Aisys CS2 Anesthesia System. Sold under the following… recallZ-0412-2019 · GE Healthcare, LLC · initiated Sep 27, 2018

Class II Terminated All recalls posted in November 2018

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.

FDA root cause category: Software design

What the company is doing

GE Healthcare notified customers on about 09/27/2018 via "URGENT MEDICAL DEVICE CORRECTION" letter. The letter informed customers that they could continue to use the anesthesia device. The issue can only occur if the patient stops spontaneous breathing in PSVPro ventilation mode after completion of lung recruitment breaths. Stopping the Cycling Procedure resolves the issue. Continue to monitor patient as is normal for Apnea. A GE Healthcare representative will contact each customer to arrange for the correction. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Recall details

Recall number
Z-0412-2019
Recall class
Class II
Status
Terminated
Initiated by the company
Sep 27, 2018
Posted by the FDA
Nov 9, 2018
Recalling firm
GE Healthcare, LLC
Quantity
4,207 units total
Product code
BSZ
FDA submissions
K110213, K170872
Distribution
Worldwide distribution - US nationwide in the states of CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, ,MA, MD, MI, MN, MO, MT, NC, NE, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. Countries of Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea (Republic of), Kuwait, Macedonia, Mexico, Netherlands, New Zealand, Nigeria, Norway, Peru, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.

Product description

Aisys CS2 Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH FOR USA, AISYS 10.X TO 11.X UPGRADE KIT FRENCH, AISYS 10.X TO 11.X UPGRADE KIT GERMAN, AISYS 10.X TO 11.X UPGRADE KIT JAPANESE. Product Usage: The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.

Code information (models, lots, serial numbers)

Show the full code information (215k characters)

Model Number, UDI Number, (Serial Number):  1011-9050-000, 010084068210229221APWX0132811180608, (APWX01328);  1011-9050-000, 010084068210229221APWX0048411180208, (APWX00484);  1011-9050-000, 010084068210229221APWW0214911171214, (APWW02149); 1011-9050-000, 10084068210229221APWW0215011171214, (APWW02150); 1011-9050-000, 010084068210229221APWX0026111180118, (APWX00261); 1011-9050-000, 010084068210229221APWX0026211180118, (APWX00262); 1011-9050-000, 010084068210229221APWX0140911180618, (APWX01409); 1011-9050-000, 010084068210229221APWX0141011180618, (APWX01410); 1011-9050-000, 010084068210229221APWX0141111180618, (APWX01411); 1011-9050-000, 010084068210229221APWX0141211180618, (APWX01412); 1011-9050-000, 010084068210229221APWX0141311180618, (APWX01413); 1011-9050-000, 010084068210229221APWX0141411180618, (APWX01414); 1011-9050-000, 010084068210229221APWX0141511180618, (APWX01415); 1011-9050-000, 010084068210229221APWX0141611180618, (APWX01416); 1011-9050-000, 010084068210229221APWX0141711180618, (APWX01417); 1011-9050-000, 010084068210229221APWX0141811180618, (APWX01418); 1011-9050-000, 010084068210229221APWX0141911180618, (APWX01419); 1011-9050-000, 010084068210229221APWX0142011180618, (APWX01420); 1011-9050-000, 010084068210229221APWX0142111180618, (APWX01421); 1011-9050-000, 010084068210229221APWX0142211180618, (APWX01422); 1011-9050-000, 010084068210229221APWX0026311180118, (APWX00263); 1011-9050-000, 010084068210229221APWX0026411180118, (APWX00264); 1011-9050-000, 010084068210229221APWX0026511180118, (APWX00265); 1011-9050-000, 010084068210229221APWW0214011171214, (APWW02140); 1011-9050-000, 010084068210229221APWW0215611171214, (APWW02156); 1011-9050-000, 010084068210229221APWW0215711171214, (APWW02157); 1011-9050-000, 010084068210229221APWX0158911180703, (APWX01589);  1011-9050-000, 010084068210229221APWX0159011180703, (APWX01590);  1011-9050-000, 010084068210229221APWX0159111180703, (APWX01591);  1011-9050-000, 010084068210229221APWX0159211180703, (APWX01592);  1011-9050-000, 010084068210229221APWX0159311180703, (APWX01593);  1011-9050-000, 010084068210229221APWX0159411180703, (APWX01594);  1011-9050-000, 010084068210229221APWW0146411170823, (APWW01464); 1011-9050-000, 10084068210229221APWW0162511170927, (APWW01625); 1011-9050-000, 10084068210229221APWW0162611170927, (APWW01626); 1011-9050-000, 10084068210229221APWW0162811170927, (APWW01628); 1011-9050-000, 10084068210229221APWW0208811171206, (APWW02088); 1011-9050-000, 10084068210229221APWW0208911171206, (APWW02089); 1011-9050-000, 10084068210229221APWW0209011171206, (APWW02090); 1011-9050-000, 10084068210229221APWW0209111171212, (APWW02091); 1011-9050-000, 10084068210229221APWW0209211171212, (APWW02092); 1011-9050-000, 10084068210229221APWW0209311171212, (APWW02093); 1011-9050-000, 10084068210229221APWW0209411171212, (APWW02094); 1011-9050-000, 10084068210229221APWW0209511171212, (APWW02095); 1011-9050-000, 10084068210229221APWW0209611171212, (APWW02096); 1011-9050-000, 10084068210229221APWW0209711171212, (APWW02097); 1011-9050-000, 10084068210229221APWW0209811171212, (APWW02098); 1011-9050-000, 10084068210229221APWW0209911171212, (APWW02099); 1011-9050-000, 10084068210229221APWW0210011171212, (APWW02100); 1011-9050-000, 10084068210229221APWW0210111171212, (APWW02101); 1011-9050-000, 10084068210229221APWW0210211171212, (APWW02102); 1011-9050-000, 10084068210229221APWW0210311171212, (APWW02103); 1011-9050-000, 010084068210229221APWX0086711180328, (APWX00867); 1011-9050-000, 010084068210229221APWX0086811180328, (APWX00868); 1011-9050-000, 010084068210229221APWX0086911180328, (APWX00869); 1011-9050-000, 010084068210229221APWX0087011180328, (APWX00870); 1011-9050-000, 010084068210229221APWX0087111180328, (APWX00871); 1011-9050-000, 010084068210229221APWX0087211180328, (APWX00872); 1011-9050-000, 010084068210229221APWX0036911180131, (APWX00369); 1011-9050-000, 010084068210229221APWX0143411180618, (APWX01434); 1011-9050-000, 010084068210229221APWX0143511180618, (APWX01435); 1011-9050-000, 010084068210229221APWX0105311180427, (APWX01053); 1011-9050-000, 010084068210229221APWX0165911180716, (APWX01659); 1011-9050-000, 010084068210229221APWX0166011180716, (APWX01660); 1011-9050-000, 010084068210229221APWX0166111180716, (APWX01661); 1011-9050-000, 010084068210229221APWX0166211180716, (APWX01662); 1011-9050-000, 010084068210229221APWX0166311180716, (APWX01663); 1011-9050-000, 010084068210229221APWX0166411180716, (APWX01664); 1011-9050-000, 010084068210229221APWX0166511180716, (APWX01665); 1011-9050-000, 010084068210229221APWX0166611180716, (APWX01666); 2097675-001-US, Not Applicable, (4651698-2); 2097675-001-US, Not Applicable, (4666280-1); 2097675-001-US, Not Applicable, (4666282-1); 2097675-001-US, Not Applicable, (4666283-1); 2097675-001-US, Not Applicable, (4666288-1); 2097675-001-US, Not Applicable, (4666291-1); 2097675-001-US, Not Applicable, (4666300-1); 2097675-001-US, Not Applicable, (4666308-1); 2097675-001-US, Not Applicable, (4666314-1); 2097675-001-US, Not Applicable, (4666324-1); 2097675-001-US, Not Applicable, (4666335-1); 2097675-001-US, Not Applicable, (4666567-1); 1011-9050-000, 010084068210229221APWX0067511180315, (APWX00675); 1011-9050-000, 010084068210229221APWX0102611180417, (APWX01026); 1011-9050-000, 010084068210229221APWW0221511171219, (APWW02215); 1011-9050-000, 10084068210229221APWW0221611171219, (APWW02216); 1011-9050-000, 10084068210229221APWW0221711171219, (APWW02217); 1011-9050-000, 10084068210229221APWW0221811171219, (APWW02218); 1011-9050-000, 10084068210229221APWW0221911171219, (APWW02219); 1011-9050-000, 10084068210229221APWW0222011171219, (APWW02220); 1011-9050-000, 10084068210229221APWW0222111171219, (APWW02221); 1011-9050-000, 10084068210229221APWW0222211171219, (APWW02222); 1011-9050-000, 10084068210229221APWW0222311171219, (APWW02223); 1011-9050-000, 10084068210229221APWW0222411171219, (APWW02224); 1011-9050-000, 10084068210229221APWW0222511171219, (APWW02225); 1011-9050-000, 10084068210229221APWW0222611171219, (APWW02226); 1011-9050-000, 10084068210229221APWW0222711171219, (APWW02227); 1011-9050-000, 10084068210229221APWW0222811171219, (APWW02228); 1011-9050-000, 10084068210229221APWW0222911171219, (APWW02229); 1011-9050-000, 10084068210229221APWW0223011171219, (APWW02230); 1011-9050-000, 010084068210229221APWX0112211180507, (APWX01122); 1011-9050-000, 010084068210229221APWX0155711180627, (APWX01557); 1011-9050-000, 010084068210229221APWX0014411180105, (APWX00144); 1011-9050-000, 010084068210229221APWW0215111171214, (APWW02151); 1011-9050-000, 010084068210229221APWX0043411180206, (APWX00434); 1011-9050-000, 010084068210229221APWX0038711180201, (APWX00387); 1011-9050-000, 010084068210229221APWX0038811180201, (APWX00388); 1011-9050-000, 010084068210229221APWX0038911180201, (APWX00389); 1011-9050-000, 010084068210229221APWX0039011180201, (APWX00390); 1011-9050-000, 010084068210229221APWX0039111180201, (APWX00391); 1011-9050-000, 010084068210229221APWX0039211180201, (APWX00392); 1011-9050-000, 010084068210229221APWX0039311180201, (APWX00393); 1011-9050-000, 010084068210229221APWW0212811171212, (APWW02128); 1011-9050-000, 010084068210229221APWW0212911171212, (APWW02129); 1011-9050-000, 010084068210229221APWW0213011171212, (APWW02130); 1011-9050-000, 10084068210229221APWW0213111171214, (APWW02131); 1011-9050-000, 10084068210229221APWW0213211171214, (APWW02132); 1011-9050-000, 10084068210229221APWW0213311171214, (APWW02133);  1011-9050-000, 010084068210229221APWW0213411171214, (APWW02134);  1011-9050-000, 010084068210229221APWW0213511171214, (APWW02135);  1011-9050-000, 010084068210229221APWW0213611171214, (APWW02136);  1011-9050-000, 010084068210229221APWX0104411180427, (APWX01044);  1011-9050-000, 010084068210229221APWX0078811180323, (APWX00788);  1011-9050-000, 010084068210229221APWX0078911180323, (APWX00789);  1011-9050-000, 010084068210229221APWX0079011180323, (APWX00790);  1011-9050-000, 010084068210229221APWX0079111180328, (APWX00791);  1011-9050-000, 010084068210229221APWX0079211180328, (APWX00792);  1011-9050-000, 010084068210229221APWX0079311180328, (APWX00793);  1011-9050-000, 010084068210229221APWX0079411180328, (APWX00794);  1011-9050-000, 010084068210229221APWX0079511180328, (APWX00795);  1011-9050-000, 010084068210229221APWX0079611180328, (APWX00796);  1011-9050-000, 010084068210229221APWX0079711180328, (APWX00797);  1011-9050-000, 010084068210229221APWX0079811180328, (APWX00798);  1011-9050-000, 010084068210229221APWX0079911180328, (APWX00799);  1011-9050-000, 010084068210229221APWX0080011180328, (APWX00800);  1011-9050-000, 010084068210229221APWX0080111180328, (APWX00801);  1011-9050-000, 010084068210229221APWX0080211180328, (APWX00802);  1011-9050-000, 010084068210229221APWX0080311180328, (APWX00803);  1011-9050-000, 010084068210229221APWX0080411180328, (APWX00804);  1011-9050-000, 010084068210229221APWX0080511180328, (APWX00805);  1011-9050-000, 010084068210229221APWX0080611180328, (APWX00806);  1011-9050-000, 010084068210229221APWX0080711180328, (APWX00807);  1011-9050-000, 010084068210229221APWX0080811180328, (APWX00808);  1011-9050-000, 010084068210229221APWX0080911180328, (APWX00809);  1011-9050-000, 010084068210229221APWX0162811180711, (APWX01628);  1011-9050-000, 010084068210229221APWX0021611180115, (APWX00216);  1011-9050-000, 010084068210229221APWX0132311180608, (APWX01323);  1011-9050-000, 010084068210229221APWX0132411180608, (APWX01324);  1011-9050-000, 010084068210229221APWW0137311170802, (APWW01373);  1011-9050-000, 010084068210229221APWX0064811180309, (APWX00648);  1011-9050-000, 010084068210229221APWX0106011180427, (APWX01060);  1011-9050-000, 010084068210229221APWX0106111180427, (APWX01061);  1011-9050-000, 010084068210229221APWX0106211180427, (APWX01062);  1011-9050-000, 010084068210229221APWX0106311180427, (APWX01063);  1011-9050-000, 010084068210229221APWW0214111171214, (APWW02141);  1011-9050-000, 010084068210229221APWW0223111171219, (APWW02231);  1011-9050-000, 010084068210229221APWX0066111180309, (APWX00661);  1011-9050-000, 010084068210229221APWX0066211180309, (APWX00662);  1011-9050-000, 010084068210229221APWX0066311180309, (APWX00663);  1011-9050-000, 010084068210229221APWX0066411180309, (APWX00664);  1011-9050-000, 010084068210229221APWX0066511180309, (APWX00665);  1011-9050-000, 010084068210229221APWX0066611180309, (APWX00666);  1011-9050-000, 010084068210229221APWX0066711180309, (APWX00667);  1011-9050-000, 010084068210229221APWX0066811180309, (APWX00668);  1011-9050-000, 010084068210229221APWX0066911180309, (APWX00669);  1011-9050-000, 010084068210229221APWX0067011180309, (APWX00670);  1011-9050-000, 010084068210229221APWX0067111180315, (APWX00671);  1011-9050-000, 010084068210229221APWX0067211180315, (APWX00672);  1011-9050-000, 010084068210229221APWX0067311180315, (APWX00673);  1011-9050-000, 010084068210229221APWX0067611180315, (APWX00676);  1011-9050-000, 010084068210229221APWX0067711180315, (APWX00677);  1011-9050-000, 010084068210229221APWX0067811180315, (APWX00678);  1011-9050-000, 010084068210229221APWX0067911180315, (APWX00679);  1011-9050-000, 010084068210229221APWX0068011180315, (APWX00680);  1011-9050-000, 010084068210229221APWX0068111180315, (APWX00681);  1011-9050-000, 010084068210229221APWX0068211180315, (APWX00682);  1011-9050-000, 010084068210229221APWX0068311180315, (APWX00683);  1011-9050-000, 010084068210229221APWX0068411180315, (APWX00684);  1011-9050-000, 010084068210229221APWX0068511180315, (APWX00685);  1011-9050-000, 010084068210229221APWX0068611180315, (APWX00686);  1011-9050-000, 010084068210229221APWX0068711180315, (APWX00687);  1011-9050-000, 010084068210229221APWX0069111180315, (APWX00691);  1011-9050-000, 010084068210229221APWX0069211180315, (APWX00692);  1011-9050-000, 010084068210229221APWX0069311180315, (APWX00693);  1011-9050-000, 010084068210229221APWX0069411180315, (APWX00694);  1011-9050-000, 010084068210229221APWX0069511180315, (APWX00695);  1011-9050-000, 010084068210229221APWX0069611180315, (APWX00696);  1011-9050-000, 010084068210229221APWX0069711180315, (APWX00697);  1011-9050-000, 010084068210229221APWX0069811180315, (APWX00698);  1011-9050-000, 010084068210229221APWX0069911180315, (APWX00699);  1011-9050-000, 010084068210229221APWX0070011180315, (APWX00700);  1011-9050-000, 010084068210229221APWX0070111180315, (APWX00701);  1011-9050-000, 010084068210229221APWX0070211180315, (APWX00702);  1011-9050-000, 010084068210229221APWX0070311180315, (APWX00703);  1011-9050-000, 010084068210229221APWX0070411180315, (APWX00704);  1011-9050-000, 010084068210229221APWX0070511180315, (APWX00705);  1011-9050-000, 010084068210229221APWX0070611180315, (APWX00706);  1011-9050-000, 010084068210229221APWX0070711180315, (APWX00707);  1011-9050-000, 010084068210229221APWX0070811180315, (APWX00708);  1011-9050-000, 010084068210229221APWX0070911180315, (APWX00709);  1011-9050-000, 010084068210229221APWX0071011180315, (APWX00710);  1011-9050-000, 010084068210229221APWX0071311180319, (APWX00713);  1011-9050-000, 010084068210229221APWX0071411180319, (APWX00714);  1011-9050-000, 010084068210229221APWX0071511180319, (APWX00715);  1011-9050-000, 010084068210229221APWX0071611180319, (APWX00716);  1011-9050-000, 010084068210229221APWX0071711180319, (APWX00717);  1011-9050-000, 010084068210229221APWX0071811180319, (APWX00718);  1011-9050-000, 010084068210229221APWX0071911180319, (APWX00719);  1011-9050-000, 010084068210229221APWX0072011180319, (APWX00720);  1011-9050-000, 010084068210229221APWX0072111180319, (APWX00721);  1011-9050-000, 010084068210229221APWX0072211180319, (APWX00722);  1011-9050-000, 010084068210229221APWX0072311180319, (APWX00723);  1011-9050-000, 010084068210229221APWX0072411180319, (APWX00724);  1011-9050-000, 010084068210229221APWX0075111180323, (APWX00751);  1011-9050-000, 010084068210229221APWX0075211180323, (APWX00752);  1011-9050-000, 010084068210229221APWX0075311180323, (APWX00753);  1011-9050-000, 010084068210229221APWX0075411180323, (APWX00754);  1011-9050-000, 010084068210229221APWX0075511180323, (APWX00755);  1011-9050-000, 010084068210229221APWX0075611180323, (APWX00756);  1011-9050-000, 010084068210229221APWX0075711180323, (APWX00757);  1011-9050-000, 010084068210229221APWX0075811180323, (APWX00758);  1011-9050-000, 010084068210229221APWX0075911180323, (APWX00759);  1011-9050-000, 010084068210229221APWX0076011180323, (APWX00760);  1011-9050-000, 010084068210229221APWX0076111180323, (APWX00761);  1011-9050-000, 010084068210229221APWX0076211180323, (APWX00762);  1011-9050-000, 010084068210229221APWX0076311180323, (APWX00763);  1011-9050-000, 010084068210229221APWX0112111180507, (APWX01121);  1011-9050-000, 010084068210229221APWX0052911180222, (APWX00529);  1011-9050-000, 010084068210229221APWX0063411180309, (APWX00634);  1011-9050-000, 010084068210229221APWX0063511180309, (APWX00635);  1011-9050-000, 010084068210229221APWX0063611180309, (APWX00636);  1011-9050-000, 010084068210229221APWX0063711180309, (APWX00637);  1011-9050-000, 010084068210229221APWX0063811180309, (APWX00638);  1011-9050-000, 010084068210229221APWX0063911180309, (APWX00639);  1011-9050-000, 010084068210229221APWX0064011180309, (APWX00640);  1011-9050-000, 010084068210229221APWX0064111180309, (APWX00641);  1011-9050-000, 010084068210229221APWX0064211180309, (APWX00642);  1011-9050-000, 010084068210229221APWX0064311180309, (APWX00643);  1011-9050-000, 010084068210229221APWX0088111180328, (APWX00881);  1011-9050-000, 010084068210229221APWX0088211180328, (APWX00882);  1011-9050-000, 010084068210229221APWX0088311180328, (APWX00883);  1011-9050-000, 010084068210229221APWX0088411180328, (APWX00884);  1011-9050-000, 010084068210229221APWX0088511180328, (APWX00885);  1011-9050-000, 010084068210229221APWX0088611180328, (APWX00886);  1011-9050-000, 010084068210229221APWX0088711180328, (APWX00887);  1011-9050-000, 010084068210229221APWX0088811180328, (APWX00888);  1011-9050-000, 010084068210229221APWX0088911180328, (APWX00889);  1011-9050-000, 010084068210229221APWX0089011180328, (APWX00890);  1011-9050-000, 010084068210229221APWX0101311180417, (APWX01013);  1011-9050-000, 010084068210229221APWX0101411180417, (APWX01014);  1011-9050-000, 010084068210229221APWX0101511180417, (APWX01015);  1011-9050-000, 010084068210229221APWX0101611180417, (APWX01016);  1011-9050-000, 010084068210229221APWX0101711180417, (APWX01017);  1011-9050-000, 010084068210229221APWX0101811180417, (APWX01018);  1011-9050-000, 010084068210229221APWW0214211171214, (APWW02142);  1011-9050-000, 010084068210229221APWW0214311171214, (APWW02143);  1011-9050-000, 010084068210229221APWW0214411171214, (APWW02144);  1011-9050-000, 010084068210229221APWW0216411171214, (APWW02164);  1011-9050-000, 010084068210229221APWW0216511171214, (APWW02165);  1011-9050-000, 010084068210229221APWW0216611171214, (APWW02166);  1011-9050-000, 010084068210229221APWW0216711171214, (APWW02167);  1011-9050-000, 010084068210229221APWW0216811171214, (APWW02168);  1011-9050-000, 010084068210229221APWW0216911171214, (APWW02169);  1011-9050-000, 010084068210229221APWW0217011171214, (APWW02170);  1011-9050-000, 010084068210229221APWW0217111171218, (APWW02171);  1011-9050-000, 010084068210229221APWW0217211171218, (APWW02172);  1011-9050-000, 010084068210229221APWW0217311171218, (APWW02173);  1011-9050-000, 010084068210229221APWW0217411171218, (APWW02174);  1011-9050-000, 010084068210229221APWW0217511171218, (APWW02175);  1011-9050-000, 010084068210229221APWW0217611171218, (APWW02176);  1011-9050-000, 010084068210229221APWW0217711171218, (APWW02177);  1011-9050-000, 010084068210229221APWW0217811171218, (APWW02178);  1011-9050-000, 010084068210229221APWW0217911171218, (APWW02179);  1011-9050-000, 010084068210229221APWW0218011171218, (APWW02180);  1011-9050-000, 010084068210229221APWW0218111171218, (APWW02181);  1011-9050-000, 010084068210229221APWW0218211171218, (APWW02182);  1011-9050-000, 010084068210229221APWW0218311171218, (APWW02183);  1011-9050-000, 010084068210229221APWW0218411171218, (APWW02184);  1011-9050-000, 010084068210229221APWW0218511171218, (APWW02185);  