Why it was recalled

Reports where patients who recently switched from the current Shiley Neonatal and Pediatric products to the affected products experienced discomfort and in some cases, breathing difficulties, corresponding to a negative effect on oxygen levels.

FDA root cause category: Device Design

What the company is doing

Consignees were notified 5/8/15 via letter. Consignees were asked to return products.

Recall details

Recall number
Z-1951-2015
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
May 8, 2015
Posted by the FDA
Jul 9, 2015
Recalling firm
Covidien LP (formerly Nellcor Puritan Bennett Inc.)
Quantity
69,461 units total
Product code
JOH
FDA submissions
K122531
Distribution
Worldwide Distribution. US nationwide, Canada, Australia, Belgium, Chile, Germany, Israel, Italy, Japan, Saudi Arabia, Singapore, South Africa, Turkey, and Uruguay.

Product description

Shiley Neonatal, Pediatric and Long Pediatric Tracheostomy Tube with TaperGuard Cuff. Intended for use in providing tracheal access for airway managements.

Code information (models, lots, serial numbers)

All products manufactured since 2/2012.

Other devices under the same FDA submission

26 devices cleared under K122531 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.