1.5 X 6 Cort Screw Hex/Se recallZ-1042-2014 · Zimmer, Inc. · initiated Oct 11, 2013
Class II Terminated All recalls posted in February 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
FDA root cause category: Packaging
What the company is doing
Distributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.
Recall details
- Recall number
- Z-1042-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 11, 2013
- Posted by the FDA
- Feb 20, 2014
- Recalling firm
- Zimmer, Inc.
- Quantity
- 7,044,680 total
- Product code
- HWC
- FDA submissions
- K060710, K063303, K112885
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
Product description
1.5 X 6 CORT SCREW HEX/SE 2.0 X 6 CORT SCREW HEX/SE 2.0 X 10 CORT SCREW CRUCI 2.0 X 12 CORT SCREW CRUCI 2.0 X 14 CORT SCREW CRUCI 2.0 X 16 CORT SCREW HEX/S 2.0 X 16 CORT SCREW CRUCI 2.0 X 18 CORT SCREW CRUCI 2.0 X 20 CORT SCREW CRUCI 2.0 X 26 CORT SCREW CRUCI 2.0 X 28 CORT SCREW CRUCI 2.0 X 30 CORT SCREW CRUCI 2.0 X 32 CORT SCREW CRUCI 2.0 X 34 CORT SCREW CRUCI 2.0 X 36 CORT SCREW CRUCI 2.0 X 38 CORT SCREW CRUCI 2.0 X 40 CORT SCREW CRUCI 2.0 X 42 CORT SCREW CRUCI 2.0 X 44 CORT SCREW CRUCI 2.0 X 46 CORT SCREW CRUCI 2.0 X 48 CORT SCREW CRUCI 2.0 X 50 CORT SCREW CRUCI 2.7MM CORT. SCREW 6MM, ST 2.7MM CORT. SCREW 7MM, ST 2.7MM CORT. SCREW 8MM, ST 2.7MM CORT. SCREW 9MM, ST 2.7MM CORT. SCREW 10MM, S 2.7MM CORT. SCREW 12MM, S 2.7MM CORT. SCREW 14MM, S 2.7MM CORT. SCREW 16MM, S 2.7MM CORT. SCREW 18MM, S 2.7MM CORT. SCREW 20MM, S 2.7MM CORT. SCREW 22MM, S 2.7MM CORT. SCREW 24MM, S 2.7MM CORT. SCREW 26MM, S 2.7MM CORT. SCREW 28MM, S 2.7MM CORT. SCREW 30MM, S 2.7MM CORT. SCREW 32MM, S 2.7MM CORT. SCREW 34MM, S 2.7MM CORT. SCREW 36MM, S 2.7MM CORT. SCREW 38MM, S 2.7MM CORT. SCREW 40MM, S 2.7MM CORT. SCREW 42MM, S 2.7MM CORT. SCREW 44MM, S 2.7MM CORT. SCREW 45MM, S 2.7MM CORT. SCREW 46MM, S 2.7MM CORT. SCREW 48MM, S 2.7MM CORT. SCREW 50MM, S 2.7X52 CORT SCREW/SELFTAP 2.7X54 CORT SCREW/SELFTAP 2.7X55 CORT SCREW/SELFTAP 2.7X56 CORT SCREW/SELFTAP 2.7X58 CORT SCREW/SELFTAP 2.7X60 CORT SCREW/SELFTAP 2.7X65 CORT SCREW/SELFTAP 2.7X70 CORT SCREW/SELFTAP 3.5MM CORT. SCREW 10MM LN 3.5MM CORT. SCREW 12MM LN 3.5X14 CORT SCREW 3.5MM CORT. SCREW 14MM LN 3.5X16 CORT SCREW 3.5MM CORT. SCREW 16MM LN 3.5 X 16 SHAFT SCREW 3.5X18 CORT SCREW 3.5MM CORT. SCREW 18MM LN 3.5 X 18 SHAFT SCREW 3.5MM CORT. SCREW 20MM LN 3.5 X 20 SHAFT SCREW 3.5MM CORT. SCREW 22MM LN 3.5 X 22 SHAFT SCREW 3.5MM CORT. SCREW 24MM LN 3.5 X 24 SHAFT SCREW 3.5MM CORT. SCREW 26MM LN 3.5 X 26 SHAFT SCREW 3.5MM CORT. SCREW 28MM LN 3.5 X 28 SHAFT SCREW 3.5MM CORT. SCREW 30MM LN 3.5 X 30 SHAFT SCREW 3.5MM CORT. SCREW 32MM LN 3.5 X 32 SHAFT SCREW 3.5MM CORT. SCREW 34MM LN 3.5 X 34 SHAFT SCREW 3.5MM CORT. SCREW 36MM LN 3.5 X 36 SHAFT SCREW 3.5MM CORT. SCREW 38MM LN 3.5 X 38 SHAFT SCREW 3.5MM CORT. SCREW 40MM LN 3.5 X 40 SHAFT SCREW 3.5MM CORT. SCREW 45MM LN 3.5MM CORT. SCREW 50MM LN 3.5 X 50 SHAFT SCREW 3.5MM CORT. SCREW 55MM LN 3.5MM CORT. SCREW 60MM LN 3.5MM CORT. SCREW 65MM LN 3.5MM CORT. SCREW 70MM LN 3.5MM CORT. SCREW 75MM LN 3.5MM CORT. SCREW 80MM LN 3.5X85 CORT SCREW 3.5X90 CORT SCREW 3.5X95 CORT SCREW 3.5X100 CORT SCREW 3.5X105 CORT SCREW 3.5X110 CORT SCREW 3.5X115 CORT SCREW 3.5X120 CORT SCREW 3.5X125 CORT SCREW 3.5X130 CORT SCREW 