1011-9050-000, 010084068210229221APWW0218611171218, (APWW02186);  1011-9050-000, 010084068210229221APWW0218711171218, (APWW02187);  1011-9050-000, 010084068210229221APWW0218811171218, (APWW02188);  1011-9050-000, 010084068210229221APWW0218911171218, (APWW02189);  1011-9050-000, 010084068210229221APWW0219011171218, (APWW02190);  1011-9050-000, 010084068210229221APWW0219111171218, (APWW02191);  1011-9050-000, 010084068210229221APWW0219211171218, (APWW02192);  1011-9050-000, 010084068210229221APWW0219311171218, (APWW02193);  1011-9050-000, 010084068210229221APWW0219411171218, (APWW02194);  1011-9050-000, 010084068210229221APWW0219511171218, (APWW02195);  1011-9050-000, 010084068210229221APWW0219611171218, (APWW02196);  1011-9050-000, 010084068210229221APWW0219711171218, (APWW02197);  1011-9050-000, 010084068210229221APWW0219811171218, (APWW02198);  1011-9050-000, 010084068210229221APWW0219911171218, (APWW02199);  1011-9050-000, 010084068210229221APWW0220011171218, (APWW02200);  1011-9050-000, 010084068210229221APWW0220111171218, (APWW02201);  1011-9050-000, 010084068210229221APWW0220211171218, (APWW02202);  1011-9050-000, 010084068210229221APWX0043711180206, (APWX00437);  1011-9050-000, 010084068210229221APWX0043811180206, (APWX00438);  1011-9050-000, 010084068210229221APWX0043911180206, (APWX00439);  1011-9050-000, 010084068210229221APWX0044011180206, (APWX00440);  1011-9050-000, 010084068210229221APWX0044111180206, (APWX00441);  1011-9050-000, 010084068210229221APWX0044211180206, (APWX00442);  1011-9050-000, 010084068210229221APWX0045211180206, (APWX00452);  1011-9050-000, 010084068210229221APWX0045311180206, (APWX00453);  1011-9050-000, 010084068210229221APWX0094211180410, (APWX00942);  1011-9050-000, 010084068210229221APWX0094311180410, (APWX00943);  1011-9050-000, 010084068210229221APWX0094411180410, (APWX00944);  1011-9050-000, 010084068210229221APWX0094511180410, (APWX00945);  1011-9050-000, 010084068210229221APWX0094611180410, (APWX00946);  1011-9050-000, 010084068210229221APWX0094711180410, (APWX00947);  1011-9050-000, 010084068210229221APWX0127911180604, (APWX01279);  1011-9050-000, 010084068210229221APWX0128011180604, (APWX01280);  1011-9050-000, 010084068210229221APWX0128111180604, (APWX01281);  1011-9050-000, 010084068210229221APWX0128211180604, (APWX01282);  1011-9050-000, 010084068210229221APWX0128311180604, (APWX01283);  1011-9050-000, 010084068210229221APWX0128411180604, (APWX01284);  1011-9050-000, 010084068210229221APWX0128511180604, (APWX01285);  1011-9050-000, 010084068210229221APWX0128611180604, (APWX01286);  1011-9050-000, 010084068210229221APWX0128711180604, (APWX01287);  1011-9050-000, 010084068210229221APWX0128811180604, (APWX01288);  1011-9050-000, 010084068210229221APWX0128911180604, (APWX01289);  1011-9050-000, 010084068210229221APWX0129011180604, (APWX01290);  1011-9050-000, 010084068210229221APWX0129111180604, (APWX01291);  1011-9050-000, 010084068210229221APWX0129211180604, (APWX01292);  1011-9050-000, 010084068210229221APWX0129311180604, (APWX01293);  1011-9050-000, 010084068210229221APWX0129411180604, (APWX01294);  1011-9050-000, 010084068210229221APWX0129511180604, (APWX01295);  1011-9050-000, 010084068210229221APWX0129611180604, (APWX01296);  1011-9050-000, 010084068210229221APWX0129711180604, (APWX01297);  1011-9050-000, 010084068210229221APWX0129811180604, (APWX01298);  1011-9050-000, 010084068210229221APWX0129911180604, (APWX01299);  1011-9050-000, 010084068210229221APWX0130011180604, (APWX01300);  1011-9050-000, 010084068210229221APWX0130111180608, (APWX01301);  1011-9050-000, 010084068210229221APWX0130211180608, (APWX01302);  1011-9050-000, 010084068210229221APWX0130311180608, (APWX01303);  1011-9050-000, 010084068210229221APWX0130411180608, (APWX01304);  1011-9050-000, 010084068210229221APWX0130511180608, (APWX01305);  1011-9050-000, 010084068210229221APWX0130611180608, (APWX01306);  1011-9050-000, 010084068210229221APWX0130711180608, (APWX01307);  1011-9050-000, 010084068210229221APWX0130811180608, (APWX01308);  1011-9050-000, 010084068210229221APWX0130911180608, (APWX01309);  1011-9050-000, 010084068210229221APWX0131011180608, (APWX01310);  1011-9050-000, 010084068210229221APWX0131111180608, (APWX01311);  1011-9050-000, 010084068210229221APWX0131211180608, (APWX01312);  1011-9050-000, 010084068210229221APWX0131311180608, (APWX01313);  1011-9050-000, 010084068210229221APWX0131411180608, (APWX01314);  1011-9050-000, 010084068210229221APWX0131511180608, (APWX01315);  1011-9050-000, 010084068210229221APWX0131611180608, (APWX01316);  1011-9050-000, 010084068210229221APWX0131711180608, (APWX01317);  1011-9050-000, 010084068210229221APWX0131811180608, (APWX01318);  1011-9050-000, 010084068210229221APWX0131911180608, (APWX01319);  1011-9050-000, 010084068210229221APWX0140311180618, (APWX01403);  1011-9050-000, 010084068210229221APWX0140411180618, (APWX01404);  1011-9050-000, 010084068210229221APWX0176811180720, (APWX01768);  1011-9050-000, 010084068210229221APWX0043511180206, (APWX00435);  1011-9050-000, 010084068210229221APWX0043611180206, (APWX00436);  1011-9050-000, 010084068210229221APWX0154511180627, (APWX01545);  1011-9050-000, 010084068210229221APWX0154611180627, (APWX01546);  1011-9050-000, 010084068210229221APWX0154711180627, (APWX01547);  1011-9050-000, 010084068210229221APWX0154811180627, (APWX01548);  1011-9050-000, 010084068210229221APWX0154911180627, (APWX01549);  1011-9050-000, 010084068210229221APWX0155011180627, (APWX01550);  1011-9050-000, 010084068210229221APWX0035311180131, (APWX00353);  1011-9050-000, 010084068210229221APWX0035411180131, (APWX00354);  1011-9050-000, 010084068210229221APWX0035511180131, (APWX00355);  1011-9050-000, 010084068210229221APWX0035611180131, (APWX00356);  2097675-001-US, Not Applicable, (5481199-1);  1011-9050-000, 010084068210229221APWX0179111180720, (APWX01791);  1011-9050-000, 010084068210229221APWX0179211180720, (APWX01792);  1011-9050-000, 010084068210229221APWX0179311180720, (APWX01793);  1011-9050-000, 010084068210229221APWX0179411180720, (APWX01794);  1011-9050-000, 010084068210229221APWX0179511180720, (APWX01795);  1011-9050-000, 010084068210229221APWX0179611180720, (APWX01796);  1011-9050-000, 010084068210229221APWX0179711180720, (APWX01797);  1011-9050-000, 010084068210229221APWX0179811180720, (APWX01798);  1011-9050-000, 010084068210229221APWX0179911180720, (APWX01799);  1011-9050-000, 010084068210229221APWX0180011180720, (APWX01800);  1011-9050-000, 010084068210229221APWX0180111180720, (APWX01801);  1011-9050-000, 010084068210229221APWX0180211180720, (APWX01802);  1011-9050-000, 010084068210229221APWX0180311180720, (APWX01803);  1011-9050-000, 010084068210229221APWX0117311180517, (APWX01173);  1011-9050-000, 010084068210229221APWX0117411180517, (APWX01174);  1011-9050-000, 010084068210229221APWX0132911180608, (APWX01329);  1011-9050-000, 010084068210229221APWX0133011180608, (APWX01330);  1011-9050-000, 010084068210229221APWX0133111180608, (APWX01331);  1011-9050-000, 010084068210229221APWX0133211180608, (APWX01332);  1011-9050-000, 010084068210229221APWX0133311180608, (APWX01333);  1011-9050-000, 010084068210229221APWX0133411180608, (APWX01334);  1011-9050-000, 010084068210229221APWX0133511180608, (APWX01335);  1011-9050-000, 010084068210229221APWX0133611180608, (APWX01336);  1011-9050-000, 010084068210229221APWX0133711180608, (APWX01337);  1011-9050-000, 010084068210229221APWX0133811180608, (APWX01338);  1011-9050-000, 010084068210229221APWX0133911180608, (APWX01339);  2097675-001-US, Not Applicable, (5479630-2);  1011-9050-000, 010084068210229221APWW0227811171222, (APWW02278);  1011-9050-000, 010084068210229221APWX0156911180703, (APWX01569);  1011-9050-000, 010084068210229221APWX0120111180524, (APWX01201);  1011-9050-000, 010084068210229221APWX0134011180608, (APWX01340);  1011-9050-000, 010084068210229221APWX0134111180611, (APWX01341);  1011-9050-000, 010084068210229221APWX0134211180611, (APWX01342);  1011-9050-000, 010084068210229221APWX0134311180611, (APWX01343);  1011-9050-000, 010084068210229221APWX0134411180611, (APWX01344);  1011-9050-000, 010084068210229221APWX0134511180611, (APWX01345);  1011-9050-000, 010084068210229221APWX0134611180611, (APWX01346);  1011-9050-000, 010084068210229221APWX0134711180611, (APWX01347);  1011-9050-000, 010084068210229221APWX0134811180611, (APWX01348);  1011-9050-000, 010084068210229221APWX0134911180611, (APWX01349);  1011-9050-000, 010084068210229221APWX0135011180611, (APWX01350);  1011-9050-000, 010084068210229221APWX0135111180611, (APWX01351);  1011-9050-000, 010084068210229221APWX0135211180611, (APWX01352);  1011-9050-000, 010084068210229221APWX0135311180611, (APWX01353);  1011-9050-000, 010084068210229221APWX0135411180611, (APWX01354);  1011-9050-000, 010084068210229221APWX0120611180524, (APWX01206);  1011-9050-000, 010084068210229221APWX0120711180524, (APWX01207);  1011-9050-000, 010084068210229221APWX0092711180410, (APWX00927);  1011-9050-000, 010084068210229221APWX0092811180410, (APWX00928);  1011-9050-000, 010084068210229221APWX0092911180410, (APWX00929);  1011-9050-000, 010084068210229221APWX0093011180410, (APWX00930);  1011-9050-000, 010084068210229221APWX0093111180410, (APWX00931);  1011-9050-000, 010084068210229221APWX0093211180410, (APWX00932);  1011-9050-000, 010084068210229221APWX0093311180410, (APWX00933);  1011-9050-000, 010084068210229221APWX0093411180410, (APWX00934);  1011-9050-000, 010084068210229221APWX0093511180410, (APWX00935);  1011-9050-000, 010084068210229221APWX0093611180410, (APWX00936);  1011-9050-000, 010084068210229221APWX0093711180410, (APWX00937);  1011-9050-000, 010084068210229221APWX0093811180410, (APWX00938);  1011-9050-000, 010084068210229221APWX0186911180803, (APWX01869);  1011-9050-000, 010084068210229221APWX0017311180105, (APWX00173);  1011-9050-000, 010084068210229221APWX0048511180208, (APWX00485);  1011-9050-000, 010084068210229221APWW0220311171218, (APWW02203);  1011-9050-000, 010084068210229221APWW0220411171218, (APWW02204);  1011-9050-000, 010084068210229221APWW0220511171218, (APWW02205);  1011-9050-000, 010084068210229221APWW0220611171218, (APWW02206);  1011-9050-000, 010084068210229221APWW0220711171218, (APWW02207);  1011-9050-000, 010084068210229221APWW0220811171218, (APWW02208);  1011-9050-000, 010084068210229221APWW0220911171218, (APWW02209);  1011-9050-000, 010084068210229221APWW0221011171218, (APWW02210);  1011-9050-000, 010084068210229221APWW0221111171219, (APWW02211);  1011-9050-000, 010084068210229221APWX0078411180323, (APWX00784);  1011-9050-000, 010084068210229221APWX0078511180323, (APWX00785);  1011-9050-000, 010084068210229221APWX0078611180323, (APWX00786);  1011-9050-000, 010084068210229221APWX0078711180323, (APWX00787);  1011-9050-000, 010084068210229221APWX0107411180502, (APWX01074);  1011-9050-000, 010084068210229221APWX0107511180502, (APWX01075);  1011-9050-000, 010084068210229221APWX0107711180502, (APWX01077);  1011-9050-000, 010084068210229221APWX0107811180502, (APWX01078);  1011-9050-000, 010084068210229221APWX0140511180618, (APWX01405);  1011-9050-000, 010084068210229221APWX0180611180720, (APWX01806);  1011-9050-000, 010084068210229221APWX0180711180720, (APWX01807);  1011-9050-000, 010084068210229221APWX0180811180720, (APWX01808);  1011-9050-000, 010084068210229221APWX0180911180720, (APWX01809);  1011-9050-000, 010084068210229221APWX0181011180720, (APWX01810);  1011-9050-000, 010084068210229221APWX0065611180309, (APWX00656);  1011-9050-000, 010084068210229221APWX0065711180309, (APWX00657);  1011-9050-000, 010084068210229221APWX0065811180309, (APWX00658);  1011-9050-000, 010084068210229221APWX0065911180309, (APWX00659);  1011-9050-000, 010084068210229221APWX0066011180309, (APWX00660);  1011-9050-000, 010084068210229221APWX0036311180131, (APWX00363);  1011-9050-000, 010084068210229221APWX0176911180720, (APWX01769);  1011-9050-000, 010084068210229221APWX0193611180815, (APWX01936);  1011-9050-000, 010084068210229221APWX0124211180531, (APWX01242);  1011-9050-000, 010084068210229221APWX0124311180531, (APWX01243);  1011-9050-000, 010084068210229221APWX0124411180531, (APWX01244);  1011-9050-000, 010084068210229221APWX0063311180309, (APWX00633);  1011-9050-000, 010084068210229221APWX0157211180703, (APWX01572);  1011-9050-000, 010084068210229221APWX0157311180703, (APWX01573);  1011-9050-000, 010084068210229221APWX0037011180131, (APWX00370);  1011-9050-000, 010084068210229221APWV0132511160622, (APWV01325);  1011-9050-000, 010084068210229221APWX0016811180105, (APWX00168);  1011-9050-000, 010084068210229221APWX0016911180105, (APWX00169);  1011-9050-000, 010084068210229221APWX0017011180105, (APWX00170);  1011-9050-000, 010084068210229221APWX0017111180105, (APWX00171);  1011-9050-000, 010084068210229221APWX0017211180105, (APWX00172);  1011-9050-000, 010084068210229221APWX0096111180417, (APWX00961);  1011-9050-000, 010084068210229221APWX0132511180608, (APWX01325);  1011-9050-000, 010084068210229221APWX0132611180608, (APWX01326);  1011-9050-000, 010084068210229221APWX0112611180507, (APWX01126);  1011-9050-000, 010084068210229221APWX0112711180507, (APWX01127);  1011-9050-000, 010084068210229221APWX0112811180507, (APWX01128);  1011-9050-000, 010084068210229221APWX0112911180507, (APWX01129);  1011-9050-000, 010084068210229221APWX0113011180507, (APWX01130);  1011-9050-000, 010084068210229221APWX0113111180507, (APWX01131);  1011-9050-000, 010084068210229221APWX0113211180507, (APWX01132);  1011-9050-000, 010084068210229221APWX0113311180507, (APWX01133);  1011-9050-000, 010084068210229221APWX0113411180507, (APWX01134);  1011-9050-000, 010084068210229221APWX0113611180507, (APWX01136);  1011-9050-000, 010084068210229221APWX0113711180507, (APWX01137);  1011-9050-000, 010084068210229221APWX0113811180507, (APWX01138);  1011-9050-000, 010084068210229221APWX0113911180507, (APWX01139);  1011-9050-000, 010084068210229221APWX0114011180507, (APWX01140);  1011-9050-000, 010084068210229221APWX0114111180507, (APWX01141);  1011-9050-000, 010084068210229221APWX0114211180507, (APWX01142);  1011-9050-000, 010084068210229221APWX0114311180507, (APWX01143);  1011-9050-000, 010084068210229221APWX0114411180507, (APWX01144);  1011-9050-000, 010084068210229221APWX0114511180507, (APWX01145);  1011-9050-000, 010084068210229221APWX0114611180507, (APWX01146);  1011-9050-000, 010084068210229221APWX0114711180507, (APWX01147);  1011-9050-000, 010084068210229221APWX0114811180507, (APWX01148);  1011-9050-000, 010084068210229221APWX0114911180507, (APWX01149);  1011-9050-000, 010084068210229221APWX0115011180507, (APWX01150);  1011-9050-000, 010084068210229221APWX0115111180517, (APWX01151);  1011-9050-000, 010084068210229221APWX0115211180517, (APWX01152);  1011-9050-000, 010084068210229221APWX0115311180517, (APWX01153);  1011-9050-000, 010084068210229221APWX0115411180517, (APWX01154);  1011-9050-000, 010084068210229221APWX0115511180517, (APWX01155);  1011-9050-000, 010084068210229221APWX0115611180517, (APWX01156);  1011-9050-000, 010084068210229221APWX0115711180517, (APWX01157);  1011-9050-000, 010084068210229221APWX0115811180517, (APWX01158);  1011-9050-000, 010084068210229221APWX0115911180517, (APWX01159);  1011-9050-000, 010084068210229221APWX0116011180517, (APWX01160);  1011-9050-000, 010084068210229221APWX0116111180517, (APWX01161);  1011-9050-000, 010084068210229221APWX0116211180517, (APWX01162);  1011-9050-000, 010084068210229221APWX0116311180517, (APWX01163);  1011-9050-000, 010084068210229221APWX0116411180517, (APWX01164);  1011-9050-000, 010084068210229221APWX0116511180517, (APWX01165);  1011-9050-000, 010084068210229221APWX0117511180517, (APWX01175);  1011-9050-000, 010084068210229221APWX0187111180803, (APWX01871);  1011-9050-000, 010084068210229221APWX0136311180612, (APWX01363);  1011-9050-000, 010084068210229221APWX0136411180612, (APWX01364);  1011-9050-000, 010084068210229221APWX0136511180612, (APWX01365);  1011-9050-000, 010084068210229221APWX0136611180612, (APWX01366);  1011-9050-000, 010084068210229221APWX0136711180612, (APWX01367);  1011-9050-000, 010084068210229221APWX0136811180612, (APWX01368);  1011-9050-000, 010084068210229221APWX0136911180612, (APWX01369);  1011-9050-000, 010084068210229221APWX0137011180612, (APWX01370);  1011-9050-000, 010084068210229221APWX0137111180612, (APWX01371);  1011-9050-000, 010084068210229221APWX0137211180612, (APWX01372);  1011-9050-000, 010084068210229221APWX0137311180612, (APWX01373);  1011-9050-000, 010084068210229221APWX0137411180612, (APWX01374);  1011-9050-000, 010084068210229221APWX0137511180612, (APWX01375);  1011-9050-000, 010084068210229221APWX0137611180612, (APWX01376);  1011-9050-000, 010084068210229221APWX0137711180612, (APWX01377);  1011-9050-000, 010084068210229221APWX0137811180612, (APWX01378);  1011-9050-000, 010084068210229221APWX0137911180612, (APWX01379);  1011-9050-000, 010084068210229221APWX0138011180612, (APWX01380);  1011-9050-000, 010084068210229221APWX0138111180612, (APWX01381);  1011-9050-000, 010084068210229221APWX0138211180612, (APWX01382);  1011-9050-000, 010084068210229221APWX0138311180612, (APWX01383);  1011-9050-000, 010084068210229221APWX0138411180612, (APWX01384);  1011-9050-000, 010084068210229221APWX0138511180612, (APWX01385);  1011-9050-000, 010084068210229221APWX0138611180612, (APWX01386);  1011-9050-000, 010084068210229221APWX0138711180612, (APWX01387);  1011-9050-000, 010084068210229221APWX0138811180612, (APWX01388);  1011-9050-000, 010084068210229221APWX0138911180612, (APWX01389);  1011-9050-000, 010084068210229221APWX0139011180612, (APWX01390);  1011-9050-000, 010084068210229221APWX0139111180612, (APWX01391);  1011-9050-000, 010084068210229221APWX0139211180612, (APWX01392);  1011-9050-000, 010084068210229221APWX0139311180612, (APWX01393);  1011-9050-000, 010084068210229221APWX0139411180612, (APWX01394);  1011-9050-000, 010084068210229221APWX0139511180612, (APWX01395);  1011-9050-000, 010084068210229221APWX0139611180612, (APWX01396);  1011-9050-000, 010084068210229221APWX0139711180612, (APWX01397);  1011-9050-000, 010084068210229221APWX0139811180612, (APWX01398);  1011-9050-000, 010084068210229221APWX0139911180612, (APWX01399);  1011-9050-000, 010084068210229221APWX0142411180618, (APWX01424);  1011-9050-000, 010084068210229221APWX0025811180116, (APWX00258);  1011-9050-000, 010084068210229221APWX0025911180116, (APWX00259);  1011-9050-000, 010084068210229221APWX0026011180116, (APWX00260);  1011-9050-000, 010084068210229221APWX0064911180309, (APWX00649);  1011-9050-000, 010084068210229221APWW0225711171222, (APWW02257);  1011-9050-000, 010084068210229221APWW0225811171222, (APWW02258);  1011-9050-000, 010084068210229221APWW0225911171222, (APWW02259);  1011-9050-000, 010084068210229221APWW0226011171222, (APWW02260);  1011-9050-000, 010084068210229221APWW0226111171222, (APWW02261);  1011-9050-000, 010084068210229221APWW0226211171222, (APWW02262);  1011-9050-000, 010084068210229221APWW0226311171222, (APWW02263);  1011-9050-000, 010084068210229221APWW0226411171222, (APWW02264);  1011-9050-000, 010084068210229221APWX0036511180131, (APWX00365);  1011-9050-000, 010084068210229221APWX0036611180131, (APWX00366);  1011-9050-000, 010084068210229221APWX0036711180131, (APWX00367);  1011-9050-000, 010084068210229221APWX0036811180131, (APWX00368);  1011-9050-000, 010084068210229221APWX0180411180720, (APWX01804);  1011-9050-000, 010084068210229221APWX0182711180802, (APWX01827);  1011-9050-000, 010084068210229221APWW0226511171222, (APWW02265);  1011-9050-000, 010084068210229221APWW0226611171222, (APWW02266);  1011-9050-000, 010084068210229221APWW0226711171222, (APWW02267);  1011-9050-000, 010084068210229221APWW0226811171222, (APWW02268);  1011-9050-000, 010084068210229221APWW0226911171222, (APWW02269);  1011-9050-000, 010084068210229221APWW0227011171222, (APWW02270);  1011-9050-000, 010084068210229221APWW0227111171222, (APWW02271);  1011-9050-000, 010084068210229221APWW0227211171222, (APWW02272);  1011-9050-000, 010084068210229221APWW0227311171222, (APWW02273);  1011-9050-000, 