Bipolar liners and Multipolar cups with the following… recallZ-1019-2014 · Zimmer, Inc. · initiated Oct 11, 2013
Class II Terminated All recalls posted in February 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
FDA root cause category: Packaging
What the company is doing
Distributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.
Recall details
- Recall number
- Z-1019-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 11, 2013
- Posted by the FDA
- Feb 20, 2014
- Recalling firm
- Zimmer, Inc.
- Quantity
- 7,044,680 total
- Product code
- KWY
- FDA submissions
- K833991
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
Product description
Bipolar liners and Multipolar cups with the following dimensions in MM (38, 39, 40, 41, 42, 43, 44, 45, 46, 47, 48, 49, 50, 53, 54, 55, 57, 58, 60, 64, 66, 68, 70, 72) and several OD
Code information (models, lots, serial numbers)
Show the full code information (96k characters)
369276 369524 60291473 60310301 60342024 60357090 60399558 60425903 60452829 60558758 60570479 60584084 60590812 60599272 60611906 60620113 60657785 60682169 60687824 60708283 60711104 60821371 60841560 60907093 60984963 61022804 61070950 61077603 61134756 61154119 61175581 61209547 61228859 61270001 61287201 61368475 61368477 61530412 61533227 61533467 61692052 61732327 61774928 61783774 61824402 61870194 61873807 61879216 61910402 61977973 62063907 62075334 62106829 62146121 62150826 62240863 62291985 60563149 60566878 60600179 60630361 60815719 60828775 61066145 61144414 61162703 61213636 61261189 61313751 61408337 61461416 61504803 61522363 61756178 61783396 61847265 61923132 61998438 62133204 62298309 369341 370194 60222129 60353227 60368431 60373709 60399559 60466343 60507563 60554703 60579220 60584085 60599273 60623449 60656522 60661145 60662839 60687825 60708286 60709627 60709632 60711101 60841591 60867231 60884747 60976986 60976987 61008062 61036699 61070953 61115030 61135548 61135550 61175582 61200729 61200730 61260738 61279518 61283870 61283871 61345873 61382663 61401473 61453030 61504719 61549273 61549274 61549275 61549276 61677022 61692027 61722491 61742558 61779295 61810839 61824403 61859392 61882807 61887464 61895659 61977974 61993270 62002515 62118749 62148588 62183205 62262698 62291986 60581741 60581742 60612133 60643129 60670842 60885034 61006076 61017248 61135870 61151284 61151285 61162712 61191580 61191582 61213637 61296090 61313753 61388431 61408356 61461420 61554436 61695657 61728055 61751742 61803721 61847257 61899761 61985050 62046280 62103973 62182475 62275050 62311689 369521 370193 370858 370945 60222133 60266021 60291474 60310305 60325143 60325144 60333414 60342026 60353228 60357681 60361328 60368433 60383211 60392168 60392169 60421527 60439918 60439919 60452834 60452835 60472566 60499470 60507564 60515205 60537020 60557545 60563123 60575106 60581719 60584086 60595310 60599274 60611914 60630347 60630619 60643072 60643073 60656529 60661146 60666934 60674054 60687826 60702902 60704516 60708288 60761049 60761050 60778162 60788342 60802051 60807907 60815683 60815684 60815685 60815686 60815688 60838441 60847607 60862248 60867234 60867235 60877580 60884750 60907095 60926685 60945942 60970688 60970689 60970690 60970691 61006226 61006227 61006234 61006235 61006236 61052314 61066114 61066115 61066116 61070954 61081445 61081449 61115068 61115069 61115070 61135554 61135555 61135556 61135557 61161634 61161635 61175585 61175586 61175587 61175588 61175589 61208545 61208553 61208557 61228861 61242738 61242739 61242740 61270003 61270004 61292656 61303971 61321909 61333808 61345874 61361607 61363350 61366868 61375005 61382666 61382667 61399358 61411930 61427145 61442434 61442436 61453079 61453080 61467465 61486352 61486353 61513114 61513116 61570263 61570264 61570266 61570267 61570268 61605272 61605296 61605297 61605298 61605299 61605301 61629339 61651244 61666034 61676996 61686646 61687561 61687565 61687567 61726167 61734191 61738760 61750704 61750706 61760436 61762021 61770214 61776750 61779304 61783776 61786879 61805392 61805393 61812076 61817904 61817905 61831361 61831362 61847227 61851996 61855296 61863206 61880798 61892522 61893817 61898116 61902150 61908927 61913837 61918152 61920651 61923955 61928928 61932515 61935516 61962922 61964016 61987248 61989374 61996829 61998897 61999827 62015975 62024653 62035427 62043516 62046445 62046446 62057489 62057492 62060719 62067595 62080059 62093427 62093428 62105215 62105217 62114977 62119028 62133333 62138609 62140362 62167399 62178655 62181489 62181491 62193446 62210498 62212160 62226881 62236556 62240864 62249054 62268438 62289670 62291987 60566879 60581743 60588820 60594567 60600182 60612134 60623547 60630362 60656847 60664271 60682816 60687491 60713777 60780808 60796184 60800936 60817922 60828746 60841705 60844040 60857256 60863774 60901556 60919125 60939959 60950554 60977351 60990620 60997525 61018612 61018613 61019174 61066146 61066147 61091390 61135873 61135874 61162715 61162716 61162718 61162720 61183089 61183092 61183095 61191586 61213640 61243207 61243208 61261190 61261191 61261192 61261194 61296091 61313754 61344440 61344441 61344442 61344443 61385726 61385727 61385728 61385729 61408338 61408339 61408340 61408341 61408342 61427252 61427254 61427255 61461386 61461387 61461388 61474658 61482967 61486430 61489763 61489764 61500175 61506142 61516401 61520368 61538756 61547909 61547910 61547911 61561683 61593418 61593419 61593421 61610827 61639948 61658226 61658227 61658228 61684179 61697415 61697416 61697417 61713003 61713004 61732433 61732434 61742468 61751735 61751736 61751737 61764745 61766065 61778329 61778330 61783397 61783398 61783399 61799659 61799660 61803729 61815250 61819685 61833503 61841944 61847255 61852596 61852599 61854558 61857298 61868626 61868627 61882647 61891261 61895577 61902083 61913502 61913503 61923134 61931746 61931748 61931749 61935584 61939320 61976181 61980214 61985048 61987219 61987220 61993179 62010712 62033252 62046329 62046332 62057514 62057515 62064030 62085382 62085383 62097277 62097278 62100713 62100714 62115177 62115178 62149793 62162439 62166295 62167439 62170685 62183233 62284437 62285317 62296733 62310254 62321737 62327059 62334141 62334142 60557560 60902516 60967583 61066148 61091394 61094797 61144285 61144287 61162721 61344444 61427251 61463598 61616398 61660045 61689324 61736450 61764747 61819687 61907746 61928307 61982570 61998446 62103975 62128872 62167438 62293248 369343 370192 370866 60266024 60277249 60280134 60298323 60310307 60310309 60325150 60325152 60333415 60333417 60342027 60345115 60353229 60353230 60361329 60363815 60368435 60373710 60373712 60388293 60392170 60399560 60415974 60421528 60425904 60439920 60452838 60456402 60472569 60476181 60507565 60515207 60535546 60537021 60557546 60560679 60566663 60579221 60581721 60584087 60590816 60599275 60604266 60615722 60615724 60620115 60623452 60630348 60630620 60643074 60643075 60653302 60656531 60669118 60669119 60678667 60678668 60689290 60689291 60704517 60715757 60728393 60728395 60728396 60761052 60761053 60763086 60769408 60769409 60778163 60780765 60780766 60785187 60785188 60795747 60807908 60825096 60825097 60828923 60837635 60838442 60857205 60867238 60867239 60877581 60877582 60899748 60907098 60907101 60907102 60926686 60926687 60945944 60945945 60945946 60945947 60976990 61006252 61006255 61006261 61006263 61052319 61052322 61052325 61066118 61066119 61070970 61076284 61076285 61083431 61114993 61114994 61114996 61134757 61134758 61135558 61135559 61135560 61135561 61154147 61154149 61154155 61175590 61175591 61175592 61175593 61175594 61184018 61184019 61188701 61208566 61208572 61208575 61208580 61228862 61228863 61228864 61270009 61270010 61270011 61270012 61270013 61303982 61303983 61310070 61310071 61321910 61327932 61333840 61345875 61351733 61351734 61354884 61366866 61375006 61375007 61379427 61382669 61386027 61394196 61396404 61401482 61411948 61427148 61427151 61442438 61453031 61453034 61467466 61476062 61476063 61486354 61504729 61504730 61510357 61530405 61530406 61567170 61570231 61570233 61570240 61570257 61570259 61570261 61570262 61618273 61618274 61618275 61618280 61629343 61629344 61640738 61640739 61650968 61660351 61666685 61666687 61692047 61695252 61697801 61711753 61728418 61744639 61750676 61750677 61750678 61758616 61767868 61786881 61786882 61803828 61805385 61815844 61815845 61824406 61824407 61824408 61833443 61840617 61847224 61847225 61857258 61859394 61859396 61870218 61870219 61873808 61879915 61881522 61890359 61891561 61909551 61915753 61918778 61920652 61922674 61925110 61927014 61930499 61932519 61940073 61947761 61949873 61955231 61957533 61958878 61987249 61987250 61993271 61997559 62002516 62004192 62039658 62039659 62041553 62046447 62046448 62052955 62057502 62067597 62067599 62080044 62080046 62080047 62080058 62106830 62106831 62113231 62113232 62114958 62114960 62121289 62127575 62137180 62156232 62163481 62173347 62181492 62190366 62193455 62202262 62207244 62219239 62226887 62233562 62234520 62264204 62285278 62320117 370453 60222137 60237591 60266031 60277250 60277253 60298324 60298325 60310311 60310320 60310322 60333420 60333421 60333422 60342028 60345120 60346514 60353232 60361330 60361331 60363816 60367196 60368437 60373713 60373714 60373715 60373716 60388298 60388299 60392174 60399561 60399562 60399563 60401134 60421529 60421531 60421533 60425905 60425906 60425907 60452845 60452848 60456462 60460140 60461996 60461997 60461999 60466344 60477994 60477995 60503662 60503663 60503664 60508416 60517252 60517253 60517255 60527047 60535548 60537022 60566665 60579222 60584088 60593479 60599277 60599278 60599279 60599280 60615728 60615730 60620116 60620117 60623456 60623457 60630349 60636951 60639445 60639446 60646293 60648069 60653303 60656534 60661147 60662840 60662841 60669120 60686733 60689292 60691341 60696905 60698981 60709765 60715758 60715759 60715760 60715762 60721162 60721163 60728398 60728399 60752329 60752703 60752704 60761054 60761056 60769410 60774398 60778164 60778165 60780773 60785193 60785194 60788343 60788344 60788345 60807909 60807910 60807911 60815653 60815654 60815655 60815656 60815657 60838445 60847610 60847611 60857207 60857208 60859180 60867242 60867244 60877583 60877585 60884751 60907104 60907106 60926688 60926689 60945948 60945950 60945951 60945954 60945958 60945959 60973070 60996276 60996277 60996280 60996282 60996283 61022805 61022810 61022811 61022814 61022816 61022818 61022820 61036715 61036716 61036717 61036718 61052541 61052549 61063502 61063503 61063504 61063505 61063516 61070967 61070969 61076298 61076299 61081452 61115055 61115061 61115062 61120240 61135563 61135564 61135565 61135567 61135568 61154156 61154157 61154161 61154162 61154164 61175595 61175597 61175599 61175601 61184021 61184022 61184023 61184025 61184027 61184032 61208592 61208594 61208601 61208602 61208604 61208607 61224328 61224329 61224332 61224334 61224335 61228866 61228868 61228869 61242744 61242745 61242747 61260747 61270014 61270018 61270019 61270020 61279520 61279522 61295915 61295918 61295919 61295920 61295922 61312950 61312951 61327844 61327847 61337731 61342380 61345877 61354887 61361609 61366867 61379436 61379437 61382673 61386031 61394207 61399388 61401484 61401485 61427153 61427154 61427155 61442442 61442446 61442448 61467468 61467473 61467474 61476064 61476068 61486355 61486357 61510380 61510387 61510392 61510394 61537490 61537491 61538731 61538734 61538735 61553146 61558513 61558519 61562553 61562554 61567186 61567196 61567204 61600240 61600243 61600244 61600245 61608108 61608109 61608111 61608115 61608116 61629345 61637626 61643257 61647708 61647712 61647714 61654590 61666692 61666694 61671978 61673848 61676964 61682212 61684818 61687550 61687555 61687557 61695254 61697857 61711775 61711776 61711778 61711779 61727751 61727755 61734203 61734206 61750718 61750721 61762034 61767857 61774925 61779301 61779302 61779303 61783785 61783786 61786892 61786893 61790401 61809198 61809202 61812077 61812079 61817906 61820406 61824410 61824412 61840631 61840633 61852002 61852004 61855303 61855304 61857265 61859398 61863207 61863209 61863210 61864116 61864117 61864123 61871788 61876369 61879916 61893819 61899251 61904421 61909552 61914586 61926343 61930500 61932522 61936078 61938449 61939307 61941934 61949874 61952696 61957534 61957615 61965993 61966658 61968070 61968071 61974190 61979985 61987251 61996831 62006653 62006654 62006655 62010588 62010590 62010591 62010592 62012008 62012009 