CP5 centrifugal pump system recallZ-0857-2014 · Sorin Group Deutschland GmbH · initiated Dec 4, 2013
Class II Terminated All recalls posted in January 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Sorin Group Deutschland GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sorin is initiating a field correction on S5/C5 Heart-lung machine and CP5 centrifugal pump system due to reports of pump speed control knob failures resulting in no response.
FDA root cause category: Device Design
What the company is doing
Sarin Group US distributed a Field Safety Notice dated December 5, 2013 to all affected customers and distributors. The letter informed users of the potential for shaft encoder failures and notified them that they would be contacted by the Sorin Service Group to arrange an appointment to repair the units onsite. Customers and distributors were also provided with a Response form to confirm they had received, read and understood the Field Notice. They were instructed to return the completed form to assist in monitoring the effectiveness of the communication. For questions call 303-467-6527 or 303-467-6503.
Recall details
- Recall number
- Z-0857-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 4, 2013
- Posted by the FDA
- Jan 28, 2014
- Recalling firm
- Sorin Group Deutschland GmbH
- Quantity
- 75
- Product code
- DWA
- FDA submissions
- K112225
- Distribution
- Worldwide Distribution - US (Nationwide) and PR; and Internationally to:BE, ES, FI, GB, NL, SE, AR, AT, AU, CA, CH, CN, CR, EG, HK, ID, JO, JP, KR, KW, LB, LY, MM, PL, RO, RS, RU, SG. TH, TN, TR, TW, UA, VN, and ZA.
Product description
CP5 centrifugal pump system Product Usage: The Stockert S5/ Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.
Code information (models, lots, serial numbers)
Item Number: 60-02-60
Other devices under the same FDA submission
5 devices cleared under K112225 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CP5 Pump Control PanelLivaNova Deutschland GmbH · Control, pump speed, cardiopulmonary bypass1 recall53 reports
- CP5 Drive UnitLivaNova Deutschland GmbH · Control, pump speed, cardiopulmonary bypass1 recall32 reports
- CP5 Emergency SystemLivaNova Deutschland GmbH · Control, pump speed, cardiopulmonary bypass1 recall2 reports
- CP5 Pump Control PanelLivaNova Deutschland GmbH · Control, pump speed, cardiopulmonary bypass9 reports
- Flow Measurement Sensor ModuleLivaNova Deutschland GmbH · Control, pump speed, cardiopulmonary bypass1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 67199).
- Z-0856-2014 · TerminatedS5/C5 Heart-lung machine Product Usage: The StOckert S5/ Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.