010084068210229221APWX0154111180627, (APWX01541);  1011-9050-000, 010084068210229221APWX0140211180618, (APWX01402);  2097675-001-US, Not Applicable, (5474392-1);  2097675-001-US, Not Applicable, (5474392-10);  2097675-001-US, Not Applicable, (5474392-11);  2097675-001-US, Not Applicable, (5474392-12);  2097675-001-US, Not Applicable, (5474392-13);  2097675-001-US, Not Applicable, (5474392-14);  2097675-001-US, Not Applicable, (5474392-15);  2097675-001-US, Not Applicable, (5474392-16);  2097675-001-US, Not Applicable, (5474392-17);  2097675-001-US, Not Applicable, (5474392-18);  2097675-001-US, Not Applicable, (5474392-19);  2097675-001-US, Not Applicable, (5474392-2);  2097675-001-US, Not Applicable, (5474392-20);  2097675-001-US, Not Applicable, (5474392-3);  2097675-001-US, Not Applicable, (5474392-4);  2097675-001-US, Not Applicable, (5474392-5);  2097675-001-US, Not Applicable, (5474392-6);  2097675-001-US, Not Applicable, (5474392-7);  2097675-001-US, Not Applicable, (5474392-8);  2097675-001-US, Not Applicable, (5474392-9);  2097675-001-US, Not Applicable, (5479053-1);  2097675-001-US, Not Applicable, (5479053-10);  2097675-001-US, Not Applicable, (5479053-11);  2097675-001-US, Not Applicable, (5479053-12);  2097675-001-US, Not Applicable, (5479053-13);  2097675-001-US, Not Applicable, (5479053-14);  2097675-001-US, Not Applicable, (5479053-15);  2097675-001-US, Not Applicable, (5479053-16);  2097675-001-US, Not Applicable, (5479053-17);  2097675-001-US, Not Applicable, (5479053-18);  2097675-001-US, Not Applicable, (5479053-19);  2097675-001-US, Not Applicable, (5479053-2);  2097675-001-US, Not Applicable, (5479053-20);  2097675-001-US, Not Applicable, (5479053-3);  2097675-001-US, Not Applicable, (5479053-4);  2097675-001-US, Not Applicable, (5479053-5);  2097675-001-US, Not Applicable, (5479053-6);  2097675-001-US, Not Applicable, (5479053-7);  2097675-001-US, Not Applicable, (5479053-8);  2097675-001-US, Not Applicable, (5479053-9);  1011-9050-000, 010084068210229221APWW0147911170823, (APWW01479);  1011-9050-000, 010084068210229221APWX0040011180201, (APWX00400);  1011-9050-000, 010084068210229221APWX0040111180201, (APWX00401);  1011-9050-000, 010084068210229221APWX0040411180201, (APWX00404);  1011-9050-000, 010084068210229221APWX0040511180201, (APWX00405);  1011-9050-000, 010084068210229221APWX0040611180201, (APWX00406);  1011-9050-000, 010084068210229221APWX0040711180201, (APWX00407);  1011-9050-000, 010084068210229221APWX0040811180201, (APWX00408);  1011-9050-000, 010084068210229221APWX0040911180201, (APWX00409);  1011-9050-000, 010084068210229221APWX0041011180201, (APWX00410);  1011-9050-000, 010084068210229221APWX0041111180201, (APWX00411);  1011-9050-000, 010084068210229221APWX0041211180201, (APWX00412);  1011-9050-000, 010084068210229221APWX0041311180201, (APWX00413);  1011-9050-000, 010084068210229221APWX0041411180201, (APWX00414);  1011-9050-000, 010084068210229221APWX0041511180201, (APWX00415);  1011-9050-000, 010084068210229221APWX0041611180201, (APWX00416);  1011-9050-000, 010084068210229221APWX0041711180201, (APWX00417);  1011-9050-000, 010084068210229221APWX0041811180201, (APWX00418);  1011-9050-000, 010084068210229221APWX0041911180201, (APWX00419);  1011-9050-000, 010084068210229221APWX0042011180201, (APWX00420);  1011-9050-000, 010084068210229221APWX0042111180206, (APWX00421);  1011-9050-000, 010084068210229221APWX0042211180206, (APWX00422);  1011-9050-000, 010084068210229221APWX0042311180206, (APWX00423);  1011-9050-000, 010084068210229221APWX0042411180206, (APWX00424);  1011-9050-000, 010084068210229221APWX0042511180206, (APWX00425);  1011-9050-000, 010084068210229221APWX0042611180206, (APWX00426);  1011-9050-000, 010084068210229221APWX0042711180206, (APWX00427);  1011-9050-000, 010084068210229221APWX0060411180307, (APWX00604);  1011-9050-000, 010084068210229221APWX0060511180307, (APWX00605);  1011-9050-000, 010084068210229221APWX0060611180307, (APWX00606);  1011-9050-000, 010084068210229221APWX0060711180307, (APWX00607);  1011-9050-000, 010084068210229221APWX0060811180307, (APWX00608);  1011-9050-000, 010084068210229221APWX0060911180307, (APWX00609);  1011-9050-000, 010084068210229221APWX0061011180307, (APWX00610);  1011-9050-000, 010084068210229221APWX0061111180307, (APWX00611);  1011-9050-000, 010084068210229221APWX0061211180307, (APWX00612);  1011-9050-000, 010084068210229221APWX0061311180307, (APWX00613);  1011-9050-000, 010084068210229221APWX0061411180307, (APWX00614);  1011-9050-000, 010084068210229221APWX0061511180307, (APWX00615);  1011-9050-000, 010084068210229221APWX0061611180307, (APWX00616);  1011-9050-000, 010084068210229221APWX0061711180307, (APWX00617);  1011-9050-000, 010084068210229221APWX0061811180307, (APWX00618);  1011-9050-000, 010084068210229221APWX0061911180307, (APWX00619);  1011-9050-000, 010084068210229221APWX0062011180307, (APWX00620);  1011-9050-000, 010084068210229221APWX0062311180309, (APWX00623);  1011-9050-000, 010084068210229221APWX0062411180309, (APWX00624);  1011-9050-000, 010084068210229221APWX0062511180309, (APWX00625);  1011-9050-000, 010084068210229221APWX0062611180309, (APWX00626);  1011-9050-000, 010084068210229221APWX0062711180309, (APWX00627);  1011-9050-000, 010084068210229221APWX0073311180319, (APWX00733);  1011-9050-000, 010084068210229221APWX0073411180319, (APWX00734);  1011-9050-000, 010084068210229221APWX0073511180319, (APWX00735);  1011-9050-000, 010084068210229221APWX0073611180319, (APWX00736);  1011-9050-000, 010084068210229221APWX0073711180319, (APWX00737);  1011-9050-000, 010084068210229221APWX0073811180319, (APWX00738);  1011-9050-000, 010084068210229221APWX0073911180319, (APWX00739);  1011-9050-000, 010084068210229221APWX0074011180319, (APWX00740);  1011-9050-000, 010084068210229221APWX0074111180319, (APWX00741);  1011-9050-000, 010084068210229221APWX0074211180319, (APWX00742);  1011-9050-000, 010084068210229221APWX0074311180319, (APWX00743);  1011-9050-000, 010084068210229221APWX0074411180319, (APWX00744);  1011-9050-000, 010084068210229221APWX0081011180328, (APWX00810);  1011-9050-000, 010084068210229221APWX0081111180328, (APWX00811);  1011-9050-000, 010084068210229221APWX0081211180328, (APWX00812);  1011-9050-000, 010084068210229221APWX0081311180328, (APWX00813);  1011-9050-000, 010084068210229221APWX0081411180328, (APWX00814);  1011-9050-000, 010084068210229221APWX0081511180328, (APWX00815);  1011-9050-000, 010084068210229221APWX0081611180328, (APWX00816);  1011-9050-000, 010084068210229221APWX0081711180328, (APWX00817);  1011-9050-000, 010084068210229221APWX0081811180328, (APWX00818);  1011-9050-000, 010084068210229221APWX0081911180328, (APWX00819);  1011-9050-000, 010084068210229221APWX0082011180328, (APWX00820);  1011-9050-000, 010084068210229221APWX0082111180328, (APWX00821);  1011-9050-000, 010084068210229221APWX0082211180328, (APWX00822);  2097675-001-US, Not Applicable, (5474296-10);  2097675-001-US, Not Applicable, (5474296-11);  2097675-001-US, Not Applicable, (5474296-12);  2097675-001-US, Not Applicable, (5474296-13);  2097675-001-US, Not Applicable, (5474296-14);  2097675-001-US, Not Applicable, (5474296-15);  2097675-001-US, Not Applicable, (5474296-16);  2097675-001-US, Not Applicable, (5474296-17);  2097675-001-US, Not Applicable, (5474296-18);  2097675-001-US, Not Applicable, (5474296-19);  2097675-001-US, Not Applicable, (5474296-2);  2097675-001-US, Not Applicable, (5474296-20);  2097675-001-US, Not Applicable, (5474296-21);  2097675-001-US, Not Applicable, (5474296-22);  2097675-001-US, Not Applicable, (5474296-23);  2097675-001-US, Not Applicable, (5474296-24);  2097675-001-US, Not Applicable, (5474296-25);  2097675-001-US, Not Applicable, (5474296-26);  2097675-001-US, Not Applicable, (5474296-27);  2097675-001-US, Not Applicable, (5474296-28);  2097675-001-US, Not Applicable, (5474296-29);  2097675-001-US, Not Applicable, (5474296-3);  2097675-001-US, Not Applicable, (5474296-30);  2097675-001-US, Not Applicable, (5474296-31);  2097675-001-US, Not Applicable, (5474296-32);  2097675-001-US, Not Applicable, (5474296-33);  2097675-001-US, Not Applicable, (5474296-34);  2097675-001-US, Not Applicable, (5474296-35);  2097675-001-US, Not Applicable, (5474296-36);  2097675-001-US, Not Applicable, (5474296-37);  2097675-001-US, Not Applicable, (5474296-38);  2097675-001-US, Not Applicable, (5474296-4);  2097675-001-US, Not Applicable, (5474296-5);  2097675-001-US, Not Applicable, (5474296-6);  2097675-001-US, Not Applicable, (5474296-7);  2097675-001-US, Not Applicable, (5474296-8);  2097675-001-US, Not Applicable, (5474296-9);  1011-9050-000, 010084068210229221APWX0145411180619, (APWX01454);  1011-9050-000, 010084068210229221APWX0145511180619, (APWX01455);  1011-9050-000, 010084068210229221APWX0145611180619, (APWX01456);  1011-9050-000, 010084068210229221APWX0145711180619, (APWX01457);  1011-9050-000, 010084068210229221APWX0145811180619, (APWX01458);  1011-9050-000, 010084068210229221APWX0145911180619, (APWX01459);  1011-9050-000, 010084068210229221APWX0146011180619, (APWX01460);  1011-9050-000, 010084068210229221APWX0146111180619, (APWX01461);  1011-9050-000, 010084068210229221APWX0146211180619, (APWX01462);  1011-9050-000, 010084068210229221APWX0146311180619, (APWX01463);  1011-9050-000, 010084068210229221APWX0146411180619, (APWX01464);  1011-9050-000, 010084068210229221APWX0146511180619, (APWX01465);  1011-9050-000, 010084068210229221APWX0146611180619, (APWX01466);  1011-9050-000, 010084068210229221APWX0146711180619, (APWX01467);  1011-9050-000, 010084068210229221APWX0146811180619, (APWX01468);  1011-9050-000, 010084068210229221APWX0146911180619, (APWX01469);  1011-9050-000, 010084068210229221APWX0147011180619, (APWX01470);  1011-9050-000, 010084068210229221APWX0147111180619, (APWX01471);  1011-9050-000, 010084068210229221APWX0147211180619, (APWX01472);  1011-9050-000, 010084068210229221APWX0147311180619, (APWX01473);  1011-9050-000, 010084068210229221APWX0147411180619, (APWX01474);  1011-9050-000, 010084068210229221APWX0147511180619, (APWX01475);  1011-9050-000, 010084068210229221APWX0147611180619, (APWX01476);  1011-9050-000, 010084068210229221APWX0147711180619, (APWX01477);  1011-9050-000, 010084068210229221APWX0147811180619, (APWX01478);  1011-9050-000, 010084068210229221APWX0148311180625, (APWX01483);  1011-9050-000, 010084068210229221APWX0148411180625, (APWX01484);  1011-9050-000, 010084068210229221APWX0190311180803, (APWX01903);  1011-9050-000, 010084068210229221APWX0190411180803, (APWX01904);  1011-9050-000, 010084068210229221APWW0158011170925, (APWW01580);  1011-9050-000, 010084068210229221APWW0158111170925, (APWW01581);  1011-9050-000, 010084068210229221APWW0165711171004, (APWW01657);  1011-9050-000, 010084068210229221APWW0168811171004, (APWW01688);  1011-9050-000, 010084068210229221APWW0168911171004, (APWW01689);  1011-9050-000, 010084068210229221APWW0169011171004, (APWW01690);  1011-9050-000, 010084068210229221APWW0169111171012, (APWW01691);  1011-9050-000, 010084068210229221APWW0169211171012, (APWW01692);  1011-9050-000, 010084068210229221APWW0169311171012, (APWW01693);  1011-9050-000, 010084068210229221APWW0169411171012, (APWW01694);  1011-9050-000, 010084068210229221APWW0169511171012, (APWW01695);  1011-9050-000, 010084068210229221APWW0169611171012, (APWW01696);  1011-9050-000, 010084068210229221APWX0056311180301, (APWX00563);  1011-9050-000, 010084068210229221APWX0056411180301, (APWX00564);  1011-9050-000, 010084068210229221APWX0056511180301, (APWX00565);  1011-9050-000, 010084068210229221APWX0056611180301, (APWX00566);  1011-9050-000, 010084068210229221APWX0090211180328, (APWX00902);  1011-9050-000, 010084068210229221APWX0029611180118, (APWX00296);  1011-9050-000, 010084068210229221APWX0110211180502, (APWX01102);  1011-9050-000, 010084068210229221APWX0148511180625, (APWX01485);  1011-9050-000, 010084068210229221APWX0148611180625, (APWX01486);  1011-9050-000, 010084068210229221APWX0148711180625, (APWX01487);  1011-9050-000, 010084068210229221APWX0148811180625, (APWX01488);  1011-9050-000, 010084068210229221APWX0148911180625, (APWX01489);  1011-9050-000, 010084068210229221APWX0149011180625, (APWX01490);  1011-9050-000, 010084068210229221APWX0149111180625, (APWX01491);  1011-9050-000, 010084068210229221APWX0149211180625, (APWX01492);  1011-9050-000, 010084068210229221APWX0149311180625, (APWX01493);  1011-9050-000, 010084068210229221APWX0149411180625, (APWX01494);  1011-9050-000, 010084068210229221APWX0149511180625, (APWX01495);  1011-9050-000, 010084068210229221APWX0149611180625, (APWX01496);  1011-9050-000, 010084068210229221APWX0149711180625, (APWX01497);  1011-9050-000, 010084068210229221APWX0149811180625, (APWX01498);  1011-9050-000, 010084068210229221APWX0149911180625, (APWX01499);  1011-9050-000, 010084068210229221APWX0150011180625, (APWX01500);  1011-9050-000, 010084068210229221APWX0150111180625, (APWX01501);  1011-9050-000, 010084068210229221APWX0150211180625, (APWX01502);  1011-9050-000, 010084068210229221APWX0150311180625, (APWX01503);  1011-9050-000, 010084068210229221APWX0150411180625, (APWX01504);  1011-9050-000, 010084068210229221APWX0150511180625, (APWX01505);  1011-9050-000, 010084068210229221APWX0150611180625, (APWX01506);  1011-9050-000, 010084068210229221APWX0150711180625, (APWX01507);  1011-9050-000, 010084068210229221APWX0150811180625, (APWX01508);  1011-9050-000, 010084068210229221APWX0150911180625, (APWX01509);  1011-9050-000, 010084068210229221APWX0151011180625, (APWX01510);  1011-9050-000, 010084068210229221APWX0151111180625, (APWX01511);  1011-9050-000, 010084068210229221APWX0151211180625, (APWX01512);  1011-9050-000, 010084068210229221APWX0151311180625, (APWX01513);  1011-9050-000, 010084068210229221APWX0151411180625, (APWX01514);  1011-9050-000, 010084068210229221APWX0151511180625, (APWX01515);  1011-9050-000, 010084068210229221APWX0151611180625, (APWX01516);  1011-9050-000, 010084068210229221APWX0151711180625, (APWX01517);  1011-9050-000, 010084068210229221APWX0117711180517, (APWX01177);  1011-9050-000, 010084068210229221APWX0164811180716, (APWX01648);  1011-9050-000, 010084068210229221APWX0164911180716, (APWX01649);  1011-9050-000, 010084068210229221APWX0165011180716, (APWX01650);  1011-9050-000, 010084068210229221APWX0165111180716, (APWX01651);  1011-9050-000, Not Applicable, (APWT02419);  1011-9050-000, 010084068210229221APWW0214811171214, (APWW02148);  1011-9050-000, 010084068210229221APWX0098511180417, (APWX00985);  1011-9050-000, 010084068210229221APWX0098611180417, (APWX00986);  1011-9050-000, 010084068210229221APWX0098711180417, (APWX00987);  1011-9050-000, 010084068210229221APWX0124511180531, (APWX01245);  1011-9050-000, 010084068210229221APWX0169411180718, (APWX01694);  1011-9050-000, 010084068210229221APWX0169511180718, (APWX01695);  1011-9050-000, 010084068210229221APWX0169611180718, (APWX01696);  1011-9050-000, 010084068210229221APWX0169711180718, (APWX01697);  1011-9050-000, 010084068210229221APWX0014811180105, (APWX00148);  1011-9050-000, 010084068210229221APWX0014911180105, (APWX00149);  1011-9050-000, 010084068210229221APWX0104911180427, (APWX01049);  1011-9050-000, 010084068210229221APWX0105011180427, (APWX01050);  1011-9050-000, 010084068210229221APWX0068911180315, (APWX00689);  1011-9050-000, 010084068210229221APWX0069011180315, (APWX00690);  1011-9050-000, 010084068210229221APWX0136111180612, (APWX01361);  1011-9050-000, 010084068210229221APWX0136211180612, (APWX01362);  1011-9050-000, 010084068210229221APWX0059611180307, (APWX00596);  1011-9050-000, 010084068210229221APWX0059711180307, (APWX00597);  1011-9050-000, 010084068210229221APWX0059811180307, (APWX00598);  1011-9050-000, 010084068210229221APWX0059911180307, (APWX00599);  1011-9050-000, 010084068210229221APWX0060011180307, (APWX00600);  1011-9050-000, 010084068210229221APWX0060111180307, (APWX00601);  1011-9050-000, 010084068210229221APWX0060211180307, (APWX00602);  1011-9050-000, 010084068210229221APWX0060311180307, (APWX00603);  1011-9050-000, 010084068210229221APWX0132711180608, (APWX01327);  1011-9050-000, 010084068210229221APWX0054511180301, (APWX00545);  1011-9050-000, 010084068210229221APWX0054611180301, (APWX00546);  1011-9050-000, 010084068210229221APWX0054711180301, (APWX00547);  1011-9050-000, 010084068210229221APWX0054811180301, (APWX00548);  1011-9050-000, 010084068210229221APWX0054911180301, (APWX00549); 1011-9050-000, 010084068210229221APWX0055011180301, (APWX00550);  1011-9050-000, 010084068210229221APWX0055111180301, (APWX00551);  1011-9050-000, 010084068210229221APWX0055211180301, (APWX00552);  1011-9050-000, 010084068210229221APWX0132011180608, (APWX01320);  1011-9050-000, 010084068210229221APWX0132111180608, (APWX01321);  1011-9050-000, 010084068210229221APWX0132211180608, (APWX01322);  1011-9050-000, 010084068210229221APWX0067411180315, (APWX00674);  1011-9050-000, 010084068210229221APWX0093911180410, (APWX00939);  1011-9050-000, 010084068210229221APWX0094011180410, (APWX00940);  1011-9050-000, 010084068210229221APWX0094111180410, (APWX00941);  1011-9050-000, 010084068210229221APWW0213711171214, (APWW02137);  1011-9050-000, 010084068210229221APWW0213811171214, (APWW02138);  1011-9050-000, 010084068210229221APWW0213911171214, (APWW02139);  1011-9050-000, 010084068210229221APWX0013511180104, (APWX00135);  1011-9050-000, 010084068210229221APWX0013611180104, (APWX00136);  1011-9050-000, 010084068210229221APWX0028511180118, (APWX00285);  1011-9050-000, 010084068210229221APWX0028611180118, (APWX00286);  1011-9050-000, 010084068210229221APWX0028711180118, (APWX00287);  1011-9050-000, 010084068210229221APWX0028811180118, (APWX00288);  1011-9050-000, 010084068210229221APWX0028911180118, (APWX00289);  1011-9050-000, 010084068210229221APWX0029011180118, (APWX00290);  1011-9050-000, 010084068210229221APWX0029111180118, (APWX00291);  1011-9050-000, 010084068210229221APWX0029211180118, (APWX00292);  1011-9050-000, 010084068210229221APWX0029311180118, (APWX00293);  1011-9050-000, 010084068210229221APWX0040211180201, (APWX00402);  1011-9050-000, 010084068210229221APWX0040311180201, (APWX00403);  1011-9050-000, 010084068210229221APWX0175711180720, (APWX01757);  1011-9050-000, 010084068210229221APWW0225311171222, (APWW02253);  1011-9050-000, 010084068210229221APWW0225411171222, (APWW02254);  1011-9050-000, 010084068210229221APWW0225511171222, (APWW02255);  1011-9050-000, 010084068210229221APWW0225611171222, (APWW02256);  1011-9050-000, 010084068210229221APWX0143311180618, (APWX01433);  1011-9050-000, 010084068210229221APWX0122011180530, (APWX01220);  1011-9050-000, 010084068210229221APWX0122111180530, (APWX01221);  1011-9050-000, 010084068210229221APWX0122211180530, (APWX01222);  1011-9050-000, 010084068210229221APWX0122311180530, (APWX01223);  1011-9050-000, 010084068210229221APWX0122411180530, (APWX01224);  1011-9050-000, 010084068210229221APWX0155211180627, (APWX01552);  1011-9050-000, 010084068210229221APWX0155311180627, (APWX01553);  1011-9050-000, 010084068210229221APWX0155411180627, (APWX01554);  1011-9050-000, 010084068210229221APWX0155511180627, (APWX01555);  1011-9050-000, 010084068210229221APWX0155611180627, (APWX01556);  1011-9050-000, 010084068210229221APWX0087711180328, (APWX00877);  1011-9050-000, 010084068210229221APWX0087811180328, (APWX00878);  1011-9050-000, 010084068210229221APWX0062911180309, (APWX00629);  1011-9050-000, 010084068210229221APWX0063011180309, (APWX00630);  1011-9050-000, 010084068210229221APWX0063111180309, (APWX00631);  1011-9050-000, 010084068210229221APWX0063211180309, (APWX00632);  2097675-001-US, Not Applicable, (5480217-1);  2097675-001-US, Not Applicable, (5480217-10);  2097675-001-US, Not Applicable, (5480217-11);  2097675-001-US, Not Applicable, (5480217-12);  2097675-001-US, Not Applicable, (5480217-2);  2097675-001-US, Not Applicable, (5480217-3);  2097675-001-US, Not Applicable, (5480217-4);  2097675-001-US, Not Applicable, (5480217-5);  2097675-001-US, Not Applicable, (5480217-6);  2097675-001-US, Not Applicable, (5480217-7);  