62027840 62027841 62027843 62030091 62035434 62042629 62060733 62060734 62060735 62076948 62079215 62079216 62097176 62109136 62109137 62109138 62109139 62111337 62111338 62111339 62111340 62113225 62113227 62121281 62127579 62142889 62144974 62144975 62144976 62146123 62154578 62154579 62163482 62169345 62171354 62171356 62187690 62188310 62190368 62197320 62197321 62198606 62208787 62209914 62210499 62212164 62216725 62216726 62220983 62224947 62224948 62239370 62251869 62254304 62255230 62266457 62268440 62268460 62282944 62288140 62291988 62294360 62309114 62312536 60535628 60557563 60564438 60566880 60573244 60581744 60584878 60588822 60588823 60600197 60604276 60611409 60611412 60612137 60612139 60630363 60636405 60638552 60643130 60648358 60653884 60661211 60664273 60664276 60664278 60670843 60670844 60676308 60680896 60687494 60699133 60713779 60713780 60722501 60727127 60732986 60746795 60746797 60752727 60752768 60774418 60780809 60786102 60791418 60800937 60807746 60815720 60815721 60821456 60821457 60822431 60828747 60834403 60841706 60844042 60844043 60854338 60857257 60863753 60863776 60896648 60896649 60901560 60901561 60919126 60930822 60934068 60967585 60977352 60985277 60990617 60990621 60998595 61006089 61018615 61018616 61018617 61018619 61018620 61019176 61066149 61066150 61066151 61066152 61066153 61091395 61091396 61094798 61135879 61135880 61144323 61144328 61144329 61144332 61144336 61144339 61156654 61156656 61157492 61157493 61162722 61162723 61183102 61183105 61183106 61183107 61183108 61183109 61191590 61191592 61191594 61213641 61213643 61213644 61228417 61228418 61243210 61243211 61243212 61243213 61243214 61261195 61261196 61261198 61261199 61261201 61261202 61296093 61296094 61296095 61313755 61313756 61313757 61356075 61356076 61356077 61356078 61356081 61356082 61356083 61356084 61356085 61356086 61356087 61385698 61385704 61385708 61385715 61400606 61400607 61400608 61400619 61400623 61400624 61400630 61408333 61408334 61408335 61408336 61450374 61450375 61450376 61450377 61450379 61469281 61469282 61469284 61469285 61469286 61469288 61469289 61482935 61482936 61482939 61482940 61483027 61483028 61489767 61489769 61500171 61500173 61500174 61514244 61514245 61514246 61514247 61538757 61538758 61538759 61561684 61561685 61561686 61561687 61561688 61561689 61593422 61593423 61593425 61610829 61637761 61637762 61648351 61648352 61658229 61658230 61658231 61666586 61666587 61666588 61666589 61666590 61666591 61695660 61695662 61700473 61713005 61718650 61728062 61728063 61732440 61732442 61736447 61747730 61747731 61747732 61747733 61751726 61751727 61751728 61764742 61767887 61767888 61767889 61767890 61772451 61772452 61772453 61778319 61778320 61778322 61798246 61798247 61803715 61803716 61809247 61810035 61810036 61810037 61815227 61819689 61833486 61833487 61833489 61835075 61837205 61857310 61857311 61857313 61868636 61868640 61871947 61871948 61873845 61875680 61881266 61895575 61899759 61907744 61910270 61910272 61937075 61939322 61939395 61941050 61941051 61953144 61976184 61980216 61980217 61980219 61982560 61987223 61987224 61987225 61992546 61992548 62010714 62010715 62036987 62036988 62042788 62042789 62064038 62073539 62073540 62077403 62082219 62082220 62082221 62091035 62091036 62092935 62092937 62100715 62112193 62112194 62115170 62115172 62133210 62137201 62137202 62137203 62163530 62163531 62163532 62164629 62164643 62166299 62166301 62170687 62170689 62174662 62174663 62174664 62182871 62310253 62317219 62317221 62320032 62320842 62322507 62326206 62326208 62329475 62329486 62334143 62334144 60687495 60727129 60807747 60807748 61018621 61066154 61162728 61243215 61243217 61296096 61445070 61445072 61538761 61608136 61610831 61713008 61764740 61767883 61819692 61906156 61910262 61928310 61987229 62010718 62060180 62060181 62167437 62293258 60237595 60257242 60257243 60266035 60266037 60277258 60310326 60310329 60314503 60333425 60333426 60345134 60353235 60353236 60357093 60357094 60357687 60361332 60361333 60363818 60367197 60368440 60368444 60373718 60373719 60373721 60373722 60373723 60392175 60392176 60399566 60406139 60412185 60421535 60421537 60425933 60425934 60425935 60452851 60452856 60460143 60460144 60462001 60462002 60462003 60462005 60466347 60477996 60477997 60503666 60508418 60508419 60508421 60517256 60517257 60517258 60517259 60517260 60527049 60527050 60535553 60539549 60547631 60563125 60563126 60566666 60575108 60579223 60581724 60581725 60589099 60589100 60589101 60599281 60599282 60599283 60599284 60615737 60615738 60620119 60620120 60620121 60623459 60623460 60630350 60639447 60639448 60639449 60639450 60643077 60643079 60656537 60661148 60662842 60662843 60669121 60670134 60674056 60674057 60674059 60686739 60689293 60691342 60696907 60702905 60708107 60708108 60709768 60709769 60709770 60709773 60715764 60715765 60721165 60740786 60752332 60752705 60752706 60752707 60761058 60763089 60763090 60764198 60769412 60778167 60778168 60780775 60780776 60785196 60785199 60785203 60788348 60802015 60821377 60821379 60825099 60833934 60833935 60833936 60838447 60841564 60841594 60854017 60857210 60859183 60867251 60877587 60877588 60877589 60884755 60907111 60907113 60907114 60907115 60926690 60945964 60945967 60945969 60945970 60945971 60945972 60984965 60984968 60984969 60996287 61022834 61022837 61022838 61022839 61022841 61022842 61022843 61022844 61036752 61036753 61036754 61036758 61036759 61052553 61052555 61052561 61052563 61066122 61066123 61070972 61070974 61070975 61070976 61076287 61076288 61081454 61081455 61083434 61110889 61110892 61110893 61120244 61134763 61134765 61135569 61135572 61154172 61154173 61154216 61154218 61175606 61175609 61184034 61184036 61184039 61184040 61184041 61184042 61188705 61188706 61188710 61200735 61200740 61200746 61200747 61200748 61228871 61228875 61234686 61234687 61234688 61234690 61242751 61242752 61242753 61242754 61242755 61242759 61242760 61260742 61270024 61279524 61279525 61279526 61284342 61295929 61295930 61295942 61295943 61303985 61303986 61321912 61321914 61327848 61337742 61337750 61345878 61345879 61345880 61354889 61361610 61363354 61363357 61375012 61379444 61382679 61386042 61386044 61394210 61396408 61399391 61399392 61401488 61411952 61421199 61427157 61427159 61441741 61441742 61442451 61442457 61442459 61453065 61467482 61467483 61467485 61467487 61476072 61486359 61504701 61504705 61504708 61504711 61513081 61513082 61522284 61522286 61530396 61530397 61536148 61536149 61538736 61538738 61553148 61555111 61555112 61558506 61567268 61567269 61567274 61600166 61600168 61600170 61600173 61600179 61608104 61608105 61608106 61608107 61629353 61629357 61637642 61637643 61637644 61640747 61640748 61643249 61643251 61647681 61654623 61660347 61660349 61660350 61666677 61666680 61671976 61676949 61682210 61684824 61686644 61687539 61687542 61697851 61711764 61711765 61711766 61711773 61728419 61728420 61728423 61738765 61738766 61738768 61744664 61750697 61750698 61750702 61758622 61767870 61767871 61767873 61776749 61782385 61782386 61782387 61790405 61790406 61790407 61795037 61798118 61798697 61803808 61803809 61803810 61805362 61809191 61820409 61820410 61831385 61833444 61841678 61841680 61847230 61847233 61847234 61855301 61859399 61859401 61859407 61861649 61861650 61861651 61879127 61881361 61883801 61885507 61888823 61890361 61897111 61903291 61904424 61914588 61926345 61926346 61926347 61934153 61936079 61941024 61945714 61951307 61952697 61957619 61957624 61957625 61969208 61976209 61976210 61976211 61976789 61980758 61993215 61998901 62004193 62005431 62006657 62006658 62010594 62015981 62015982 62017539 62017540 62026452 62026453 62039669 62046449 62046450 62050998 62057507 62059767 62082194 62082195 62082196 62082197 62082198 62082199 62082201 62090612 62090614 62090615 62090616 62090617 62097142 62119030 62119031 62119032 62119033 62119036 62124077 62125342 62125343 62126369 62126370 62127577 62142883 62142884 62154580 62154581 62158171 62163484 62163485 62166273 62166274 62166275 62177352 62177353 62178304 62178658 62178659 62183207 62188313 62190369 62199452 62204607 62210501 62216721 62222809 62224949 62224950 62233564 62234522 62236558 62240867 62251879 62256069 62262703 62279792 62284376 62291990 62294363 62294380 62300237 62300239 62304842 62334124 60277263 60298294 60298295 60298296 60310333 60310334 60310363 60310365 60314505 60314506 60314507 60325156 60325163 60333430 60333431 60342030 60342031 60345181 60345182 60345183 60353237 60353238 60357096 60357688 60357689 60357690 60361334 60363820 60368446 60368447 60368449 60368450 60368453 60373724 60373725 60373726 60373727 60373729 60373730 60373731 60373733 60373734 60388301 60388302 60392178 60392179 60392180 60392181 60406140 60406141 60412186 60412187 60412188 60415975 60415976 60415978 60421539 60421541 60421542 60421543 60425937 60425938 60425940 60434153 60434154 60434155 60434156 60434157 60434159 60434160 60439922 60439923 60439924 60439925 60439926 60442577 60456486 60456487 60462011 60462013 60462014 60462015 60462016 60466348 60466349 60466351 60466352 60477999 60478001 60478002 60478003 60478004 60492467 60492469 60492471 60492472 60492473 60492474 60508422 60517277 60517278 60517279 60517280 60517281 60517282 60525147 60525148 60527052 60527054 60527055 60527057 60531796 60531797 60531798 60531799 60545146 60547635 60547643 60547644 60549930 60549931 60549932 60557547 60558766 60558767 60558769 60560680 60563128 60566668 60575110 60579225 60579226 60579227 60581726 60581727 60581728 60581729 60584111 60584113 60584114 60584116 60589102 60593481 60593482 60599286 60599287 60604269 60611916 60611917 60611918 60611920 60620123 60620124 60620125 60623464 60630351 60630721 60636958 60636961 60636962 60636963 60636965 60639451 60646296 60648071 60656575 60656576 60656577 60656578 60661150 60669122 60669139 60669143 60669146 60669149 60669152 60674061 60678670 60678671 60682182 60685523 60689294 60696908 60696909 60702908 60702909 60708109 60709777 60709781 60709784 60709787 60709791 60709793 60709795 60709796 60709798 60709803 60709804 60709805 60709806 60709807 60715768 60715770 60721166 60721167 60721168 60728409 60728410 60728411 60728412 60728414 60728416 60728419 60752333 60752334 60752338 60752340 60752709 60752710 60752711 60761061 60761062 60761063 60763081 60763084 60763091 60763092 60764199 60764202 60767148 60769416 60769417 60778152 60780779 60785214 60788350 60788351 60788352 60788353 60788354 60795756 60802016 60802019 60802055 60802056 60802057 60802059 60807915 60812451 60812452 60812458 60812461 60812462 60812463 60812464 60812465 60812467 60812469 60828926 60833937 60833939 60833940 60837631 60837633 60841595 60841596 60844745 60844746 60847616 60854018 60857217 60857218 60859172 60859191 60867254 60877591 60877594 60877596 60877598 60884758 60884761 60899756 60899757 60899758 60899759 60907117 60907119 60907123 60926695 60926696 60926698 60926699 60926704 60926705 60926706 60926707 60945981 60945988 60945989 60945990 60945995 60945997 60945999 60946001 60973093 60973095 60976994 60977015 60977026 60977033 61006285 61006286 61006289 61006292 61006301 61006303 61006305 61006306 61006307 61006308 61006324 61006327 61006328 61006346 61006348 61006349 61006350 61006353 61006357 61006358 61036771 61036772 61036773 61036774 61036775 61036776 61036780 61036782 61052576 61052582 61052587 61052595 61052605 61052615 61052617 61077673 61077676 61077678 61077681 61077684 61077688 61077689 61077690 61077691 61077692 61077695 61089206 61089207 61089208 61089211 61089216 61089217 61102821 61102823 61102825 61102826 61102827 61110859 61110860 61120252 61120253 61120255 61134769 61134773 61134774 61134777 61134778 61134779 61134782 61134783 61134784 61135575 61135580 61135582 61135583 61135584 61175614 61175615 61175617 61175619 61175620 61175621 61175623 61184045 61184050 61188714 61188717 61188718 61188720 61200769 61200784 61200788 61200805 61200847 61200866 61208614 61208615 61208619 61228877 61228879 61228880 61228881 61234693 61234695 61234697 61234698 61234701 61234702 61234703 61242765 61242769 61242775 61242780 61242789 61242796 61242798 61242799 61242801 61242802 61242805 61252227 61252229 61252230 61252235 61252236 61252237 61252238 61260732 61260734 61270029 61270031 61270033 61270038 61270041 61270043 61270047 61279527 61279528 61279531 61279533 61284345 61295949 61295953 61295954 61295956 61310072 61310073 61310078 