Skip to content

Box Tgglogowx1 Do Total Gastrectomiii recallZ-1862-2018 · Covidien Medtronic · initiated Apr 26, 2018

Class II Terminated All recalls posted in May 2018

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Covidien Medtronic or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

FDA root cause category: Process control

What the company is doing

On April 26, 2108, all consignees were issued were issued Urgent Medical Device Recall notices and Return Forms via Federal Express or certified mail. On May 22, 2018, all consignees received an UPDATED URGENT MEDICAL DEVICE RECALL notice. The updated notice included NEW product which has been processed under FDA Ref # 80423. The notice informs customers of the potential for use of a device with a missing sled and the actions they should take to discontinue use of the product and return all inventory of the affected lots to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. All in-control inventory and devices returned from customers will be reworked. **FOR INTERNATIONAL CUSTOMERS** Please be aware that the recalled units are packed within various convenience kits as identified on the Attachment B provided with your customer letter. Customers are also encouraged to EMAIL OR FAX THE ACKNOWLEDGEMENT FORM TO: - "Product purchased directly from Medtronic": [email protected] or fax to (800) 895-6140. - "Product purchased through distributor": [email protected] or fax it to (203) 492-7719.

Recall details

Recall number
Z-1862-2018
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 26, 2018
Posted by the FDA
May 16, 2018
Recalling firm
Covidien Medtronic
Quantity
169,775 units in total
Product code
GDW
FDA submissions
K111825
Distribution
Worldwide and US Nationwide

Product description

BOX TGGLOGOWX1 DO TOTAL GASTRECTOMIII Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and simultaneously divides the tissue between the two, triple-staggered rows. The size of the staples is determined by the selection of the 2.0 mm, 2.5 mm, 3.5 mm or 4.8 mm single use loading unit. The Endo GIA universal staplers will accommodate any of the single use loading unit sizes that are available in the 30 mm, 45 mm and 60 mm lines.

Code information (models, lots, serial numbers)

Lot 0214726876

Other devices under the same FDA submission

77 devices cleared under K111825 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 79974).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page