2097675-001-US, Not Applicable, (5480217-8);  2097675-001-US, Not Applicable, (5480217-9);  1011-9050-000, 010084068210229221APWX0071111180319, (APWX00711);  1011-9050-000, 010084068210229221APWX0071211180319, (APWX00712);  1011-9050-000, 010084068210229221APWX0072511180319, (APWX00725);  1011-9050-000, 010084068210229221APWX0072611180319, (APWX00726);  1011-9050-000, 010084068210229221APWX0072711180319, (APWX00727);  1011-9050-000, 010084068210229221APWX0072811180319, (APWX00728);  1011-9050-000, 010084068210229221APWX0072911180319, (APWX00729);  1011-9050-000, 010084068210229221APWX0073011180319, (APWX00730);  1011-9050-000, 010084068210229221APWX0073111180319, (APWX00731);  1011-9050-000, 010084068210229221APWX0073211180319, (APWX00732);  1011-9050-000, 010084068210229221APWX0175011180720, (APWX01750);  1011-9050-000, 010084068210229221APWW0223611171219, (APWW02236);  1011-9050-000, 010084068210229221APWX0030811180123, (APWX00308);  1011-9050-000, 010084068210229221APWX0030911180123, (APWX00309);  1011-9050-000, 010084068210229221APWW0223211171219, (APWW02232);  1011-9050-000, 010084068210229221APWW0223311171219, (APWW02233);  1011-9050-000, 010084068210229221APWW0223411171219, (APWW02234);  1011-9050-000, 010084068210229221APWW0223511171219, (APWW02235);  1011-9050-000, 010084068210229221APWX0178511180720, (APWX01785);  1011-9050-000, 010084068210229221APWX0178611180720, (APWX01786);  1011-9050-000, 010084068210229221APWX0178711180720, (APWX01787);  1011-9050-000, 010084068210229221APWX0178811180720, (APWX01788);  1011-9050-000, 010084068210229221APWX0178911180720, (APWX01789);  1011-9050-000, 010084068210229221APWX0179011180720, (APWX01790);  2097675-001-US, Not Applicable, (5477453-1);  2097675-001-US, Not Applicable, (5477453-2);  1011-9050-000, 010084068210229221APWX0064411180309, (APWX00644);  1011-9050-000, 010084068210229221APWX0064511180309, (APWX00645);  1011-9050-000, 010084068210229221APWX0064611180309, (APWX00646);  1011-9050-000, 010084068210229221APWX0064711180309, (APWX00647);  1011-9050-000, 010084068210229221APWX0035811180131, (APWX00358);  1011-9050-000, 010084068210229221APWX0029711180118, (APWX00297);  1011-9050-000, 010084068210229221APWX0029811180118, (APWX00298);  1011-9050-000, 010084068210229221APWW0152911170913, (APWW01529);  1011-9050-000, 010084068210229221APWW0153011170913, (APWW01530);  1011-9050-000, 010084068210229221APWW0153111170913, (APWW01531);  1011-9050-000, 010084068210229221APWW0153211170913, (APWW01532);  1011-9050-000, 010084068210229221APWW0153311170913, (APWW01533);  1011-9050-000, 010084068210229221APWW0154411170913, (APWW01544);  1011-9050-000, 010084068210229221APWW0154711170913, (APWW01547);  1011-9050-000, 010084068210229221APWW0155211170913, (APWW01552);  1011-9050-000, 010084068210229221APWW0155411170913, (APWW01554);  1011-9050-000, 010084068210229221APWW0155611170913, (APWW01556);  1011-9050-000, 010084068210229221APWW0155711170913, (APWW01557);  1011-9050-000, 010084068210229221APWW0155811170913, (APWW01558);  1011-9050-000, 010084068210229221APWW0155911170913, (APWW01559);  1011-9050-000, 010084068210229221APWW0156011170913, (APWW01560);  1011-9050-000, 010084068210229221APWW0156411170925, (APWW01564);  1011-9050-000, 010084068210229221APWW0156511170925, (APWW01565);  1011-9050-000, 010084068210229221APWW0156811170925, (APWW01568);  1011-9050-000, 010084068210229221APWW0156911170925, (APWW01569);  1011-9050-000, 010084068210229221APWW0157011170925, (APWW01570);  1011-9050-000, 010084068210229221APWW0157111170925, (APWW01571);  1011-9050-000, 010084068210229221APWW0157211170925, (APWW01572);  1011-9050-000, 010084068210229221APWW0157311170925, (APWW01573);  1011-9050-000, 010084068210229221APWW0157411170925, (APWW01574);  1011-9050-000, 010084068210229221APWW0157511170925, (APWW01575);  1011-9050-000, 010084068210229221APWW0157611170925, (APWW01576);  1011-9050-000, 010084068210229221APWW0157711170925, (APWW01577);  1011-9050-000, 010084068210229221APWW0157811170925, (APWW01578);  1011-9050-000, 010084068210229221APWW0154311170913, (APWW01543);  1011-9050-000, 010084068210229221APWW0154511170913, (APWW01545);  1011-9050-000, 010084068210229221APWW0154611170913, (APWW01546);  1011-9050-000, 010084068210229221APWW0154811170913, (APWW01548);  1011-9050-000, 010084068210229221APWW0154911170913, (APWW01549);  1011-9050-000, 010084068210229221APWW0155011170913, (APWW01550);  1011-9050-000, 010084068210229221APWW0155111170913, (APWW01551);  1011-9050-000, 010084068210229221APWW0155311170913, (APWW01553);  1011-9050-000, 010084068210229221APWW0155511170913, (APWW01555);  1011-9050-000, 010084068210229221APWW0156111170925, (APWW01561);  1011-9050-000, 010084068210229221APWW0156211170925, (APWW01562);  1011-9050-000, 010084068210229221APWW0156311170925, (APWW01563);  1011-9050-000, 010084068210229221APWW0156611170925, (APWW01566);  1011-9050-000, 010084068210229221APWW0156711170925, (APWW01567);  1011-9050-000, 010084068210229221APWX0077311180323, (APWX00773);  1011-9050-000, 010084068210229221APWX0077411180323, (APWX00774);  1011-9050-000, 010084068210229221APWX0077511180323, (APWX00775);  1011-9050-000, 010084068210229221APWX0077611180323, (APWX00776);  1011-9050-000, 010084068210229221APWX0077711180323, (APWX00777);  1011-9050-000, 010084068210229221APWX0077811180323, (APWX00778);  2097675-001-US, Not Applicable, (5457466-11);  2097675-001-US, Not Applicable, (5457466-12);  2097675-001-US, Not Applicable, (5457466-13);  2097675-001-US, Not Applicable, (5457466-14);  1011-9050-000, 010084068210229221APWW0169711171012, (APWW01697);  1011-9050-000, 010084068210229221APWW0169811171012, (APWW01698);  1011-9050-000, 010084068210229221APWW0169911171012, (APWW01699);  1011-9050-000, 010084068210229221APWW0170011171012, (APWW01700);  1011-9050-000, 010084068210229221APWW0170111171012, (APWW01701);  1011-9050-000, 010084068210229221APWW0091011170501, (APWW00910);  1011-9050-000, 010084068210229221APWW0189411171109, (APWW01894);  1011-9050-000, 010084068210229221APWW0189511171109, (APWW01895);  1011-9050-000, 010084068210229221APWW0189611171109, (APWW01896);  1011-9050-000, 010084068210229221APWW0189711171109, (APWW01897);  1011-9050-000, 010084068210229221APWW0189811171109, (APWW01898);  1011-9050-000, 010084068210229221APWW0189911171109, (APWW01899);  1011-9050-000, 010084068210229221APWW0190011171109, (APWW01900);  1011-9050-000, 010084068210229221APWW0190111171109, (APWW01901);  1011-9050-000, 010084068210229221APWW0190211171109, (APWW01902);  1011-9050-000, 010084068210229221APWW0190311171109, (APWW01903);  1011-9050-000, 010084068210229221APWW0190411171109, (APWW01904);  1011-9050-000, 010084068210229221APWX0029911180118, (APWX00299);  1011-9050-000, 010084068210229221APWX0030011180118, (APWX00300);  1011-9050-000, 010084068210229221APWX0030111180123, (APWX00301);  1011-9050-000, 010084068210229221APWX0030211180123, (APWX00302);  1011-9050-000, 010084068210229221APWX0030311180123, (APWX00303);  1011-9050-000, 010084068210229221APWX0030411180123, (APWX00304);  1011-9050-000, 010084068210229221APWX0030511180123, (APWX00305);  1011-9050-000, 010084068210229221APWX0030611180123, (APWX00306);  1011-9050-000, 010084068210229221APWX0030711180123, (APWX00307);  1011-9050-000, 010084068210229221APWX0028411180118, (APWX00284);  1011-9050-000, 010084068210229221APWX0140811180618, (APWX01408);  1011-9050-000, 010084068210229221APWX0110311180502, (APWX01103);  1011-9050-000, 010084068210229221APWW0215811171214, (APWW02158);  1011-9050-000, 010084068210229221APWW0215911171214, (APWW02159);  1011-9050-000, 010084068210229221APWW0216011171214, (APWW02160);  1011-9050-000, 010084068210229221APWW0216111171214, (APWW02161);  1011-9050-000, 010084068210229221APWW0216211171214, (APWW02162);  1011-9050-000, 010084068210229221APWW0216311171214, (APWW02163);  1011-9050-000, 010084068210229221APWX0020511180115, (APWX00205);  1011-9050-000, 010084068210229221APWX0020611180115, (APWX00206);  1011-9050-000, 010084068210229221APWX0020711180115, (APWX00207);  1011-9050-000, 010084068210229221APWX0020811180115, (APWX00208);  1011-9050-000, 010084068210229221APWX0022111180116, (APWX00221);  1011-9050-000, 010084068210229221APWX0022211180116, (APWX00222);  1011-9050-000, 010084068210229221APWX0022311180116, (APWX00223);  1011-9050-000, 010084068210229221APWX0022411180116, (APWX00224);  1011-9050-000, 010084068210229221APWX0022511180116, (APWX00225);  1011-9050-000, 010084068210229221APWX0022611180116, (APWX00226);  1011-9050-000, 010084068210229221APWX0022711180116, (APWX00227);  1011-9050-000, 010084068210229221APWX0022811180116, (APWX00228);  1011-9050-000, 010084068210229221APWX0022911180116, (APWX00229);  1011-9050-000, 010084068210229221APWX0023011180116, (APWX00230);  1011-9050-000, 010084068210229221APWX0023111180116, (APWX00231);  1011-9050-000, 010084068210229221APWX0023211180116, (APWX00232);  1011-9050-000, 010084068210229221APWX0023311180116, (APWX00233);  1011-9050-000, 010084068210229221APWX0023411180116, (APWX00234);  1011-9050-000, 010084068210229221APWX0023511180116, (APWX00235);  1011-9050-000, 010084068210229221APWX0023611180116, (APWX00236);  1011-9050-000, 010084068210229221APWX0023711180116, (APWX00237);  1011-9050-000, 010084068210229221APWX0023811180116, (APWX00238);  1011-9050-000, 010084068210229221APWX0023911180116, (APWX00239);  1011-9050-000, 010084068210229221APWX0024011180116, (APWX00240);  1011-9050-000, 010084068210229221APWX0021111180115, (APWX00211);  1011-9050-000, 010084068210229221APWX0021211180115, (APWX00212);  1011-9050-000, 010084068210229221APWX0021311180115, (APWX00213);  1011-9050-000, 010084068210229221APWX0068811180315, (APWX00688);  2097675-001-US, Not Applicable, (5471472-1);  1011-9050-000, Not Applicable, (APWT00238);  1011-9050-000, Not Applicable, (APWT02122);  1011-9050-000, Not Applicable, (APWT02126);  1011-9050-000, Not Applicable, (APWT02127);  1011-9050-000, Not Applicable, (APWU00522);  1011-9050-000, Not Applicable, (APWU00525);  1011-9050-000, 010084068210229221APWU0137411150624, (APWU01374);  1011-9050-000, 010084068210229221APWU0183811150819, (APWU01838);  1011-9050-000, 010084068210229221APWU0258811151210, (APWU02588);  1011-9050-000, 010084068210229221APWV0192211160831, (APWV01922);  1011-9050-000, 010084068210229221APWV0206511160927, (APWV02065);  1011-9050-000, 010084068210229221APWV0206611160927, (APWV02066);  1011-9050-000, 010084068210229221APWV0206811160927, (APWV02068);  1011-9050-000, 010084068210229221APWW0027111170117, (APWW00271);  1011-9050-000, 010084068210229221APWW0131311170721, (APWW01313);  1011-9050-000, 010084068210229221APWW0151811170830, (APWW01518);  1011-9050-000, 010084068210229221APWW0160111170927, (APWW01601);  1011-9050-000, 010084068210229221APWX0010611180104, (APWX00106);  1011-9050-000, 010084068210229221APWX0026611180118, (APWX00266);  1011-9050-000, 010084068210229221APWX0026711180118, (APWX00267);  1011-9050-000, 010084068210229221APWX0026811180118, (APWX00268);  1011-9050-000, 010084068210229221APWX0026911180118, (APWX00269);  1011-9050-000, 010084068210229221APWX0027011180118, (APWX00270);  1011-9050-000, 010084068210229221APWX0027111180118, (APWX00271);  1011-9050-000, 010084068210229221APWX0027211180118, (APWX00272);  1011-9050-000, 010084068210229221APWX0027311180118, (APWX00273);  1011-9050-000, 010084068210229221APWX0027411180118, (APWX00274);  1011-9050-000, 010084068210229221APWX0027511180118, (APWX00275);  1011-9050-000, 010084068210229221APWX0107611180502, (APWX01076);  1011-9050-000, 010084068210229221APWX0113511180507, (APWX01135);  1011-9050-000, 010084068210229221APWX0147911180619, (APWX01479);  1011-9050-000, 010084068210229221APWX0162911180711, (APWX01629);  1011-9050-000, 010084068210229221APWX0163011180711, (APWX01630);  1011-9050-000, 010084068210229221APWX0163111180711, (APWX01631);  1011-9050-000, 010084068210229221APWX0163211180711, (APWX01632);  1011-9050-000, 010084068210229221APWX0163311180711, (APWX01633);  1011-9050-000, 010084068210229221APWX0163411180711, (APWX01634);  1011-9050-000, 010084068210229221APWX0163511180711, (APWX01635);  1011-9050-000, 010084068210229221APWX0163611180711, (APWX01636);  1011-9050-000, 010084068210229221APWX0163711180711, (APWX01637);  1011-9050-000, 010084068210229221APWX0163811180711, (APWX01638);  1011-9050-000, 010084068210229221APWW0221411171219, (APWW02214);  1011-9050-000, 010084068210229221APWX0056111180301, (APWX00561);  1011-9050-000, 010084068210229221APWX0056211180301, (APWX00562);  1011-9050-000, Not Applicable, (APWS00101);  1011-9050-000, Not Applicable, (APWT00792);  1011-9050-000, Not Applicable, (APWT02123);  1011-9050-000, Not Applicable, (APWT02124);  1011-9050-000, Not Applicable, (APWT02125);  1011-9050-000, Not Applicable, (APWT02193);  1011-9050-000, Not Applicable, (APWU00518);  1011-9050-000, Not Applicable, (APWU00521);  1011-9050-000, Not Applicable, (APWU00523);  1011-9050-000, 010084068210229221APWU0137111150624, (APWU01371);  1011-9050-000, 010084068210229221APWU0137211150624, (APWU01372);  1011-9050-000, 010084068210229221APWU0183911150819, (APWU01839);  1011-9050-000, 010084068210229221APWU0184411150831, (APWU01844);  1011-9050-000, 010084068210229221APWU0258511151210, (APWU02585);  1011-9050-000, 010084068210229221APWU0258611151210, (APWU02586);  1011-9050-000, 010084068210229221APWV0192311160831, (APWV01923);  1011-9050-000, 010084068210229221APWV0192411160831, (APWV01924);  1011-9050-000, 010084068210229221APWV0192611160831, (APWV01926);  1011-9050-000, 010084068210229221APWV0192711160831, (APWV01927);  1011-9050-000, 010084068210229221APWV0196911160831, (APWV01969);  1011-9050-000, 010084068210229221APWV0197011160831, (APWV01970);  1011-9050-000, 010084068210229221APWW0026711170117, (APWW00267);  1011-9050-000, 010084068210229221APWW0026811170117, (APWW00268);  1011-9050-000, 010084068210229221APWW0026911170117, (APWW00269);  1011-9050-000, 010084068210229221APWW0027011170117, (APWW00270);  1011-9050-000, 010084068210229221APWW0131111170721, (APWW01311);  1011-9050-000, 010084068210229221APWW0131211170721, (APWW01312);  1011-9050-000, 010084068210229221APWW0131411170721, (APWW01314);  1011-9050-000, 010084068210229221APWW0131511170721, (APWW01315);  1011-9050-000, 010084068210229221APWW0134611170724, (APWW01346);  1011-9050-000, 010084068210229221APWW0151711170830, (APWW01517);  1011-9050-000, 010084068210229221APWW0160011170925, (APWW01600);  1011-9050-000, 010084068210229221APWW0160211170927, (APWW01602);  1011-9050-000, 010084068210229221APWX0120311180524, (APWX01203);  1011-9050-000, 010084068210229221APWX0120411180524, (APWX01204);  1011-9050-000, 010084068210229221APWX0120511180524, (APWX01205);  1011-9050-000, 010084068210229221APWX0048111180208, (APWX00481);  1011-9050-000, 010084068210229221APWW0212511171212, (APWW02125);  1011-9050-000, 010084068210229221APWW0212611171212, (APWW02126);  1011-9050-000, 010084068210229221APWW0212711171212, (APWW02127);  1011-9050-000, 010084068210229221APWX0176111180720, (APWX01761);  1011-9050-000, 010084068210229221APWX0095511180417, (APWX00955);  1011-9050-000, 010084068210229221APWX0049111180208, (APWX00491);  2097675-001, Not Applicable, (5471991-1);  2097675-001, Not Applicable, (5478965-1);  1011-9050-000, 010084068210229221APWW0161011170927, (APWW01610);  2097675-001-US, Not Applicable, (4628596-1);  2097675-001-US, Not Applicable, (4628596-2);  2097675-001-US, Not Applicable, (4628596-3);  2097675-001-US, Not Applicable, (4628596-4);  2097675-001-US, Not Applicable, (4628596-5);  2097675-001-US, Not Applicable, (4628596-6);  2097675-001, Not Applicable, (5447955-1);  2097675-001, Not Applicable, (5447958-1);  2097675-001, Not Applicable, (5447959-1);  2097675-001, Not Applicable, (5447960-1);  2097675-001, Not Applicable, (5447961-1);  2097675-001, Not Applicable, (5447962-1);  2097675-001, Not Applicable, (5447963-1);  2097675-001, Not Applicable, (5457385-1);  2097675-001, Not Applicable, (5460715-1);  2097675-001, Not Applicable, (5460715-2);  2097675-001, Not Applicable, (5460715-3);  2097675-001, Not Applicable, (5460715-4);  2097675-001, Not Applicable, (5460715-5);  2097675-001, Not Applicable, (5460715-6);  2097675-001, Not Applicable, (5467309-1);  2097675-001, Not Applicable, (5467309-10);  2097675-001, Not Applicable, (5467309-11);  2097675-001, Not Applicable, (5467309-12);  2097675-001, Not Applicable, (5467309-13);  2097675-001, Not Applicable, (5467309-14);  2097675-001, Not Applicable, (5467309-15);  2097675-001, Not Applicable, (5467309-2);  2097675-001, Not Applicable, (5467309-3);  2097675-001, Not Applicable, (5467309-4);  2097675-001, Not Applicable, (5467309-5);  2097675-001, Not Applicable, (5467309-6);  2097675-001, Not Applicable, (5467309-7);  2097675-001, Not Applicable, (5467309-8);  2097675-001, Not Applicable, (5467309-9);  2097675-001, Not Applicable, (5475515-3);  2097675-001, Not Applicable, (5475515-4);  2097675-001, Not Applicable, (5475516-11);  2097675-001, Not Applicable, (5475516-12);  2097675-001, Not Applicable, (5475516-13);  2097675-001, Not Applicable, (5475516-14);  2097675-001, Not Applicable, (5475516-15);  2097675-001, Not Applicable, (5475516-16);  2097675-001, Not Applicable, (5475516-17);  2097675-001, Not Applicable, (5475516-18);  2097675-001, Not Applicable, (5475516-19);  2097675-001, Not Applicable, (5475516-20);  2097675-001, Not Applicable, (5475528-20);  2097675-001, Not Applicable, (5475528-21);  2097675-001, Not Applicable, (5475528-22);  2097675-001, Not Applicable, (5475528-23);  2097675-001, Not Applicable, (5475528-24);  2097675-001, Not Applicable, (5475528-25);  2097675-001, Not Applicable, (5475528-26);  2097675-001, Not Applicable, (5475528-27);  2097675-001, Not Applicable, (5475528-28);  2097675-001, Not Applicable, (5475528-29);  2097675-001, Not Applicable, (5475528-30);  2097675-001, Not Applicable, (5475528-31);  2097675-001, Not Applicable, (5475528-32);  2097675-001, Not Applicable, (5475528-33);  2097675-001, Not Applicable, (5475528-34);  2097675-001, Not Applicable, (5475528-35);  2097675-001, Not Applicable, (5475528-36);  2097675-001, Not Applicable, (5475528-37);  2097675-001, Not Applicable, (5475528-38);  2097675-001, Not Applicable, (5479475-1);  2097675-001, Not Applicable, (5481655-1);  2097675-001, Not Applicable, (5481655-2);  2097675-001, Not Applicable, (5481655-3);  2097675-001, Not Applicable, (5481655-4);  2097675-001, Not Applicable, (5481655-5);  2097675-001, Not Applicable, (5481655-6);  2097675-001, Not Applicable, (5481655-7);  2097675-001, Not Applicable, (5481656-1);  2097675-001, Not Applicable, (5481656-2);  2097675-001, Not Applicable, (5481656-3);  2097675-001, Not Applicable, (5481656-4);  2097675-001, Not Applicable, (5481906-1);  2097675-001, Not Applicable, (5481906-2);  2097675-001, Not Applicable, (5481906-3);  2097675-001, Not Applicable, (5481906-4);  1011-9050-000, 010084068210229221APWX0035911180131, (APWX00359);  1011-9050-000, 010084068210229221APWX0036011180131, (APWX00360);  1011-9050-000, 010084068210229221APWX0049811180208, (APWX00498);  1011-9050-000, 010084068210229221APWX0074511180319, (APWX00745);  1011-9050-000, 010084068210229221APWX0074611180319, (APWX00746);  1011-9050-000, 010084068210229221APWX0074711180319, (APWX00747);  1011-9050-000, 010084068210229221APWX0074811180319, (APWX00748);  1011-9050-000, 010084068210229221APWX0074911180319, (APWX00749);  1011-9050-000, 010084068210229221APWX0075011180319, (APWX00750);  1011-9050-000, 010084068210229221APWX0076411180323, (APWX00764);  1011-9050-000, 010084068210229221APWX0076511180323, (APWX00765);  1011-9050-000, 010084068210229221APWX0076611180323, (APWX00766);  1011-9050-000, 010084068210229221APWX0076711180323, (APWX00767);  1011-9050-000, 010084068210229221APWX0076811180323, (APWX00768);  1011-9050-000, 010084068210229221APWX0076911180323, (APWX00769);  1011-9050-000, 010084068210229221APWX0111511180507, (APWX01115);  1011-9050-000, 010084068210229221APWX0111611180507, (APWX01116);  