ELI 380 Electrocardiograph recallZ-1796-2015 · Mortara Instrument, Inc · initiated May 29, 2015
Class II Terminated All recalls posted in June 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Mortara Instrument, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Mortara Instrument, Inc. has recently become aware of a potential safety hazard involving our ELI 380 electrocardiograph. When used in a particular workflow, acquired ECG waveforms for one patient may become associated with the patient demographics for a different patient when the record is transmitted to a records management system.
FDA root cause category: Software design
What the company is doing
Mortara Instrument, Inc. sent an "Urgent Medical Device Correction" letter dated May 29, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised that users in their facilities be made aware of the safety notification and that required actions listed on the letter be taken immediately. A Field service Engineer will be dispatched to their facility to update the affected unit(s) on-site. For questions they can contact Technical support at 1-888-667-8272 or [email protected] or contact your local service Representative.
Recall details
- Recall number
- Z-1796-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 29, 2015
- Posted by the FDA
- Jun 12, 2015
- Recalling firm
- Mortara Instrument, Inc
- Quantity
- 176
- Product code
- DPS
- FDA submissions
- K142105
- Distribution
- Worldwide Distribution - US Distribution to the states of : AZ, CA, FL, GA, IL, MI, NE, SC, UT, WA and WI., and to the countries of : CHINA, PORTUGAL, CANADA, OMAN, MALAYSIA, NEW ZEALAND, INDIA, POLAND, LEBANON, SWITZERLAND, FINLAND, ROMANIA,UNITED ARAB EMIRATES, UNITED KINGDOM, AUSTRALIA, THE NETHERLANDS, ITALY, FRANCE, GERMANY, THAILAND,SAUDI, ARABIA, NORWAY and IRELAND.
Product description
ELI 380 Electrocardiograph. The ELI 380 is intended to be a high-performance, multichannel resting electrocardiograph. As a resting electrocardiograph, the ELI 380 simultaneously acquires data from each lead. Once the data is acquired, it can be analyzed, reviewed, stored, printed or transmitted. It is a device primarily intended for use in hospitals, but may be used in medical clinics and offices of any size.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Serial Number 114470237158 114470237159 114470237163 114470237169 114470237173 114470237176 114470237191 115040243225 115040243448 115040243449 115130250975 115130250976 115130250977 115130250978 115130250979 115130250980 115130250981 115130250982 115130250983 114470237160 114470237165 114470237172 114470237174 114470237178 114470237179 114470237184 114470237185 114470237188 114470237190 114470237161 114470237162 114470237168 114470237177 114470237183 114470237186 114470237187 115040243087 115140252395 115140252396 115150252749 115160253501 114470237164 114470237167 114470237171 114470237175 115040243229 115070245110 114470237166 114470237170 114470237180 114470237181 114470237182 114470237189 114370228740 115110250032 115110250033 115110250034 114110201576 114110201578 115140252393 115140252394 115140252399 115140252397 114360227784 115190256009 115190256010 115190256011 115140252400 115150252442 Serial Number 115170254254 114530241364 115160253238 114530241295 115070245550 115070245551 115100247989 115100247990 115190255287 115030241850 114500239441 115100247699 114510240312 114510240313 115020241387 115030241876 115040243081 114490238052 114490238053 114490238054 114490238055 114490238056 114490238057 114490238058 114490238066 114510240310 115030241851 115030241852 115030241853 115030241854 115030241855 115030241856 115030241857 115030241858 115030241859 115030241860 115030241861 115030241862 115030241863 115030241864 115030241865 115040243079 115040243080 115060244254 115060244255 115060244256 115060244257 115060244258 115160253181 115160253182 115160253183 115190255734 115190255735 115160254106 114510240309 114490238556 114490238557 114490238558 114490238559 114490238560 114490238561 115060244381 115060244382 115060244383 115060244384 115060244385 115140252387 115140252388 115140252389 115040243082 115040243085 115040243086 115120250253 115120250254 115120250255 115120250256 115170254250 115170254251 115170254252 115170254253 114360227772 114360227773 115020241385 115030242164 115120250688 115130250965 115130250966 115130250967 115130250968 114110201950 114110201952 114530241293 114530241294 114530241296 115020241383 115080245620 115110249768 115110249769 115110249770 115110249771 115110249772 115060244864 115160253239 114510240311 115020241384 115020241386 114510240314
Other devices under the same FDA submission
11 devices cleared under K142105 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ELI 380 ELI380-DCS11Mortara Instrument, Inc. · Electrocardiograph1 recall1 death70 reports
- ELI 380 ELI380-DCX11Mortara Instrument, Inc. · Electrocardiograph1 recall12 reports
- ELI 380 ELI380-ACX11Mortara Instrument, Inc. · Electrocardiograph1 recall8 reports
- ELI 380 ELI380-DCX21Mortara Instrument, Inc. · Electrocardiograph1 recall6 reports
- ELI 380 ELI380-DCS21Mortara Instrument, Inc. · Electrocardiograph1 recall5 reports
- ELI 380 ELI380-ACX21Mortara Instrument, Inc. · Electrocardiograph1 recall3 reports
- ELI 380 ELI380-ACS11Mortara Instrument, Inc. · Electrocardiograph1 recall2 reports
- ELI 380 ELI380-DCS43Mortara Instrument, Inc. · Electrocardiograph1 recall1 report
- ELI 380 ELI380-DAX3XMortara Instrument, Inc. · Electrocardiograph1 recall0 reports
- ELI 380 ELI380-DCX43Mortara Instrument, Inc. · Electrocardiograph1 recall0 reports
- ELI 380 ELI380-ACS21Mortara Instrument, Inc. · Electrocardiograph2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.