HeartWare Ventricular Assist System (HVAD) recallZ-1698-2015 · HeartWare Inc · initiated Apr 29, 2015

Class I Open, Classified All recalls posted in June 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from HeartWare Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

HeartWare has received complaints relating to damage or bent connection pins within the power supply ports of the VAD controller.

FDA root cause category: Device Design

What the company is doing

HeartWare sent an Urgent Medical Device Correction letter to customers on May 25, 2015. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: 1. Review the enclosed notice and Patient Communication. 2. Forward the notice to those individuals within their organization who need to be aware of its contents. 3. Identify the patients currently supported by the HVAD System. 4. Distribute the "Patient Communication" to their patients directly via FedEx or some other reliable means of communication. 5. Continue to reinforce the messages set forth in the notice with their patients during their regularly scheduled appointments. 6. Complete, sign and return the Acknowledgement and Completion Form to HeartWare within 30 days of receipt of this letter. For questions contact your local HeartWare representative or the 24-Hour Clinical Support personnel at 1-888-494-6365.

Recall details

Recall number
Z-1698-2015
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Apr 29, 2015
Posted by the FDA
Jun 8, 2015
Recalling firm
HeartWare Inc
Quantity
3,747 currently implanted worldwide
Product code
DSQ
FDA submissions
P100047
Distribution
Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA and countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Korea, Lebanon, Lithuania, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, and UK.

Product description

HeartWare Ventricular Assist System (HVAD). Catalog #'s US: 1101 and 1103 / OUS: 1100, 1101,1102,1104,1104JP and 1205. Product Usage: For use as a bridge-to-cardiac transplantation in patient who are at risk of death from refractory end-stage left ventricular heart failure.

Code information (models, lots, serial numbers)

Serial #: ALL HeartWare HVAD systems currently in use.

Other devices under the same FDA submission

33 devices cleared under P100047 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.