Product Catalog (part number 20-0020) utilized with the… recallZ-0200-2014 · The Anspach Effort, Inc. · initiated May 6, 2011
Class II Terminated All recalls posted in November 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from The Anspach Effort, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Anspach Effort, Inc. Palm Beach Gardens, FL initiated a correction of its medical Device labeling for its Pneumatic Systems Operating Manual, Cutting Burr Chart and Product Catalog due to a recent review of their regulations and compliance requirements that determined the claim that the device is capable of dissection of biometals, bioplastics and cement has been verified, but is is not part of th
FDA root cause category: Labeling Change Control
What the company is doing
The firm, Anspach Effort, Inc., sent an "Urgent Medical Device Labeling Correction" letter dated June 27, 2011 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: 1) Destroy or delete any items in their inventory immediately and replace with the attached revised Operating Manual, Product Catalog and/or Cutting Bur Chart. 2) Complete and return the attached Customer Reply Form indicating disposition of the labeling materials and confirming their receipt of this letter, and fax to 1-800-327-6661. 3) If they distributed any of the products to other services or facilities, please forward this information as appropriate. For additional copies, contact Anspach Customer Support at (800) 327-6887.
Recall details
- Recall number
- Z-0200-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 6, 2011
- Posted by the FDA
- Nov 13, 2013
- Recalling firm
- The Anspach Effort, Inc.
- Quantity
- 2,020
- Product code
- HBE
- FDA submissions
- K080802
- Distribution
- Worldwide Distribution: USA Nationwide and the countries of: Egypt, India, Germany, Turkey, Canada, Saudi Arabia, Greece, United Kingdom, Cayman Islands, Kazakhstan, Republic of Korea, Taiwan, Hong Kong, China, Lebanon, Argentina, Denmark, Belgium, Netherlands, Mexico, France, Australia, Panama, Italy, Guam, United Arab Emirates, Israel, Polan, Venezuela, Austria, Hungary, Lithuania, Norway, Malaysia, Colombia, Costa Rica, Ecuador, Jamaica, Peru, Thailand, Italy, Czech Republic, Cameroon, South Africa, Portugal, Finland, Sweden, Nigeria, Bahamas, Cambodia, Dominican Republic, Switzerland, Chile, Singapore, Spain, Syrian Arab Republic, Jordan, Japan.
Product description
Product Catalog (part number 20-0020) utilized with the Electric and Pneumatic Motor Systems including: MicroMax Motor Systems, XMax Motor Systems, EMax2 Motor Systems, and Emax2Plus Motor Systems, all models and serial numbers. Product Usage: Indication for use is cutting and shaping bone.
Code information (models, lots, serial numbers)
part number 20-0020
Other devices under the same FDA submission
11 devices cleared under K080802 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- NA LONG-HDSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)22 reports
- NA R-SAWAnspach Effort, The · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)14 reports
- NA O-SAWAnspach Effort, The · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)5 reports
- NA S-SAWAnspach Effort, The · MOTOR, SURGICAL INSTRUMENT, AC-POWERED4 reports
- NA SC2100Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)4 reports
- NA SC2101Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED4 reports
- NA E-FP-DIRSynthes GmbH · MOTOR, DRILL, ELECTRIC3 reports
- NA E-FPSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)3 reports
- NA XL-HDSynthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED2 reports
- NA SC2102Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 65863).
- Z-0198-2014 · TerminatedPneumatic Systems Operating Manual (part numbers 19-0005) utilized with the Pneumatic Motor Systems including: BlackMax Motor Systems, MicroMax Systems and XMax Motor Systems all m…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.