CPT 12/14 COCR (various sizes) recallZ-1032-2014 · Zimmer, Inc. · initiated Oct 11, 2013
Class II Terminated All recalls posted in February 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
FDA root cause category: Packaging
What the company is doing
Distributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.
Recall details
- Recall number
- Z-1032-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 11, 2013
- Posted by the FDA
- Feb 20, 2014
- Recalling firm
- Zimmer, Inc.
- Quantity
- 7,044,680 total
- Product code
- JDI
- FDA submissions
- K030265, K960658
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
Product description
CPT 12/14 COCR (various sizes)
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
60225078 60261757 60274775 60310767 60398002 60491038 60638468 60638470 60685165 60776190 60797319 60927586 61189804 61237349 61342952 61468762 61490103 61596784 61748948 61764725 61840597 61917999 61964635 61999588 62003977 60261761 60261763 60776227 60776234 60846362 60916726 60952935 60979483 61162865 61189800 61356814 61460831 61523005 61739387 61748949 61829563 61996730 60310780 60335521 60412478 60457571 60533531 60776287 60776288 60823773 60990055 61045309 61351320 61500146 61727738 61727739 61957411 60217292 60245136 60330281 60448728 60491098 60648441 60704737 60748770 60776312 60797325 60916737 60952946 61261797 61293863 61307997 61733348 61758312 61868121 61881187 61885896 61991679 62001331 60245153 60245154 60323912 60372490 60589471 60704749 60723802 60748790 60921196 60957269 61045290 61061843 61104280 61114610 61733350 61987756 61991676 60217518 60310793 60330358 60589477 60789763 60829712 60957183 60957252 61010390 61047782 61356826 61484547 61589156 61650372 61696379 61747548 61762055 61774068 61783714 61878465 61930415 61991684 61999776 62003973 62006549 62006552 62014206 60509163 61130729 61460829 61613256 61711999 62097052 60600577 60704770 60898905 61035924 61045315 61114612 61237346 61243405 61516824 61596779 61744474 61833302 61878466 61930418 61964629 60310799 60352497 60600588 60615058 60748834 60797333 60823790 60889814 60889815 60952953 60957186 60979502 61045316 61325399 61336846 61351344 61522157 61694927 61762060 61815831 62001327 60261985 60261986 60326340 60330332 60510215 60574913 60685247 60748844 60748850 60797335 60799614 60822508 60916748 60916749 60921201 60944305 61075599 61114637 61460856 61472388 61731627 61762053 61833303 61840602 61884450 61957405 62001329 60380023 60443946 60491173 60685261 60809480 61024361 61104288 61516832 61607601 60378595 60723862 60748854 60748855 60898910 61045291 61114614 61211994 61505499 61512550 61814507 61914812 61987765 60200828 60225125 60274811 60510226 60660270 60685253 60723871 60748857 60822510 60842663 60889819 60889820 61027639 61061835 61182969 61282937 61314041 61460835 61490099 61696376 61731630 61809118 61987773 62017570 60217523 60274812 60420451 60517574 60704826 60704827 60723878 60749232 60789777 60809481 60809482 60898877 60927619 60927620 60979509 61002485 61045318 61173152 61205952 61216685 61490100 61505498 61650364 61696373 61744461 61762059 61833301 62005396 60491175 60704843 60789778 60799619 60880782 60997430 61546126 60324107 60443947 60704844 60724240 60809483 60921203 61114628 61601357 60491179 60820144 61061846 61290086 61546128 61691513 60225115 60330349 60723884 60723885 60809484 61027640 61061855 61500145 61579869 61817968 61927037 60510237 60510239 60704831 60749247 60749248 60820124 60858530 60889822 60990042 60990043 61006802 61045292 61094975 61114629 61316307 61354857 61647194 61696371 61744471 61820815 369396 60222666 60274815 60510242 60589746 60660275 60704835 60749255 60749256 60944307 60973154 60979538 61024369 61024370 61062535 61114616 61153032 61339722 61550881 61683768 61696359 61779345 61820816 369397 60225128 60704836 61024371 61066495 61490101 61790493 60704845 60777781 60803456 61770622 60789779 60979539 61664275 60704715 61035926 61075584 61325390 61520903 61636260 61762068 61878931 61993208 60809444 61061836 61297358 61473756 61522154 61589113 61790499 61993210
Other devices under the same FDA submission
25 devices cleared under K030265, K960658 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CPT® 8114-01-10Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall24 reports
- CPT® 8114-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall11 reports
- CPT® 8114-02-10Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall11 reports
- CPT® 8114-03-10Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall11 reports
- CPT® 8114-00-10Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall10 reports
- CPT® 8114-02Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall9 reports
- CPT® 8114-02-30Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall6 reports
- CPT® 8114Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall4 reports
- CPT® 8114-02-18Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall3 reports
- CPT® 8114-03Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall2 reports
- CPT® 8114-04-10Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall2 reports
- CPT® 8114-03-18Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 66500).