61310080 61310081 61312872 61312875 61312877 61321917 61327853 61333843 61333844 61337752 61342389 61345881 61345882 61345883 61345884 61345885 61345888 61351747 61351750 61351751 61351754 61375014 61375021 61375023 61379447 61379448 61379451 61382683 61382684 61382685 61386008 61386056 61394211 61396412 61408067 61411955 61411956 61415033 61424481 61437027 61442464 61453036 61453038 61453039 61453041 61453045 61453046 61453048 61453049 61463524 61467489 61476077 61476078 61486365 61486366 61486367 61499952 61499960 61504691 61504695 61504699 61513085 61513088 61542306 61542307 61547979 61547980 61551423 61551425 61553158 61554646 61558189 61558190 61558191 61558194 61558196 61558197 61560501 61560502 61566054 61567298 61567301 61567303 61568182 61568185 61568187 61568188 61568192 61568194 61568196 61568198 61568199 61600184 61600188 61600190 61605374 61631066 61637617 61637618 61637620 61643241 61643242 61643244 61643245 61643248 61647667 61647668 61650970 61654511 61654512 61654527 61660338 61660340 61661387 61664094 61664098 61664099 61666027 61692029 61692031 61695229 61695230 61697178 61697766 61697770 61697786 61697794 61710637 61710639 61710644 61727758 61727761 61727762 61727763 61728432 61728433 61732342 61732350 61732351 61734198 61734201 61734202 61744661 61744662 61750723 61750724 61750725 61750727 61764662 61764663 61764665 61767860 61767875 61770227 61770228 61770230 61774931 61776745 61779316 61779317 61782367 61782369 61786902 61786903 61786907 61786908 61798123 61798698 61802419 61802420 61803820 61805371 61813807 61818540 61818541 61820412 61820413 61820414 61820415 61820417 61833429 61837073 61839063 61841665 61851989 61855287 61855288 61855292 61855293 61857254 61857255 61861635 61861637 61861638 61861640 61863212 61867265 61867268 61867269 61871792 61879917 61879918 61881362 61883803 61886048 61906242 61906246 61906247 61910406 61910407 61911410 61911411 61912488 61913838 61918779 61918780 61918781 61923958 61927016 61936081 61936083 61940991 61943520 61951309 61965995 61968074 61974189 61977979 61980762 61980763 61980764 61980767 61983223 61983225 61990135 61990136 61990137 61992598 61992599 61996836 61997563 61998902 61999830 62006662 62006664 62012011 62012012 62015977 62015979 62017537 62018441 62027844 62027847 62030105 62035443 62036923 62036924 62039670 62041547 62042632 62052962 62052966 62057498 62062923 62062926 62063911 62065364 62065365 62076960 62076961 62079228 62079233 62090618 62090619 62097146 62097147 62097148 62100701 62100704 62102488 62102489 62103585 62105204 62105209 62105210 62111333 62111334 62114956 62114957 62121288 62122112 62122113 62123070 62123071 62125340 62126375 62127586 62137187 62138611 62138612 62148584 62160614 62160615 62163486 62165218 62166268 62169344 62170604 62170605 62171358 62171360 62177356 62178305 62181495 62183208 62186356 62187693 62188316 62192523 62192525 62194610 62197311 62197313 62199453 62199463 62199467 62203197 62209916 62215694 62216728 62221896 62225748 62232601 62233565 62233566 62234523 62234525 62234526 62240869 62240872 62254306 62255231 62255235 62256070 62257232 62260159 62260160 62262684 62262686 62265300 62266456 62268232 62268442 62268443 62277157 62277989 62277991 62278724 62288148 62291992 62298328 62300243 62300244 62326153 60630364 60648362 60664279 60670846 60687496 60774462 60786105 60800938 61183113 61213645 61213646 61356168 61482983 62057520 60611414 60727130 60733007 60866565 60977354 60990624 61261206 61356090 61461408 61483014 61520356 61561695 61612606 61700478 61756186 61799635 61847263 61864136 61935581 61964891 61982572 62073548 62115176 62166294 62293263 62306055 62312645 60515260 60541486 60541487 60541488 60557568 60563155 60566881 60566882 60575158 60575159 60576014 60581747 60581748 60588840 60588841 60588842 60588843 60588844 60598321 60598322 60598323 60599798 60599799 60600210 60600211 60600212 60604293 60611417 60611419 60611420 60612140 60612141 60612143 60623551 60624814 60624815 60636413 60643133 60648364 60648365 60653886 60653887 60653888 60656858 60661215 60661216 60664288 60664292 60664295 60664297 60664301 60664302 60664304 60670848 60670849 60670850 60670851 60676323 60676324 60676325 60680902 60682819 60687501 60687502 60699135 60699137 60699138 60713787 60713788 60713789 60713790 60722506 60722507 60722508 60727131 60733011 60733013 60738735 60740579 60746806 60746807 60746809 60752730 60752731 60752732 60752777 60764224 60769432 60774485 60780812 60780813 60786107 60791365 60791366 60791420 60791421 60800931 60800932 60800933 60807773 60807776 60807779 60807782 60817923 60817924 60822434 60822435 60828730 60828749 60828750 60834397 60834398 60834404 60841710 60841711 60844045 60844046 60844047 60854340 60854341 60854342 60857260 60857261 60857262 60863758 60863759 60863760 60863779 60863780 60885039 60896664 60896667 60896669 60896671 60901562 60902519 60902520 60919128 60919129 60919130 60919131 60919132 60930824 60930826 60934071 60939963 60946219 60946220 60950559 60950560 60967587 60967588 60973586 60973587 60977356 60985298 60985301 60985306 60985309 60990616 60990626 60997542 60997543 60997544 61006093 61006094 61018628 61018629 61018630 61018631 61018633 61018634 61019172 61036961 61036962 61055665 61055666 61055667 61055668 61055669 61055670 61066162 61066163 61066164 61066165 61066166 61066167 61091408 61091409 61091410 61091411 61091412 61091413 61091415 61120692 61120693 61120694 61120696 61120697 61120698 61120699 61130772 61130773 61130775 61130777 61156645 61156646 61156647 61156648 61156649 61156650 61156651 61162730 61162733 61162734 61162735 61162739 61162742 61162743 61191604 61191605 61191606 61191607 61191608 61191609 61191610 61191611 61191612 61191613 61213649 61213650 61213652 61213653 61213654 61213655 61213656 61213657 61213658 61213659 61242109 61243222 61243223 61243224 61243225 61243226 61243227 61243228 61243229 61243230 61243231 61243232 61243233 61261209 61261210 61261211 61261212 61261213 61261214 61261215 61261216 61261217 61261218 61296098 61296099 61296101 61296102 61296103 61296104 61296105 61296106 61296107 61296108 61313766 61313767 61313768 61313769 61313770 61313772 61313773 61324667 61325772 61325773 61325774 61326546 61326547 61326548 61326549 61337933 61337935 61337936 61337937 61337938 61337940 61337941 61337942 61356057 61356058 61356059 61356060 61356061 61356063 61356064 61356066 61377138 61377140 61388423 61388424 61388425 61388427 61388430 61396652 61396653 61396654 61396655 61396656 61396657 61396658 61396659 61396660 61396661 61400520 61400543 61400582 61400591 61400596 61400599 61400601 61408324 61408326 61408327 61408328 61408329 61443079 61443080 61443081 61443082 61443083 61443084 61450345 61450346 61450348 61450350 61450351 61450352 61450353 61450354 61458214 61458215 61458216 61458217 61461389 61461390 61461391 61461394 61461395 61461396 61461397 61469294 61469299 61469300 61469302 61486432 61486433 61486434 61489786 61489787 61489788 61489790 61489791 61489793 61500163 61500164 61500165 61500166 61500168 61500169 61500170 61504808 61506133 61506134 61506136 61506137 61514226 61514227 61514228 61514229 61514230 61514232 61520350 61520351 61522316 61522317 61522318 61522319 61522320 61526573 61526575 61538768 61538770 61538771 61538772 61561698 61561699 61561701 61561704 61561705 61561706 61561708 61561709 61567686 61567687 61605548 61605549 61610835 61610836 61610839 61610842 61632200 61632201 61632202 61637765 61637766 61637767 61637768 61637769 61639950 61648356 61648357 61648358 61648359 61648360 61648361 61648362 61648363 61658232 61658233 61658234 61658235 61658237 61658238 61658239 61676460 61676461 61676462 61684181 61684182 61684184 61689327 61689328 61689329 61689330 61695668 61695672 61695673 61695677 61713015 61723047 61723048 61728065 61732435 61732437 61736440 61736442 61736443 61740395 61746862 61746863 61746864 61746865 61746867 61746869 61746870 61746871 61746877 61751729 61751730 61751731 61751732 61751733 61751734 61764746 61767891 61767892 61767893 61767894 61772441 61772443 61772444 61772445 61776710 61776711 61776722 61778305 61778306 61778307 61778308 61778309 61782462 61782463 61782464 61794268 61798242 61798243 61798244 61798245 61799675 61799676 61803730 61815231 61815232 61815234 61815235 61815236 61819698 61819699 61819702 61819703 61823605 61823606 61823607 61828296 61828297 61833478 61833479 61835066 61835069 61837191 61837193 61837194 61839830 61839831 61841930 61841931 61847269 61852609 61852610 61855322 61855323 61858160 61858161 61861717 61861719 61864141 61864143 61864145 61867401 61867402 61867403 61867404 61871959 61871961 61871968 61873847 61873848 61873849 61879323 61879324 61881272 61881273 61882648 61884437 61891268 61895570 61895574 61902103 61902104 61902105 61902108 61908944 61908945 61912460 61920541 61925655 61925656 61928947 61928948 61928949 61931759 61931760 61934544 61939324 61939325 61945138 61945139 61945142 61948390 61948393 61952520 61952524 61953154 61953155 61953156 61957673 61957675 61957676 61957681 61968326 61980224 61985052 61985054 61985055 61992555 61992557 61993184 61993185 61993187 61993188 62002356 62002357 62007738 62007739 62007740 62007741 62020940 62020941 62020945 62027858 62033254 62033255 62036980 62036981 62040504 62040505 62042792 62042793 62042794 62042795 62050788 62065575 62065576 62065577 62065578 62069023 62069024 62069025 62069026 62071955 62071956 62071957 62071959 62071961 62071962 62071963 62080199 62080200 62080201 62080202 62080204 62080205 62091037 62091038 62091039 62091040 62097273 62097275 62097276 62097279 62097280 62098890 62098891 62099681 62099682 62103590 62103591 62103592 62109977 62115180 62115181 62126449 62126450 62126451 62126452 62127604 62127605 62127606 62127607 62127608 62127609 62127610 62134053 62135129 62135130 62137178 62141839 62143885 62143886 62143887 62154599 62161573 62161575 62162182 62162183 62164627 62164628 62164645 62170702 62170703 62172064 62175376 62175377 62175379 62175380 62180240 62180241 62180242 62181553 62188398 62188399 62191275 62191276 62191277 62193552 62193554 62195532 62198705 62284439 62284442 62285305 62285307 62288153 62292037 62292038 62293265 62296739 62298319 62301501 62306034 62306035 62306036 62306037 62306039 62307776 62311696 62311697 62311698 62312635 62315234 62317207 62317209 62317211 62317824 62317825 62317826 62317827 62317834 62320035 62326201 62326202 62327793 62329478 62329479 62329480 62329481 62329483 62329484 62329528 62329530 62333068 366761 60277265 60277266 60277267 60277268 60298299 60314510 60314511 60314512 60333432 60333433 60333438 60333439 60333440 60345171 60345172 60346522 60346523 60346525 60346526 60346528 60353240 60357099 60357101 60357692 60357693 60357694 60361355 60361356 60363821 60367201 60367202 60368457 60368459 60368460 60368461 60373735 60373736 60373738 60373739 60373741 60373742 60388294 60388303 60388305 60392182 60392183 60399567 60399570 60399571 60399573 60403294 60406142 60406144 60412189 60414138 60421544 60421545 60421546 60425945 60425946 60425948 60425949 60434162 60434163 60436352 60436353 60436354 60439927 60439928 60442585 60462020 60462021 60462022 60462023 60462024 60466353 60466354 60466355 60478005 60478007 60478009 60478011 60492477 60492478 60492479 60492480 60492481 60508424 60508425 60508426 60517286 60517287 60527061 60527062 60527066 60531803 60531804 60537025 60537026 60539551 60539552 60547647 60556516 60557548 60557549 60558770 60563129 60563130 60566669 60569795 60575113 60579228 60579229 60581730 60581731 60581732 60584119 60584120 60589103 60589104 60593483 60599292 60599293 60599294 60599295 60599296 60599298 60604270 60611922 60611923 60611924 60611926 60611928 60620126 60623466 60623467 60623470 60630352 60630353 60630722 60630723 60636967 60636969 60636970 60639452 60642610 60646298 60653304 60653305 60656579 60661151 60661152 60661153 60666923 60669158 60669160 60674062 60678672 60678673 60678674 60682191 60682192 60685526 60686749 60691354 60696912 60698984 60702910 60702911 60708111 60708112 60709812 60709814 60709815 60709816 60709817 60715774 60715775 60715776 60715777 60715779 60721242 60721244 60728422 60728425 60728426 60728427 60752342 60752343 60752712 60752713 60752715 60761076 60763095 60764204 60764206 60764207 60769420 60778154 60778175 60780780 60780781 60785215 60785216 60788326 60788355 60788357 60795757 60795758 60795759 60802020 60802060 60802061 60802062 60807919 60815661 60815662 60815664 60815666 60815668 60815671 60815673 