Hologic Selenia Full Field Digital Mammography System:… recallZ-1700-2014 · Hologic, Inc. · initiated May 13, 2014

Class II Terminated All recalls posted in June 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hologic, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury

FDA root cause category: Component change control

What the company is doing

Hologic notified customers by letter dated 5/13/14 via certified mail or Federal Express to inform customers of the problem and provide acknowledgement of the notification and confirmation that their systems have been corrected A Dealer notification dated 5/13/14 sent by certified mail or Federal Express to inform Dealers of the need to notify their customers of the problem and to correct the identified Selenia system using Hologic Technical Bulletin TB-00806. On 5/16/14 the firm expanded the recall to an additional units Mammography System M-IV (2 units) and Selenia Mammography Systems (4) by letter dated 5/14/14. If there are any questions regarding this field correction, please call Hologic Customer Support at 877 371-4372.

Recall details

Recall number
Z-1700-2014
Recall class
Class II
Status
Terminated
Initiated by the company
May 13, 2014
Posted by the FDA
Jun 3, 2014
Recalling firm
Hologic, Inc.
Quantity
11 units
Product code
MUE
FDA submissions
P010025
Distribution
Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.

Product description

Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The Selenia Full Field Digital Mammography system generates digital mammography images that can be used for screening and diagnosis of breast cancer.

Code information (models, lots, serial numbers)

Serial Numbers: 2840314W8065, 2840314W8062W, 2840314W8066W, 2840314W8067 2840314W8073, 2840314W8064, 2840314W8056W, 28401082720 28407083579, 28401137611W, 28403083045W