S5/C5 Heart-lung machine recallZ-0856-2014 · Sorin Group Deutschland GmbH · initiated Dec 4, 2013
Class II Terminated All recalls posted in January 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Sorin Group Deutschland GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sorin is initiating a field correction on S5/C5 Heart-lung machine and CP5 centrifugal pump system due to reports of pump speed control knob failures resulting in no response.
FDA root cause category: Device Design
What the company is doing
Sarin Group US distributed a Field Safety Notice dated December 5, 2013 to all affected customers and distributors. The letter informed users of the potential for shaft encoder failures and notified them that they would be contacted by the Sorin Service Group to arrange an appointment to repair the units onsite. Customers and distributors were also provided with a Response form to confirm they had received, read and understood the Field Notice. They were instructed to return the completed form to assist in monitoring the effectiveness of the communication. For questions call 303-467-6527 or 303-467-6503.
Recall details
- Recall number
- Z-0856-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 4, 2013
- Posted by the FDA
- Jan 28, 2014
- Recalling firm
- Sorin Group Deutschland GmbH
- Quantity
- 323
- Product code
- DTQ
- FDA submissions
- K071318, K093882
- Distribution
- Worldwide Distribution - US (Nationwide) and PR; and Internationally to:BE, ES, FI, GB, NL, SE, AR, AT, AU, CA, CH, CN, CR, EG, HK, ID, JO, JP, KR, KW, LB, LY, MM, PL, RO, RS, RU, SG. TH, TN, TR, TW, UA, VN, and ZA.
Product description
S5/C5 Heart-lung machine Product Usage: The StOckert S5/ Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.
Code information (models, lots, serial numbers)
S5 Item Number: 10-80-00, 10-85-00, 28-9S-80, 28-95-8S. C5 Item Number: S8-00-00
Other devices under the same FDA submission
23 devices cleared under K071318, K093882 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- S5 Roller Pump 150 RP 150LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 death120 reports
- S5 Console 4-positionLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 death65 reports
- S5 Double Roller Pump 85 DRP 85LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS54 reports
- Electrical Remote-Controlled Tubing Clamp 620 mmLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS52 reports
- S5 Console 3-positionLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS37 reports
- Electrical Remote-Controlled Tubing Clamp 500 mmLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS31 reports
- Bubble Detector Sensor 3/8"LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 death26 reports
- S5 Console 5-positionLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS2 deaths16 reports
- Bubble Detector Sensor 1/4"LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 reports
- Display & Control Module (S5)LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 reports
- SCP Pump Control Panel for S5/C5LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 reports
- Bubble Detector Sensor 3/16"LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 67199).
- Z-0857-2014 · TerminatedCP5 centrifugal pump system Product Usage: The Stockert S5/ Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.