Skip to content

Optima MR450 1.5T recallZ-3022-2017 · GE Medical Systems, LLC · initiated May 11, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

FDA root cause category: Mixed-up of materials/components

What the company is doing

Consignees received a written recall notice dated 05/17/2017 notifying them of a correction to all units using an Uninterruptable Power Supply. Consignees were instructed not to unplug the unit and to continue normal use. Units with the Eaton 9130 Uninterruptable Power Supply connected to the MR operator console would receive a correction at no charge. This would involve possibly replacing the power cable. GE Representatives will be contacting consignees to arrange corrections. Any questions can be directed to GE Healthcare Service at 1-800-437-1171 or your local service representative.

Recall details

Recall number
Z-3022-2017
Status
Terminated
Initiated by the company
May 11, 2017
Posted by the FDA
Aug 2, 2017
Recalling firm
GE Medical Systems, LLC
Quantity
68 units in total
Product code
LNH
FDA submissions
K083147, K132376, K142085, K160618
Distribution
US Nationwide in the states of : AK, CA, CO, CT, DE, FL, IA, IL, IN, KS, KY, LA, MD, MO, ND, NE, NJ, NY, OH, PA, TX, WA, and WI

Product description

Optima MR450 1.5T Product Usage: 1 .5T Brivo MR355 and 1.5T Optima MR360 are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the 1 .5T Brivo MR355 and I .5T Optima MR360 reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Code information (models, lots, serial numbers)

Medical Device Listing No. D117354, D295466, D182389, D102432, D206140, D230310, D269569, D102430, D140572, D188736, D102434, D234841, D196247, D134019, D236543, D006275, D159181, D288890, D229859; Model No. Eaton 9130, 5323286, 5396502-5, 5499460-3, 5499460-6, 5499460-9, 5741000-3, 00840682115971, 5396502-2, 5499460-2, 5741000-2, 00840682115872, 5373011-3, 5396469-3, 5481490-2, 5481500-7, 5499460-7, 5499460-10, 5741000, 00840682103817, 5323286-101, 2377062-61, 2395001, 2395001-2, 2395001-3, 2395001-4, 00840682105699, 840682105378

Devices named in this recall

5 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

10 devices cleared under K083147, K132376, K142085, K160618 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 77488).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page