Hgp Ii Acetabular Cup Mod recallZ-1026-2014 · Zimmer, Inc. · initiated Oct 11, 2013
Class II Terminated All recalls posted in February 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
FDA root cause category: Packaging
What the company is doing
Distributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.
Recall details
- Recall number
- Z-1026-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 11, 2013
- Posted by the FDA
- Feb 20, 2014
- Recalling firm
- Zimmer, Inc.
- Quantity
- 7,044,680 total
- Product code
- LPH
- FDA submissions
- K921308
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
Product description
HGP II ACETABULAR CUP MOD
Code information (models, lots, serial numbers)
60035657 60035663 60163301 60078959 60163345
Other devices under the same FDA submission
7 devices cleared under K921308 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 6624-65-30Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED7 reports
- NA 6624-65-40Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED6 reports
- NA 6624-65-50Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED6 reports
- NA 6624-65-20Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
- NA 6624-65-25Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- NA 6624-65-60Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- NA 6624-65-35Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 66500).
- Z-1019-2014 · TerminatedBipolar liners and Multipolar cups with the following dimensions in MM (38, 39, 40, 41, 42, 43, 44, 45, 46, 47, 48, 49, 50, 53, 54, 55, 57, 58, 60, 64, 66, 68, 70, 72) and several …
- Z-1021-2014 · TerminatedI/B II KNEE DOMED PATELLA, I/B II KNEE FLUTED ROD (1 and 2), MGII KNEE 10MM PATELLA (2,3 and 4), MGII KNEE SLF-TAP BONE ST
- Z-1023-2014 · TerminatedMODULAR CUP NEUTRAL LINER MODULAR CUP 10 DEGREE LIN MODULAR CUP 20 DEGREE LIN MODULAR CUP 7MM OFFSET LI
- Z-1024-2014 · TerminatedTRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY ACET SHELL 42MM O TRILOGY ACET SHELL 44MM O TRILOGY ACET SHELL 46MM O TRILOGY ACET SHELL 48MM …
- Z-1025-2014 · TerminatedHGP II ACETABULAR CUP BON
- Z-1029-2014 · Terminated12/14 UNIPOLAR FEMORAL HE VERSYS CEM LD/FX SZ 11X12 VERSYS CEM LD/FX SZ 12X12 VERSYS CEM LD/FX SZ 14X13 VERSYS CEM LD/FX SZ 16X14 VERSYS CEM LD/FX ST 18X15 VERSYS PRESS FIT LD/FX S…
- Z-1031-2014 · Terminated12/14 COCR FEMORAL HEAD (2, 3 and 4); 12/14 FEMORAL HEAD (+7X28M and 7X32M)
- Z-1032-2014 · TerminatedCPT 12/14 COCR (various sizes)
- Z-1033-2014 · Terminated6 DEGREE COCR FEM HEAD (various sizes) and 6 DEG UNIPOLAR FEMORAL HE
- Z-1034-2014 · Terminated6 DEGREE COCR FEM HEAD 36
- Z-1036-2014 · TerminatedCPT DEGREE TAPER (various sizes) and FEMORAL HD (various sizes)
- Z-1040-2014 · TerminatedPERI SCR. 3.5MM X 10MM PERI SCR. 3.5MM X 12MM PERI SCR. 3.5MM X 14MM PERI SCR 3.5MM X 44MM STE PERI SCR. 3.5MM X 65MM PERI. SCR 4.0MM X24MM PERI. SCR 4.0MM X40MM …
- Z-1042-2014 · Terminated1.5 X 6 CORT SCREW HEX/SE 2.0 X 6 CORT SCREW HEX/SE 2.0 X 10 CORT SCREW CRUCI 2.0 X 12 CORT SCREW CRUCI 2.0 X 14 CORT SCREW CRUCI 2.0 X 16 CORT SCREW HEX/S 2.0 X 16 CORT SCREW CRUC…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page