1011-9050-000, 010084068210229221APWX0159511180703, (APWX01595);  1011-9050-000, 010084068210229221APWX0159611180703, (APWX01596);  1011-9050-000, 010084068210229221APWX0159711180703, (APWX01597);  1011-9050-000, 010084068210229221APWX0159811180703, (APWX01598);  1011-9050-000, 010084068210229221APWX0159911180703, (APWX01599);  1011-9050-000, 010084068210229221APWX0181111180802, (APWX01811);  1011-9050-000, 010084068210229221APWX0181211180802, (APWX01812);  1011-9050-000, 010084068210229221APWX0181311180802, (APWX01813);  1011-9050-000, 010084068210229221APWX0181911180802, (APWX01819);  1011-9050-000, 010084068210229221APWX0182011180802, (APWX01820);  1011-9050-000, 010084068210229221APWX0182111180802, (APWX01821);  1011-9050-000, 010084068210229221APWX0182211180802, (APWX01822);  1011-9050-000, 010084068210229221APWX0182311180802, (APWX01823);  1011-9050-000, 010084068210229221APWX0182411180802, (APWX01824);  1011-9050-000, 010084068210229221APWX0182511180802, (APWX01825);  1011-9050-000, 010084068210229221APWX0182611180802, (APWX01826);  1011-9050-000, 010084068210229221APWX0186011180803, (APWX01860);  1011-9050-000, 010084068210229221APWX0186111180803, (APWX01861);  1011-9050-000, 010084068210229221APWX0186211180803, (APWX01862);  1011-9050-000, 010084068210229221APWX0186311180803, (APWX01863);  1011-9050-000, 010084068210229221APWX0010311180104, (APWX00103);  1011-9050-000, 010084068210229221APWX0010411180104, (APWX00104);  1011-9050-000, 010084068210229221APWX0037111180131, (APWX00371);  1011-9050-000, 010084068210229221APWX0037211180131, (APWX00372);  1011-9050-000, 010084068210229221APWX0037311180131, (APWX00373);  1011-9050-000, 010084068210229221APWX0037411180131, (APWX00374);  1011-9050-000, 010084068210229221APWX0185911180803, (APWX01859);  1011-9050-000, 010084068210229221APWX0117811180517, (APWX01178);  1011-9050-000, 010084068210229221APWX0117911180517, (APWX01179);  1011-9050-000, 010084068210229221APWX0118011180517, (APWX01180);  1011-9050-000, 010084068210229221APWX0118111180517, (APWX01181);  1011-9050-000, 010084068210229221APWX0118211180517, (APWX01182);  1011-9050-000, 010084068210229221APWX0118311180517, (APWX01183);  1011-9050-000, 010084068210229221APWX0118411180517, (APWX01184);  1011-9050-000, 010084068210229221APWX0118511180517, (APWX01185);  1011-9050-000, 010084068210229221APWX0118611180517, (APWX01186);  1011-9050-000, 010084068210229221APWX0118711180517, (APWX01187);  1011-9050-000, 010084068210229221APWX0118811180517, (APWX01188);  1011-9050-000, 010084068210229221APWX0118911180517, (APWX01189);  1011-9050-000, 010084068210229221APWX0119011180517, (APWX01190);  1011-9050-000, 010084068210229221APWX0119111180524, (APWX01191);  1011-9050-000, 010084068210229221APWX0119211180524, (APWX01192);  1011-9050-000, 010084068210229221APWX0119311180524, (APWX01193);  1011-9050-000, 010084068210229221APWX0119411180524, (APWX01194);  1011-9050-000, 010084068210229221APWX0119511180524, (APWX01195);  1011-9050-000, 010084068210229221APWX0119611180524, (APWX01196);  1011-9050-000, 010084068210229221APWX0119711180524, (APWX01197);  1011-9050-000, 010084068210229221APWX0119811180524, (APWX01198);  1011-9050-000, 010084068210229221APWX0119911180524, (APWX01199);  1011-9050-000, 010084068210229221APWX0173211180720, (APWX01732);  2097675-001, Not Applicable, (5463244-1);  2097675-001, Not Applicable, (5463244-2);  2097675-001, Not Applicable, (5463244-3);  2097675-001, Not Applicable, (5463244-4);  2097675-001, Not Applicable, (5463244-5);  2097675-001, Not Applicable, (5463244-6);  2097675-001, Not Applicable, (5463244-7);  1011-9050-000, 010084068210229221APWX0168811180716, (APWX01688);  2097675-001, Not Applicable, (5472013-11);  2097675-001, Not Applicable, (5472013-13);  2097675-001, Not Applicable, (5472013-2);  2097675-001, Not Applicable, (5472013-9);  2097675-001, Not Applicable, (5477472-1);  2097675-001, Not Applicable, (5477472-2);  2097675-001, Not Applicable, (5477472-3);  2097675-001, Not Applicable, (5477472-4);  1011-9050-000, 010084068210229221APWX0084911180328, (APWX00849);  1011-9050-000, 010084068210229221APWX0048611180208, (APWX00486);  1011-9050-000, 010084068210229221APWX0048911180208, (APWX00489);  1011-9050-000, 010084068210229221APWX0049011180208, (APWX00490);  1011-9050-000, 010084068210229221APWX0037511180131, (APWX00375);  1011-9050-000, 010084068210229221APWX0157111180703, (APWX01571);  2097675-001, Not Applicable, (5473049-1);  2097675-001, Not Applicable, (5473049-2);  1011-9050-000, 010084068210229221APWX0084511180328, (APWX00845);  1011-9050-000, 010084068210229221APWX0095311180417, (APWX00953);  1011-9050-000, 010084068210229221APWX0078311180323, (APWX00783);  1011-9050-000, 010084068210229221APWX0148011180619, (APWX01480);  1011-9050-000, 010084068210229221APWX0148111180625, (APWX01481);  1011-9050-000, 010084068210229221APWX0168111180716, (APWX01681);  1011-9050-000, 010084068210229221APWX0168211180716, (APWX01682);  1011-9050-000, 010084068210229221APWX0168311180716, (APWX01683);  1011-9050-000, 010084068210229221APWX0168411180716, (APWX01684);  1011-9050-000, 010084068210229221APWX0168511180716, (APWX01685);  1011-9050-000, 010084068210229221APWX0168611180716, (APWX01686);  2097675-001, Not Applicable, (5456564-1);  2097675-001, Not Applicable, (5456564-10);  2097675-001, Not Applicable, (5456564-11);  2097675-001, Not Applicable, (5456564-12);  2097675-001, Not Applicable, (5456564-13);  2097675-001, Not Applicable, (5456564-14);  2097675-001, Not Applicable, (5456564-2);  2097675-001, Not Applicable, (5456564-3);  2097675-001, Not Applicable, (5456564-4);  2097675-001, Not Applicable, (5456564-5);  2097675-001, Not Applicable, (5456564-6);  2097675-001, Not Applicable, (5456564-7);  2097675-001, Not Applicable, (5456564-8);  2097675-001, Not Applicable, (5456564-9);  2097675-001, Not Applicable, (5463632-1);  2097675-001, Not Applicable, (5463632-10);  2097675-001, Not Applicable, (5463632-11);  2097675-001, Not Applicable, (5463632-12);  2097675-001, Not Applicable, (5463632-13);  2097675-001, Not Applicable, (5463632-14);  2097675-001, Not Applicable, (5463632-15);  2097675-001, Not Applicable, (5463632-16);  2097675-001, Not Applicable, (5463632-17);  2097675-001, Not Applicable, (5463632-18);  2097675-001, Not Applicable, (5463632-19);  2097675-001, Not Applicable, (5463632-2);  2097675-001, Not Applicable, (5463632-20);  2097675-001, Not Applicable, (5463632-21);  2097675-001, Not Applicable, (5463632-3);  2097675-001, Not Applicable, (5463632-4);  2097675-001, Not Applicable, (5463632-5);  2097675-001, Not Applicable, (5463632-6);  2097675-001, Not Applicable, (5463632-7);  2097675-001, Not Applicable, (5463632-8);  2097675-001, Not Applicable, (5463632-9);  1011-9050-000, 010084068210229221APWX0011911180104, (APWX00119);  1011-9050-000, 010084068210229221APWX0012011180104, (APWX00120);  1011-9050-000, 010084068210229221APWX0012111180104, (APWX00121);  1011-9050-000, 010084068210229221APWX0012211180104, (APWX00122);  1011-9050-000, 010084068210229221APWX0012311180104, (APWX00123);  1011-9050-000, 010084068210229221APWX0012411180104, (APWX00124);  1011-9050-000, 010084068210229221APWX0012511180104, (APWX00125);  1011-9050-000, 010084068210229221APWX0012611180104, (APWX00126);  1011-9050-000, 010084068210229221APWX0012711180104, (APWX00127);  1011-9050-000, 010084068210229221APWX0012811180104, (APWX00128);  1011-9050-000, 010084068210229221APWX0012911180104, (APWX00129);  1011-9050-000, 010084068210229221APWX0013011180104, (APWX00130);  1011-9050-000, 010084068210229221APWX0013111180104, (APWX00131);  1011-9050-000, 010084068210229221APWX0013211180104, (APWX00132);  1011-9050-000, 010084068210229221APWX0013311180104, (APWX00133);  1011-9050-000, 010084068210229221APWX0013411180104, (APWX00134);  1011-9050-000, 010084068210229221APWX0117611180517, (APWX01176);  2097675-001, Not Applicable, (5474921-1);  2097675-001, Not Applicable, (5474921-2);  2097675-001, Not Applicable, (5474921-3);  2097675-001, Not Applicable, (5474921-4);  2097675-001, Not Applicable, (5474921-5);  2097675-001, Not Applicable, (5469427-1);  2097675-001, Not Applicable, (5469427-2);  2097675-001, Not Applicable, (5469427-3);  2097675-001, Not Applicable, (5469427-4);  2097675-001, Not Applicable, (5469427-5);  2097675-001, Not Applicable, (5469427-6);  1011-9050-000, 010084068210229221APWX0087911180328, (APWX00879);  1011-9050-000, 010084068210229221APWX0039411180201, (APWX00394);  2097675-001, Not Applicable, (5466893-1);  2097675-001, Not Applicable, (5466893-10);  2097675-001, Not Applicable, (5466893-2);  2097675-001, Not Applicable, (5466893-3);  2097675-001, Not Applicable, (5466893-4);  2097675-001, Not Applicable, (5466893-5);  2097675-001, Not Applicable, (5466893-6);  2097675-001, Not Applicable, (5466893-7);  2097675-001, Not Applicable, (5466893-8);  2097675-001, Not Applicable, (5466893-9);  2097675-001, Not Applicable, (5466894-1);  2097675-001, Not Applicable, (5466894-2);  2097675-001, Not Applicable, (5466894-3);  2097675-001, Not Applicable, (5466894-4);  2097675-001, Not Applicable, (5466894-5);  1011-9050-000, 010084068210229221APWX0034811180131, (APWX00348);  1011-9050-000, 010084068210229221APWX0034911180131, (APWX00349);  1011-9050-000, 010084068210229221APWX0035011180131, (APWX00350);  1011-9050-000, 010084068210229221APWX0035111180131, (APWX00351);  1011-9050-000, 010084068210229221APWX0035211180131, (APWX00352);  1011-9050-000, 010084068210229221APWX0188111180803, (APWX01881);  1011-9050-000, 010084068210229221APWW0223711171219, (APWW02237);  1011-9050-000, 010084068210229221APWX0087411180328, (APWX00874);  1011-9050-000, 010084068210229221APWX0123311180531, (APWX01233);  1011-9050-000, 010084068210229221APWX0127511180604, (APWX01275);  1011-9050-000, 010084068210229221APWX0127611180604, (APWX01276);  2097675-001, Not Applicable, (5478740-1);  2097675-001, Not Applicable, (5478740-2);  2097675-001, Not Applicable, (5478740-3);  2097675-001, Not Applicable, (5478740-4);  2097675-001, Not Applicable, (5478740-5);  2097675-001, Not Applicable, (5478740-6);  2097675-001, Not Applicable, (5478740-7);  2097675-001, Not Applicable, (5478740-8);  2097675-001, Not Applicable, (5478740-9);  1011-9050-000, 010084068210229221APWX0152011180625, (APWX01520);  1011-9050-000, 010084068210229221APWX0052711180222, (APWX00527);  1011-9050-000, 010084068210229221APWX0052811180222, (APWX00528);  1011-9050-000, 010084068210229221APWX0180511180720, (APWX01805);  2097675-DE, Not Applicable, (5454727-1);  2097675-DE, Not Applicable, (5474323-2);  2097675-DE, Not Applicable, (5474323-3);  2097675-DE, Not Applicable, (5474323-4);  2097675-DE, Not Applicable, (5474323-5);  2097675-DE, Not Applicable, (5474323-6);  2097675-DE, Not Applicable, (5474323-7);  2097675-DE, Not Applicable, (5474323-8);  2097675-DE, Not Applicable, (5474323-9);  2097675-DE, Not Applicable, (5476483-1);  2097675-DE, Not Applicable, (5476483-2);  2097675-DE, Not Applicable, (5476483-3);  2097675-DE, Not Applicable, (5476483-4);  2097675-DE, Not Applicable, (5476483-5);  2097675-DE, Not Applicable, (5476483-6);  2097675-DE, Not Applicable, (5476483-7);  2097675-DE, Not Applicable, (5476483-8);  2097675-DE, Not Applicable, (5476965-1);  2097675-DE, Not Applicable, (5476965-2);  2097675-DE, Not Applicable, (5476965-3);  1011-9050-000, Not Applicable, (APWS00251);  1011-9050-000, Not Applicable, (APWS00252);  1011-9050-000, Not Applicable, (APWT00199);  1011-9050-000, Not Applicable, (APWT00200);  1011-9050-000, Not Applicable, (APWT00201);  1011-9050-000, Not Applicable, (APWT00927);  1011-9050-000, Not Applicable, (APWT00928);  1011-9050-000, Not Applicable, (APWT02530);  1011-9050-000, Not Applicable, (APWU01117);  1011-9050-000, Not Applicable, (APWU01118);  1011-9050-000, Not Applicable, (APWU01119);  1011-9050-000, Not Applicable, (APWU01120);  1011-9050-000, Not Applicable, (APWU01121);  1011-9050-000, Not Applicable, (APWU01122);  1011-9050-000, 010084068210229221APWV0076911160329, (APWV00769);  1011-9050-000, 010084068210229221APWW0062011170307, (APWW00620);  1011-9050-000, 010084068210229221APWW0062411170307, (APWW00624);  1011-9050-000, 010084068210229221APWW0146811170823, (APWW01468);  1011-9050-000, 010084068210229221APWW0146911170823, (APWW01469);  1011-9050-000, 010084068210229221APWW0147011170823, (APWW01470);  1011-9050-000, 010084068210229221APWW0147111170823, (APWW01471);  1011-9050-000, 010084068210229221APWW0173911171013, (APWW01739);  1011-9050-000, 010084068210229221APWW0174011171013, (APWW01740);  1011-9050-000, 010084068210229221APWW0183811171026, (APWW01838);  1011-9050-000, 010084068210229221APWW0183911171026, (APWW01839);  1011-9050-000, 010084068210229221APWX0057911180301, (APWX00579);  1011-9050-000, 010084068210229221APWX0082611180328, (APWX00826);  1011-9050-000, 010084068210229221APWX0120011180524, (APWX01200);  1011-9050-000, 010084068210229221APWX0120211180524, (APWX01202);  2097675-NL, Not Applicable, (5473625-8);  2097675-NL, Not Applicable, (5481369-10);  2097675-NL, Not Applicable, (5481369-11);  2097675-NL, Not Applicable, (5481369-12);  1011-9050-000, Not Applicable, (APWS00138);  1011-9050-000, Not Applicable, (APWS00140);  1011-9050-000, Not Applicable, (APWT00900);  1011-9050-000, Not Applicable, (APWT00901);  1011-9050-000, Not Applicable, (APWT02055);  1011-9050-000, Not Applicable, (APWT02681);  1011-9050-000, 010084068210229221APWV0016511160108, (APWV00165);  1011-9050-000, 010084068210229221APWW0067511170323, (APWW00675);  1011-9050-000, 010084068210229221APWW0107211170509, (APWW01072);  1011-9050-000, 010084068210229221APWW0107311170509, (APWW01073);  1011-9050-000, 010084068210229221APWW0107411170509, (APWW01074);  1011-9050-000, 010084068210229221APWW0107511170509, (APWW01075);  1011-9050-000, 010084068210229221APWW0107611170509, (APWW01076);  1011-9050-000, 010084068210229221APWW0107711170509, (APWW01077);  1011-9050-000, 010084068210229221APWW0107811170509, (APWW01078);  1011-9050-000, 010084068210229221APWW0107911170509, (APWW01079);  1011-9050-000, 010084068210229221APWW0108011170509, (APWW01080);  1011-9050-000, 010084068210229221APWW0131611170721, (APWW01316);  1011-9050-000, 010084068210229221APWW0131711170721, (APWW01317);  1011-9050-000, 010084068210229221APWW0141711170809, (APWW01417);  1011-9050-000, 010084068210229221APWW0142211170809, (APWW01422);  1011-9050-000, 010084068210229221APWW0158611170925, (APWW01586);  1011-9050-000, 010084068210229221APWW0158711170925, (APWW01587);  1011-9050-000, 010084068210229221APWW0176711171013, (APWW01767);  1011-9050-000, 010084068210229221APWW0184311171026, (APWW01843);  1011-9050-000, 010084068210229221APWW0190511171109, (APWW01905);  1011-9050-000, 010084068210229221APWW0190611171109, (APWW01906);  1011-9050-000, 010084068210229221APWW0196311171122, (APWW01963);  1011-9050-000, 010084068210229221APWW0196411171122, (APWW01964);  1011-9050-000, 010084068210229221APWW0196511171122, (APWW01965);  1011-9050-000, 010084068210229221APWW0199111171127, (APWW01991);  1011-9050-000, 010084068210229221APWW0199211171127, (APWW01992);  1011-9050-000, 010084068210229221APWX0059011180307, (APWX00590);  1011-9050-000, 010084068210229221APWX0065011180309, (APWX00650);  1011-9050-000, 010084068210229221APWX0065111180309, (APWX00651);  1011-9050-000, 010084068210229221APWX0065211180309, (APWX00652);  1011-9050-000, 010084068210229221APWX0065311180309, (APWX00653);  1011-9050-000, 010084068210229221APWX0065411180309, (APWX00654);  1011-9050-000, 010084068210229221APWX0065511180309, (APWX00655);  1011-9050-000, 010084068210229221APWX0083411180328, (APWX00834);  1011-9050-000, 010084068210229221APWX0083511180328, (APWX00835);  1011-9050-000, 010084068210229221APWX0083611180328, (APWX00836);  1011-9050-000, 010084068210229221APWX0083711180328, (APWX00837);  1011-9050-000, 010084068210229221APWX0083811180328, (APWX00838);  1011-9050-000, 010084068210229221APWX0083911180328, (APWX00839);  1011-9050-000, 010084068210229221APWX0084011180328, (APWX00840);  1011-9050-000, 010084068210229221APWX0084111180328, (APWX00841);  1011-9050-000, 010084068210229221APWX0096411180417, (APWX00964);  1011-9050-000, 010084068210229221APWX0096511180417, (APWX00965);  1011-9050-000, 010084068210229221APWX0096611180417, (APWX00966);  1011-9050-000, 010084068210229221APWX0096711180417, (APWX00967);  1011-9050-000, 010084068210229221APWX0101911180417, (APWX01019);  1011-9050-000, 010084068210229221APWX0102011180417, (APWX01020);  1011-9050-000, 010084068210229221APWX0102111180417, (APWX01021);  1011-9050-000, 010084068210229221APWX0102211180417, (APWX01022);  1011-9050-000, 010084068210229221APWX0102311180417, (APWX01023);  1011-9050-000, 010084068210229221APWX0102411180417, (APWX01024);  1011-9050-000, 010084068210229221APWX0187211180803, (APWX01872);  1011-9050-000, 010084068210229221APWX0187311180803, (APWX01873);  1011-9050-000, 010084068210229221APWX0187411180803, (APWX01874);  1011-9050-000, 010084068210229221APWX0187511180803, (APWX01875);  1011-9050-000, 010084068210229221APWX0187611180803, (APWX01876);  1011-9050-000, 010084068210229221APWX0187711180803, (APWX01877);  1011-9050-000, 010084068210229221APWX0187811180803, (APWX01878);  1011-9050-000, 010084068210229221APWX0187911180803, (APWX01879);  1011-9050-000, 010084068210229221APWX0090511180328, (APWX00905);  1011-9050-000, 010084068210229221APWX0188011180803, (APWX01880);  1011-9050-000, 010084068210229221APWX0163911180711, (APWX01639);  1011-9050-000, 010084068210229221APWX0164011180711, (APWX01640);  1011-9050-000, 010084068210229221APWX0164111180716, (APWX01641);  1011-9050-000, 010084068210229221APWX0164211180716, (APWX01642);  1011-9050-000, 010084068210229221APWX0164311180716, (APWX01643);  1011-9050-000, 010084068210229221APWX0164411180716, (APWX01644);  1011-9050-000, 010084068210229221APWX0164511180716, (APWX01645);  1011-9050-000, 010084068210229221APWX0036111180131, (APWX00361);  1011-9050-000, 010084068210229221APWX0036211180131, (APWX00362);  1011-9050-000, 010084068210229221APWX0055411180301, (APWX00554);  1011-9050-000, 010084068210229221APWX0055511180301, (APWX00555);  1011-9050-000, 010084068210229221APWX0055611180301, (APWX00556);  1011-9050-000, 010084068210229221APWX0055711180301, (APWX00557);  1011-9050-000, 010084068210229221APWX0055811180301, (APWX00558);  1011-9050-000, 010084068210229221APWX0160011180703, (APWX01600);  2097675-PT-BR, Not Applicable, (5478265-1);  2097675-PT-BR, Not Applicable, (5478265-2);  2097675-PT-BR, Not Applicable, (5478265-3);  2097675-PT-BR, Not Applicable, (5478265-4);  1011-9050-000, 010084068210229221APWX0013811180104, (APWX00138);  1011-9050-000, 010084068210229221APWX0013911180104, (APWX00139);  1011-9050-000, 010084068210229221APWX0188411180803, (APWX01884);  1011-9050-000, 010084068210229221APWX0014511180105, (APWX00145);  1011-9050-000, 010084068210229221APWX0014611180105.