Medtronic Launcher Coronary Guide Catheter recallZ-0205-2015 · Medtronic Vascular · initiated Oct 13, 2014
Class II Terminated All recalls posted in November 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Vascular or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Two reports have been received on the 6F launcher Guide Catheter lot (Model Number LA6JR40K; lot No 0007244836} for incorrect usable length of the Guide Catheter. It was reported that the catheters had a usable length of 90 cm instead of 55 cm, which was specified on the packaging label. Usable length is defined as the total length of catheter shaft from the distal end of the strain relief to the
FDA root cause category: Process control
What the company is doing
Beginning October 13, 2014, Medtronic field representatives will being hand delivery of the customer communication to affected customers, notifying them of the issue, and requesting return of unaffected product. Customers will be asked to sign a certificate confirming they have received the notification and have taken the requested action. 1. Identify and quarantine all listed, unused product in your inventory. 2. Return all listed product in your inventory to Medtronic. Contact Customer Service at 888-283-7868 to initiate a product return and credit. Your Medtronic sales representative can assist you in the return of this product as necessary. 3. Complete the attached Customer Confirmation Certificate and email it to [email protected] or fax it to Medtronic at 651-367-0612 to the attention of Customer Focused Quality.
Recall details
- Recall number
- Z-0205-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 13, 2014
- Posted by the FDA
- Nov 13, 2014
- Recalling firm
- Medtronic Vascular
- Quantity
- 11 devices
- Product code
- DQY
- FDA submissions
- K021256
- Distribution
- US Distribution including the states of NC, IA, IL and KS.
Product description
Medtronic Launcher Coronary Guide Catheter; MODEL: LA6JR40K
Code information (models, lots, serial numbers)
Lot Number: 0007244836 MODEL: LA6JR40K
Other devices under the same FDA submission
50 devices cleared under K021256 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Launcher LA6JR40Medtronic, Inc. · CATHETER, PERCUTANEOUS10 reports
- Launcher LA6IMAKMedtronic, Inc. · CATHETER, PERCUTANEOUS9 reports
- Launcher LA6EBU35Medtronic, Inc. · Catheter, percutaneous9 reports
- Launcher LA6EBU35Medtronic, Inc. · CATHETER, PERCUTANEOUS1 death5 reports
- LAUNCHER LA6EBU35Medtronic, Inc. · Catheter, percutaneous5 reports
- Launcher LA6JR35Medtronic, Inc. · CATHETER, PERCUTANEOUS4 reports
- Launcher LA6JL35Medtronic, Inc. · CATHETER, PERCUTANEOUS4 reports
- Launcher LA6AL75Medtronic, Inc. · CATHETER, PERCUTANEOUS4 reports
- Launcher LA6JL40Medtronic, Inc. · CATHETER, PERCUTANEOUS3 reports
- Launcher LA6AR10Medtronic, Inc. · CATHETER, PERCUTANEOUS3 reports
- Launcher LA6JR35SHMedtronic, Inc. · CATHETER, PERCUTANEOUS2 reports
- Launcher LA6MP1SHMedtronic, Inc. · CATHETER, PERCUTANEOUS2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page