AVEA GDE/UIM upgrade kit, 12283-PMN recallZ-1898-2016 · Carefusion 211 Inc dba Carefusion · initiated May 17, 2016
Class I Terminated All recalls posted in June 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Carefusion 211 Inc dba Carefusion or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
FDA root cause category: Under Investigation by firm
What the company is doing
CareFusion sent a Field Safety Notice dated May 17, 2016, to all affected customers. The notice informed customers that CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM). This malfunction creates a blank AVEA Ventilator UIM and INOP condition during pre-use check or during use with cessation of ventilation to the patient. By design an audible clinical alarm is activated. Customers were informed of the problems identified and the actions to be taken. Customers were not required to return the affected devices or suspend use, customers would receive an copy of the identified affected serial numbers, as well as a response card, customers were requested to promptly return the response card to expedite the correction process and acknowledge receipt of the notification, Customers would be contacted by CareFusion to arrange onsite remediation of the affected devices, in the interim if any AVEA ventilator unit in their facility exhibits a blank AVEA Ventilator UIM and INOP condition during pre-use check or during use with cessation of ventilation, immediately remove the ventilator from service, provide alternate ventilation and contact CareFusion Technical Support per the contact information listed below to report the issue. Customers with questions should call 888-562-6016, for technical support 800-231-2466 or [email protected].
Recall details
- Recall number
- Z-1898-2016
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- May 17, 2016
- Posted by the FDA
- Jun 13, 2016
- Recalling firm
- Carefusion 211 Inc dba Carefusion
- Quantity
- 922 units total (501 units in US)
- Product code
- CBK
- FDA submissions
- K013642, K073069, K103211
- Distribution
- Worldwide Distribution - US (Nationwide) and the countries of Australia, Hungary, Paraguay, Belgium, Indonesia, Philippines, Bolivia, Ireland, Poland, Bermuda, India, Saudi Arabia, Canada, Japan, Singapore, Chile, Jordon, South Africa, China, Kuwait, Taiwan, Ecuador, Malaysia, Thailand, Egypt, Mexico, Turkey, France, Netherlands, United Kingdom, Guatemala, and Nicaragua.
Product description
AVEA GDE/UIM upgrade kit, 12283-PMN A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxygen in the breathing gas.
Code information (models, lots, serial numbers)
n/a
Other devices under the same FDA submission
12 devices cleared under K013642, K073069, K103211 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- AVEA 2 VENTILATOR 32029-001Vyaire Medical, Inc. · Ventilator, Continuous, Facility Use1 death300 reports
- Revel 19260-001Vyaire Medical, Inc. · Ventilator, Continuous, Facility Use5 reports
- AVEA™ 17312-24Vyaire Medical, Inc. · Ventilator, continuous, facility use4 reports
- AVEA™ 17310-00Vyaire Medical, Inc. · Ventilator, continuous, facility use4 reports
- Avea 17312-00Vyaire Medical, Inc. · Ventilator, Continuous, Facility Use4 reports
- Avea 32029-001Vyaire Medical, Inc. · Ventilator, Continuous, Facility Use4 reports
- AVEA 17220Vyaire Medical, Inc. · Ventilator, continuous, facility use3 reports
- AVEA II 32193-001Vyaire Medical, Inc. · Ventilator, Continuous, Facility Use2 reports
- Vela 16530-00Vyaire Medical, Inc. · Ventilator, Continuous, Facility Use1 report
- AVEA 50000-40033Carefusion Corporation · Ventilator, Continuous, Facility Use1 report
- AVEA 12283-PMNCarefusion Corporation · Ventilator, Continuous, Facility Use1 report
- AVEA II 32078-001Vyaire Medical, Inc. · Ventilator, Continuous, Facility Use1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 74111).
- Z-1888-2016 · TerminatedAVEA Comprehensive ventilator, 17310-xx (xx = 0 - 14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predeter…
- Z-1889-2016 · TerminatedAVEA Standard ventilator, 17311-xx (xx = 0 - 14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined …
- Z-1890-2016 · TerminatedAVEA Standard with Compressor ventilator, 17312-xx (xx = 0 - 14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering …
- Z-1891-2016 · TerminatedAVEA Comprehensive ventilator-refurbished, R17310-xx (xx = 0-14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering …
- Z-1892-2016 · TerminatedAVEA Standard ventilator-refurbished, R17311-xx (xx = 0-14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a pred…
- Z-1893-2016 · TerminatedAVEA Standard with Compressor ventilator-refurbished, R17312-xx (xx = 0-14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by d…
- Z-1894-2016 · TerminatedTCA Board, 16542A A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxygen in the brea…
- Z-1895-2016 · TerminatedGas Delivery Engine (GDE)-1st Generation, 16222-001-99 A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predeterm…
- Z-1896-2016 · TerminatedGas Delivery Engine (GDE), 16650A A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of ox…
- Z-1897-2016 · TerminatedGas Delivery Engine (GDE)-refurbished, R16650A A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined per…
- Z-1899-2016 · TerminatedAVEA GDE upgrade kit, 12284-PMN A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxyg…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page