(APWX00146);  2097675-001, Not Applicable, (5465689-1);  2097675-001, Not Applicable, (5465689-10);  2097675-001, Not Applicable, (5465689-11);  2097675-001, Not Applicable, (5465689-12);  2097675-001, Not Applicable, (5465689-13);  2097675-001, Not Applicable, (5465689-2);  2097675-001, Not Applicable, (5465689-3);  2097675-001, Not Applicable, (5465689-4);  2097675-001, Not Applicable, (5465689-5);  2097675-001, Not Applicable, (5465689-6);  2097675-001, Not Applicable, (5465689-7);  2097675-001, Not Applicable, (5465689-8);  2097675-001, Not Applicable, (5465689-9);  1011-9050-000, 010084068210229221APWW0214511171214, (APWW02145);  1011-9050-000, 010084068210229221APWW0214611171214, (APWW02146);  1011-9050-000, 010084068210229221APWW0214711171214, (APWW02147);  1011-9050-000, 010084068210229221APWX0019911180115, (APWX00199);  1011-9050-000, 010084068210229221APWX0020011180115, (APWX00200);  1011-9050-000, 010084068210229221APWX0032111180123, (APWX00321);  1011-9050-000, 010084068210229221APWX0032211180123, (APWX00322);  1011-9050-000, 010084068210229221APWX0032311180123, (APWX00323);  1011-9050-000, 010084068210229221APWX0032411180123, (APWX00324);  1011-9050-000, 010084068210229221APWX0032511180123, (APWX00325);  1011-9050-000, 010084068210229221APWX0032611180123, (APWX00326);  1011-9050-000, 010084068210229221APWX0032711180123, (APWX00327);  1011-9050-000, 010084068210229221APWX0013711180104, (APWX00137);  1011-9050-000, 010084068210229221APWX0186811180803, (APWX01868);  1011-9050-000, 010084068210229221APWX0057511180301, (APWX00575);  1011-9050-000, 010084068210229221APWX0099011180417, (APWX00990);  2097675-001, Not Applicable, (5475700-1);  1011-9050-000, 010084068210229221APWX0127811180604, (APWX01278);  1011-9050-000, 010084068210229221APWW0145711170823, (APWW01457);  1011-9050-000, 010084068210229221APWX0160111180711, (APWX01601);  1011-9050-000, Not Applicable, (APWT00140);  1011-9050-000, Not Applicable, (APWT02984);  1011-9050-000, 010084068210229221APWV0220211161021, (APWV02202);  1011-9050-000, 010084068210229221APWW0205111171206, (APWW02051);  1011-9050-000, 010084068210229221APWW0205211171206, (APWW02052);  1011-9050-000, 010084068210229221APWW0090311170501, (APWW00903);  1011-9050-000, 010084068210229221APWW0145811170823, (APWW01458);  1011-9050-000, 010084068210229221APWW0145911170823, (APWW01459);  1011-9050-000, 010084068210229221APWW0146011170823, (APWW01460);  1011-9050-000, 010084068210229221APWW0146111170823, (APWW01461);  1011-9050-000, 010084068210229221APWW0146211170823, (APWW01462);  1011-9050-000, 010084068210229221APWX0140611180618, (APWX01406);  1011-9050-000, 010084068210229221APWX0165411180716, (APWX01654);  1011-9050-000, 010084068210229221APWX0165511180716, (APWX01655);  1011-9050-000, 010084068210229221APWX0165611180716, (APWX01656);  1011-9050-000, 010084068210229221APWX0165711180716, (APWX01657);  1011-9050-000, 010084068210229221APWX0165811180716, (APWX01658);  1011-9050-000, 010084068210229221APWX0104511180427, (APWX01045);  1011-9050-000, 010084068210229221APWX0031011180123, (APWX00310);  1011-9050-000, 010084068210229221APWX0031111180123, (APWX00311);  1011-9050-000, 010084068210229221APWX0031211180123, (APWX00312);  1011-9050-000, 010084068210229221APWX0031311180123, (APWX00313);  1011-9050-000, 010084068210229221APWX0031411180123, (APWX00314);  1011-9050-000, 010084068210229221APWX0031511180123, (APWX00315);  1011-9050-000, 010084068210229221APWX0031611180123, (APWX00316);  1011-9050-000, 010084068210229221APWX0031711180123, (APWX00317);  1011-9050-000, 010084068210229221APWX0031811180123, (APWX00318);  1011-9050-000, 010084068210229221APWX0031911180123, (APWX00319);  1011-9050-000, 010084068210229221APWX0032011180123, (APWX00320);  1011-9050-000, 010084068210229221APWX0032811180123, (APWX00328);  1011-9050-000, 010084068210229221APWX0032911180123, (APWX00329);  1011-9050-000, 010084068210229221APWX0033011180123, (APWX00330);  1011-9050-000, 010084068210229221APWX0033111180123, (APWX00331);  1011-9050-000, 010084068210229221APWX0033211180123, (APWX00332);  1011-9050-000, 010084068210229221APWX0033311180123, (APWX00333);  1011-9050-000, 010084068210229221APWX0033411180123, (APWX00334);  1011-9050-000, 010084068210229221APWX0033511180123, (APWX00335);  1011-9050-000, 010084068210229221APWX0034211180131, (APWX00342);  1011-9050-000, 010084068210229221APWX0034311180131, (APWX00343);  1011-9050-000, 010084068210229221APWX0034411180131, (APWX00344);  1011-9050-000, 010084068210229221APWX0034511180131, (APWX00345);  1011-9050-000, 010084068210229221APWX0034611180131, (APWX00346);  1011-9050-000, 010084068210229221APWX0034711180131, (APWX00347);  1011-9050-000, 010084068210229221APWX0016011180105, (APWX00160);  1011-9050-000, 010084068210229221APWX0016111180105, (APWX00161);  1011-9050-000, 010084068210229221APWX0016211180105, (APWX00162);  1011-9050-000, 010084068210229221APWV0120411160609, (APWV01204);  1011-9050-000, 010084068210229221APWV0120511160609, (APWV01205);  1011-9050-000, 010084068210229221APWV0120611160609, (APWV01206);  1011-9050-000, 010084068210229221APWV0120711160609, (APWV01207);  1011-9050-000, 010084068210229221APWV0120811160609, (APWV01208);  1011-9050-000, 010084068210229221APWV0120911160609, (APWV01209);  1011-9050-000, 010084068210229221APWV0121011160609, (APWV01210);  1011-9050-000, 010084068210229221APWV0121111160609, (APWV01211);  1011-9050-000, 010084068210229221APWV0121211160609, (APWV01212);  1011-9050-000, 010084068210229221APWV0121311160609, (APWV01213);  1011-9050-000, 010084068210229221APWV0121411160609, (APWV01214);  1011-9050-000, 010084068210229221APWV0121511160609, (APWV01215);  1011-9050-000, 010084068210229221APWV0121611160609, (APWV01216);  1011-9050-000, 010084068210229221APWV0121711160609, (APWV01217);  1011-9050-000, 010084068210229221APWV0175311160823, (APWV01753);  1011-9050-000, 010084068210229221APWV0175411160823, (APWV01754);  1011-9050-000, 010084068210229221APWV0175511160823, (APWV01755);  1011-9050-000, 010084068210229221APWV0175611160823, (APWV01756);  1011-9050-000, 010084068210229221APWV0175711160823, (APWV01757);  1011-9050-000, 010084068210229221APWV0175811160823, (APWV01758);  1011-9050-000, 010084068210229221APWV0175911160823, (APWV01759);  1011-9050-000, 010084068210229221APWV0176011160823, (APWV01760);  1011-9050-000, 010084068210229221APWW0191211171109, (APWW01912);  1011-9050-000, 010084068210229221APWW0191311171109, (APWW01913);  1011-9050-000, 010084068210229221APWW0191411171109, (APWW01914);  1011-9050-000, 010084068210229221APWW0191511171109, (APWW01915);  1011-9050-000, 010084068210229221APWW0191611171109, (APWW01916);  1011-9050-000, 010084068210229221APWW0191711171109, (APWW01917);  1011-9050-000, 010084068210229221APWW0191811171109, (APWW01918);  1011-9050-000, 010084068210229221APWW0191911171109, (APWW01919);  1011-9050-000, 010084068210229221APWW0192011171109, (APWW01920);  2097675-001, Not Applicable, (5474992-1);  2097675-001, Not Applicable, (5474992-10);  2097675-001, Not Applicable, (5474992-11);  2097675-001, Not Applicable, (5474992-12);  2097675-001, Not Applicable, (5474992-13);  2097675-001, Not Applicable, (5474992-14);  2097675-001, Not Applicable, (5474992-15);  2097675-001, Not Applicable, (5474992-16);  2097675-001, Not Applicable, (5474992-17);  2097675-001, Not Applicable, (5474992-18);  2097675-001, Not Applicable, (5474992-19);  2097675-001, Not Applicable, (5474992-2);  2097675-001, Not Applicable, (5474992-20);  2097675-001, Not Applicable, (5474992-21);  2097675-001, Not Applicable, (5474992-22);  2097675-001, Not Applicable, (5474992-23);  2097675-001, Not Applicable, (5474992-24);  2097675-001, Not Applicable, (5474992-25);  2097675-001, Not Applicable, (5474992-3);  2097675-001, Not Applicable, (5474992-4);  2097675-001, Not Applicable, (5474992-5);  2097675-001, Not Applicable, (5474992-6);  2097675-001, Not Applicable, (5474992-7);  2097675-001, Not Applicable, (5474992-8);  2097675-001, Not Applicable, (5474992-9);  1011-9050-000, 010084068210229221APWX0100511180417, (APWX01005);  1011-9050-000, 010084068210229221APWX0100611180417, (APWX01006);  1011-9050-000, 010084068210229221APWX0100711180417, (APWX01007);  1011-9050-000, 010084068210229221APWX0100811180417, (APWX01008);  1011-9050-000, 010084068210229221APWX0100911180417, (APWX01009);  1011-9050-000, 010084068210229221APWX0101011180417, (APWX01010);  1011-9050-000, 010084068210229221APWX0101111180417, (APWX01011);  1011-9050-000, 010084068210229221APWX0101211180417, (APWX01012);  1011-9050-000, 010084068210229221APWX0105111180427, (APWX01051);  1011-9050-000, 010084068210229221APWX0105211180427, (APWX01052);  1011-9050-000, 010084068210229221APWX0127111180604, (APWX01271);  1011-9050-000, 010084068210229221APWW0203611171130, (APWW02036);  1011-9050-000, 010084068210229221APWW0203711171130, (APWW02037);  1011-9050-000, 010084068210229221APWW0203811171130, (APWW02038);  1011-9050-000, 010084068210229221APWV0150611160719, (APWV01506);  1011-9050-000, 010084068210229221APWV0150711160719, (APWV01507);  1011-9050-000, 010084068210229221APWV0150811160719, (APWV01508);  1011-9050-000, 010084068210229221APWV0150911160719, (APWV01509);  1011-9050-000, 010084068210229221APWV0151011160719, (APWV01510);  1011-9050-000, 010084068210229221APWV0151111160719, (APWV01511);  1011-9050-000, 010084068210229221APWV0151211160719, (APWV01512);  1011-9050-000, 010084068210229221APWV0151311160719, (APWV01513);  1011-9050-000, 010084068210229221APWV0152811160719, (APWV01528);  1011-9050-000, 010084068210229221APWV0152911160719, (APWV01529);  1011-9050-000, 010084068210229221APWV0153011160719, (APWV01530);  1011-9050-000, 010084068210229221APWV0153111160719, (APWV01531);  1011-9050-000, 010084068210229221APWV0153211160719, (APWV01532);  1011-9050-000, 010084068210229221APWV0153311160719, (APWV01533);  1011-9050-000, 010084068210229221APWV0153411160719, (APWV01534);  1011-9050-000, 010084068210229221APWV0232511161102, (APWV02325);  1011-9050-000, 010084068210229221APWW0084111170412, (APWW00841);  1011-9050-000, 010084068210229221APWW0084311170412, (APWW00843);  1011-9050-000, 010084068210229221APWW0084411170412, (APWW00844);  1011-9050-000, 010084068210229221APWW0090211170501, (APWW00902);  1011-9050-000, 010084068210229221APWW0090511170501, (APWW00905);  1011-9050-000, 010084068210229221APWW0090611170501, (APWW00906);  1011-9050-000, 010084068210229221APWW0093311170505, (APWW00933);  1011-9050-000, 010084068210229221APWW0093411170505, (APWW00934);  1011-9050-000, 010084068210229221APWW0147511170823, (APWW01475);  1011-9050-000, 010084068210229221APWW0147611170823, (APWW01476);  1011-9050-000, 010084068210229221APWW0203911171130, (APWW02039);  1011-9050-000, 010084068210229221APWW0204011171130, (APWW02040);  1011-9050-000, 010084068210229221APWW0204111171130, (APWW02041);  1011-9050-000, 010084068210229221APWW0204211171130, (APWW02042);  1011-9050-000, 010084068210229221APWW0204311171130, (APWW02043);  1011-9050-000, 010084068210229221APWW0204411171130, (APWW02044);  1011-9050-000, 010084068210229221APWW0204511171130, (APWW02045);  1011-9050-000, 010084068210229221APWX0184411180803, (APWX01844);  1011-9050-000, 010084068210229221APWX0184511180803, (APWX01845);  1011-9050-000, 010084068210229221APWX0184611180803, (APWX01846);  1011-9050-000, 010084068210229221APWX0184711180803, (APWX01847);  1011-9050-000, 010084068210229221APWX0184811180803, (APWX01848);  1011-9050-000, 010084068210229221APWX0184911180803, (APWX01849);  1011-9050-000, 010084068210229221APWX0185011180803, (APWX01850);  1011-9050-000, 010084068210229221APWX0185111180803, (APWX01851);  1011-9050-000, 010084068210229221APWX0185211180803, (APWX01852);  1011-9050-000, 010084068210229221APWX0185311180803, (APWX01853);  1011-9050-000, 010084068210229221APWX0185411180803, (APWX01854);  1011-9050-000, 010084068210229221APWX0047711180208, (APWX00477);  1011-9050-000, 010084068210229221APWX0047811180208, (APWX00478);  1011-9050-000, 010084068210229221APWX0047911180208, (APWX00479);  2097675-001, Not Applicable, (5460180-1);  1011-9050-000, Not Applicable, (APWU00119);  1011-9050-000, 010084068210229221APWU0223011151028, (APWU02230);  1011-9050-000, 010084068210229221APWU0223111151028, (APWU02231);  1011-9050-000, 010084068210229221APWU0251211151202, (APWU02512);  1011-9050-000, 010084068210229221APWW0066611170322, (APWW00666);  1011-9050-000, 010084068210229221APWX0015911180105, (APWX00159);  1011-9050-000, 010084068210229221APWX0108711180502, (APWX01087);  1011-9050-000, 010084068210229221APWX0109011180502, (APWX01090);  1011-9050-000, 010084068210229221APWX0109111180502, (APWX01091);  1011-9050-000, 010084068210229221APWX0109211180502, (APWX01092);  1011-9050-000, 010084068210229221APWX0109311180502, (APWX01093);  1011-9050-000, 010084068210229221APWX0111811180507, (APWX01118);  1011-9050-000, 010084068210229221APWX0111911180507, (APWX01119);  1011-9050-000, 010084068210229221APWX0112011180507, (APWX01120);  1011-9050-000, 010084068210229221APWX0112511180507, (APWX01125);  2097675-001, Not Applicable, (5475858-1);  2097675-001, Not Applicable, (5475858-10);  2097675-001, Not Applicable, (5475858-11);  2097675-001, Not Applicable, (5475858-12);  2097675-001, Not Applicable, (5475858-13);  2097675-001, Not Applicable, (5475858-14);  2097675-001, Not Applicable, (5475858-15);  2097675-001, Not Applicable, (5475858-16);  2097675-001, Not Applicable, (5475858-17);  2097675-001, Not Applicable, (5475858-18);  2097675-001, Not Applicable, (5475858-2);  2097675-001, Not Applicable, (5475858-3);  2097675-001, Not Applicable, (5475858-4);  2097675-001, Not Applicable, (5475858-5);  2097675-001, Not Applicable, (5475858-6);  2097675-001, Not Applicable, (5475858-7);  2097675-001, Not Applicable, (5475858-8);  2097675-001, Not Applicable, (5475858-9);  1011-9050-000, 010084068210229221APWW0160511170927, (APWW01605);  1011-9050-000, 010084068210229221APWX0015811180105, (APWX00158);  1011-9050-000, 010084068210229221APWX0043311180206, (APWX00433);  1011-9050-000, 010084068210229221APWX0120911180524, (APWX01209);  1011-9050-000, 010084068210229221APWX0121011180524, (APWX01210);  1011-9050-000, 010084068210229221APWX0121111180530, (APWX01211);  2097675-001, Not Applicable, (5469262-24);  1011-9050-000, Not Applicable, (APWU00391);  1011-9050-000, 010084068210229221APWV0127511160609, (APWV01275);  1011-9050-000, 010084068210229221APWV0204311160927, (APWV02043);  1011-9050-000, 010084068210229221APWV0204411160927, (APWV02044);  1011-9050-000, 010084068210229221APWV0204511160927, (APWV02045);  1011-9050-000, 010084068210229221APWV0204611160927, (APWV02046);  1011-9050-000, 010084068210229221APWV0204711160927, (APWV02047);  1011-9050-000, 010084068210229221APWV0204811160927, (APWV02048);  1011-9050-000, 010084068210229221APWV0204911160927, (APWV02049);  1011-9050-000, 010084068210229221APWV0205011160927, (APWV02050);  1011-9050-000, 010084068210229221APWV0205111160927, (APWV02051);  1011-9050-000, 010084068210229221APWV0205211160927, (APWV02052);  1011-9050-000, 010084068210229221APWV0205311160927, (APWV02053);  1011-9050-000, 010084068210229221APWV0205411160927, (APWV02054);  1011-9050-000, 010084068210229221APWV0205511160927, (APWV02055);  1011-9050-000, 010084068210229221APWV0205611160927, (APWV02056);  1011-9050-000, 010084068210229221APWV0205711160927, (APWV02057);  1011-9050-000, 010084068210229221APWV0205811160927, (APWV02058);  1011-9050-000, 010084068210229221APWV0205911160927, (APWV02059);  1011-9050-000, 010084068210229221APWV0206011160927, (APWV02060);  1011-9050-000, 010084068210229221APWV0206111160927, (APWV02061);  1011-9050-000, 010084068210229221APWV0206211160927, (APWV02062);  1011-9050-000, 010084068210229221APWV0206311160927, (APWV02063);  1011-9050-000, 010084068210229221APWW0179411171023, (APWW01794);  1011-9050-000, 010084068210229221APWX0044611180206, (APWX00446);  1011-9050-000, 010084068210229221APWX0044711180206, (APWX00447);  1011-9050-000, 010084068210229221APWX0044811180206, (APWX00448);  1011-9050-000, 010084068210229221APWX0044911180206, (APWX00449);  1011-9050-000, 010084068210229221APWX0045011180206, (APWX00450);  1011-9050-000, 010084068210229221APWX0045111180206, (APWX00451);  1011-9050-000, 010084068210229221APWW0166011171004, (APWW01660);  1011-9050-000, 010084068210229221APWW0166111171004, (APWW01661);  1011-9050-000, 010084068210229221APWW0166211171004, (APWW01662);  1011-9050-000, 010084068210229221APWW0166311171004, (APWW01663);  1011-9050-000, 010084068210229221APWW0166411171004, (APWW01664);  1011-9050-000, 010084068210229221APWW0166511171004, (APWW01665);  1011-9050-000, 010084068210229221APWW0166611171004, (APWW01666);  1011-9050-000, 010084068210229221APWW0166711171004, (APWW01667);  1011-9050-000, 010084068210229221APWW0166811171004, (APWW01668);  1011-9050-000, 010084068210229221APWU0185011150831, (APWU01850);  1011-9050-000, 010084068210229221APWX0010711180104, (APWX00107);  1011-9050-000, 010084068210229221APWX0010811180104, (APWX00108);  1011-9050-000, 010084068210229221APWX0099111180417, (APWX00991);  1011-9050-000, 010084068210229221APWX0099211180417, (APWX00992);  1011-9050-000, 010084068210229221APWX0099311180417, (APWX00993);  1011-9050-000, 010084068210229221APWX0099411180417, (APWX00994);  1011-9050-000, 010084068210229221APWX0099511180417, (APWX00995);  1011-9050-000, 010084068210229221APWX0099611180417, (APWX00996);  1011-9050-000, 010084068210229221APWX0099711180417, (APWX00997);  1011-9050-000, 010084068210229221APWX0099811180417, (APWX00998);  1011-9050-000, 010084068210229221APWX0099911180417, (APWX00999);  1011-9050-000, 010084068210229221APWX0100011180417, (APWX01000);  1011-9050-000, 010084068210229221APWX0100111180417, (APWX01001);  1011-9050-000, 010084068210229221APWX0100211180417, (APWX01002);  1011-9050-000, 010084068210229221APWX0046111180208, (APWX00461);  1011-9050-000, 010084068210229221APWX0046211180208, (APWX00462);  1011-9050-000, 010084068210229221APWX0046311180208, (APWX00463);  1011-9050-000, 010084068210229221APWX0046411180208, (APWX00464);  1011-9050-000, 010084068210229221APWX0046511180208, (APWX00465);  1011-9050-000, 010084068210229221APWX0046611180208, (APWX00466);  1011-9050-000, 010084068210229221APWX0046711180208, (APWX00467);  1011-9050-000, 010084068210229221APWX0046811180208, (APWX00468);  1011-9050-000, 010084068210229221APWX0046911180208, (APWX00469);  1011-9050-000, 010084068210229221APWX0047011180208, (APWX00470);  1011-9050-000, 010084068210229221APWX0047111180208, (APWX00471);  2097675-001, Not Applicable, (5475713-1);  2097675-001, Not Applicable, (5475713-10);  2097675-001, Not Applicable, (5475713-11);  2097675-001, Not Applicable, (5475713-12);  2097675-001, Not Applicable, (5475713-13);  2097675-001, Not Applicable, (5475713-14);  2097675-001, Not Applicable, (5475713-15);  2097675-001, Not Applicable, (5475713-16);  2097675-001, Not Applicable, (5475713-17);  2097675-001, Not Applicable, (5475713-18);  2097675-001, Not Applicable, (5475713-19);  2097675-001, Not Applicable, (5475713-2);  2097675-001, Not Applicable, (5475713-20);  2097675-001, Not Applicable, (5475713-21);  2097675-001, Not Applicable, (5475713-22);  2097675-001, Not Applicable, (5475713-23);  2097675-001, Not Applicable, (5475713-24);  2097675-001, Not Applicable, (5475713-25);  2097675-001, Not Applicable, (5475713-26);  2097675-001, Not Applicable, (5475713-27);  2097675-001, Not Applicable, (5475713-28);  2097675-001, Not Applicable, (5475713-29);  2097675-001, Not Applicable, (5475713-3);  2097675-001, Not Applicable, (5475713-30);  2097675-001, Not Applicable, (5475713-31);  2097675-001, Not Applicable, (5475713-32);  2097675-001, Not Applicable, (5475713-33);  2097675-001, Not Applicable, (5475713-34);  2097675-001, Not Applicable, (5475713-35);  2097675-001, Not Applicable, (5475713-36);  2097675-001, Not Applicable, (5475713-37);  2097675-001, Not Applicable, (5475713-38);  2097675-001, Not Applicable, (5475713-39);  2097675-001, Not Applicable, (5475713-4);  2097675-001, Not Applicable, (5475713-40);  2097675-001, Not Applicable, (5475713-5);  2097675-001, Not Applicable, (5475713-6);  2097675-001, Not Applicable, (5475713-7);  2097675-001, Not Applicable, (5475713-8);  2097675-001, Not Applicable, (5475713-9);  2097675-001, Not Applicable, (5480829-1);  2097675-001, Not Applicable, (5480829-2);  2097675-001, Not Applicable, (5480829-3);  2097675-001, Not Applicable, (5480829-4);  2097675-001, Not Applicable, (5480829-5);  2097675-001, Not Applicable, (5480829-6);  2097675-001, Not Applicable, (5480829-7);  2097675-001, Not Applicable, (5480829-8);  1011-9050-000, 010084068210229221APWX0124111180531, (APWX01241);  1011-9050-000, 010084068210229221APWX0172911180718, (APWX01729);  1011-9050-000, 010084068210229221APWW0208211171206, (APWW02082);  1011-9050-000, 010084068210229221APWW0208311171206, (APWW02083);  1011-9050-000, 010084068210229221APWW0208411171206, (APWW02084);  1011-9050-000, 010084068210229221APWW0208511171206, (APWW02085);  1011-9050-000, 010084068210229221APWW0208611171206, (APWW02086);  