Skip to content

VIEW PLATING SYSTEM, REDUCE FRACTURE PLATING SYSTEM recallZ-2124-2018 · Orthofix, Inc · initiated Apr 6, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Orthofix, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

FDA root cause category: Device Design

What the company is doing

On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.

Recall details

Recall number
Z-2124-2018
Status
Terminated
Initiated by the company
Apr 6, 2018
Posted by the FDA
Jun 7, 2018
Recalling firm
Orthofix, Inc
Quantity
1574
Product code
HRS
FDA submissions
K140876
Distribution
U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,

Product description

Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) D 3.2MM X 10MM (002-E-32010) b) D 3.2MM X 12MM (002-E-32012) c) D 3.2MM X 14MM (002-E-32014) d) D 3.2MM X 16MM (002-E-32016) e) D 3.2MM X 18MM (002-E-32018) f) D 3.2MM X 20MM (002-E-32020) g) D 3.2MM X 22MM (002-E-32022) h) D 3.2MM X 24MM (002-E-32024) i) D 3.2MM X 26MM (002-E-32026) j) D 3.2MM X 28MM (002-E-32028) k) D 3.2MM X 30MM (002-E-32030) l) D 3.2MM X 32MM (002-E-32032) m) D 3.2MM X 34MM (002-E-32034) n) D 3.2MM X 36MM (002-E-32036) o) D 3.2MM X 38MM (002-E-32038) p) D 3.2MM X 40MM (002-E-32040) q) D 3.2MM X 42MM (002-E-32042) r) D 3.2MM X 44MM (002-E-32044) s) D 3.2MM X 46MM (002-E-32046) t) D 3.2MM X 48MM (002-E-32048) u) D 3.2MM X 50MM (002-E-32050)

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

(UDI)/Lot: a) (18056099646505) B1075433; b) (18056099646512) B1075434; c) (18056099646529) B1075844; d) (18056099646536) B1075432; e) (18056099646543) B1075436; f) (18056099646550) B1075826; g) (18056099646567) B1075837; h) (18056099646574) B1084804; i) (18056099646581) B1075835; j) (18056099646598) B1082233; k) (18056099646604) B1204272, B1075829; l) (18056099646611) B1204977, B1082245; m) (18056099646628) B1215471, B1204978, B1082247; n) (18056099646635) B1204979, B1084800; o) (18056099646642) B1204582, B1082248; p) (18056099646659) B1218842, B1204980; q) (18056099646666) B1218840, B1204271, B1089889; r) (18056099646673) B1218843, B1195750, B1083640; s) (18056099646680) B1204981, B1083837; t) (18056099646697) B1204982, B1083846; u) (18056099646703) B1204584, B1083852, B1071600, B1071601

Devices named in this recall

21 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

108 devices cleared under K140876 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 79819).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page