60815674 60815678 60815679 60815680 60815681 60815682 60838449 60838450 60838451 60847618 60857220 60857222 60857223 60859195 60867259 60867260 60867261 60877602 60884771 60884773 60884774 60899761 60899762 60899764 60899767 60907128 60907129 60907130 60907131 60907133 60907134 60907135 60926708 60926719 60926720 60926721 60926722 60946004 60946006 60946008 60946011 60946015 60946017 60946088 60946089 60946093 60946094 60946096 60946098 60977036 60977039 60977040 60977041 60977042 60977043 60977045 60996260 60996262 60996263 60996267 60996269 60998715 60998716 60998719 60998720 61022849 61022850 61022851 61022852 61022855 61022856 61022857 61022859 61022860 61022861 61036761 61036762 61036763 61036764 61036765 61036768 61036769 61052627 61052630 61052635 61052636 61052637 61052640 61052641 61052642 61077696 61077697 61077703 61077704 61077705 61077710 61077725 61077726 61077728 61077729 61089218 61089221 61089222 61089225 61089226 61089227 61102811 61102812 61102814 61102817 61102820 61110871 61110878 61120258 61120259 61120260 61120262 61130577 61130578 61130580 61130581 61130582 61135586 61135587 61135588 61135592 61154297 61154299 61154304 61154316 61154339 61154346 61175624 61175626 61184060 61184064 61184065 61184066 61188726 61188727 61188728 61200895 61200898 61200902 61200904 61200906 61200910 61200920 61200925 61209501 61209503 61209506 61209961 61209963 61228882 61228884 61234710 61234711 61234714 61234716 61234717 61234720 61234721 61234723 61242807 61242808 61242809 61242811 61242812 61242813 61242815 61242817 61242819 61256442 61256444 61256445 61256446 61256449 61256450 61260737 61270050 61270051 61270052 61270055 61270056 61270057 61270059 61279536 61279539 61279541 61279543 61292646 61292647 61292648 61292649 61292650 61292654 61295961 61295962 61295963 61295964 61303990 61310085 61310087 61310089 61310092 61312939 61321919 61321920 61327854 61327856 61327857 61333821 61333823 61333826 61337772 61337782 61342396 61342397 61345889 61351757 61351758 61351759 61351762 61351764 61354892 61354894 61361611 61361612 61361613 61363358 61363359 61363360 61363361 61366875 61375026 61375028 61382689 61386064 61394198 61394213 61394214 61396413 61399393 61399394 61399395 61401493 61401495 61421207 61427164 61427165 61427166 61441733 61441734 61441736 61441737 61453067 61453070 61467496 61467497 61476106 61486376 61486377 61486378 61486380 61486382 61486384 61499966 61499972 61499974 61499978 61529584 61529585 61537494 61537875 61542257 61557053 61557055 61558469 61562556 61562557 61562558 61562559 61562562 61567277 61567278 61567279 61567280 61567281 61567282 61567283 61567284 61567285 61567286 61567289 61581706 61581708 61581716 61581719 61608085 61608086 61608089 61608092 61608096 61608097 61618293 61618308 61618312 61618313 61618314 61631045 61631047 61637627 61637628 61640754 61650971 61650973 61654618 61654620 61654621 61654622 61664101 61664106 61666674 61676974 61682203 61682205 61691367 61691378 61691379 61691381 61707319 61707321 61710655 61710658 61710659 61710660 61710662 61711763 61726170 61738770 61738771 61738772 61738773 61738777 61744654 61744655 61744656 61744657 61750686 61750687 61750690 61758614 61758615 61764668 61770238 61774934 61776748 61783787 61783788 61786926 61790411 61790414 61790415 61795032 61795033 61798114 61798687 61798690 61798691 61803804 61803806 61805355 61809189 61824427 61824429 61828148 61828149 61828150 61833436 61833438 61833439 61837075 61837076 61840625 61840626 61841654 61841655 61851063 61851064 61851065 61857259 61861306 61861307 61861308 61870197 61870198 61870199 61870200 61870221 61871795 61871797 61876374 61876376 61879217 61883804 61883805 61885508 61885509 61890364 61900836 61900838 61900840 61903288 61905809 61905810 61906248 61914591 61914592 61916895 61918782 61919598 61920653 61928929 61930501 61932525 61932526 61936086 61938450 61945716 61945717 61948427 61951311 61955920 61957633 61957636 61957637 61957638 61966659 61969366 61974191 61974192 61980760 61982539 61983226 61985271 61989377 61989378 61993274 61998903 61998904 62004196 62009806 62018445 62018446 62020924 62024654 62024655 62024657 62030958 62030959 62036921 62036922 62039676 62046437 62046438 62046439 62052967 62052968 62059769 62063912 62066611 62076962 62076963 62080050 62080051 62080052 62080054 62082188 62082189 62082190 62090607 62090610 62111343 62111344 62113234 62113235 62114962 62114963 62114965 62114974 62114976 62119855 62121284 62127573 62131450 62137183 62141816 62141817 62143878 62156234 62158168 62158169 62160232 62166278 62167404 62167405 62173351 62174657 62176449 62177358 62181232 62190371 62193450 62193454 62199456 62204602 62210503 62213276 62215043 62215691 62216719 62216720 62233568 62234527 62234528 62234530 62239373 62240335 62240873 62240874 62240875 62240877 62240911 62249061 62257233 62262544 62262687 62262688 62267521 62268234 62268445 62268465 62277158 62277987 62284380 62291129 62298330 62298338 62299155 62304847 62305924 62311775 62312523 62317235 62317811 62326158 62330824 62334090 366717 369348 370119 60290885 60290888 60290891 60290893 60297073 60297074 60314513 60314520 60325173 60325176 60333444 60333446 60333448 60333449 60342032 60342033 60342034 60345177 60346530 60346558 60346559 60353243 60357695 60357696 60361357 60361358 60363822 60367203 60367204 60368462 60368463 60368464 60374114 60374115 60374116 60374118 60374119 60374120 60388295 60392185 60392186 60392188 60399574 60399575 60399576 60399577 60399578 60403295 60403296 60406145 60406146 60412193 60412194 60412195 60421547 60421549 60421550 60425952 60425953 60425954 60425955 60434169 60436355 60436356 60436357 60436358 60436359 60439929 60442586 60442587 60442588 60442589 60462025 60462026 60462027 60462029 60462030 60462032 60462033 60478012 60478013 60478014 60478015 60478016 60478017 60492486 60492487 60492488 60492489 60492490 60492491 60492492 60508427 60517288 60525149 60527069 60527070 60527071 60527072 60527073 60527074 60527075 60527076 60527077 60531805 60537029 60539554 60541433 60541434 60541435 60549933 60549934 60549935 60549937 60563131 60563132 60566671 60566672 60575114 60575115 60579230 60579231 60581733 60581734 60584123 60584124 60593484 60595314 60595315 60604271 60611929 60611931 60611932 60611933 60620129 60620130 60620131 60623472 60623474 60630355 60630724 60630725 60639453 60641954 60641955 60641956 60641957 60642612 60642613 60656582 60656583 60656584 60657786 60660568 60660575 60660576 60662844 60662845 60662847 60662848 60669161 60669162 60669163 60674063 60678688 60682195 60685527 60689311 60691357 60696914 60696915 60698986 60702912 60702913 60709818 60715783 60715786 60715787 60721176 60721177 60721180 60721181 60721182 60721183 60728439 60728445 60728448 60728449 60728456 60740788 60752344 60752716 60752717 60752718 60761082 60761085 60761086 60763096 60764208 60764209 60764210 60769421 60769422 60769423 60778176 60778178 60780782 60780783 60780784 60785218 60785220 60788359 60795760 60795762 60802022 60802063 60802064 60807922 60821382 60821383 60825101 60825102 60825103 60828899 60828901 60828928 60828929 60838452 60838453 60838454 60838455 60838456 60841602 60841603 60841604 60847622 60847623 60847624 60847625 60847626 60857224 60857225 60857226 60857228 60859197 60859198 60859201 60867264 60867266 60877606 60877607 60877608 60877610 60877611 60877612 60884777 60884778 60899771 60899773 60907137 60907138 60907139 60907140 60907141 60907144 60926724 60926729 60926730 60926731 60926733 60926734 60946101 60946102 60946104 60946105 60946107 60946111 60946112 60946113 60946114 60946115 60946117 60973101 60973104 60977064 60977065 60977066 60977068 60977074 60977075 60996237 60996255 60996256 60996257 60998724 60998725 60998727 60998730 61022867 61022868 61022869 61022870 61022871 61022873 61022874 61022875 61022876 61022878 61022880 61022882 61022883 61036784 61036787 61036788 61036791 61052644 61052647 61052651 61052652 61052653 61052654 61077756 61077757 61077758 61077759 61077760 61077763 61077766 61077768 61077769 61077772 61089229 61089233 61089235 61089236 61089238 61102830 61102833 61102834 61102836 61110894 61110895 61110899 61120266 61130565 61130566 61130572 61135593 61135594 61135597 61135598 61135599 61169846 61169847 61169848 61169850 61169851 61170401 61170402 61170404 61170405 61170407 61170411 61170412 61170416 61184075 61188730 61188732 61188737 61200930 61200931 61200933 61200934 61200939 61200941 61200943 61200947 61200955 61200956 61209965 61209966 61209967 61228887 61228888 61228890 61228891 61234724 61234752 61234754 61234756 61234757 61234758 61234760 61234762 61234763 61234764 61242821 61242822 61242823 61242824 61242826 61242828 61242831 61242850 61256454 61256456 61256458 61256466 61256468 61270060 61279545 61279547 61279548 61279551 61279552 61284347 61295967 61295968 61295970 61295971 61295972 61295973 61295974 61303993 61303994 61303995 61312926 61312928 61312929 61312932 61312933 61312934 61321923 61321924 61321925 61321926 61327861 61333835 61333836 61333837 61337783 61337784 61342399 61342400 61345891 61345892 61345894 61351778 61351795 61351798 61354871 61354898 61361614 61363363 61363364 61363365 61363366 61363367 61370566 61370567 61370568 61370569 61375037 61379466 61379476 61382693 61382694 61382698 61386071 61386245 61391374 61396415 61399399 61399402 61401496 61408071 61408072 61411962 61427169 61427170 61427171 61442485 61442499 61442501 61442502 61453052 61453054 61453056 61453057 61453058 61467507 61467508 61467511 61476322 61486385 61486387 61486389 61504731 61504733 61504736 61504740 61504742 61504743 61504745 61504747 61513164 61513169 61513179 61513180 61513186 61557063 61557065 61557068 61557070 61557083 61557089 61562569 61562570 61562573 61562574 61562578 61562579 61567304 61567309 61567310 61567311 61581720 61581724 61581726 61581728 61581729 61605221 61605237 61608122 61608124 61608126 61608127 61608129 61608132 61634122 61640756 61643293 61651239 61651240 61654615 61654616 61654617 61660343 61660344 61661390 61661391 61661392 61664070 61664072 61664074 61666017 61682243 61692066 61692068 61692074 61692076 61697829 61697835 61697837 61697840 61707314 61711760 61711761 61711762 61722493 61722494 61726162 61726163 61726164 61727764 61727765 61727766 61727768 61728425 61728427 61734193 61734194 61734195 61750731 61750734 61750735 61760272 61760273 61760275 61764659 61764660 61767856 61770219 61770222 61774921 61779305 61779309 61782382 61782383 61786910 61786912 61790409 61802422 61802423 61802425 61812080 61815852 61817909 61817911 61820420 61820422 61820423 61828139 61833423 61833426 61837080 61840627 61840628 61840629 61841666 61841668 61844248 61847219 61851067 61851068 61861644 61861645 61861648 61863222 61863223 61867272 61867273 61867274 61867275 61867276 61870879 61870880 61876377 61876378 61876379 61876598 61880966 61881364 61881533 61885511 61886050 61889518 61891570 61892533 61900097 61905811 61905812 61909905 61909906 61909907 61910409 61911412 61913839 61914593 61920654 61921837 61923960 61927017 61927018 61930503 61937204 61951313 61951314 61951315 61955921 61957538 61962914 61962915 61966660 61976212 61980761 61983227 61985276 62002517 62002518 62004198 62005439 62012015 62012016 62012017 62012020 62015987 62017541 62017542 62018443 62018444 62028906 62030963 62039661 62039664 62043519 62043520 62046441 62052969 62052970 62055937 62057503 62057504 62060728 62060729 62060730 62062919 62062921 62065370 62065374 62088206 62088208 62091830 62091831 62094990 62097168 62097169 62097171 62102497 62105218 62105219 62111341 62113228 62113229 62114959 62121291 62128897 62128898 62140370 62141811 62141813 62143872 62143873 62143875 62148587 62154248 62154330 62154331 62154332 62154584 62165212 62165213 62165214 62177359 62178307 62181496 62183209 62188320 62190363 62190364 62192521 62193448 62194613 62203192 62203999 62210504 62219240 62219243 62220987 62233569 62234531 62236560 62239374 62240336 62240337 62240878 62240880 62240881 62250441 62255236 62258969 62260162 62262545 62264206 62265303 62268235 62268236 62277988 62284382 62288144 62288146 62289678 62291131 62291995 62291996 62304849 62307674 62309103 62311779 62312524 62317233 62343740 369349 369510 60291475 60291476 60291477 60291478 60298330 60314521 60314522 60333452 60333454 60333455 60342035 60342036 60342037 60342038 60345184 60345191 60345193 60346531 60353245 60353246 60353247 60357697 60357698 60361359 60361360 60367205 60367206 60368467 60368468 60374121 60374122 