Cardiac Resynchronization Therapy Defibrillators (CRT-Ds)… recallZ-0583-2018 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · initiated Dec 19, 2017
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
FDA root cause category: Process design
What the company is doing
On 12/19/2017, Medtronic verbally notified consignees and retrieved eight devices at sites that had affected inventory on hand. On 01/22/2018, Medtronic provided a recall notification letter to physicians following 48 patients implanted with affected devices and risk managers of those medical facilities, recommending that physicians strongly consider prophylactic device replacement for patients implanted with an affected device. Medtronic Technical Services is available to assist physicians with questions at 800-723-4636. Medtronic Patient Service is available to assist patients at 800-551-5544 (Monday-Friday, 8am-5pm Central Time). The recall was expanded on 03/15/2018 and customers were notified by letter. Affected product, not yet implanted, was retrieved on about 03/08/2018. Instructions in the physicians following patients implanted with the additional affected devices included prophylactic device replacement for those patients at higher risk, including patients whose clinical history indicates prior need for high-voltage therapy and/or for pacemaker-dependent patients. The physicians should carefully weigh the risks and benefits of device replacement. If it is determined that replacement is not warranted, instructions included the following: programming changes to reduce the potential for high-voltage charges associated with arrhythmia detection and therapies, such as enabling ATP before charging for fast ventricular rhythms or programming a separate fast VT via VF zone with ATP; Continue three-month in-clinic or remote follow-ups to verify device functionality because inability to interrogate a device or a failed remote monitoring transmission may be an indication that internal arcing has occurred; and to advice patients to seek medical attention immediately if they experience new or unexpected symptoms suspicious for a ventricular arrhythmia. Medtronic Patient Services is available to assist patients at 800-551-5544 (Monday-Friday, 8am-5pm Centra
Recall details
- Recall number
- Z-0583-2018
- Status
- Terminated
- Initiated by the company
- Dec 19, 2017
- Posted by the FDA
- Feb 1, 2018
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Quantity
- 951 devices total
- Product code
- NIK
- FDA submissions
- P010031S550, P010031S513
- Distribution
- US Nationwide Distribution.
Product description
Cardiac Resynchronization Therapy Defibrillators (CRT-Ds), Compia MRI. Labeled as the following: a. Compia MRI CRT-D DF1 (Product No. DTMC1D1); b. Compia MRI Quad CRT-D DF4 (Product No.DTMC1QQ ) Implantable devices that automatically detect and treat episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia. Some devices can also provide resynchronization therapy, including sequential biventricular pacing (CRT-D devices).
Code information (models, lots, serial numbers)
a. Product No. DTMC1D1: UDI 00643169705807 (Serial No. RPW200289H); b. Product No. DTMC1QQ: UDI 00643169565272 (Serial No. RPL201034H) (EXPANSION: Serial No. RPL200052H, RPL200534H, RPL201039H, RPL201059H, RPL201235H, RPL201236H, RPL201454H, RPL201645H, RPL201664H, RPL201234H)
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Compia MRI™ Quad CRT-D SureScan™ DTMC1QQMedtronic, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)3 recalls2 deaths59 reports
- Compia MRI™ CRT-D SureScan™ DTMC1D1Medtronic, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)2 recalls11 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 78888).
- Z-0582-2018 · TerminatedCardiac Resynchronization Therapy Defibrillators (CRT-Ds), Amplia MRI. Labeled as the following: a. Amplia MRI CRT-D DF4 (Product No. DTMB1D4); b. Amplia MRI Quad CRT-D DF1 (Pr…
- Z-0584-2018 · TerminatedCardiac Resynchronization Therapy Defibrillators (CRT-Ds), Viva. Labeled as the following: a. Viva Quad S CRT-D DF4 (Product No. DTBB1QQ); b. Viva Quad XT CRT-D DF4 (Product N…
- Z-0585-2018 · TerminatedImplantable Cardioverter Defibrillators (ICDs), Evera MRI. Labeled as the following: a. Evera MRI XT DR ICD DF1 (Product No. DDMB1D1); b. Evera MRI XT DR ICD DF4 (Product No. …
- Z-0586-2018 · TerminatedCardiac Resynchronization Therapy Defibrillators (CRT-Ds), Claria MRI. Labeled as the following: a. Claria MRI CRT-D DF1 (Product No. DTMA1D1); b. Claria MRI Quad CRT-D DF4 (Pr…
- Z-0587-2018 · TerminatedImplantable Cardioverter Defibrillators (ICDs), Evera. Labeled as the following: a. Evera S DR ICD DF1 (Product No. DDBC3D1); b. Evera S VR ICD DF1 (Product No. DVBC3D1); c. E…
- Z-0588-2018 · TerminatedImplantable Cardioverter Defibrillators (ICDs), Visia AF MRI. Labeled as the following: a. Visia AF MRI VR SureScan ICD DF1 (Product No. DVFB1D1); b. Visia AF MRI VR SureScan IC…
- Z-0589-2018 · TerminatedImplantable Cardioverter Defibrillators (ICDs), Visia AF. Labeled as: a. Visia AF VR ICD DF1 (Product No. DVAB1D1); EXPANSION: b. Visia AF" VR ICD DF4 (Product No. DVAB1D4) …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page