1011-9050-000, 010084068210229221APWW0208711171206, (APWW02087);  1011-9050-000, 010084068210229221APWX0104611180427, (APWX01046);  1011-9050-000, 010084068210229221APWX0104711180427, (APWX01047);  1011-9050-000, 010084068210229221APWX0104811180427, (APWX01048);  1011-9050-000, 010084068210229221APWX0057611180301, (APWX00576);  1011-9050-000, 010084068210229221APWX0100311180417, (APWX01003);  1011-9050-000, 010084068210229221APWX0100411180417, (APWX01004);  1011-9050-000, 010084068210229221APWX0090311180328, (APWX00903);  1011-9050-000, 010084068210229221APWX0090411180328, (APWX00904);  1011-9050-000, 010084068210229221APWV0026611160119, (APWV00266);  1011-9050-000, 010084068210229221APWV0026711160119, (APWV00267);  1011-9050-000, 010084068210229221APWV0026811160119, (APWV00268);  1011-9050-000, 010084068210229221APWV0026911160119, (APWV00269);  1011-9050-000, 010084068210229221APWV0027011160119, (APWV00270);  1011-9050-000, 010084068210229221APWV0027111160202, (APWV00271);  1011-9050-000, 010084068210229221APWV0027211160202, (APWV00272);  1011-9050-000, 010084068210229221APWV0027311160202, (APWV00273);  1011-9050-000, 010084068210229221APWV0027411160202, (APWV00274);  1011-9050-000, 010084068210229221APWV0027511160202, (APWV00275);  1011-9050-000, 010084068210229221APWV0027611160202, (APWV00276);  1011-9050-000, 010084068210229221APWV0027711160202, (APWV00277);  1011-9050-000, 010084068210229221APWX0056711180301, (APWX00567);  1011-9050-000, 010084068210229221APWW0224911171219, (APWW02249);  1011-9050-000, 010084068210229221APWW0225011171219, (APWW02250);  1011-9050-000, 010084068210229221APWX0097311180417, (APWX00973);  1011-9050-000, 010084068210229221APWX0097411180417, (APWX00974);  1011-9050-000, 010084068210229221APWX0097511180417, (APWX00975);  1011-9050-000, 010084068210229221APWX0097611180417, (APWX00976);  1011-9050-000, 010084068210229221APWX0097711180417, (APWX00977);  1011-9050-000, 010084068210229221APWX0097811180417, (APWX00978);  1011-9050-000, 010084068210229221APWX0097911180417, (APWX00979);  1011-9050-000, 010084068210229221APWX0098011180417, (APWX00980);  1011-9050-000, 010084068210229221APWX0098111180417, (APWX00981);  1011-9050-000, 010084068210229221APWX0098211180417, (APWX00982);  1011-9050-000, 010084068210229221APWX0098311180417, (APWX00983);  1011-9050-000, 010084068210229221APWX0098411180417, (APWX00984);  1011-9050-000, Not Applicable, (APWU00120);  1011-9050-000, 010084068210229221APWW0160311170927, (APWW01603);  1011-9050-000, 010084068210229221APWX0019211180115, (APWX00192);  1011-9050-000, 010084068210229221APWX0019311180115, (APWX00193);  1011-9050-000, 010084068210229221APWX0019411180115, (APWX00194);  1011-9050-000, 010084068210229221APWX0019511180115, (APWX00195);  1011-9050-000, 010084068210229221APWX0019611180115, (APWX00196);  1011-9050-000, 010084068210229221APWX0019711180115, (APWX00197);  1011-9050-000, 010084068210229221APWX0019811180115, (APWX00198);  1011-9050-000, 010084068210229221APWX0037611180131, (APWX00376);  1011-9050-000, 010084068210229221APWX0037711180131, (APWX00377);  1011-9050-000, 010084068210229221APWX0037811180131, (APWX00378);  1011-9050-000, 010084068210229221APWX0037911180131, (APWX00379);  1011-9050-000, 010084068210229221APWX0038011180131, (APWX00380);  1011-9050-000, 010084068210229221APWX0038111180201, (APWX00381);  1011-9050-000, 010084068210229221APWX0038211180201, (APWX00382);  1011-9050-000, 010084068210229221APWX0038311180201, (APWX00383);  1011-9050-000, 010084068210229221APWX0108511180502, (APWX01085);  1011-9050-000, 010084068210229221APWX0178411180720, (APWX01784);  1011-9050-000, 010084068210229221APWX0038411180201, (APWX00384);  1011-9050-000, 010084068210229221APWX0038511180201, (APWX00385);  1011-9050-000, 010084068210229221APWX0038611180201, (APWX00386);  1011-9050-000, 010084068210229221APWX0170111180718, (APWX01701);  1011-9050-000, 010084068210229221APWW0160411170927, (APWW01604);  1011-9050-000, 010084068210229221APWX0103911180427, (APWX01039);  1011-9050-000, 010084068210229221APWX0104011180427, (APWX01040);  1011-9050-000, 010084068210229221APWX0104111180427, (APWX01041);  1011-9050-000, 010084068210229221APWX0104211180427, (APWX01042);  1011-9050-000, 010084068210229221APWX0104311180427, (APWX01043);  2097675-001, Not Applicable, (5463535-1);  2097675-001, Not Applicable, (5463535-2);  2097675-001, Not Applicable, (5463535-3);  1011-9050-000, 010084068210229221APWW0147211170823, (APWW01472);  1011-9050-000, 010084068210229221APWX0116611180517, (APWX01166);  1011-9050-000, 010084068210229221APWX0122511180530, (APWX01225);  1011-9050-000, 010084068210229221APWX0188511180803, (APWX01885);  1011-9050-000, 010084068210229221APWX0188611180803, (APWX01886);  1011-9050-000, 010084068210229221APWX0188711180803, (APWX01887);  1011-9050-000, 010084068210229221APWX0188811180803, (APWX01888);  1011-9050-000, 010084068210229221APWX0188911180803, (APWX01889);  1011-9050-000, 010084068210229221APWX0189011180803, (APWX01890);  1011-9050-000, 010084068210229221APWX0189111180803, (APWX01891);  1011-9050-000, 010084068210229221APWX0189211180803, (APWX01892);  1011-9050-000, 010084068210229221APWX0189311180803, (APWX01893);  2097675-ES, Not Applicable, (5480862-1);  1011-9050-000, 010084068210229221APWW0196011171122, (APWW01960);  1011-9050-000, 010084068210229221APWX0106411180427, (APWX01064);  1011-9050-000, 010084068210229221APWX0059211180307, (APWX00592);  1011-9050-000, 010084068210229221APWX0059311180307, (APWX00593);  1011-9050-000, 010084068210229221APWX0059411180307, (APWX00594);  1011-9050-000, 010084068210229221APWX0059511180307, (APWX00595);  2097675-CS, Not Applicable, (5479436-1);  2097675-CS, Not Applicable, (5479436-2);  2097675-CS, Not Applicable, (5479436-3);  2097675-CS, Not Applicable, (5479436-4);  2097675-CS, Not Applicable, (5479436-5);  2097675-CS, Not Applicable, (5480342-1);  2097675-CS, Not Applicable, (5480342-2);  2097675-CS, Not Applicable, (5480342-3);  1011-9050-000, 010084068210229221APWX0107111180502, (APWX01071);  1011-9050-000, 010084068210229221APWX0168911180716, (APWX01689);  1011-9050-000, 010084068210229221APWX0169011180716, (APWX01690);  1011-9050-000, 010084068210229221APWX0169111180718, (APWX01691);  1011-9050-000, 010084068210229221APWX0169211180718, (APWX01692);  1011-9050-000, 010084068210229221APWX0169311180718, (APWX01693);  2097675-DA, Not Applicable, (5453900-2);  1011-9050-000, 010084068210229221APWW0135411170802, (APWW01354);  1011-9050-000, 010084068210229221APWW0135511170802, (APWW01355);  1011-9050-000, 010084068210229221APWW0199311171127, (APWW01993);  1011-9050-000, 010084068210229221APWW0199411171127, (APWW01994);  1011-9050-000, 010084068210229221APWW0199511171127, (APWW01995);  1011-9050-000, 010084068210229221APWX0045611180206, (APWX00456);  1011-9050-000, 010084068210229221APWX0156411180703, (APWX01564);  1011-9050-000, 010084068210229221APWX0156511180703, (APWX01565);  1011-9050-000, 010084068210229221APWX0156611180703, (APWX01566);  1011-9050-000, 010084068210229221APWX0160211180711, (APWX01602);  1011-9050-000, 010084068210229221APWX0160311180711, (APWX01603);  1011-9050-000, 010084068210229221APWX0160411180711, (APWX01604);  1011-9050-000, 010084068210229221APWX0160511180711, (APWX01605);  1011-9050-000, 010084068210229221APWX0160611180711, (APWX01606);  1011-9050-000, 010084068210229221APWX0160711180711, (APWX01607);  1011-9050-000, 010084068210229221APWX0160811180711, (APWX01608);  1011-9050-000, 010084068210229221APWX0160911180711, (APWX01609);  1011-9050-000, 010084068210229221APWX0161011180711, (APWX01610);  1011-9050-000, 010084068210229221APWX0161111180711, (APWX01611);  1011-9050-000, 010084068210229221APWX0161211180711, (APWX01612);  1011-9050-000, 010084068210229221APWX0161311180711, (APWX01613);  1011-9050-000, 010084068210229221APWX0161411180711, (APWX01614);  1011-9050-000, 010084068210229221APWX0161511180711, (APWX01615);  1011-9050-000, 010084068210229221APWX0161611180711, (APWX01616);  1011-9050-000, 010084068210229221APWX0161711180711, (APWX01617);  1011-9050-000, 010084068210229221APWX0161811180711, (APWX01618);  1011-9050-000, 010084068210229221APWX0161911180711, (APWX01619);  1011-9050-000, 010084068210229221APWX0162011180711, (APWX01620);  1011-9050-000, 010084068210229221APWX0183111180802, (APWX01831);  2097675-FI, Not Applicable, (5453842-1);  1011-9050-000, 010084068210229221APWW0148311170830, (APWW01483);  1011-9050-000, 010084068210229221APWW0164311171003, (APWW01643);  1011-9050-000, 010084068210229221APWW0164411171003, (APWW01644);  1011-9050-000, 010084068210229221APWW0175711171013, (APWW01757);  1011-9050-000, 010084068210229221APWW0179511171023, (APWW01795);  1011-9050-000, 010084068210229221APWW0179611171023, (APWW01796);  1011-9050-000, 010084068210229221APWW0181211171026, (APWW01812);  1011-9050-000, 010084068210229221APWW0181311171026, (APWW01813);  1011-9050-000, 010084068210229221APWW0181411171026, (APWW01814);  1011-9050-000, 010084068210229221APWW0186911171109, (APWW01869);  1011-9050-000, 010084068210229221APWW0187011171109, (APWW01870);  1011-9050-000, 010084068210229221APWW0187111171109, (APWW01871);  1011-9050-000, 010084068210229221APWW0187211171109, (APWW01872);  1011-9050-000, 010084068210229221APWW0187311171109, (APWW01873);  1011-9050-000, 010084068210229221APWW0187411171109, (APWW01874);  1011-9050-000, 010084068210229221APWW0199911171127, (APWW01999);  1011-9050-000, 010084068210229221APWW0201111171130, (APWW02011);  1011-9050-000, 010084068210229221APWX0059111180307, (APWX00591);  1011-9050-000, 010084068210229221APWX0102511180417, (APWX01025);  1011-9050-000, 010084068210229221APWX0109611180502, (APWX01096);  1011-9050-000, 010084068210229221APWX0109711180502, (APWX01097);  1011-9050-000, 010084068210229221APWX0112311180507, (APWX01123);  1011-9050-000, 010084068210229221APWX0155111180627, (APWX01551);  1011-9050-000, 010084068210229221APWX0176611180720, (APWX01766);  1011-9050-000, 010084068210229221APWX0176711180720, (APWX01767);  2097675-FI, Not Applicable, (5462161-1);  2097675-FI, Not Applicable, (5462161-2);  2097675-FI, Not Applicable, (5467778-1);  2097675-FI, Not Applicable, (5467938-1);  2097675-FI, Not Applicable, (5467938-2);  2097675-FI, Not Applicable, (5469377-1);  2097675-FI, Not Applicable, (5469377-10);  2097675-FI, Not Applicable, (5469377-11);  2097675-FI, Not Applicable, (5469377-12);  2097675-FI, Not Applicable, (5469377-13);  2097675-FI, Not Applicable, (5469377-14);  2097675-FI, Not Applicable, (5469377-15);  2097675-FI, Not Applicable, (5469377-2);  2097675-FI, Not Applicable, (5469377-3);  2097675-FI, Not Applicable, (5469377-4);  2097675-FI, Not Applicable, (5469377-5);  2097675-FI, Not Applicable, (5469377-6);  2097675-FI, Not Applicable, (5469377-7);  2097675-FI, Not Applicable, (5469377-8);  2097675-FI, Not Applicable, (5469377-9);  1011-9050-000, 010084068210229221APWX0049211180208, (APWX00492);  1011-9050-000, 010084068210229221APWX0049311180208, (APWX00493);  1011-9050-000, 010084068210229221APWX0049411180208, (APWX00494);  1011-9050-000, 010084068210229221APWX0049511180208, (APWX00495);  1011-9050-000, 010084068210229221APWX0049611180208, (APWX00496);  1011-9050-000, 010084068210229221APWX0049711180208, (APWX00497);  2097675-FI, Not Applicable, (5462301-1);  2097675-FI, Not Applicable, (5472372-1);  2097675-FI, Not Applicable, (5472372-2);  2097675-FI, Not Applicable, (5471824-1);  2097675-FI, Not Applicable, (5471824-2);  1011-9050-000, 010084068210229221APWX0057711180301, (APWX00577);  1011-9050-000, 010084068210229221APWX0057811180301, (APWX00578);  1011-9050-000, 010084068210229221APWX0122811180530, (APWX01228);  2097675-FR, Not Applicable, (5481462-10);  2097675-FR, Not Applicable, (5481462-11);  2097675-IT, Not Applicable, (5465405-1);  1011-9050-000, 010084068210229221APWW0201711171130, (APWW02017);  1011-9050-000, 010084068210229221APWW0201811171130, (APWW02018);  1011-9050-000, 010084068210229221APWW0192611171109, (APWW01926);  1011-9050-000, 010084068210229221APWW0181611171026, (APWW01816);  1011-9050-000, 010084068210229221APWW0181711171026, (APWW01817);  1011-9050-000, 010084068210229221APWW0181811171026, (APWW01818);  1011-9050-000, 010084068210229221APWW0195911171122, (APWW01959);  1011-9050-000, 010084068210229221APWX0107011180427, (APWX01070);  1011-9050-000, 010084068210229221APWX0153211180627, (APWX01532);  1011-9050-000, 010084068210229221APWX0174411180720, (APWX01744);  1011-9050-000, 010084068210229221APWX0174511180720, (APWX01745);  1011-9050-000, 010084068210229221APWX0174611180720, (APWX01746);  1011-9050-000, 010084068210229221APWX0174711180720, (APWX01747);  1011-9050-000, 010084068210229221APWX0174811180720, (APWX01748);  1011-9050-000, 010084068210229221APWX0174911180720, (APWX01749);  1011-9050-000, 010084068210229221APWX0105811180427, (APWX01058);  1011-9050-000, 010084068210229221APWX0105911180427, (APWX01059);  1011-9050-000, 010084068210229221APWX0077911180323, (APWX00779);  1011-9050-000, 010084068210229221APWV0215411161018, (APWV02154);  1011-9050-000, 010084068210229221APWX0045811180206, (APWX00458);  1011-9050-000, 010084068210229221APWX0045911180206, (APWX00459);  1011-9050-000, 010084068210229221APWX0046011180206, (APWX00460);  1011-9050-000, 010084068210229221APWW0176911171013, (APWW01769);  1011-9050-000, 010084068210229221APWW0177011171013, (APWW01770);  1011-9050-000, 010084068210229221APWV0011311160105, (APWV00113);  2097675-FR, Not Applicable, (5478514-1);  1011-9050-000, 010084068210229221APWW0223911171219, (APWW02239);  1011-9050-000, 010084068210229221APWW0180911171023, (APWW01809);  1011-9050-000, 010084068210229221APWX0152911180627, (APWX01529);  1011-9050-000, 010084068210229221APWX0153011180627, (APWX01530);  1011-9050-000, 010084068210229221APWX0084811180328, (APWX00848);  1011-9050-000, 010084068210229221APWX0048011180208, (APWX00480);  1011-9050-000, 010084068210229221APWX0094811180410, (APWX00948);  1011-9050-000, 010084068210229221APWX0094911180410, (APWX00949);  1011-9050-000, Not Applicable, (APWT00133);  1011-9050-000, 010084068210229221APWV0222511161026, (APWV02225);  1011-9050-000, 010084068210229221APWV0222611161026, (APWV02226);  1011-9050-000, 010084068210229221APWW0172311171013, (APWW01723);  1011-9050-000, 010084068210229221APWW0172411171013, (APWW01724);  1011-9050-000, 010084068210229221APWW0172511171013, (APWW01725);  1011-9050-000, 010084068210229221APWW0172611171013, (APWW01726);  1011-9050-000, 010084068210229221APWW0172711171013, (APWW01727);  1011-9050-000, 010084068210229221APWX0183011180802, (APWX01830);  1011-9050-000, 010084068210229221APWX0153111180627, (APWX01531);  1011-9050-000, 010084068210229221APWX0095411180417, (APWX00954);  1011-9050-000, 010084068210229221APWX0095011180410, (APWX00950);  1011-9050-000, 010084068210229221APWX0095111180417, (APWX00951);  1011-9050-000, 010084068210229221APWX0095211180417, (APWX00952);  1011-9050-000, 010084068210229221APWX0085911180328, (APWX00859);  1011-9050-000, 010084068210229221APWX0121411180530, (APWX01214);  1011-9050-000, 010084068210229221APWX0121511180530, (APWX01215);  1011-9050-000, 010084068210229221APWX0121611180530, (APWX01216);  1011-9050-000, 010084068210229221APWX0121711180530, (APWX01217);  1011-9050-000, 010084068210229221APWX0121811180530, (APWX01218);  2097675-FR, Not Applicable, (5444103-1);  2097675-FR, Not Applicable, (5447048-1);  2097675-FR, Not Applicable, (5447048-2);  2097675-FR, Not Applicable, (5447048-3);  2097675-FR, Not Applicable, (5447048-4);  2097675-FR, Not Applicable, (5447048-5);  2097675-FR, Not Applicable, (5447048-6);  2097675-FR, Not Applicable, (5452713-1);  2097675-FR, Not Applicable, (5452713-2);  2097675-FR, Not Applicable, (5455188-1);  2097675-FR, Not Applicable, (5455188-2);  2097675-FR, Not Applicable, (5455188-3);  2097675-FR, Not Applicable, (5455188-4);  2097675-FR, Not Applicable, (5458018-1);  1011-9050-000, 010084068210229221APWW0160611170927, (APWW01606);  1011-9050-000, 010084068210229221APWW0160811170927, (APWW01608);  1011-9050-000, 010084068210229221APWW0160911170927, (APWW01609);  1011-9050-000, 010084068210229221APWX0105511180427, (APWX01055);  1011-9050-000, 010084068210229221APWX0105611180427, (APWX01056);  1011-9050-000, 010084068210229221APWX0105711180427, (APWX01057);  1011-9050-000, 010084068210229221APWX0108811180502, (APWX01088);  1011-9050-000, 010084068210229221APWX0112411180507, (APWX01124);  1011-9050-000, 010084068210229221APWX0177011180720, (APWX01770);  1011-9050-000, 010084068210229221APWX0177111180720, (APWX01771);  1011-9050-000, 010084068210229221APWX0177211180720, (APWX01772);  1011-9050-000, 010084068210229221APWX0177311180720, (APWX01773);  1011-9050-000, 010084068210229221APWX0177411180720, (APWX01774);  1011-9050-000, 010084068210229221APWX0177511180720, (APWX01775);  1011-9050-000, 010084068210229221APWX0177611180720, (APWX01776);  1011-9050-000, 010084068210229221APWV0222411161026, (APWV02224);  1011-9050-000, 010084068210229221APWX0156711180703, (APWX01567);  1011-9050-000, 010084068210229221APWX0156811180703, (APWX01568);  1011-9050-000, 010084068210229221APWX0108911180502, (APWX01089);  1011-9050-000, 010084068210229221APWW0176111171013, (APWW01761);  1011-9050-000, 010084068210229221APWX0095911180417, (APWX00959);  1011-9050-000, 010084068210229221APWX0096011180417, (APWX00960);  1011-9050-000, Not Applicable, (APWS00158);  1011-9050-000, Not Applicable, (APWS00159);  1011-9050-000, Not Applicable, (APWS00160);  1011-9050-000, Not Applicable, (APWS00161);  1011-9050-000, 010084068210229221APWX0177711180720, (APWX01777);  1011-9050-000, 010084068210229221APWX0177811180720, (APWX01778);  1011-9050-000, 010084068210229221APWW0176511171013, (APWW01765);  1011-9050-000, 010084068210229221APWX0120811180524, (APWX01208);  1011-9050-000, 010084068210229221APWW0192511171109, (APWW01925);  2097675-FR, Not Applicable, (5461581-9);  2097675-FR, Not Applicable, (5479017-1);  2097675-FR, Not Applicable, (5479017-2);  2097675-FR, Not Applicable, (5479017-3);  1011-9050-000, Not Applicable, (APWT00634);  1011-9050-000, Not Applicable, (APWT00636);  1011-9050-000, Not Applicable, (APWT00637);  1011-9050-000, Not Applicable, (APWT00638);  1011-9050-000, Not Applicable, (APWT00639);  1011-9050-000, Not Applicable, (APWT00640);  1011-9050-000, Not Applicable, (APWT00641);  1011-9050-000, Not Applicable, (APWT00642);  1011-9050-000, Not Applicable, (APWT00643);  1011-9050-000, Not Applicable, (APWT00644);  1011-9050-000, Not Applicable, (APWT00645);  1011-9050-000, Not Applicable, (APWT00646);  1011-9050-000, Not Applicable, (APWT00647);  1011-9050-000, Not Applicable, (APWT00648);  1011-9050-000, Not Applicable, (APWT00649);  1011-9050-000, Not Applicable, (APWT00650);  1011-9050-000, Not Applicable, (APWT00651);  1011-9050-000, Not Applicable, (APWT00652);  1011-9050-000, Not Applicable, (APWT00653);  1011-9050-000, Not Applicable, (APWT00654);  1011-9050-000, Not Applicable, (APWT00655);  1011-9050-000, Not Applicable, (APWT00656);  1011-9050-000, Not Applicable, (APWT00657);  1011-9050-000, Not Applicable, (APWT00658);  1011-9050-000, Not Applicable, (APWT00660);  1011-9050-000, Not Applicable, (APWT00663);  1011-9050-000, Not Applicable, (APWT00665);  1011-9050-000, Not Applicable, (APWT00666);  1011-9050-000, Not Applicable, (APWT00667);  1011-9050-000, 010084068210229221APWW0224411171219, (APWW02244);  1011-9050-000, 010084068210229221APWW0192711171109, (APWW01927);  1011-9050-000, 010084068210229221APWW0201911171130, (APWW02019);  1011-9050-000, 010084068210229221APWX0078011180323, (APWX00780);  1011-9050-000, 010084068210229221APWX0078111180323, (APWX00781);  1011-9050-000, 010084068210229221APWX0078211180323, (APWX00782);  1011-9050-000, 010084068210229221APWW0160711170927, (APWW01607);  1011-9050-000, 010084068210229221APWW0179711171023, (APWW01797);  1011-9050-000, 010084068210229221APWW0179811171023, (APWW01798);  1011-9050-000, 010084068210229221APWW0182111171026, (APWW01821);  