60374123 60374124 60374125 60374126 60383215 60383216 60388307 60392190 60392192 60398717 60399579 60399580 60399581 60399583 60406147 60406148 60406149 60412196 60421551 60421553 60421554 60421555 60425978 60425979 60436360 60436363 60440730 60440731 60440733 60452883 60452885 60452886 60462034 60462035 60466356 60472572 60472574 60472576 60476186 60478018 60499473 60499474 60499475 60499476 60499477 60506629 60515208 60515210 60527078 60527079 60527080 60527081 60531809 60541436 60541437 60549939 60549940 60563133 60563134 60566674 60566676 60573640 60575116 60581735 60581736 60584129 60589106 60589107 60595316 60604272 60611934 60611935 60611937 60620132 60623475 60623476 60630356 60630621 60636976 60642614 60643080 60643081 60643082 60643083 60653309 60656585 60656586 60656587 60660553 60660558 60660559 60660560 60666925 60666927 60669165 60669166 60674064 60678675 60678676 60682225 60686751 60691373 60696918 60698990 60702914 60702915 60708113 60708114 60715789 60715791 60721190 60721192 60721193 60721194 60740791 60740792 60740793 60740794 60740795 60752347 60752719 60752720 60752721 60761089 60767146 60767150 60767151 60769425 60774410 60778179 60778180 60780785 60785222 60788362 60795594 60795764 60795765 60802068 60821388 60821389 60821390 60821392 60828906 60828907 60828908 60833945 60833946 60838457 60838458 60844749 60847627 60857231 60857232 60859202 60859203 60862257 60867267 60867268 60867269 60877613 60877615 60884780 60884782 60884784 60899774 60899775 60899776 60899777 60907191 60907192 60907193 60907194 60907198 60926735 60926736 60926738 60926739 60926740 60926741 60926742 60946121 60946122 60946126 60946128 60946129 60946130 60946132 60996421 60996429 60996431 60996445 60996451 61006380 61006391 61006392 61006397 61006399 61006401 61006402 61006403 61006405 61006407 61006408 61036794 61036795 61036802 61036804 61052656 61052657 61052659 61052660 61063550 61063551 61063552 61063555 61063556 61066143 61070979 61070980 61070981 61070982 61076303 61076304 61076306 61077615 61081458 61081459 61083438 61114997 61114998 61115005 61115025 61115029 61130561 61135603 61135604 61135605 61135608 61135609 61154390 61154399 61154426 61169852 61169853 61170463 61170518 61170520 61184079 61184081 61184092 61184093 61184096 61188742 61188744 61188745 61188748 61209522 61209523 61209524 61209525 61209526 61209968 61209971 61209973 61224313 61224314 61224316 61224317 61224318 61224319 61224321 61224322 61224324 61228892 61228894 61228895 61228897 61242859 61242864 61242865 61242866 61242867 61242869 61242879 61242882 61252186 61252187 61256470 61256471 61256472 61256474 61270065 61270066 61270067 61270070 61279553 61279555 61279557 61279558 61292629 61292631 61292632 61292633 61292635 61295975 61295976 61303996 61312840 61312858 61312860 61312862 61312864 61313923 61321931 61321933 61327862 61327865 61333791 61333792 61333794 61333795 61345896 61345897 61345898 61351825 61351826 61354901 61354902 61354903 61363368 61363369 61363370 61366879 61366880 61375046 61375047 61382699 61382701 61386247 61386250 61386251 61386254 61396418 61396419 61408044 61408045 61408074 61415036 61415037 61415038 61427172 61427173 61442506 61442508 61442509 61442510 61467512 61467513 61467515 61467516 61467518 61467520 61467521 61467522 61476325 61476328 61486390 61486391 61486393 61510397 61510398 61510400 61510404 61510423 61510425 61510426 61510427 61529580 61529581 61538741 61538742 61538744 61538745 61538746 61538747 61562580 61562581 61562582 61562586 61565657 61567291 61567292 61567293 61567294 61567295 61608383 61608385 61608386 61608387 61608392 61637651 61643288 61643289 61643290 61647673 61647676 61647677 61647679 61650790 61654470 61664090 61666699 61673843 61673846 61676986 61684821 61684822 61686642 61686643 61691334 61691337 61691350 61697806 61697808 61697813 61707324 61707325 61710647 61710649 61738778 61738779 61738781 61744646 61744647 61750679 61750683 61758598 61758599 61758607 61762030 61762031 61762032 61767863 61774930 61776747 61783790 61783791 61786914 61786915 61786918 61786920 61790398 61790399 61795024 61795026 61798685 61798686 61803821 61805358 61809174 61809175 61809176 61824415 61824418 61824420 61824421 61830031 61841674 61841676 61852007 61855306 61859403 61861652 61861653 61861654 61863228 61873809 61873810 61876592 61879919 61881525 61883809 61887469 61897112 61898118 61899141 61899252 61902153 61906252 61907868 61914594 61923961 61926332 61926333 61926335 61926336 61926337 61926338 61931868 61934157 61935519 61937621 61941025 61945718 61947762 61949875 61955924 61957642 61962916 61962918 61962919 61969368 61986479 61993275 61993276 62006667 62018447 62018448 62020927 62030965 62030967 62035438 62036925 62041551 62041552 62042623 62046453 62046454 62055935 62059771 62073537 62080056 62080057 62082184 62082185 62082187 62088211 62088213 62088214 62088215 62091825 62097158 62097159 62109130 62109131 62109134 62114979 62114980 62114981 62119029 62127584 62127585 62136714 62136715 62141818 62141819 62142881 62143877 62146126 62146127 62151894 62151895 62160617 62160618 62161499 62161501 62165209 62178662 62178663 62180239 62183210 62187697 62197317 62198604 62198605 62204001 62212178 62213279 62215040 62215690 62226902 62234532 62236561 62238096 62240338 62240883 62240913 62249056 62250443 62251882 62260165 62277993 62278728 62284383 62298332 62299158 62299168 62304854 62305921 62307675 62317813 62320251 62330829 60752791 60990627 61213660 61504813 61658241 61787314 61803722 60857263 60977358 61144301 61396636 61435736 61561712 61561713 61654437 61658243 61732443 61943354 62301497 60515272 60535635 60539991 60563159 60564441 60566884 60573247 60573248 60576015 60581752 60581753 60588847 60588848 60588849 60588850 60588851 60594568 60602844 60602845 60602847 60604297 60604298 60612147 60612149 60624816 60624817 60643135 60648366 60653889 60653890 60656859 60657828 60661217 60664320 60664321 60668019 60670852 60670853 60676326 60680904 60682820 60687506 60699141 60699142 60699143 60713794 60713795 60713796 60713797 60733018 60733020 60738736 60738737 60746811 60746812 60752733 60764225 60769434 60774489 60780814 60786108 60791369 60791423 60800940 60800941 60807790 60807796 60807797 60817925 60822437 60828731 60828752 60834399 60834405 60841712 60844049 60844050 60854346 60854347 60857264 60857265 60863762 60863763 60863782 60896674 60896675 60902523 60902524 60919133 60919135 60919136 60930828 60930829 60934072 60939964 60946223 60946224 60973588 60973589 60977359 60985316 60990618 60997533 60997536 60998596 61006098 61006099 61018637 61018640 61018642 61018647 61018648 61018649 61019175 61036969 61055662 61055663 61055664 61066170 61066171 61066172 61066173 61091419 61091420 61091421 61091423 61091426 61091427 61119668 61119669 61135888 61135894 61156638 61156639 61156640 61156641 61156642 61162746 61162747 61162748 61162749 61162750 61162751 61162752 61162753 61183125 61183127 61183129 61183131 61183134 61183135 61183141 61183236 61183237 61191670 61191671 61191676 61191678 61191679 61213662 61213663 61213664 61213665 61243237 61243238 61243239 61243240 61243241 61261227 61261228 61261229 61261230 61261231 61261232 61261233 61261234 61296111 61296112 61296113 61296114 61296115 61296116 61296117 61313764 61313765 61313774 61313775 61325778 61326550 61337950 61337951 61337952 61337953 61337954 61356099 61356102 61356104 61356105 61356112 61356117 61356120 61356121 61396431 61396432 61396433 61396434 61396435 61396436 61396437 61396438 61396440 61400490 61400491 61400493 61400497 61400502 61400509 61408319 61408320 61408321 61427257 61427258 61427259 61427261 61445061 61445062 61445064 61445065 61445067 61445068 61445069 61458506 61474668 61474670 61478450 61478451 61478452 61478453 61483031 61486435 61486436 61489814 61489815 61489816 61496773 61496774 61496775 61496776 61496777 61496779 61504816 61506138 61506139 61514240 61514241 61520365 61520366 61520367 61526580 61526581 61526583 61538775 61538776 61561714 61561715 61561716 61561718 61593413 61593414 61593416 61593417 61610847 61610848 61610849 61637770 61637771 61637772 61648365 61648366 61648367 61656001 61658244 61658245 61658246 61666578 61666579 61666580 61666581 61670952 61670954 61689332 61689333 61689334 61689335 61695678 61695679 61695681 61713022 61713023 61713025 61724967 61736437 61736438 61736439 61742456 61742457 61742459 61742460 61742461 61747726 61751714 61751715 61751716 61751717 61760279 61760280 61760281 61767897 61767898 61772457 61772459 61776714 61776715 61782458 61782459 61782460 61782461 61787294 61799667 61799668 61799669 61803724 61809249 61809250 61810042 61810043 61810045 61819714 61819715 61823615 61823616 61828304 61835079 61835080 61839836 61839837 61839838 61841936 61841937 61852601 61852602 61852603 61852604 61855312 61858152 61858153 61867409 61867410 61867411 61871969 61871971 61875692 61875693 61875694 61875696 61875697 61881278 61882649 61891273 61895566 61895568 61895569 61902114 61902116 61908947 61912458 61912459 61923147 61937078 61937079 61939328 61939330 61939332 61945132 61945133 61945134 61948398 61952512 61952514 61957661 61957663 61980228 61980229 61980230 61980231 61982568 61992552 61992553 61992554 61993195 61993196 62002347 62010725 62017519 62017520 62024693 62024694 62024695 62027866 62027867 62027868 62036973 62042809 62042810 62042811 62062931 62062932 62064032 62064033 62064034 62065563 62065564 62065565 62065566 62065568 62069035 62069037 62069038 62071945 62071946 62071948 62103976 62118735 62121301 62122150 62122151 62122152 62123074 62127598 62127599 62127601 62134050 62134051 62135125 62136786 62136787 62143881 62143882 62143883 62147371 62147373 62148643 62148646 62149791 62156254 62170694 62170695 62170696 62170697 62170698 62172060 62180244 62181551 62182879 62193555 62193557 62258980 62260187 62266490 62266491 62268519 62285320 62285321 62292040 62296746 62299204 62306048 62306050 62306061 62306062 62306063 62308581 62308582 62311682 62312631 62312632 62315241 62317829 62317831 62317832 62320036 62320037 62324226 62324228 62325626 62334133 62334134 62334135 62334136 62334137 62334138 60844050R 61602901 61784007 366570 366575 366667 369511 60277270 60277271 60277302 60297070 60298301 60298302 60310368 60314528 60314532 60325188 60325189 60325191 60333456 60333457 60346536 60346537 60353249 60357699 60357701 60361361 60363823 60367207 60368469 60368471 60374127 60374128 60374129 60374130 60374132 60388296 60388316 60392194 60392196 60399584 60399586 60399587 60403297 60406150 60406151 60412198 60421556 60421557 60425981 60425982 60425983 60425989 60436366 60439930 60439931 60440734 60452887 60452889 60452890 60452891 60452892 60466357 60466358 60466359 60472578 60472579 60476190 60478019 60478020 60499479 60499481 60506630 60515211 60515212 60525150 60527082 60527083 60527084 60531810 60541439 60541441 60549942 60549943 60557550 60558774 60564433 60573642 60575118 60579234 60584131 60589110 60593485 60595318 60611938 60611939 60611940 60611943 60623477 60623478 60636978 60636979 60643084 60643086 60643087 60643088 60648072 60656588 60656589 60661154 60669167 60674066 60676281 60676282 60676283 60678689 60682197 60682204 60685534 60689314 60691391 60696922 60696923 60696924 60702916 60709819 60715795 60715797 60721196 60721197 60721198 60721199 60740796 60740797 60740799 60740803 60752348 60752349 60752723 60761094 60761095 60761096 60761097 60763098 60764211 60764212 60774411 60778182 60780788 60785225 60788364 60795768 60795769 60807925 60807929 60807931 60807935 60807937 60821396 60825107 60828909 60828910 60828911 60833948 60838459 60841570 60841608 60844751 60847629 60847631 60854020 60857235 60859205 60859206 60867270 60867273 60877617 60877618 60884787 60884789 60884791 60899781 60899782 60899783 60907204 60907205 60907208 60926743 60926744 60926746 60926747 60926748 60926751 60946133 60946134 60946136 60946137 60946138 60946139 60946141 60970698 60970701 60970705 60977147 60977148 60996476 60996478 60996481 60996482 60996512 61006458 61006459 61006464 61006465 61022886 61022888 61022890 61022902 61022915 61022916 61022917 61022918 61036805 61036806 61036808 61036809 61052671 61052672 61052675 61063492 61063493 61063494 61063495 61063496 61063497 61063498 61070983 61070984 61070985 61070987 61076292 61076293 61076294 61083443 61083444 61083447 61115036 61115037 61115038 61115041 61115042 61115046 61130563 61134787 61134789 61134790 61135612 61135613 61161609 61161613 61161615 61161616 61161617 