Pressure Monitoring sets with VAMP Plus closed blood… recallZ-1146-2013 · Edwards Lifesciences, LLC · initiated Jan 30, 2013
Class II Terminated All recalls posted in April 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Edwards Lifesciences, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The device is being recalled because the firm has received multiple complaints related to incorrect assembly that may result in dilution of blood samples.
FDA root cause category: Process change control
What the company is doing
The firm initiated a recall in Europe via customer letters starting January 30, 2013 to notify the recall of model number T001691A lot number 59326333. The recall was expanded and a worldwide recall via customer letters was initiated on February 11, 2013. The letter identified the product, problem, and actions to be taken by the customers.
Recall details
- Recall number
- Z-1146-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jan 30, 2013
- Posted by the FDA
- Apr 18, 2013
- Recalling firm
- Edwards Lifesciences, LLC
- Quantity
- 30,467
- Product code
- KRA
- FDA submissions
- K896819, K925638
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AR, CA, FL, HI, ID, IN, KS, KY, LA, MA, MI, MN, MO, NC, ND, NM, NV, NY, OH, OK, PA, SD, TN, TX, VA, WA, and WI) and Europe.
Product description
Pressure Monitoring sets with VAMP Plus closed blood sampling system model # T001691A, PXVP0566, PXVP2260, PXVP0826, PXVP0827, PXVP2284AT3, PXVP23X3AT3, T398807D, T100605B, T001762A, T391T00A, VP2. Blood sampling system used on patients requiring periodic withdrawal of blood samples.
Code information (models, lots, serial numbers)
59184554, 59187809, 59191725, 59221801, 59227844, 59238613, 59326333, 59356987, 59401041, 59222349, 59227178, 59234808, 59284693, 59328047, 59270408, 59284617, 59276238, 59401734, 59270380, 59319038, 59306624, 59390840, 59195810, 59331224, 59397181, 59291446, 59272771.
Other devices under the same FDA submission
40 devices cleared under K896819, K925638 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- TRUWAVE, VAMP PLUS PXVP2284AT3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall9 reports
- TRUWAVE, VAMP PXAVMPEdwards Lifesciences LLC · Catheter, continuous flush1 recall7 reports
- TRUWAVE, VAMP PLUS PXVP2284Edwards Lifesciences LLC · Catheter, continuous flush1 recall6 reports
- TRUWAVE, VAMP PLUS PXVP0550Edwards Lifesciences LLC · Catheter, continuous flush1 recall5 reports
- TRUWAVE, VAMP PXAVMP3Edwards Lifesciences LLC · Catheter, continuous flush1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP23X3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP2260Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP23X3AT3Edwards Lifesciences LLC · Catheter, continuous flush1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP2272Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall1 report
- TRUWAVE, VAMP PLUS PXVP2X2AT3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall1 report
- TRUWAVE, VAMP PXVMP2X22Edwards Lifesciences LLC · Catheter, continuous flush1 recall0 reports
- TRUWAVE, VAMP PXVMP3X32Edwards Lifesciences LLC · Catheter, continuous flush1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page