HeartWare Ventricular Assist System (HVAD) recallZ-1698-2015 · HeartWare Inc · initiated Apr 29, 2015
Class I Open, Classified All recalls posted in June 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from HeartWare Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
HeartWare has received complaints relating to damage or bent connection pins within the power supply ports of the VAD controller.
FDA root cause category: Device Design
What the company is doing
HeartWare sent an Urgent Medical Device Correction letter to customers on May 25, 2015. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: 1. Review the enclosed notice and Patient Communication. 2. Forward the notice to those individuals within their organization who need to be aware of its contents. 3. Identify the patients currently supported by the HVAD System. 4. Distribute the "Patient Communication" to their patients directly via FedEx or some other reliable means of communication. 5. Continue to reinforce the messages set forth in the notice with their patients during their regularly scheduled appointments. 6. Complete, sign and return the Acknowledgement and Completion Form to HeartWare within 30 days of receipt of this letter. For questions contact your local HeartWare representative or the 24-Hour Clinical Support personnel at 1-888-494-6365.
Recall details
- Recall number
- Z-1698-2015
- Recall class
- Class I (most serious)
- Status
- Open, Classified
- Initiated by the company
- Apr 29, 2015
- Posted by the FDA
- Jun 8, 2015
- Recalling firm
- HeartWare Inc
- Quantity
- 3,747 currently implanted worldwide
- Product code
- DSQ
- FDA submissions
- P100047
- Distribution
- Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA and countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Korea, Lebanon, Lithuania, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, and UK.
Product description
HeartWare Ventricular Assist System (HVAD). Catalog #'s US: 1101 and 1103 / OUS: 1100, 1101,1102,1104,1104JP and 1205. Product Usage: For use as a bridge-to-cardiac transplantation in patient who are at risk of death from refractory end-stage left ventricular heart failure.
Code information (models, lots, serial numbers)
Serial #: ALL HeartWare HVAD systems currently in use.
Other devices under the same FDA submission
33 devices cleared under P100047 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HeartWare® Controller 1403USHeartware Inc. · Ventricular (Assisst) Bypass3 recalls10 deaths1,495 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assisst) Bypass1 recall1,126 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass2 recalls606 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall615 reports
- HVAD® Pump Implant Kit MCS1705PUHeartware Inc. · Ventricular (Assist) Bypass7 recalls36 deaths203 reports
- HVAD® Controller 1403USHeartware Inc. · Ventricular (Assist) Bypass548 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall487 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assist) Bypass1 recall281 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall159 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page