61161618 61175630 61175631 61175632 61184180 61184184 61184186 61184187 61184192 61188751 61188753 61209513 61209514 61209515 61209516 61209520 61209521 61209975 61209977 61209979 61209980 61209981 61228898 61228899 61228901 61243176 61243177 61243179 61243181 61243186 61243188 61243189 61243190 61256478 61256481 61256482 61256483 61270073 61270075 61270076 61270080 61279560 61279561 61279562 61279564 61279566 61284349 61292614 61292615 61292616 61292619 61304002 61304007 61304009 61312892 61312896 61312897 61321935 61321937 61327927 61327928 61333798 61333799 61333801 61342411 61345899 61345900 61351837 61351851 61354905 61354906 61361617 61363372 61363374 61363375 61370594 61370595 61379482 61379484 61382703 61382704 61382706 61382707 61394216 61394217 61396420 61396421 61401502 61401503 61408077 61411937 61424489 61424490 61442511 61442514 61442516 61453074 61453075 61453076 61467526 61467533 61476334 61476335 61486397 61513204 61513205 61513206 61513209 61513211 61522274 61522275 61522276 61537530 61542294 61555095 61555099 61555101 61555103 61555104 61555105 61555106 61555107 61558504 61562588 61562589 61562590 61562594 61567313 61567314 61567316 61605359 61605360 61605361 61608487 61608488 61608489 61608491 61608493 61637657 61647692 61647695 61647697 61647701 61647705 61650964 61650965 61654510 61660345 61661393 61661404 61664076 61664077 61664081 61666037 61673838 61676958 61676959 61682192 61684806 61687515 61687517 61687519 61722506 61726165 61726166 61727771 61728430 61732331 61734189 61734190 61750714 61750716 61758628 61762029 61767853 61767854 61770218 61776744 61779310 61779311 61779313 61786924 61795035 61803823 61809182 61809183 61815847 61830034 61830036 61831367 61851992 61851993 61851994 61863229 61863231 61863233 61863234 61867279 61876380 61876381 61879920 61881526 61885513 61891571 61892535 61893824 61898119 61898396 61902159 61902262 61903681 61915755 61920655 61934161 61941026 61943522 61945720 61949876 61952698 61957627 61957628 61957630 61957631 61964018 61964020 61965999 61980768 61982540 61983230 61987253 61987254 61992604 61992606 62005440 62005441 62005443 62006672 62006674 62008860 62008861 62010581 62017532 62020931 62026457 62026459 62028903 62028904 62030971 62030972 62042625 62042626 62042627 62046455 62050999 62052973 62055934 62059773 62060725 62065359 62065360 62065361 62065391 62065392 62065393 62067402 62094995 62094996 62097179 62097181 62106832 62106833 62109123 62109124 62109125 62109126 62109127 62111335 62112184 62126371 62126372 62137184 62138623 62140365 62149783 62154325 62154326 62154327 62154328 62154587 62158173 62159866 62159867 62160620 62171364 62171365 62173352 62176453 62181498 62181501 62186364 62192529 62195524 62200355 62207247 62208791 62213271 62215035 62215036 62215687 62226904 62234533 62236562 62238097 62240885 62240886 62240914 62249057 62256075 62260167 62266454 62268448 62277994 62284385 62285291 62291134 62294372 62294373 62294381 62299160 62299169 62300232 62314576 62320249 62332996 62338834 62338835 60276633 60276634 60276635 60276637 60276638 60276640 60276641 60325193 60333458 60333459 60333460 60346538 60346539 60346540 60357702 60357703 60363825 60367208 60368472 60374133 60374134 60374135 60374136 60383222 60388317 60392199 60392202 60403298 60403299 60403301 60403302 60406152 60412199 60421558 60421559 60425990 60434172 60434173 60434174 60436367 60440740 60440746 60452894 60452895 60456489 60476195 60492483 60499935 60499936 60515214 60527085 60527086 60531812 60541442 60541443 60557551 60558775 60563136 60566678 60575119 60579235 60581738 60590820 60593486 60593487 60611962 60611969 60630622 60636980 60638568 60638569 60653313 60660579 60660583 60669168 60669169 60678677 60682208 60682209 60682212 60685535 60686753 60689413 60691398 60691404 60708115 60708117 60715799 60715801 60715802 60728458 60728459 60761099 60761103 60764213 60769428 60778183 60778184 60780789 60780790 60788368 60795598 60795771 60795772 60802071 60802072 60812438 60812439 60812440 60812442 60812443 60841572 60841573 60841610 60841611 60847632 60857238 60857239 60859208 60867274 60867275 60867277 60877620 60884793 60884794 60899784 60907209 60907210 60907211 60907213 60907214 60907215 60926752 60926755 60926756 60941602 60941605 60941606 60941609 60941614 60946146 60984974 60984975 60984978 60992272 60992273 60992279 61006416 61006429 61006431 61006437 61006444 61006445 61022920 61022922 61022923 61022925 61022927 61036811 61036813 61052677 61052682 61063544 61063545 61063546 61063548 61063549 61070990 61070991 61070992 61070993 61070995 61076301 61076302 61081461 61081463 61081464 61110907 61110909 61120270 61130535 61130537 61130539 61130541 61134791 61134792 61135616 61135618 61135619 61161622 61161626 61161627 61161629 61175635 61175636 61184196 61184198 61184204 61184206 61184209 61188756 61188758 61188762 61200961 61200964 61209529 61209530 61209531 61209532 61209533 61209534 61228904 61228905 61228906 61228907 61243191 61243192 61243194 61243196 61243197 61252188 61252189 61252191 61260743 61260744 61260745 61270082 61270084 61279568 61279571 61295977 61295978 61295980 61295981 61310103 61313926 61321939 61321940 61333803 61333804 61333805 61337791 61337792 61345901 61351864 61351867 61354873 61361619 61361620 61363376 61366882 61379487 61382709 61382710 61386259 61386271 61394219 61396422 61399417 61408046 61408079 61411938 61411968 61411969 61421209 61424485 61442518 61442519 61442520 61442521 61467554 61467560 61467563 61467567 61467569 61476374 61476375 61486411 61513117 61513119 61513129 61513130 61513143 61533231 61533487 61558471 61558472 61558473 61558474 61558475 61558476 61558495 61558498 61558501 61565659 61565660 61565662 61565663 61570269 61570270 61581736 61581744 61581745 61608518 61608519 61608603 61608604 61637653 61637654 61643298 61643299 61647684 61647688 61654612 61671981 61673874 61676979 61682217 61684812 61691325 61691326 61695222 61697105 61710652 61738763 61744636 61744637 61750673 61750675 61758624 61762022 61762025 61770232 61774927 61783796 61786894 61786895 61786896 61786897 61795027 61798111 61803827 61805365 61805366 61809192 61809194 61815849 61817913 61824423 61824424 61833441 61841669 61841670 61841671 61855302 61857264 61868623 61868624 61870213 61870214 61873813 61879921 61882815 61884680 61888825 61891572 61893825 61898120 61900843 61904425 61904426 61907383 61907870 61909379 61909560 61910410 61915413 61918161 61920656 61925115 61928598 61934163 61937206 61940078 61943523 61945721 61947764 61949877 61952699 61962920 61962921 61966661 61969213 61982542 62002519 62004202 62010586 62010587 62012002 62012003 62024666 62024667 62024669 62043521 62043522 62043523 62043524 62046457 62046458 62059775 62067602 62067603 62080048 62080049 62088216 62088217 62088218 62091829 62097149 62097150 62097151 62097152 62121292 62121293 62121294 62127582 62127583 62137185 62141806 62143871 62154589 62156236 62158174 62160621 62160622 62165222 62165223 62176454 62178660 62178661 62181493 62187701 62192527 62192528 62204007 62210507 62219245 62220988 62222815 62226905 62236564 62240340 62240887 62240888 62240916 62249059 62250445 62262695 62267526 62281857 62282949 62299161 62320253 62334131 369226 370184 60297071 60298304 60314537 60314538 60325199 60333462 60346543 60346544 60346546 60353250 60353251 60357704 60367209 60368475 60368476 60368477 60378622 60378624 60378625 60378626 60378627 60388318 60392206 60392208 60392211 60403304 60426003 60426005 60426009 60439932 60460149 60460151 60460153 60460154 60460155 60466362 60466363 60466364 60472586 60478023 60499483 60499492 60499493 60499494 60499495 60499496 60515215 60515216 60515217 60515218 60517102 60517103 60517104 60531815 60535595 60541444 60556518 60558776 60563137 60584163 60584166 60584171 60593488 60595319 60611983 60611984 60611985 60623481 60630358 60630624 60636981 60636982 60646301 60646302 60656590 60661156 60669171 60669172 60674068 60674069 60676285 60686754 60691406 60696925 60696927 60696929 60696931 60702918 60708120 60709822 60709823 60715803 60721201 60721202 60721203 60740804 60740809 60740810 60740811 60740812 60752724 60763099 60764214 60778185 60780769 60780770 60780791 60785227 60785228 60788369 60795774 60795776 60802073 60802074 60807952 60807953 60821397 60821398 60821399 60825110 60828932 60838461 60841575 60841576 60841612 60844753 60847633 60854022 60857241 60857242 60857243 60877621 60877622 60877623 60877624 60877626 60884795 60884797 60884798 60884803 60884804 60899786 60899787 60907219 60907221 60907222 60907223 60907224 60907225 60907226 60907227 60926759 60926761 60941622 60941623 60941625 60946151 60973047 60973048 60973049 60973051 60973052 60973053 60977151 60977152 60992258 60992260 60992262 60992263 60992265 60992267 60996272 60996273 60996274 61020007 61020015 61020019 61036818 61036821 61036824 61052684 61052687 61052689 61052695 61052697 61052698 61052699 61063583 61063584 61063585 61070996 61070997 61070999 61076290 61083449 61083450 61083451 61083453 61102841 61102843 61102849 61110879 61110880 61110882 61110883 61134795 61134797 61134798 61154451 61154454 61154460 61170522 61170523 61170524 61175638 61175639 61175641 61184972 61184973 61184980 61188766 61188767 61188768 61209493 61209495 61209496 61209982 61209984 61209985 61209986 61209987 61209989 61228969 61228972 61228985 61228988 61228989 61229001 61229002 61229003 61229007 61229010 61234765 61234768 61243199 61243201 61243202 61260680 61270085 61270087 61270091 61270093 61270094 61270095 61279573 61279574 61279576 61284352 61292586 61292587 61295982 61304012 61305722 61305725 61305729 61305732 61312831 61312836 61321942 61321944 61327923 61327924 61333811 61333812 61333814 61342418 61351871 61354878 61361622 61361624 61363377 61366883 61370610 61379490 61379491 61386009 61386275 61386277 61394221 61396424 61399420 61399421 61401478 61411941 61411971 61415074 61424488 61427177 61442522 61442523 61463534 61463540 61463541 61476377 61476378 61476380 61476381 61486413 61486414 61504722 61504724 61513090 61513091 61513092 61530398 61530399 61537532 61537546 61538751 61538752 61553170 61555071 61555072 61562596 61567317 61567318 61567319 61581750 61581755 61581758 61581766 61605370 61631057 61631058 61637638 61637639 61637640 61647682 61673943 61673950 61677019 61682214 61682215 61684809 61684810 61686639 61686640 61718323 61718324 61722510 61722511 61722513 61732339 61732340 61744668 61744669 61750713 61758617 61783780 61783781 61786885 61786886 61786887 61798107 61798108 61803812 61803813 61803814 61803815 61809179 61812086 61820425 61820426 61831376 61833442 61839059 61847228 61847229 61851997 61851998 61855297 61855298 61868620 61870866 61870868 61873814 61882816 61889171 61890366 61907384 61907871 61908928 61915414 61920657 61923962 61926340 61935521 61937623 61941027 61943524 61948430 61954130 61958879 61982543 61986485 61986486 61986487 61992601 61992602 61998911 61999833 62006675 62024659 62026461 62026462 62026463 62026464 62030099 62035428 62035429 62046460 62052958 62052959 62052961 62060721 62065355 62065357 62076956 62079212 62079213 62080043 62091832 62097164 62097165 62097166 62097167 62102490 62102491 62102492 62103586 62114953 62123066 62137186 62142885 62142886 62142887 62142888 62152931 62154577 62158175 62159868 62171367 62176455 62178308 62183211 62186363 62188321 62188322 62192531 62197319 62204008 62210510 62224951 62224952 62226906 62226907 62234535 62236565 62240342 62240889 62258961 62266455 62289682 62292002 62292004 62299162 62305927 62316129 62320114 62330843 62334109 60680905 60800934 61144377 61144380 61261235 61561721 61593407 61734115 61803731 62170701 60515282 60564444 60566885 60573249 60575161 60588870 60588871 60588872 60588873 60598753 60600219 60611424 60612154 60612155 60624809 60636421 60636423 60648368 60653891 60661239 60664323 60664327 60664328 60670855 60670856 60676327 60682821 60687512 60699144 60699145 60713802 60713803 60722511 60727139 60733022 60752785 60769435 60786111 60791424 60800942 60815723 60817926 60822438 60828754 60834401 60841713 60844052 60854336 60854348 60857266 60863783 60885044 60896676 60896677 60902525 60919137 60919138 60930831 60934073 60946225 60950562 60967591 60973590 60977361 60997528 60997529 61006108 61018655 61018656 61019173 61036972 61055657 61055658 61055659 61066177 61066178 61066179 61080036 61091432 61091434 61091436 61091439 61091440 61120685 61120686 61120687 61120688 61120689 61135902 61135905 61135907 61162758 61162759 