S5 Heart-lung machine, Stockert S5 System. Catalogue No:… recallZ-0142-2014 · Sorin Group Deutschland GmbH · initiated Sep 24, 2013
Class II Terminated All recalls posted in November 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Sorin Group Deutschland GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sorin Group is recalling certain models of S5 Perfusion Systems due to malfunction of cardioplegia control.
FDA root cause category: Software Manufacturing/Software Deployment
What the company is doing
Sorin Group sent an Urgent Field Safety Notice dated September 24, 2013, to all affected customers. The letter informed users of the potential for failing of automatic cardioplegia delivery function and notified them that they would be contacted by the Sorin Service Group to arrange an appointment to update the firmware on site. Customers were asked to complete the Response form to confirm they had received, read and understood the Field Notice. For questions regarding this recall call 303-467-6527.
Recall details
- Recall number
- Z-0142-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 24, 2013
- Posted by the FDA
- Nov 4, 2013
- Recalling firm
- Sorin Group Deutschland GmbH
- Quantity
- 84
- Product code
- DTQ
- FDA submissions
- K071318
- Distribution
- Worldwide Distribution - USA including WA, NY, VA, TN, SC, TX, CA, CO, MA, WI, MD, AL, ND, and Puerto Rico and Internationally to Italy, Finland, Germany, Sweden, and Great Britain.
Product description
S5 Heart-lung machine, Stockert S5 System. Catalogue No: 48-30/40/50-00. The Stockert S5 System is intended to be used during cardiopulmonary bypass for procedures lasting six(6) hours or less.
Code information (models, lots, serial numbers)
Serial Numbers: 48E00235,48E00339,48E00342,48E01753,48E02767,48E02753,48E02745,48E02746,48E02748,48E02416,48E01033,480E1038,48E01040,48E01041,48E01043,48E01044,48E01051,48E02631,48E01183,48E02758,48E02764,48E01832,48E01834,48E02278,48E02722,48E02732,48E02326,48E02797,48E02793,48E02791,48E02781,48E02785,48E02786,48E02783,48E02772,48E02802,48E02778,48E02784,48E02788,48E02789,48E02779,48E02792,48E02790,48E02771,48E02773,48E02776,48E02799,48E02801
Other devices under the same FDA submission
22 devices cleared under K071318 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- S5 Roller Pump 150 RP 150LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 death120 reports
- S5 Console 4-positionLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 death65 reports
- S5 Double Roller Pump 85 DRP 85LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS54 reports
- Electrical Remote-Controlled Tubing Clamp 620 mmLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS52 reports
- S5 Console 3-positionLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS37 reports
- Electrical Remote-Controlled Tubing Clamp 500 mmLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS31 reports
- Bubble Detector Sensor 3/8"LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 death26 reports
- S5 Console 5-positionLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS2 deaths16 reports
- Bubble Detector Sensor 1/4"LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 reports
- Display & Control Module (S5)LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 reports
- SCP Pump Control Panel for S5/C5LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 reports
- Bubble Detector Sensor 3/16"LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page