S5/C5 Heart-lung machine recallZ-0856-2014 · Sorin Group Deutschland GmbH · initiated Dec 4, 2013

Class II Terminated All recalls posted in January 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Sorin Group Deutschland GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Sorin is initiating a field correction on S5/C5 Heart-lung machine and CP5 centrifugal pump system due to reports of pump speed control knob failures resulting in no response.

FDA root cause category: Device Design

What the company is doing

Sarin Group US distributed a Field Safety Notice dated December 5, 2013 to all affected customers and distributors. The letter informed users of the potential for shaft encoder failures and notified them that they would be contacted by the Sorin Service Group to arrange an appointment to repair the units onsite. Customers and distributors were also provided with a Response form to confirm they had received, read and understood the Field Notice. They were instructed to return the completed form to assist in monitoring the effectiveness of the communication. For questions call 303-467-6527 or 303-467-6503.

Recall details

Recall number
Z-0856-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 4, 2013
Posted by the FDA
Jan 28, 2014
Recalling firm
Sorin Group Deutschland GmbH
Quantity
323
Product code
DTQ
FDA submissions
K071318, K093882
Distribution
Worldwide Distribution - US (Nationwide) and PR; and Internationally to:BE, ES, FI, GB, NL, SE, AR, AT, AU, CA, CH, CN, CR, EG, HK, ID, JO, JP, KR, KW, LB, LY, MM, PL, RO, RS, RU, SG. TH, TN, TR, TW, UA, VN, and ZA.

Product description

S5/C5 Heart-lung machine Product Usage: The StOckert S5/ Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.

Code information (models, lots, serial numbers)

S5 Item Number: 10-80-00, 10-85-00, 28-9S-80, 28-95-8S. C5 Item Number: S8-00-00

Other devices under the same FDA submission

23 devices cleared under K071318, K093882 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 67199).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page