61162760 61162761 61191697 61191698 61191702 61191703 61191711 61191712 61213551 61213552 61213553 61213554 61243244 61243245 61243246 61243247 61243248 61261236 61261237 61261238 61261239 61261240 61296134 61296139 61296141 61296144 61313776 61325782 61325783 61326553 61337944 61337945 61337947 61356046 61356049 61356050 61356056 61377145 61377146 61396623 61396624 61396625 61396626 61408314 61408315 61408316 61408317 61408318 61427235 61427238 61427239 61427240 61427242 61461398 61461399 61461401 61461402 61461403 61461404 61461405 61469290 61469291 61469292 61482963 61483008 61486437 61489821 61500176 61500177 61500179 61514233 61514234 61514235 61514236 61520353 61526577 61526579 61538777 61538778 61561722 61561724 61561725 61593427 61593428 61593429 61610851 61610852 61610853 61610854 61648368 61648369 61648370 61658248 61658249 61666584 61697464 61697465 61697466 61697467 61713030 61718644 61718646 61718647 61718648 61723050 61723051 61723052 61723053 61745783 61747727 61747728 61747729 61760471 61760474 61764744 61767896 61772439 61776718 61776719 61778316 61782457 61796059 61803717 61803718 61809253 61809254 61810050 61810051 61815229 61823619 61833496 61833497 61839820 61839821 61855316 61855318 61858156 61864139 61867394 61867395 61867396 61871943 61871944 61873855 61882650 61889290 61889291 61889292 61889295 61896827 61902120 61902124 61902126 61920538 61928955 61943358 61943359 61943360 61948401 61952508 61957693 61957695 61968318 61968319 61968320 61968321 61980206 61987234 61987235 61993202 62010720 62010721 62017515 62017516 62027881 62036976 62036977 62040498 62040500 62042801 62042802 62064040 62064041 62073544 62073545 62077393 62077394 62077395 62085388 62085389 62085390 62085391 62091033 62091034 62118740 62118741 62122153 62122154 62122155 62123076 62123079 62127602 62137208 62141840 62141841 62141842 62143879 62143880 62154597 62156256 62166296 62166297 62166298 62175392 62175397 62183237 62188404 62268521 62293240 62304884 62310245 62310248 62315874 62315875 62317215 62325639 62325642 62327791 62333071 62333072 62333073 61474661 62103978 369352 60277282 60277284 60277285 60277286 60297072 60314541 60314542 60333464 60333465 60333466 60346547 60346548 60357705 60363826 60368479 60368480 60374137 60374138 60374139 60392213 60392214 60403305 60406155 60406156 60415993 60421561 60426017 60426020 60460158 60460159 60460160 60466365 60466366 60472587 60472589 60478025 60492484 60506631 60506632 60517105 60531817 60549945 60549946 60584177 60584182 60590826 60595320 60611987 60611993 60630359 60630630 60636983 60639458 60656591 60656592 60662849 60662850 60676287 60676288 60678678 60678679 60678680 60683488 60686755 60689414 60691411 60698991 60708121 60715806 60721205 60721206 60728466 60728467 60728469 60728470 60752725 60764216 60764217 60769429 60780792 60785230 60788370 60788371 60802076 60821401 60821402 60825112 60825113 60828913 60833952 60837637 60841615 60844755 60857244 60857245 60859209 60859210 60867278 60877627 60877628 60884805 60884806 60899789 60907229 60907230 60907231 60907232 60907233 60926765 60926768 60926769 60941633 60946152 60946153 60946154 60977154 60977155 60977156 60998731 60998733 60998734 60998736 60998737 60998738 61022933 61022934 61022935 61022936 61022937 61036828 61036829 61052702 61052703 61052704 61052705 61066127 61066129 61066131 61066133 61066134 61066136 61076295 61076296 61081465 61081468 61083454 61083455 61083456 61083457 61134803 61134805 61134806 61134807 61134808 61135620 61135622 61135624 61161641 61175646 61175647 61175648 61175649 61175650 61175651 61188769 61209538 61209539 61209540 61209541 61209543 61209544 61209990 61209991 61209992 61209993 61209994 61209995 61209996 61243204 61243205 61270096 61270098 61270099 61279577 61279578 61279579 61292612 61295984 61295985 61295986 61321945 61321946 61321948 61333848 61337794 61345904 61351874 61354879 61363378 61366887 61370613 61375057 61379492 61382713 61386280 61394222 61396425 61401479 61401480 61411942 61411943 61427179 61427181 61441739 61441740 61453081 61453082 61453083 61463546 61463547 61476383 61476384 61476385 61486415 61513094 61513097 61513099 61513101 61530411 61533489 61553176 61557091 61557093 61557094 61562597 61565665 61565666 61565668 61565669 61565670 61581776 61581778 61581780 61581784 61581786 61608080 61608081 61637633 61637634 61643263 61643267 61647666 61650789 61654331 61664066 61671960 61671969 61691372 61691373 61691376 61691377 61692028 61695226 61697131 61697133 61707307 61711754 61732334 61732335 61732336 61744640 61758625 61767865 61767866 61770234 61779298 61783799 61786889 61795028 61798113 61805387 61805388 61805390 61817914 61817915 61817916 61817917 61831374 61841684 61841685 61855307 61855308 61855309 61870215 61870216 61870870 61880802 61884682 61888826 61890367 61892538 61894774 61908929 61915756 61919603 61921839 61925116 61926331 61935522 61938453 61940080 61945724 61954131 61966662 61976794 61976796 61976797 61985278 61996833 61998912 62017526 62020933 62026455 62026456 62030101 62036919 62036920 62063909 62063910 62067397 62075412 62075413 62076946 62076947 62080041 62102498 62109128 62114961 62121280 62123062 62123063 62142882 62143876 62146128 62146129 62156239 62160233 62167409 62167410 62171370 62188323 62194615 62200358 62210513 62213278 62216724 62224954 62229920 62236567 62239376 62240343 62260169 62267529 62268471 62292005 62299163 62309142 62314591 62320115 62320258 369353 60403306 60403307 60406157 60415996 60421562 60426023 60460161 60460162 60466367 60472607 60499939 60499941 60506633 60506634 60517106 60517107 60531818 60535596 60547650 60549947 60566683 60584197 60584202 60584204 60590830 60590831 60595321 60623485 60630631 60636984 60636985 60656594 60661159 60669173 60670135 60674070 60676289 60678681 60686766 60689415 60691412 60698994 60698995 60702919 60702920 60715809 60721207 60728479 60728480 60728485 60728487 60728492 60728494 60761125 60761127 60769430 60769431 60788372 60788373 60802077 60802078 60807954 60812447 60812448 60812449 60812450 60833953 60841617 60841618 60844756 60854023 60857247 60857248 60859212 60867279 60867280 60877630 60884807 60884808 60899790 60907234 60907236 60907237 60941634 60941635 60941636 60941637 60941638 60970706 60970707 60970708 60970709 60970710 60996514 60996515 60996517 60996518 60996520 60996521 60996523 61020037 61020045 61022939 61022940 61052706 61052707 61052708 61052711 61066108 61066110 61066111 61066112 61066113 61071001 61076280 61076281 61077617 61083460 61083461 61083462 61083463 61102856 61102858 61110868 61110870 61115065 61115066 61115067 61154486 61154488 61154490 61154492 61154494 61154496 61175652 61175653 61175654 61175655 61175656 61185033 61185044 61185048 61185054 61188771 61188773 61200970 61200975 61209507 61209508 61209509 61209510 61229068 61229069 61229074 61229075 61234771 61234772 61234773 61234774 61234775 61252219 61252221 61256488 61279580 61279581 61279582 61279583 61284353 61295988 61312945 61321949 61327925 61333816 61333817 61337798 61337799 61345905 61351877 61354880 61354911 61363380 61363381 61366888 61375062 61382714 61382715 61386281 61386282 61396426 61396427 61399428 61399429 61408080 61408081 61421211 61427183 61437036 61437037 61437038 61437039 61467585 61467586 61467589 61467591 61467593 61486416 61513108 61513109 61513110 61513111 61522281 61522282 61522283 61533232 61533490 61533491 61562599 61562601 61562602 61562603 61562604 61562605 61618261 61618263 61618264 61618267 61618268 61618270 61618271 61618272 61634113 61650541 61650544 61650545 61650547 61653566 61687511 61687512 61692039 61697135 61697136 61707304 61711752 61722514 61732328 61732329 61742539 61750666 61750667 61760277 61776742 61779293 61786898 61789197 61795022 61795023 61798110 61803798 61803801 61812088 61828128 61831354 61841673 61847222 61847223 61851995 61855295 61857256 61867262 61867263 61867264 61879515 61892539 61893827 61897116 61908930 61910412 61915757 61919604 61925117 61928599 61934164 61937207 61940082 61945730 61951319 61966000 61968076 61976801 61987255 61987256 61987257 61987258 61996832 62012004 62012006 62015976 62020934 62020935 62024661 62024662 62024663 62043525 62043526 62046461 62057501 62060727 62067606 62073535 62082203 62082204 62093431 62093432 62094991 62097154 62097156 62109129 62128902 62128903 62141807 62154320 62154321 62154322 62167398 62171375 62182860 62182861 62182862 62183212 62188325 62188326 62194616 62199455 62204595 62210517 62220990 62234537 62236568 62240894 62240895 62258964 62267531 62268472 62299164 62307731 62320261 62333445 62333447 369355 370947 60277291 60277293 60314554 60314555 60333470 60333472 60346552 60346554 60361388 60368482 60374144 60388320 60392220 60403308 60403309 60412203 60421565 60426058 60439933 60460163 60462036 60472610 60499943 60499944 60506635 60527088 60531819 60541446 60556519 60566684 60575120 60581740 60589111 60612006 60620145 60630726 60636986 60643091 60643092 60646304 60661160 60661161 60662851 60676291 60678682 60678683 60686767 60689416 60691413 60702921 60702922 60708122 60715816 60728503 60728504 60728505 60728506 60788374 60802035 60807955 60807956 60821403 60825114 60828914 60828915 60833955 60841578 60841620 60844757 60854024 60857249 60857250 60862267 60867281 60867282 60877632 60884809 60907241 60926772 60926774 60926775 60926776 60941639 60941640 60946157 60946159 60984988 60984989 60984990 60998739 60998740 60998742 61020062 61020066 61020069 61020071 61020072 61036834 61036836 61052713 61052714 61052718 61071003 61071004 61076282 61076283 61077619 61077620 61115048 61115050 61115052 61115053 61135626 61161596 61161599 61161602 61161603 61161605 61185062 61185063 61185065 61185066 61185067 61200981 61200988 61201000 61201003 61229076 61229077 61229078 61234776 61234777 61234778 61234779 61234780 61243206 61252181 61252182 61252183 61270101 61270102 61279585 61279586 61295989 61295990 61295991 61333806 61337801 61337802 61337803 61351878 61354882 61363382 61370617 61370618 61382716 61382717 61394223 61394224 61399434 61411945 61424483 61424484 61442532 61442564 61463542 61463543 61463544 61467608 61467609 61513214 61513215 61513216 61522279 61522280 61530402 61530403 61537686 61537687 61537689 61605254 61605256 61605261 61605263 61605264 61605265 61637632 61650966 61666030 61671958 61684814 61684815 61687529 61687530 61687532 61687535 61744651 61744652 61744653 61750685 61774936 61783802 61786883 61798109 61803802 61803803 61815850 61817920 61817921 61828125 61828126 61828127 61831356 61855310 61857269 61868616 61868617 61884686 61887472 61890368 61891998 61892540 61904429 61910413 61913842 61915415 61920658 61928601 61938454 61976802 61976803 61989380 61993277 61998914 62017527 62017528 62036917 62041542 62042617 62050993 62050994 62060723 62075416 62091826 62091827 62091828 62097138 62097139 62097140 62105198 62105199 62105201 62144973 62146130 62149784 62149785 62159865 62166267 62178664 62178665 62181487 62181488 62207251 62212144 62219246 62226912 62226914 62229921 62234538 62240896 62240897 62267533 62289684 62292007 62299165 62316136 62320262 62333448 60857267 60563161 60800930 60885047 61132154 61144291 61261241 61337932 61522321 61579995 61751722 61920528 61964896 62162191 60535638 60563162 60566886 60575162 60581755 60588874 60598751 60611426 60612163 60623595 60648369 60653892 60661241 60664329 60687513 60699146 60699147 60713804 60722877 60752790 60780823 60800935 60815737 60821461 60828733 60834402 60844053 60854337 60857268 60863784 60896679 60901594 60901595 60930832 60959521 60967592 60985327 60997527 60998599 61018659 61036974 61066180 61066181 61066182 61091443 61091444 61094796 61119664 61119665 61144347 61144349 61144351 61162767 61162768 61191713 61191714 61213667 61213669 61213671 61243249 61261242 61261243 61261244 61261245 61296151 61296154 61313777 61337924 61337925 61337926 61337927 61337928 61337929 61337930 61408344 61408345 61408347 61408348 61408349 61445073 61445074 61445075 61445076 61445077 61474662 61474663 61474664 61478443 61478444 61489823 61504820 61504821 61506144 61506145 61520357 61520358 61520359 61561726 61561727 61605539 61605540 61605541 61610855 61654440 61666582 61684185 61697437 61713034 61713035 61724957 61728058 61732430 61732432 61751738 61751739 61751740 61767885 61767886 61782466 61787321 61796060 61799649 61799651 61803726 61815247 61823623 61828307 61841940 61858158 61861715 61871953 61871954 61889293 61889294 61895564 61908948 61908949 61925654 61931753 61931754 