the same FDA submission
13 devices cleared under K050451, K051665, K071667 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VECTRA 04.613.251Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY3 reports
- VECTRA 04.613.016Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- VECTRA 04.613.130Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- VECTRA 04.613.132Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- VECTRA 04.613.136Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- VECTRA 04.613.138Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- VECTRA 04.613.254Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- VECTRA 04.613.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 report
- VECTRA 04.613.134Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 report
- VECTRA 04.613.248Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 report
- VECTRA 04.613.384Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 report
- VECTRA-T 450.571Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 62943).
- Z-0781-2013 · TerminatedSynthes(R) Spine Vectra, Vectra-T and Vectra-One Anterior Cervical Plating for Spinal Fusion - 2.5mm Drill Bit, graduated 12mm - 26mm adjustable depth, Part Number 03.613.011. Sy…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page