HIgH-Flo Subcutaneous Safety Needle Sets, 2-Needle Set/ 24… recallZ-0318-2018 · Repro-Med Systems, Inc. · initiated Mar 10, 2016
Class II Terminated All recalls posted in January 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Repro-Med Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
On 2/24/2016 during post sterilization inspection of part number F180 Precision Flow Rate Tubing, Lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the supplier s manufacturing seal for Part number #317036 5 X 6.5 ST Bag Combo. the firm decided to recall Precision Flow Rate Tubing and RMS HIgH-Flo Subcutaneous Safety Needles Sets due to the sterile barrier packaging being found to have gaps during post-sterilization inspection of one lot of flow tubing, in which 701 out of 3,195 units were rejected for not passing inspection. The firm determined that the issue was large than just the one lot and recalled the product via March 10, 2016 Voluntary Medical Device Corrections and Removal Notification to customers. The recall was conducted without FDA notification.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On March 10, 2016 RMS Medical Products distributed Urgent Voluntary Medical Device Corrections and Removal notices to their customers. On March 12, 2016 & on March 18, 2016, RMS Medical Products distributed a 2nd Urgent Voluntary Medical Device Corrections and Removal notices to their customers advising them that affected products list was reduced after a comprehensive risk assessment was performed. Customers are advised to check stock, quarantine, cease distribution and remove product from use. If customers further distributed product, they are also encouraged to inform the affected parties. Customers are encouraged to complete and return the acknowledgement form via fax to: 845-469-5518.
Recall details
- Recall number
- Z-0318-2018
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Mar 10, 2016
- Posted by the FDA
- Jan 5, 2018
- Recalling firm
- Repro-Med Systems, Inc.
- Product code
- FPA
- FDA submissions
- K122404
- Distribution
- Nationwide in US
Product description
HIgH-Flo Subcutaneous Safety Needle Sets, 2-Needle Set/ 24 Gauge 9mm Needles, REF/Part Numbers RMS22606, RMS22609, RMS22612, RMS12606, RMS22409, RMS32609, RMS32609, RMS12409, RMS12609, RMS62612, RMS62606, RMS12406, RMS22412, RMS12412, RMS42606, RMS62609, RMS52612, RMS42612, RMS32609, RMS22609, RMS22606, RMS42609 & RMS32612.
Code information (models, lots, serial numbers)
Lot Numbers: 7.069/16, 7.068/16, 7.066/16, 7.065/16, 7.064/16, 7.063/16, 7.062/16, 7.061/16, 7.060/16, 7.059/16, 7.058/16, 7.057/16, 7.056/16, 7.055/16, 7.054/16, 7.053/16, 7.052/16, 7.051/16, 7.050/16, 7.049/16, 7.048/16 & 7.046/16. Expiration: 2019-02
Other devices under the same FDA submission
11 devices cleared under K122404 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- HIgH-Flo™ Needle Set RMS22606Repro Med Systems, Inc. · Set, administration, intravascular1 recall0 reports
- HIgH-Flo™ Needle Set RMS22609Repro Med Systems, Inc. · Set, administration, intravascular2 reports
- HIgH-Flo™ Needle Set RMS52609Repro Med Systems, Inc. · Set, administration, intravascular2 reports
- HIgH-Flo™ Needle Set RMS12414Repro Med Systems, Inc. · Set, administration, intravascular2 reports
- HIgH-Flo™ Needle Set RMS32409Repro Med Systems, Inc. · Set, administration, intravascular2 reports
- HIgH-Flo™ Needle Set RMS32614Repro Med Systems, Inc. · Set, administration, intravascular1 report
- HIgH-Flo™ Needle Set RMS12409Repro Med Systems, Inc. · Set, administration, intravascular1 report
- HIgH-Flo™ Needle Set RMS22409Repro Med Systems, Inc. · Set, administration, intravascular1 report
- HIgH-Flo™ Needle Set RMS22414Repro Med Systems, Inc. · Set, administration, intravascular1 report
- HIgH-Flo™ Needle Set RMS32414Repro Med Systems, Inc. · Set, administration, intravascular1 report
- HIgH-Flo™ Needle Set RMS42412Repro Med Systems, Inc. · Set, administration, intravascular1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 76569).
- Z-0317-2018 · TerminatedFreedom60 Precision Flow Rate Tubing Sets, REF/Part Numbers F900, F2400, F60, F10, F275, F600, F45, F1200 F180.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page