3.5X135 CORT SCREW 3.5X140 CORT SCREW 3.5X145 CORT SCREW 3.5X150 CORT SCREW 4.0 X 10 CANCELL. SCREW 4.0 X 10 PART THD CANC SC 4.0 X 12 CANCELL. SCREW 4.0 X 12 PART THD CANC SC 4.0 X 14 CANCELL. SCREW 4.0 X 14 PART THD CANC SC 4.0 X 16 CANCELL. SCREW 4.0 X 16 PART THD CANC SC 4.0 X 18 CANCELL. SCREW 4.0 X 18 PART THD CANC SC 4.0 X 20 CANCELL. SCREW 4.0 X 20 PART THD CANC SC 4.0 X 22 CANCELL. SCREW 4.0 X 22 PART THD CANC SC 4.0 X 24 CANCELL. SCREW 4.0 X 24 PART THD CANC SC 4.0 X 26 CANCELL. SCREW 4.0 X 26 PART THD CANC SC 4.0 X 28 CANCELL. SCREW 4.0 X 28 PART THD CANC SC 4.0 X 30 CANCELL. SCREW 4.0 X 30 PART THD CANC SC 4.0 X 32 CANCELL. SCREW 4.0 X 34 CANCELL. SCREW 4.0 X 35 PART THD CANC SC 4.0 X 36 CANCELL. SCREW 4.0 X 40 CANCELL. SCREW 4.0 X 40 PART THD CANC SC 4.0 X 45 CANCELL. SCREW 4.0 X 45 PART THD CANC SC 4.0 X 50 CANCELL. SCREW 4.0 X 50 PART THD CANC SC 4.0 X 55 CANCELL. SCREW 4.0 X 55 PART THD CANC SC 4.0 X 60 CANCELL. SCREW 4.0 X 60 PART THD CANC SC 4.0 X 65 CANCELL. SCREW 4.0
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
61267037 61195515 61233513 61255911 61196113 61255912 61255913 61196115 61233517 61255914 61255915 61255916 61196121 61196122 61759822 61196123 61582530 61605810 61196124 61582531 61196125 61196126 61582532 61196127 61759823 61196128 61196129 61196131 61196132 61196133 61196134 61196460 61196461 61168863 61234474 61196462 61196463 61234476 61247538 61196464 61225322 61234477 61196465 61234479 61168867 61196466 61234480 61196467 61234481 61196468 61234482 61196469 61234484 61168870 61196470 61168871 61196471 61196472 61168873 61196473 61168874 61196474 61196475 61168875 61196476 61168876 61196477 60654904 61168877 61196478 61196479 60654906 61196480 61168878 61196482 61196483 61196484 61168880 61196485 61256597 61256599 61256600 61256601 61256603 61310792 61344225 61530374 61654821 61256605 61256606 61216613 61247253 61557215 61881416 61216616 61247255 61263780 61585001 61247492 61247493 61251362 61251363 61322883 61326924 61473762 61873515 61247496 61247272 61247273 61247275 61775835 61835769 61196225 61247504 61247505 61241617 61337168 61339774 61775836 61196226 61247277 61263781 61263782 61263783 61196227 61234462 61247278 61263784 61196228 61263785 61196229 61247280 61263786 61196230 61234465 61263787 61640844 61775837 61196231 60619377 61216644 61234466 61196232 61216646 61234467 61196233 61585002 61196234 61216649 61296496 61196235 60619404 61216651 61635840 61196236 61234469 61420532 61614769 61196237 61216653 61234470 61585003 61640846 61234471 61640847 61196239 60619379 61216656 60619382 60564048 60619380 61216659 61234473 60564052 60553867 60564055 60619373 60564056 60619372 61196212 61196213 61196214 61196215 61196216 61196217 61196218 61196219 61196220 61196221 61196222 61196223 61196224 61251335 61378853 61296498 61359796 61553346 61574969 61216664 61234486 61296499 61420534 61234487 61234488 61247467 61263788 61234489 61234491 61567476 61247468 61378854 61447036 61553315 61565530 61567477 61234505 61241624 61234494 61420533 61473763 61557223 61268670 61268671 61557234 61251336 61553318 61593247 61263789 61216671 61234496 61553324 61263790 61553354 61216672 61234497 61234507 61234498 61359797 61557240 61337169 61234508 61553360 61574970 61359792 61557227 61234499 61296492 