61935586 61935587 61964903 61980212 61987239 61987240 61993203 62030974 62030975 62030976 62050784 62062935 62062936 62064035 62064036 62100716 62100717 62109975 62109976 62115173 62115174 62159895 62161574 62162172 62162173 62172065 62172066 62175381 62175382 62188406 62268522 62300805 62300806 62311680 62318371 62326204 62327792 61036975 366610 366718 369358 370948 60298331 60314556 60314557 60333473 60346555 60353253 60368485 60374145 60374146 60383324 60392225 60392227 60406158 60426059 60426740 60439935 60452898 60472614 60476198 60499945 60525151 60527089 60531820 60541447 60558777 60564434 60575121 60589112 60589113 60595322 60604275 60643093 60674071 60676293 60682213 60682214 60686771 60689417 60708124 60709629 60721208 60721209 60740813 60740814 60761132 60763101 60767147 60780771 60785264 60795786 60795787 60807957 60807959 60812446 60844758 60844759 60857799 60857800 60857803 60862268 60880712 60880714 60880715 60880716 60880718 60899792 60899793 60907242 60907243 60907244 60907245 60926777 60926778 60941641 60941642 60958775 60958776 60961258 60961259 60973039 60984992 60984993 60984994 60984995 60996258 60996259 61022943 61022944 61022945 61022946 61036837 61036838 61036839 61052719 61052720 61066138 61077621 61077622 61077623 61102862 61102865 61102868 61102870 61114970 61114972 61114973 61134810 61134812 61161632 61161633 61175657 61175658 61185068 61185069 61209536 61209997 61224308 61224309 61224310 61224311 61224312 61252203 61252205 61252213 61252215 61252217 61260667 61260668 61270105 61270106 61270107 61270109 61284355 61284356 61284357 61292607 61292608 61310097 61310098 61313929 61321951 61327921 61327922 61337805 61351880 61363383 61370619 61370620 61382719 61399444 61399446 61411946 61424467 61424480 61437040 61437041 61463548 61467610 61467611 61513106 61513107 61530404 61533492 61533493 61542297 61542298 61558510 61562606 61562607 61567323 61629376 61629377 61640737 61647672 61660332 61664065 61673841 61684803 61684804 61687513 61691329 61697124 61707308 61732332 61732333 61744633 61744635 61750668 61750670 61770216 61774922 61774923 61786878 61798106 61803796 61805352 61828123 61831351 61833421 61841651 61841652 61847236 61847237 61864121 61870871 61870873 61873815 61888827 61892541 61895263 61895661 61906255 61914612 61916897 61920659 61928930 61937208 61945734 61955927 61964118 61977982 61991431 61996834 61997567 61999834 62024664 62024665 62030098 62050995 62050996 62050997 62052956 62060720 62076950 62076954 62090604 62094992 62103587 62113233 62121290 62127574 62140361 62148583 62154323 62154324 62167408 62173357 62183214 62193445 62212155 62216715 62226915 62240346 62240898 62240899 62255241 62265310 62267534 62268455 62289685 62291132 62307716 369271 369496 370180 370949 60277294 60298332 60305721 60305723 60346556 60353254 60357103 60363827 60368486 60388297 60398728 60398729 60406159 60416001 60426062 60452901 60472615 60497074 60499946 60527090 60531821 60563142 60563145 60573645 60579236 60620149 60630360 60636987 60657788 60666931 60674072 60675798 60678684 60686772 60698997 60709631 60715818 60721211 60728507 60761040 60761045 60761046 60778188 60785270 60802036 60807960 60821404 60821405 60838463 60838464 60859175 60859176 60859177 60862269 60877633 60877634 60884811 60884812 60884813 60899796 60899797 60926779 60926780 60941644 60941645 60941646 60961254 60961255 60961256 60961257 60998743 60998744 60998746 60998747 60998748 61022947 61022948 61036847 61052721 61052722 61071009 61071010 61071011 61077627 61115033 61115034 61115035 61130573 61143228 61143229 61143236 61170527 61170528 61170529 61185070 61185071 61185073 61209545 61209546 61209998 61209999 61246291 61246292 61246293 61246294 61246295 61252239 61252240 61252241 61260749 61260750 61279587 61292641 61292642 61292643 61292644 61295992 61312942 61312943 61312944 61321952 61337806 61361627 61366889 61386013 61396428 61399450 61415030 61427185 61467612 61467617 61467628 61513102 61513103 61513104 61533233 61533494 61533495 61537695 61553216 61553218 61558507 61567324 61605382 61629380 61629381 61677020 61682240 61684807 61687528 61695223 61726158 61726159 61726160 61734192 61750729 61767869 61774918 61779296 61804972 61805374 61805376 61830039 61830040 61841661 61857266 61864122 61873816 61887473 61892547 61894775 61908931 61913843 61916898 61918785 61922680 61928931 61939306 61941028 61951320 61964115 61977983 61989381 61990139 61996835 61997568 62035445 62043527 62046463 62050991 62057497 62067398 62092915 62094988 62097145 62105212 62136717 62136718 62141808 62154318 62154319 62154575 62154576 62194617 62194618 62198607 62229922 62240347 62251878 62294378 62294379 62304860 62305919 369356 369495 370859 370950 60305720 60314562 60346557 60367211 60383361 60416004 60426063 60452902 60497078 60507566 60533776 60570481 60603091 60603093 60683489 60687827 60715819 60728508 60761135 60802038 60802039 60807962 60833930 60844760 60854016 60862270 60876783 60884816 60884819 60884821 60884824 60884826 60907246 60946163 60946164 60998749 60998750 60998751 61020073 61022949 61036848 61063586 61063587 61071013 61102874 61102875 61102876 61114977 61134813 61161631 61185074 61185075 61201019 61209535 61210000 61224325 61252192 61252193 61252194 61260666 61287183 61287184 61310099 61310101 61321953 61321954 61333788 61453028 61453029 61499961 61499962 61499963 61533496 61533497 61553220 61553221 61618285 61618286 61692048 61710636 61732330 61750684 61798119 61803811 61830038 61873817 61873818 61895662 61908932 61918164 61936088 61940085 61948434 61977984 61993278 61999835 61999836 62075414 62079208 62080042 62103588 62114950 62126365 62173358 62194619 62213268 62215038 62240902 62262704 62268457 62288151 61296157 61976233 61620568 60576018 60581756 60595142 60600227 60611427 60623596 60630366 60656863 60682822 60687521 60967593 60997526 61018660 61018661 61060255 61066183 61091447 61119666 61119667 61131870 61131871 61213672 61243250 61261246 61261247 61261248 61261249 61296158 61296163 61296166 61296167 61344438 61344439 61408351 61461417 61461418 61461419 61462838 61489833 61567691 61593408 61593409 61684187 61700481 61742492 61764739 61778327 61792462 61792467 61847267 61998471 62103974 62160409 62172415 62268523 61036976 367257 369357 369497 370177 370457 370860 370895 370911 370925 60282133 60421567 60446638 60452904 60472618 60497083 60525152 60531822 60560745 60564436 60570482 60584232 60595329 60612009 60623486 60639462 60656597 60657789 60678685 60682216 60685536 60685537 60687828 60699001 60709824 60709826 60721215 60721216 60725149 60740815 60761145 60780772 60785271 60807963 60807964 60821411 60821412 60834490 60837639 60859178 60862271 60884831 60884834 60926783 60946166 60992257 61006469 61006470 61052725 61071015 61071016 61077632 61114991 61120272 61143227 61170530 61185076 61201023 61209511 61210001 61224307 61252242 61252243 61270111 61304014 61310102 61549290 61549292 61774933 61857261 61873819 61873820 61902156 61911414 61927020 61941944 61968077 61977985 61985279 61991433 62031764 62041540 62050992 62068498 62073532 62092913 62097137 62099609 62124074 62134048 62163488 62173359 62199460 62255242 62277995 62278732 62294314 367218 369272 369474 370176 370665 370810 370893 370910 370924 370986 60282134 60374147 60388321 60452907 60472619 60497084 60527091 60557552 60560682 60580978 60595338 60620157 60639465 60653321 60653322 60656608 60678687 60682217 60682218 60686774 60687829 60699002 60709830 60709836 60709848 60715820 60718476 60721218 60728509 60761156 60785040 60795607 60805894 60807965 60812436 60815648 60833932 60834491 60841583 60854014 60859214 60877635 60998752 61022950 61052726 61071018 61083464 61102877 61110865 61114974 61134814 61143226 61161591 61170531 61185077 61209537 61295993 61549293 61549295 61549296 61873821 61873822 61907872 61964978 61977986 62003797 62010580 62011999 62036509 62039583 62043528 62063738 62068497 62073533 62083343 62109667 62109810 62114949 62118746 62121295 62123056 62144977 62156241 62173362 62197315 62215039 62285294 62300223 61547907 61608138 61620569 60581757 60595144 60607614 60620593 60630367 60664330 60681615 60681616 60699149 61191715 61243251 61261250 61261251 61261252 61296170 61296173 61296174 61313779 61430330 61430332 61462839 61522314 61567690 61605545 61697438 61756176 61771939 61799310 61803727 61911180 61998475 62103972 62167432 60531769 60534916 60766656 60766658 60766659 60766660 60766661 60766666 369273 369499 370175 370373 370867 370894 370912 60298334 60314565 60426064 60580984 60593490 60639472 60657790 60676295 60686777 60702923 60708125 60711105 60785283 60807966 60815690 60837634 61134815 61154498 61161590 61549302 61549303 61549304 61873824 61897938 62043529 62068490 62073534 62099610 62148582 62199462 62257231 62282954 369274 369512 370174 370868 370920 60282135 60580988 60620168 60662854 60702924 60711086 60785293 60821414 61076291 61873826 61930504 61943526 61990144 61997569 62098853 62171376 62199459 62262682 62291133 61608134 60576020 60638556 60764315 60764316 60764317 61602916 369275 369513 370173 370871 370909 370954 60580989 60615740 60740816 60805895 61130584 61966002 62018105 62032352 62063906 62083344 62099611 62171377 62200362 62289687 369514 370875 370952 60581026 60620174 60656996 60662859 60785045 60812432 60821415 60841588 62008854 62025226 62099612 62171378 62281861 61654442 60588882 60656866 60712817 61182982 61190184 61243252 61261253 61482942 61632206 60718764 366665 369515 370171 370908 60581027 60821416 60835761 60844762 61130531 61143224 62025227 62099613 62118750 370751 370953 60581028 60620180 60655960 60663038 61134816 61948163 61993616 62025228 62146120 61640785 61807462 61261254 62046384 62122979
Other devices under the same FDA submission
23 devices cleared under K833991 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- MultiPolar® 5001-47Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented4 reports
- MultiPolar® 5001-50Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented4 reports
- MultiPolar® 5001-50-28Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented4 reports
- MultiPolar® 5001-52Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented3 reports
- MultiPolar® 5001-43Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 reports
- MultiPolar® 5001-44Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 reports
- MultiPolar® 5001-45Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 reports
- MultiPolar® 5001-46Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 reports
- MultiPolar® 5001-49Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 reports
- MultiPolar® 5001-42-22Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 reports
- MultiPolar® 5001-44-28Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 reports
- MultiPolar® 5001-50-28Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 66500).
- Z-1021-2014 · TerminatedI/B II KNEE DOMED PATELLA, I/B II KNEE FLUTED ROD (1 and 2), MGII KNEE 10MM PATELLA (2,3 and 4), MGII KNEE SLF-TAP BONE ST
- Z-1023-2014 · TerminatedMODULAR CUP NEUTRAL LINER MODULAR CUP 10 DEGREE LIN MODULAR CUP 20 DEGREE LIN MODULAR CUP 7MM OFFSET LI
- Z-1024-2014 · TerminatedTRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY ACET SHELL 42MM O TRILOGY ACET SHELL 44MM O TRILOGY ACET SHELL 46MM O TRILOGY ACET SHELL 48MM …
- Z-1025-2014 · TerminatedHGP II ACETABULAR CUP BON
- Z-1026-2014 · TerminatedHGP II ACETABULAR CUP MOD
- Z-1029-2014 · Terminated12/14 UNIPOLAR FEMORAL HE VERSYS CEM LD/FX SZ 11X12 VERSYS CEM LD/FX SZ 12X12 VERSYS CEM LD/FX SZ 14X13 VERSYS CEM LD/FX SZ 16X14 VERSYS CEM LD/FX ST 18X15 VERSYS PRESS FIT LD/FX S…
- Z-1031-2014 · Terminated12/14 COCR FEMORAL HEAD (2, 3 and 4); 12/14 FEMORAL HEAD (+7X28M and 7X32M)
- Z-1032-2014 · TerminatedCPT 12/14 COCR (various sizes)
- Z-1033-2014 · Terminated6 DEGREE COCR FEM HEAD (various sizes) and 6 DEG UNIPOLAR FEMORAL HE
- Z-1034-2014 · Terminated6 DEGREE COCR FEM HEAD 36
- Z-1036-2014 · TerminatedCPT DEGREE TAPER (various sizes) and FEMORAL HD (various sizes)
- Z-1040-2014 · TerminatedPERI SCR. 3.5MM X 10MM PERI SCR. 3.5MM X 12MM PERI SCR. 3.5MM X 14MM PERI SCR 3.5MM X 44MM STE PERI SCR. 3.5MM X 65MM PERI. SCR 4.0MM X24MM PERI. SCR 4.0MM X40MM …
- Z-1042-2014 · Terminated1.5 X 6 CORT SCREW HEX/SE 2.0 X 6 CORT SCREW HEX/SE 2.0 X 10 CORT SCREW CRUCI 2.0 X 12 CORT SCREW CRUCI 2.0 X 14 CORT SCREW CRUCI 2.0 X 16 CORT SCREW HEX/S 2.0 X 16 CORT SCREW CRUC…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page