- Z-1019-2014 · TerminatedBipolar liners and Multipolar cups with the following dimensions in MM (38, 39, 40, 41, 42, 43, 44, 45, 46, 47, 48, 49, 50, 53, 54, 55, 57, 58, 60, 64, 66, 68, 70, 72) and several …
- Z-1021-2014 · TerminatedI/B II KNEE DOMED PATELLA, I/B II KNEE FLUTED ROD (1 and 2), MGII KNEE 10MM PATELLA (2,3 and 4), MGII KNEE SLF-TAP BONE ST
- Z-1023-2014 · TerminatedMODULAR CUP NEUTRAL LINER MODULAR CUP 10 DEGREE LIN MODULAR CUP 20 DEGREE LIN MODULAR CUP 7MM OFFSET LI
- Z-1024-2014 · TerminatedTRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY ACET SHELL 42MM O TRILOGY ACET SHELL 44MM O TRILOGY ACET SHELL 46MM O TRILOGY ACET SHELL 48MM …
- Z-1025-2014 · TerminatedHGP II ACETABULAR CUP BON
- Z-1026-2014 · TerminatedHGP II ACETABULAR CUP MOD
- Z-1029-2014 · Terminated12/14 UNIPOLAR FEMORAL HE VERSYS CEM LD/FX SZ 11X12 VERSYS CEM LD/FX SZ 12X12 VERSYS CEM LD/FX SZ 14X13 VERSYS CEM LD/FX SZ 16X14 VERSYS CEM LD/FX ST 18X15 VERSYS PRESS FIT LD/FX S…
- Z-1031-2014 · Terminated12/14 COCR FEMORAL HEAD (2, 3 and 4); 12/14 FEMORAL HEAD (+7X28M and 7X32M)
- Z-1033-2014 · Terminated6 DEGREE COCR FEM HEAD (various sizes) and 6 DEG UNIPOLAR FEMORAL HE
- Z-1034-2014 · Terminated6 DEGREE COCR FEM HEAD 36
- Z-1036-2014 · TerminatedCPT DEGREE TAPER (various sizes) and FEMORAL HD (various sizes)
- Z-1040-2014 · TerminatedPERI SCR. 3.5MM X 10MM PERI SCR. 3.5MM X 12MM PERI SCR. 3.5MM X 14MM PERI SCR 3.5MM X 44MM STE PERI SCR. 3.5MM X 65MM PERI. SCR 4.0MM X24MM PERI. SCR 4.0MM X40MM …
- Z-1042-2014 · Terminated1.5 X 6 CORT SCREW HEX/SE 2.0 X 6 CORT SCREW HEX/SE 2.0 X 10 CORT SCREW CRUCI 2.0 X 12 CORT SCREW CRUCI 2.0 X 14 CORT SCREW CRUCI 2.0 X 16 CORT SCREW HEX/S 2.0 X 16 CORT SCREW CRUC…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page