Selenia Full Field Digital Mammography System with software… recallZ-2301-2013 · Hologic, Inc. · initiated Sep 6, 2013
Class II Terminated All recalls posted in September 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Hologic, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software: When the Selenia upon meeting the hard drive threshold of 4,000 images, may delete the most recently dated images
FDA root cause category: Software design
What the company is doing
Hologic sent an Urgent Medical Device Recall letter dated September 5, 2013, by Certified Mail to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised how to avoid loss of images until Hologic corrects the problem. A software upgrade to customers who are connected to Hologic Connect who have already been connected remotely and customers who have already been connected via a Service Call will receive notification of the problem and the date the system was remotely corrected notified by Certified Mail. A Dealer notification issued by certified mail to inform Dealers of the need to notify their customers of the problem and to correct the identified Selenia system using Hologic Technical Bulletin TB-00072. Selenia software version 5.2 will implement a double safeguard to both guarantee a 28 day retention period, as well as ensure images are removed in a least recently used order. If there are any questions regarding this field correction, please call Hologic Customer Support at 877 371-4372.
Recall details
- Recall number
- Z-2301-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 6, 2013
- Posted by the FDA
- Sep 30, 2013
- Recalling firm
- Hologic, Inc.
- Quantity
- 62 systems
- Product code
- MUE
- FDA submissions
- P010025
- Distribution
- Worldwide Distribution - USA including the states of AL, AR, CA, CO, FL, IL, KY, LA, NE, NC, NJ, NY, OH, PA, TN, TX, UT, WA and WV., and the countries of Canada, Denmark, Iran, Italy, Mexico and Taiwan.
Product description
Selenia Full Field Digital Mammography System with software versions 5.0.x and 5.1.x The Selenia Full Field Digital Mammography system generates digital mammography images that can be used for screening and diagnosis of breast cancer. The Selenia Full Field Digital Mammography System is intended for use in the same clinical applications as traditional screen-film mammography systems. Mammography images can be interpreted on either hard copy film or soft copy review workstations, cleared for mammography
Code information (models, lots, serial numbers)
Software versions 5.0.x and 5.1.x
Other devices under the same FDA submission
5 devices cleared under P010025 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Selenia Dimensions ASY-04160Hologic, Inc. · Full field digital,system,x-ray,mammographic2 recalls47 reports
- Selenia Dimensions ASY-01966Hologic, Inc. · Full field digital,system,x-ray,mammographic131 reports
- Selenia Dimensions ASY-07460Hologic, Inc. · Full field digital,system,x-ray,mammographic6 reports
- Selenia Dimensions ASY-08483Hologic, Inc. · Full field digital,system,x-ray,mammographic5 reports
- Selenia FFDM ASY-00676Hologic, Inc. · Full field digital,system,x-ray,mammographic5 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page