1.5T Signa HDx, 3.0T Signa HDx, 1.5T Signa HDxt, 3.0T Signa… recallZ-3027-2017 · GE Medical Systems, LLC · initiated May 11, 2017
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
FDA root cause category: Mixed-up of materials/components
What the company is doing
Consignees received a written recall notice dated 05/17/2017 notifying them of a correction to all units using an Uninterruptable Power Supply. Consignees were instructed not to unplug the unit and to continue normal use. Units with the Eaton 9130 Uninterruptable Power Supply connected to the MR operator console would receive a correction at no charge. This would involve possibly replacing the power cable. GE Representatives will be contacting consignees to arrange corrections. Any questions can be directed to GE Healthcare Service at 1-800-437-1171 or your local service representative.
Recall details
- Recall number
- Z-3027-2017
- Status
- Terminated
- Initiated by the company
- May 11, 2017
- Posted by the FDA
- Aug 2, 2017
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 68 units in total
- Product code
- LNH
- FDA submissions
- K052293
- Distribution
- US Nationwide in the states of : AK, CA, CO, CT, DE, FL, IA, IL, IN, KS, KY, LA, MD, MO, ND, NE, NJ, NY, OH, PA, TX, WA, and WI
Product description
1.5T Signa HDx, 3.0T Signa HDx, 1.5T Signa HDxt, 3.0T Signa HDxt, Signa HDi, Signa Vibrant Product Usage: The GE Signa HDx MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa HDx MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa eHDx system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Code information (models, lots, serial numbers)
Medical Device Listing No. D117354, D295466, D182389, D102432, D206140, D230310, D269569, D102430, D140572, D188736, D102434, D234841, D196247, D134019, D236543, D006275, D159181, D288890, D229859; Model No. Eaton 9130, 5323286, 5396502-5, 5499460-3, 5499460-6, 5499460-9, 5741000-3, 00840682115971, 5396502-2, 5499460-2, 5741000-2, 00840682115872, 5373011-3, 5396469-3, 5481490-2, 5481500-7, 5499460-7, 5499460-10, 5741000, 00840682103817, 5323286-101, 2377062-61, 2395001, 2395001-2, 2395001-3, 2395001-4, 00840682105699, 840682105378
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- NA Discovery MR750w 3.0TGE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging13 recalls0 reports
- NA SIGNA PET MRGE Medical Systems, L.L.C. · Tomographic imager combining emission computed tomography with nuclear magnetic resonance11 recalls1 report
- Not Applicable Optima MR450w 1.5TGE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging11 recalls0 reports
- Not Applicable Discovery MR750 3.0TGE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging10 recalls1 report
- Not Applicable 5494440GE Medical Systems, L.L.C. · Tomographic imager combining emission computed tomography with nuclear magnetic resonance9 recalls0 reports
Other devices under the same FDA submission
1 device cleared under K052293 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- NA 1.5T SPLIT HEAD COILGE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 77488).
- Z-3021-2017 · TerminatedBrivo MR355, Optima MR360 Product Usage: The Brivo MR355/Optima MR360 is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, a…
- Z-3022-2017 · TerminatedOptima MR450 1.5T Product Usage: 1 .5T Brivo MR355 and 1.5T Optima MR360 are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio…
- Z-3023-2017 · TerminatedOptima MR750 3.0T Product Usage: The Optima MR450 1.5T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to noise ratio, and shor…
- Z-3024-2017 · TerminatedOptima MR450w 1.5T Product Usage: The Optima MR450w 1.5T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and sho…
- Z-3025-2017 · TerminatedDiscovery MR750 3.0T Product Usage: Intended use: The Discovery MR750 3.0Tsystem is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noi…
- Z-3026-2017 · TerminatedSIGNA Creator, SIGNA ExplorerProduct Usage: 1.5T SIGNA Creator and 1.5T SIGNA Explorer is a whole body magnetic resonance scanner designed to support high resolution, high signal-t…
- Z-3028-2017 · Terminated1.5T Signa HDxt, 3.0T Signa HDxt, Signa HDi, Signa Vibrant Product Usage: The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed…
- Z-3029-2017 · TerminatedSigna PET/MR (K163619, K142098) Product Usage: The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that p…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page