Continuum Acetabular System Trabecular Metal Shell with… recallZ-0094-2015 · Zimmer, Inc. · initiated Oct 1, 2014
Class II Terminated All recalls posted in October 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The affected products are missing polar boss threads.
FDA root cause category: Equipment maintenance
What the company is doing
Zimmer sent an "URGENT MEDICAL DEVICE RECALL- LOT SPECIFIC" notifications dated October 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Your Responsibilities 1. Review the notification and ensure affected personnel are aware of the contents. 2. If you find any product from the affected lot , quarantine the product and notify your Zimmer sales representative. 3. Your Zimmer sales representative will remove the recalled product from your facility. 4. For patients that previously had this product implanted, it is recommended that you continue your normal post operative follow up routine. 5. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-800-348-2759 between 8:00 am and 5:00pm EST.
Recall details
- Recall number
- Z-0094-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 1, 2014
- Posted by the FDA
- Oct 20, 2014
- Recalling firm
- Zimmer, Inc.
- Quantity
- 19
- Product code
- JDI
- FDA submissions
- K091508
- Distribution
- Worldwide Distribution: US (nationwide) in the states of : IA, CA, NC, AL, WA, FL, TN, MA, AZ, TX, UT, and NV; and Internationally to Japan.
Product description
Continuum Acetabular System Trabecular Metal Shell with Multi Holes, Porous 68mm, Prosthesis, Hip, Semi-Constrained, Metal/Polymer
Code information (models, lots, serial numbers)
Part Number: 00-8757-068-02; Lot Number: 62207029
Other devices under the same FDA submission
40 devices cleared under K091508 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Continuum® Trilogy® Allofit® Longevity® 00-8751-010-36Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented8 reports
- Continuum® Trabecular Metal™ 00-8757-056-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented8 reports
- Continuum® Trabecular Metal™ 00-8757-048-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented5 reports
- Continuum® Trabecular Metal™ 00-8757-052-01Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED5 reports
- Continuum® Trabecular Metal™ 00-8757-054-01Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED5 reports
- Continuum® Trilogy® Allofit® Longevity® 00-8751-009-32Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED4 reports
- Continuum® Trilogy® Allofit® Longevity® 00-8751-011-36Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented4 reports
- Continuum® Trilogy® Allofit® Longevity® 00-8751-013-36Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented4 reports
- Trilogy® 00-8753-050-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented4 reports
- Trilogy® 00-8753-052-01Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED4 reports
- Continuum® Trabecular Metal™ 00-8757-050-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented4 reports
- Continuum® Trabecular Metal™ 00-8757-060-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented4 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page