61553325 61567479 61234509 61337170 61339782 61574968 61234500 61337171 61553337 61567481 61597163 61234510 61234511 61241623 61247481 61251337 61234512 61234513 61216677 61234502 61234514 61234515 61247475 61473764 61567482 61231084 61231085 61231088 61378855 61231091 61231093 61369555 61567484 61231095 61231097 61567485 61231108 61359794 61553343 61359795 61854115 61251502 61225496 61251503 61251504 61225498 61225503 61247289 61225576 61225505 61225510 61251507 61251508 61225513 61251509 61251514 61247293 61225578 61251517 61251520 61225542 61225543 61225544 61225547 61237540 61225549 61225551 61225552 61251532 61251533 61251535 61553371 61553376 61337178 61225599 61225601
Other devices under the same FDA submission
19 devices cleared under K060710, K063303, K112885 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 47-4835-016-01Zimmer, Inc. · Screw, fixation, bone5 reports
- NA 47-4828-016-02Zimmer, Inc. · Screw, fixation, bone4 reports
- NA 47-4835-024-01Zimmer, Inc. · Screw, fixation, bone2 reports
- NA 47-4835-045-01Zimmer, Inc. · PLATE, FIXATION, BONE2 reports
- NA 47-4828-014-02Zimmer, Inc. · Screw, fixation, bone1 report
- NA 47-4828-018-02Zimmer, Inc. · Screw, fixation, bone1 report
- NA 47-4835-014-01Zimmer, Inc. · PLATE, FIXATION, BONE1 report
- NA 47-4835-018-01Zimmer, Inc. · PLATE, FIXATION, BONE1 report
- NA 47-4835-020-01Zimmer, Inc. · Screw, fixation, bone1 report
- NA 47-4835-028-01Zimmer, Inc. · Screw, fixation, bone1 report
- NA 47-4835-060-01Zimmer, Inc. · PLATE, FIXATION, BONE1 report
- NA 47-4840-014-00Zimmer, Inc. · WASHER, BOLT NUT1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 66500).
- Z-1019-2014 · TerminatedBipolar liners and Multipolar cups with the following dimensions in MM (38, 39, 40, 41, 42, 43, 44, 45, 46, 47, 48, 49, 50, 53, 54, 55, 57, 58, 60, 64, 66, 68, 70, 72) and several …
- Z-1021-2014 · TerminatedI/B II KNEE DOMED PATELLA, I/B II KNEE FLUTED ROD (1 and 2), MGII KNEE 10MM PATELLA (2,3 and 4), MGII KNEE SLF-TAP BONE ST
- Z-1023-2014 · TerminatedMODULAR CUP NEUTRAL LINER MODULAR CUP 10 DEGREE LIN MODULAR CUP 20 DEGREE LIN MODULAR CUP 7MM OFFSET LI
- Z-1024-2014 · TerminatedTRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY ACET SHELL 42MM O TRILOGY ACET SHELL 44MM O TRILOGY ACET SHELL 46MM O TRILOGY ACET SHELL 48MM …
- Z-1025-2014 · TerminatedHGP II ACETABULAR CUP BON
- Z-1026-2014 · TerminatedHGP II ACETABULAR CUP MOD
- Z-1029-2014 · Terminated12/14 UNIPOLAR FEMORAL HE VERSYS CEM LD/FX SZ 11X12 VERSYS CEM LD/FX SZ 12X12 VERSYS CEM LD/FX SZ 14X13 VERSYS CEM LD/FX SZ 16X14 VERSYS CEM LD/FX ST 18X15 VERSYS PRESS FIT LD/FX S…
- Z-1031-2014 · Terminated12/14 COCR FEMORAL HEAD (2, 3 and 4); 12/14 FEMORAL HEAD (+7X28M and 7X32M)
- Z-1032-2014 · TerminatedCPT 12/14 COCR (various sizes)
- Z-1033-2014 · Terminated6 DEGREE COCR FEM HEAD (various sizes) and 6 DEG UNIPOLAR FEMORAL HE
- Z-1034-2014 · Terminated6 DEGREE COCR FEM HEAD 36
- Z-1036-2014 · TerminatedCPT DEGREE TAPER (various sizes) and FEMORAL HD (various sizes)
- Z-1040-2014 · TerminatedPERI SCR. 3.5MM X 10MM PERI SCR. 3.5MM X 12MM PERI SCR. 3.5MM X 14MM PERI SCR 3.5MM X 44MM STE PERI SCR. 3.5MM X 65MM PERI. SCR 4.0MM X24MM PERI. SCR 4.0MM X40MM …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page