HeartWare Ventricular Assist System (HeartWare Controller) recallZ-1131-2015 · HeartWare Inc · initiated Jan 16, 2015

Class I Terminated All recalls posted in February 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from HeartWare Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The affected clinical trial Controllers exhibit a higher susceptibility to ESD than newer commercial Controllers.

FDA root cause category: Device Design

What the company is doing

HeartWare sent an URGENT MEDICAL DEVICE RECALL dated January 16, 2015 asking clinicians to quarantine the affected controllers and complete and return the acknowledgement form. HeartWare also asks clinicians to identify patients with affected controllers and review the applicable risks with the patient and as soon as reasonably possible if medically advisable, exchange the affected controller under medical supervision with a new controller. HeartWare also asks that for each patient using an affected controller as their backup to contact the patient and arrange to have the backup controller replaced with a new controller. Lastly, HeartWare asks the firm to complete the patient tracking form and return all quarantined controllers to HeartWare along with the completion form. Should you have any questions or concerns, please contact your local HeartWare representative. Our 24-Hour Clinical Support personnel are also available at 1-888-494-6365.

Recall details

Recall number
Z-1131-2015
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Jan 16, 2015
Posted by the FDA
Feb 25, 2015
Recalling firm
HeartWare Inc
Quantity
4,845
Product code
DSQ
FDA submissions
P100047
Distribution
Worldwide Distribution - US Nationwide in the states of: AL, AZ, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NY, OH, OR, PA, TX, UT, VA, WA and District of Columbia and countries of: Canada, Poland, Australia, Austria, Italy, DE, New Zealand, South Africa, France, Switzerland, Germany, Brazil, UK, Belgium, Netherlands, Israel, Malaysia, Turkey, Greece, Japan, Singapore, Poland, Lithuania, Sweden, Norway, Finland, and India.

Product description

HeartWare Ventricular Assist System (HeartWare Controller) Product Usage: For use as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. The HeartWare System is designed for in-hospital and out-of-hospital settings, including transportation via fix-wing aircraft or helicopter.

Code information (models, lots, serial numbers)

Product Codes- US product: Model No.: 1400, 1401US; International product: Model No.: 1400, 1401XX ('XX' represents country designation) Serial No.: CON000001 through CON005472

Other devices under the same FDA submission

33 devices cleared under P100047 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.