Trigen Intertan Nail recallZ-1636-2014 · Smith & Nephew Inc · initiated Apr 4, 2014
Class II Terminated All recalls posted in May 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A manufacturing error resulted in the set screw being inserted too deep in the affected nails.
FDA root cause category: Process control
What the company is doing
The firm initiated the recall via an "Urgent - Product Recall 1st Notification - Urgent" letter and e-mail on 04/04/2014, identifying the product, explaining the problem, and the potential risk associated with using the product. Customers are to inspect their inventory, and locate and quarantine any affected product. Additionally, the Inventory Return Certification form should be completed and returned along with any affected product identified, once a Return Authorization number has been obtained from Smith & Nephew's Safety Affairs Department.
Recall details
- Recall number
- Z-1636-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 4, 2014
- Posted by the FDA
- May 20, 2014
- Recalling firm
- Smith & Nephew Inc
- Quantity
- 1,253 units total
- Product code
- JDS
- FDA submissions
- K040212
- Distribution
- Worldwide Distribution -- USA, France, Germany, Sweden, Switzerland, UK, Canada, Australia, Japan, China, India, Dubai, South Africa, South Korea, Mexico, and Venezuela.
Product description
TRIGEN INTERTAN NAIL: (1) REF 71675201, 10 MM X 18 CM, 125 degree, (2) REF 71675202, 11.5 MM X 18 CM, 125 degree, (3) REF 71675203, 13 MM X 18 CM, 125 degree, (4) REF 71675204, 10 MM X 20 CM, 125 degree, (5) REF 71675205, 11.5 MM X 20 CM, 125 degree, (6) REF 71675207, 10 MM X 18 CM, 130 degree, (7) REF 71675208, 11.5 MM X 18 CM, 130 degree, (8) REF 71675211, 11.5 MM X 20 CM, 130 degree, Smith & Nephew, Inc. Orthopedic fixation device.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
TRIGEN INTERTAN NAIL: (1) REF 71675201, Lot Numbers: 13LT36305, 13LT36306, 13LT36547, 13LT36548, 13LT37009, 13LT37010, 13MT37347, 13MT37665, 13MT37667, 13MT37668, 14AT38179, 14AT38180, 14AT38181, 14T38182, 14AT38183, 14AT38189, 14AT38190 (2) REF 71675202, Lot Numbers: 13LT36308, 13LT36549, 13LT37011, 13MT37348, 14AT38750 (3) REF 71675203, Lot Numbers: 13KT36106, 13LT36596, 14AT38494 (4) REF 71675204, Lot Numbers: 13LT36597, 13LT37012, 14AT38495, 14AT38660 (5) REF 71675205, Lot Number 13LT36598 (6) REF 71675207, Lot Numbers: 13LT363089, 13LT36310, 13LT36599, 13LT37058, 13LT37059, 13LT37063, 13MT37264, 13MT37265, 13MT37266, 13MT37329, 13MT37349, 13MT37350, 13MT37669, 13MT37671, 13MT37672 (7) REF 71675208, Lot Numbers: 13LT36311, 13LT36312, 13LT36600, 13LT37060, 13LT37061, 13MT37351, 13MT37666, 13MT37963 (8) REF 71675211, Lot Number 13LT36313
Other devices under the same FDA submission
84 devices cleared under K040212 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- TRIGEN 71676627Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall2 reports
- TRIGEN 71676628Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall0 reports
- TRIGEN 71675383Smith & Nephew, Inc. · NAIL, FIXATION, BONE29 reports
- TRIGEN 71677100Smith & Nephew, Inc. · NAIL, FIXATION, BONE20 reports
- TRIGEN 71675385Smith & Nephew, Inc. · NAIL, FIXATION, BONE18 reports
- TRIGEN 71677095Smith & Nephew, Inc. · NAIL, FIXATION, BONE18 reports
- TRIGEN 71675202Smith & Nephew, Inc. · NAIL, FIXATION, BONE15 reports
- INTERTAN 71934479Smith & Nephew, Inc. · NAIL, FIXATION, BONE12 reports
- TRIGEN 71677090Smith & Nephew, Inc. · NAIL, FIXATION, BONE11 reports
- TRIGEN 71675384Smith & Nephew, Inc. · NAIL, FIXATION, BONE10 reports
- TRIGEN 71675501Smith & Nephew, Inc. · NAIL, FIXATION, BONE8 reports
- TRIGEN 71675208Smith & Nephew, Inc. · NAIL, FIXATION, BONE8 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 68043).
- Z-1637-2014 · TerminatedTRIGEN INTERTAN 1.5 NAIL: (1) REF 71676515, 10 MM X 10 CM, 125 degree, LEFT, (2) REF 71676556, 11.5 MM X 36 CM, 125 degree, RIGHT, (3) REF 71676558, 11.5 MM X 38 CM, 125 degree…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page