MAGNETOM systems, magnetic resonance diagnostic devices… recallZ-0467-2016 · Siemens Medical Solutions USA, Inc · initiated Dec 7, 2015
Class II Terminated All recalls posted in December 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Possible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the helium gas to be blown directly into the magnet room or other areas. This could lead to displacement of oxygen, cause cold burns, and/or pose a risk to users, patient or others.
FDA root cause category: Other
What the company is doing
Siemens sent an Customer Advisory letter dated December 7, 2015, certified mail to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter notifies customers of the potential injuries, which may result of improperly installed quench lines. The letter indicates it is the system's owner who is responsible for installing the quench line per the requirements. If Siemens installed the quench lines, the consignee is directed to contact Siemens, who will inspect the quench lines during next maintenance visit. Siemens will monitor service reports for affected sites. For further questions please call ( 610) 219-6300.
Recall details
- Recall number
- Z-0467-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 7, 2015
- Posted by the FDA
- Dec 22, 2015
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- in US 3752 units
- Product code
- LNH
- FDA submissions
- K002179, K032428, K130885, K132951, K133226
- Distribution
- Nationwide including : DC, PR, Virgin Islands and Bermuda
Product description
MAGNETOM systems, magnetic resonance diagnostic devices (MRDD).
Code information (models, lots, serial numbers)
all units distributed since 1991
Other devices under the same FDA submission
7 devices cleared under K002179, K032428, K130885, K132951, K133226 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- MAGNETOM Skyra 10432915Siemens Healthcare GmbH · System, nuclear magnetic resonance imaging2 recalls8 reports
- MAGNETOM Skyra 10849580Siemens Healthcare GmbH · Coil, magnetic resonance, specialty2 recalls2 reports
- MAGNETOM Aera 10432914Siemens Healthcare GmbH · System, nuclear magnetic resonance imaging1 recall25 reports
- MAGNETOM Avanto 10849579Siemens Healthcare GmbH · System, nuclear magnetic resonance imaging1 recall1 report
- MAGNETOM Avanto 10849578Siemens Healthcare GmbH · System, nuclear magnetic resonance spectroscopic1 recall0 reports
- Biograph mMR 10433372Siemens Healthcare GmbH · Tomographic imager combining emission computed tomography with nuclear magnetic resonance1 recall0 reports
- MAGNETOM Avanto 10684332Siemens Healthcare GmbH · System, nuclear magnetic resonance imaging3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.