1011-9050-000, 010084068210229221APWW0184511171026, (APWW01845);  1011-9050-000, 010084068210229221APWW0184711171026, (APWW01847);  1011-9050-000, 010084068210229221APWW0184811171026, (APWW01848);  1011-9050-000, 010084068210229221APWW0192311171109, (APWW01923);  1011-9050-000, 010084068210229221APWW0192411171109, (APWW01924);  1011-9050-000, 010084068210229221APWX0143611180618, (APWX01436);  1011-9050-000, 010084068210229221APWX0143711180618, (APWX01437);  1011-9050-000, 010084068210229221APWX0143811180618, (APWX01438);  1011-9050-000, 010084068210229221APWX0143911180618, (APWX01439);  1011-9050-000, 010084068210229221APWX0152111180627, (APWX01521);  1011-9050-000, 010084068210229221APWX0152211180627, (APWX01522);  1011-9050-000, 010084068210229221APWX0152311180627, (APWX01523);  1011-9050-000, 010084068210229221APWW0184911171026, (APWW01849);  1011-9050-000, 010084068210229221APWW0192211171109, (APWW01922);  1011-9050-000, 010084068210229221APWW0201411171130, (APWW02014);  1011-9050-000, 010084068210229221APWW0201511171130, (APWW02015);  1011-9050-000, 010084068210229221APWW0201611171130, (APWW02016);  1011-9050-000, 010084068210229221APWX0177911180720, (APWX01779);  2097675-FR, Not Applicable, (5465772-5);  1011-9050-000, 010084068210229221APWW0184611171026, (APWW01846);  1011-9050-000, 010084068210229221APWW0185011171026, (APWW01850);  1011-9050-000, 010084068210229221APWW0185111171109, (APWW01851);  1011-9050-000, 010084068210229221APWX0024411180116, (APWX00244);  1011-9050-000, 010084068210229221APWX0025011180116, (APWX00250);  1011-9050-000, 010084068210229221APWX0165311180716, (APWX01653);  1011-9050-000, 010084068210229221APWX0024311180116, (APWX00243);  1011-9050-000, 010084068210229221APWX0024511180116, (APWX00245);  1011-9050-000, 010084068210229221APWX0024611180116, (APWX00246);  1011-9050-000, 010084068210229221APWX0024711180116, (APWX00247);  1011-9050-000, 010084068210229221APWX0024811180116, (APWX00248);  1011-9050-000, 010084068210229221APWX0024911180116, (APWX00249);  1011-9050-000, 010084068210229221APWX0084611180328, (APWX00846);  1011-9050-000, 010084068210229221APWX0084711180328, (APWX00847);  1011-9050-000, 010084068210229221APWX0127411180604, (APWX01274);  1011-9050-000, 010084068210229221APWX0178011180720, (APWX01780);  1011-9050-000, 010084068210229221APWW0190711171109, (APWW01907);  1011-9050-000, 010084068210229221APWX0185611180803, (APWX01856);  1011-9050-000, 010084068210229221APWX0185711180803, (APWX01857);  1011-9050-000, 010084068210229221APWV0222711161026, (APWV02227);  1011-9050-000, 010084068210229221APWV0222811161026, (APWV02228);  1011-9050-000, 010084068210229221APWW0172811171013, (APWW01728);  1011-9050-000, 010084068210229221APWW0172911171013, (APWW01729);  1011-9050-000, 010084068210229221APWX0053211180222, (APWX00532);  1011-9050-000, 010084068210229221APWX0123811180531, (APWX01238);  1011-9050-000, Not Applicable, (APWS00335);  1011-9050-000, Not Applicable, (APWS00336);  1011-9050-000, Not Applicable, (APWS00337);  1011-9050-000, Not Applicable, (APWS00338);  1011-9050-000, Not Applicable, (APWT00501);  1011-9050-000, 010084068210229221APWU0202411150929, (APWU02024);  1011-9050-000, 010084068210229221APWU0202811150929, (APWU02028);  1011-9050-000, 010084068210229221APWU0202911150929, (APWU02029);  1011-9050-000, 010084068210229221APWW0181911171026, (APWW01819);  1011-9050-000, 010084068210229221APWW0182011171026, (APWW01820);  1011-9050-000, 010084068210229221APWX0084211180328, (APWX00842);  1011-9050-000, 010084068210229221APWX0084311180328, (APWX00843);  1011-9050-000, 010084068210229221APWX0084411180328, (APWX00844);  1011-9050-000, 010084068210229221APWX0169911180718, (APWX01699);  1011-9050-000, 010084068210229221APWX0170011180718, (APWX01700);  1011-9050-000, 010084068210229221APWX0152811180627, (APWX01528);  1011-9050-000, 010084068210229221APWV0149111160719, (APWV01491);  1011-9050-000, 010084068210229221APWV0149211160719, (APWV01492);  1011-9050-000, 010084068210229221APWV0149311160719, (APWV01493);  1011-9050-000, 010084068210229221APWV0149411160719, (APWV01494);  1011-9050-000, 010084068210229221APWV0149511160719, (APWV01495);  1011-9050-000, 010084068210229221APWV0149611160719, (APWV01496);  1011-9050-000, 010084068210229221APWV0151411160719, (APWV01514);  1011-9050-000, 010084068210229221APWV0151511160719, (APWV01515);  1011-9050-000, 010084068210229221APWV0151611160719, (APWV01516);  2097675-FR, Not Applicable, (5454898-1);  2097675-FR, Not Applicable, (5454898-2);  1011-9050-000, 010084068210229221APWW0138111170802, (APWW01381);  1011-9050-000, 010084068210229221APWW0138311170802, (APWW01383);  1011-9050-000, 010084068210229221APWW0158211170925, (APWW01582);  1011-9050-000, 010084068210229221APWW0189311171109, (APWW01893);  1011-9050-000, 010084068210229221APWW0167511171004, (APWW01675);  1011-9050-000, 010084068210229221APWW0181511171026, (APWW01815);  1011-9050-000, 010084068210229221APWW0158311170925, (APWW01583);  1011-9050-000, 010084068210229221APWW0223811171219, (APWW02238);  1011-9050-000, 010084068210229221APWX0027611180118, (APWX00276);  1011-9050-000, 010084068210229221APWW0131011170721, (APWW01310);  1011-9050-000, 010084068210229221APWX0062811180309, (APWX00628);  1011-9050-000, 010084068210229221APWV0051811160302, (APWV00518);  1011-9050-000, 010084068210229221APWV0237411161108, (APWV02374);  1011-9050-000, 010084068210229221APWW0197911171127, (APWW01979);  1011-9050-000, 010084068210229221APWW0198011171127, (APWW01980);  1011-9050-000, 010084068210229221APWW0198111171127, (APWW01981);  1011-9050-000, 010084068210229221APWW0198211171127, (APWW01982);  1011-9050-000, 010084068210229221APWW0198311171127, (APWW01983);  1011-9050-000, 010084068210229221APWW0198411171127, (APWW01984);  1011-9050-000, 010084068210229221APWX0107911180502, (APWX01079);  1011-9050-000, 010084068210229221APWX0108011180502, (APWX01080);  1011-9050-000, 010084068210229221APWX0108111180502, (APWX01081);  1011-9050-000, 010084068210229221APWX0108211180502, (APWX01082);  1011-9050-000, 010084068210229221APWX0108311180502, (APWX01083);  1011-9050-000, 010084068210229221APWX0108411180502, (APWX01084);  1011-9050-000, 010084068210229221APWX0135811180611, (APWX01358);  1011-9050-000, 010084068210229221APWX0135911180611, (APWX01359);  1011-9050-000, 010084068210229221APWX0136011180611, (APWX01360);  1011-9050-000, 010084068210229221APWX0140111180618, (APWX01401);  1011-9050-000, 010084068210229221APWX0191611180815, (APWX01916);  1011-9050-000, Not Applicable, (APWT00461);  1011-9050-000, Not Applicable, (APWT00462);  1011-9050-000, Not Applicable, (APWT00463);  1011-9050-000, Not Applicable, (APWT00464);  1011-9050-000, Not Applicable, (APWT00465);  1011-9050-000, Not Applicable, (APWT00466);  1011-9050-000, Not Applicable, (APWT00468);  1011-9050-000, Not Applicable, (APWT00469);  1011-9050-000, Not Applicable, (APWT00470);  1011-9050-000, Not Applicable, (APWT00471);  1011-9050-000, Not Applicable, (APWT00494);  1011-9050-000, Not Applicable, (APWT00496);  1011-9050-000, 010084068210229221APWW0215511171214, (APWW02155);  1011-9050-000, 010084068210229221APWX0034111180131, (APWX00341);  1011-9050-000, 010084068210229221APWX0085111180328, (APWX00851);  1011-9050-000, 010084068210229221APWX0185811180803, (APWX01858);  1011-9050-000, 010084068210229221APWV0085211160418, (APWV00852);  1011-9050-000, 010084068210229221APWV0085311160418, (APWV00853);  1011-9050-000, 010084068210229221APWV0085411160418, (APWV00854);  1011-9050-000, 010084068210229221APWW0109311170607, (APWW01093);  1011-9050-000, 010084068210229221APWV0051911160302, (APWV00519);  1011-9050-000, 010084068210229221APWV0275411161221, (APWV02754);  1011-9050-000, 010084068210229221APWV0275511161221, (APWV02755);  1011-9050-000, 010084068210229221APWX0121911180530, (APWX01219);  1011-9050-000, 010084068210229221APWV0051711160302, (APWV00517);  1011-9050-000, 010084068210229221APWW0052211170215, (APWW00522);  1011-9050-000, 010084068210229221APWW0052411170215, (APWW00524);  1011-9050-000, 010084068210229221APWW0052611170215, (APWW00526);  1011-9050-000, 010084068210229221APWW0151211170830, (APWW01512);  1011-9050-000, 010084068210229221APWW0151311170830, (APWW01513);  1011-9050-000, 010084068210229221APWX0087311180328, (APWX00873);  1011-9050-000, 010084068210229221APWX0089911180328, (APWX00899);  1011-9050-000, 010084068210229221APWX0153311180627, (APWX01533);  1011-9050-000, 010084068210229221APWX0153411180627, (APWX01534);  1011-9050-000, 010084068210229221APWX0153511180627, (APWX01535);  1011-9050-000, 010084068210229221APWW0148511170830, (APWW01485);  1011-9050-000, 010084068210229221APWW0195611171122, (APWW01956);  1011-9050-000, 010084068210229221APWW0195711171122, (APWW01957);  1011-9050-000, 010084068210229221APWX0124611180531, (APWX01246);  1011-9050-000, 010084068210229221APWX0124711180531, (APWX01247);  1011-9050-000, 010084068210229221APWX0124811180531, (APWX01248);  1011-9050-000, 010084068210229221APWX0124911180531, (APWX01249);  1011-9050-000, 010084068210229221APWX0125011180531, (APWX01250);  1011-9050-000, 010084068210229221APWX0125111180531, (APWX01251);  1011-9050-000, 010084068210229221APWX0125211180604, (APWX01252);  1011-9050-000, 010084068210229221APWX0125311180604, (APWX01253);  1011-9050-000, 010084068210229221APWX0125411180604, (APWX01254);  1011-9050-000, 010084068210229221APWX0125511180604, (APWX01255);  1011-9050-000, 010084068210229221APWX0125611180604, (APWX01256);  1011-9050-000, 010084068210229221APWX0125711180604, (APWX01257);  1011-9050-000, Not Applicable, (APWT03158);  1011-9050-000, 010084068210229221APWW0176311171013, (APWW01763);  1011-9050-000, 010084068210229221APWW0176411171013, (APWW01764);  1011-9050-000, 010084068210229221APWX0135511180611, (APWX01355);  1011-9050-000, 010084068210229221APWX0135611180611, (APWX01356);  1011-9050-000, 010084068210229221APWX0140011180612, (APWX01400);  1011-9050-000, 010084068210229221APWX0169811180718, (APWX01698);  1011-9050-000, 010084068210229221APWW0195811171122, (APWW01958);  1011-9050-000, 010084068210229221APWV0269511161220, (APWV02695);  1011-9050-000, 010084068210229221APWX0044411180206, (APWX00444);  1011-9050-000, 010084068210229221APWX0045511180206, (APWX00455);  1011-9050-000, 010084068210229221APWW0197111171127, (APWW01971);  1011-9050-000, 010084068210229221APWW0197211171127, (APWW01972);  1011-9050-000, 010084068210229221APWW0197311171127, (APWW01973);  1011-9050-000, 010084068210229221APWW0200011171127, (APWW02000);  1011-9050-000, 010084068210229221APWW0200111171127, (APWW02001);  1011-9050-000, 010084068210229221APWW0200211171127, (APWW02002);  1011-9050-000, 010084068210229221APWW0200311171127, (APWW02003);  1011-9050-000, 010084068210229221APWW0200411171127, (APWW02004);  1011-9050-000, 010084068210229221APWW0200511171127, (APWW02005);  1011-9050-000, 010084068210229221APWW0200611171127, (APWW02006);  1011-9050-000, 010084068210229221APWW0200711171127, (APWW02007);  1011-9050-000, 010084068210229221APWW0200811171127, (APWW02008);  1011-9050-000, 010084068210229221APWW0202111171130, (APWW02021);  1011-9050-000, 010084068210229221APWW0202211171130, (APWW02022);  1011-9050-000, 010084068210229221APWW0202311171130, (APWW02023);  1011-9050-000, 010084068210229221APWW0206611171206, (APWW02066);  1011-9050-000, 010084068210229221APWX0017611180105, (APWX00176);  1011-9050-000, 010084068210229221APWX0017711180105, (APWX00177);  1011-9050-000, 010084068210229221APWX0017811180105, (APWX00178);  1011-9050-000, 010084068210229221APWX0029411180118, (APWX00294);  1011-9050-000, 010084068210229221APWX0029511180118, (APWX00295);  1011-9050-000, 010084068210229221APWX0034011180123, (APWX00340);  1011-9050-000, 010084068210229221APWW0196911171122, (APWW01969);  1011-9050-000, 010084068210229221APWW0197011171122, (APWW01970);  1011-9050-000, 010084068210229221APWW0197711171127, (APWW01977);  1011-9050-000, 010084068210229221APWW0197811171127, (APWW01978);  1011-9050-000, 010084068210229221APWX0102711180417, (APWX01027);  1011-9050-000, 010084068210229221APWX0102811180417, (APWX01028);  1011-9050-000, 010084068210229221APWX0102911180417, (APWX01029);  1011-9050-000, 010084068210229221APWX0103011180417, (APWX01030);  1011-9050-000, 010084068210229221APWW0013311170103, (APWW00133);  1011-9050-000, 010084068210229221APWW0013511170103, (APWW00135);  1011-9050-000, 010084068210229221APWW0197411171127, (APWW01974);  1011-9050-000, 010084068210229221APWW0197511171127, (APWW01975);  1011-9050-000, 010084068210229221APWW0197611171127, (APWW01976);  1011-9050-000, 010084068210229221APWX0082511180328, (APWX00825);  1011-9050-000, 010084068210229221APWX0123911180531, (APWX01239);  1011-9050-000, 010084068210229221APWX0124011180531, (APWX01240);  1011-9050-000, 010084068210229221APWX0125811180604, (APWX01258);  1011-9050-000, 010084068210229221APWX0125911180604, (APWX01259);  1011-9050-000, 010084068210229221APWX0045411180206, (APWX00454);  1011-9050-000, 010084068210229221APWX0109811180502, (APWX01098);  1011-9050-000, 010084068210229221APWX0109911180502, (APWX01099);  1011-9050-000, 010084068210229221APWX0110011180502, (APWX01100);  1011-9050-000, 010084068210229221APWX0110111180502, (APWX01101);  1011-9050-000, 010084068210229221APWX0121311180530, (APWX01213);  1011-9050-000, 010084068210229221APWX0140711180618, (APWX01407);  1011-9050-000, 010084068210229221APWX0148211180625, (APWX01482);  1011-9050-000, 010084068210229221APWX0154211180627, (APWX01542);  1011-9050-000, 010084068210229221APWX0154311180627, (APWX01543);  1011-9050-000, 010084068210229221APWX0170211180718, (APWX01702);  1011-9050-000, 010084068210229221APWX0170311180718, (APWX01703);  1011-9050-000, 010084068210229221APWX0170411180718, (APWX01704);  1011-9050-000, 010084068210229221APWX0170511180718, (APWX01705);  1011-9050-000, 010084068210229221APWX0170611180718, (APWX01706);  1011-9050-000, 010084068210229221APWX0170711180718, (APWX01707);  1011-9050-000, 010084068210229221APWX0189811180803, (APWX01898);  1011-9050-000, 010084068210229221APWX0189911180803, (APWX01899);  1011-9050-000, 010084068210229221APWW0184111171026, (APWW01841);  1011-9050-000, 010084068210229221APWW0184211171026, (APWW01842);  1011-9050-000, 010084068210229221APWW0225111171222, (APWW02251);  1011-9050-000, 010084068210229221APWW0188211171109, (APWW01882);  1011-9050-000, 010084068210229221APWX0158611180703, (APWX01586);  1011-9050-000, 010084068210229221APWX0158711180703, (APWX01587);  1011-9050-000, 010084068210229221APWX0158811180703, (APWX01588);  1011-9050-000, 010084068210229221APWX0153611180627, (APWX01536);  1011-9050-000, 010084068210229221APWX0153711180627, (APWX01537);  1011-9050-000, 010084068210229221APWX0153811180627, (APWX01538);  1011-9050-000, 010084068210229221APWX0153911180627, (APWX01539);  1011-9050-000, 010084068210229221APWX0175911180720, (APWX01759);  1011-9050-000, 010084068210229221APWX0176011180720, (APWX01760);  1011-9050-000, 010084068210229221APWW0136011170802, (APWW01360);  1011-9050-000, 010084068210229221APWW0136111170802, (APWW01361);  1011-9050-000, 010084068210229221APWX0024111180116, (APWX00241);  1011-9050-000, 010084068210229221APWX0095811180417, (APWX00958);  1011-9050-000, 010084068210229221APWX0057011180301, (APWX00570);  1011-9050-000, 010084068210229221APWX0057111180301, (APWX00571);  1011-9050-000, 010084068210229221APWW0056911170224, (APWW00569);  1011-9050-000, 010084068210229221APWX0058111180307, (APWX00581);  1011-9050-000, 010084068210229221APWX0058211180307, (APWX00582);  1011-9050-000, 010084068210229221APWX0058311180307, (APWX00583);  1011-9050-000, 010084068210229221APWX0058411180307, (APWX00584);  1011-9050-000, 010084068210229221APWX0058511180307, (APWX00585);  1011-9050-000, 010084068210229221APWX0058611180307, (APWX00586);  1011-9050-000, Not Applicable, (APWU00880);  1011-9050-000, 010084068210229221APWV0186611160831, (APWV01866);  1011-9050-000, 010084068210229221APWW0120311170608, (APWW01203);  1011-9050-000, 010084068210229221APWW0124711170608, (APWW01247);  1011-9050-000, 010084068210229221APWW0124811170608, (APWW01248);  1011-9050-000, 010084068210229221APWW0124911170608, (APWW01249);  1011-9050-000, 010084068210229221APWW0125011170608, (APWW01250);  1011-9050-000, 010084068210229221APWW0125111170609, (APWW01251);  1011-9050-000, 010084068210229221APWW0125211170609, (APWW01252);  1011-9050-000, 010084068210229221APWW0125311170609, (APWW01253);  1011-9050-000, 010084068210229221APWW0125411170609, (APWW01254);  1011-9050-000, 010084068210229221APWW0126211170609, (APWW01262);  1011-9050-000, 010084068210229221APWW0126311170609, (APWW01263);  1011-9050-000, 010084068210229221APWW0126411170609, (APWW01264);  1011-9050-000, 010084068210229221APWW0126511170609, (APWW01265);  1011-9050-000, 010084068210229221APWW0126611170609, (APWW01266);  1011-9050-000, 010084068210229221APWW0126711170609, (APWW01267);  1011-9050-000, 010084068210229221APWW0126811170609, (APWW01268);  1011-9050-000, 010084068210229221APWW0126911170609, (APWW01269);  1011-9050-000, 010084068210229221APWX0056811180301, (APWX00568);  1011-9050-000, 010084068210229221APWX0056911180301, (APWX00569);  1011-9050-000, 010084068210229221APWX0057211180301, (APWX00572);  1011-9050-000, 010084068210229221APWX0057311180301, (APWX00573);  1011-9050-000, 010084068210229221APWW0201211171130, (APWW02012);  1011-9050-000, 010084068210229221APWW0201311171130, (APWW02013);  2097675-NL, Not Applicable, (5476622-1);  2097675-NL, Not Applicable, (5476622-10);  2097675-NL, Not Applicable, (5476622-11);  2097675-NL, Not Applicable, (5476622-12);  2097675-NL, Not Applicable, (5476622-13);  2097675-NL, Not Applicable, (5476622-14);  2097675-NL, Not Applicable, (5476622-15);  2097675-NL, Not Applicable, (5476622-16);  2097675-NL, Not Applicable, (5476622-17);  2097675-NL, Not Applicable, (5476622-18);  2097675-NL, Not Applicable, (5476622-19);  2097675-NL, Not Applicable, (5476622-2);  2097675-NL, Not Applicable, (5476622-20);  2097675-NL, Not Applicable, (5476622-3);  2097675-NL, Not Applicable, (5476622-4);  2097675-NL, Not Applicable, (5476622-5);  2097675-NL, Not Applicable, (5476622-6);  2097675-NL, Not Applicable, (5476622-7);  2097675-NL, Not Applicable, (5476622-8);  2097675-NL, Not Applicable, (5476622-9);  1011-9050-000, 010084068210229221APWX0181411180802, (APWX01814);  1011-9050-000, 010084068210229221APWX0181511180802, (APWX01815);  1011-9050-000, 010084068210229221APWX0181611180802, (APWX01816);  1011-9050-000, 010084068210229221APWX0181711180802, (APWX01817);  1011-9050-000, 010084068210229221APWX0181811180802, (APWX01818);  1011-9050-000, 010084068210229221APWX0157411180703, (APWX01574);  1011-9050-000, 010084068210229221APWX0157511180703, (APWX01575);  1011-9050-000, 010084068210229221APWX0157611180703, (APWX01576);  1011-9050-000, 010084068210229221APWX0157711180703, (APWX01577);  2097675-001, Not Applicable, (5456935-1);  2097675-001, Not Applicable, (5453833-1);  2097675-001, Not Applicable, (5453833-10);  2097675-001, Not Applicable, (5453833-11);  2097675-001, Not Applicable, (5453833-12);  2097675-001, Not Applicable, (5453833-13);  2097675-001, Not Applicable, (5453833-14);  2097675-001, Not Applicable, (5453833-15);  2097675-001, Not Applicable, (5453833-16);  2097675-001, Not Applicable, (5453833-17);  2097675-001, Not Applicable, (5453833-18);  2097675-001, Not Applicable, (5453833-19);  2097675-001, Not Applicable, (5453833-2);  2097675-001, Not Applicable, (5453833-20);  2097675-001, Not Applicable, (5453833-3);  2097675-001, Not Applicable, (5453833-4);  2097675-001, Not Applicable, (5453833-5);  2097675-001, Not Applicable, (5453833-6);  2097675-001, Not Applicable, (5453833-7);  2097675-001, Not Applicable, (5453833-8);  2097675-001, Not Applicable, (5453833-9);  2097675-001, Not Applicable, (5459047-1);  2097675-001, Not Applicable, (5459047-2);  2097675-001, Not Applicable, (5459047-3);  2097675-001, Not Applicable, (5459047-4);  1011-9050-000, 010084068210229221APWX0010511180104, (APWX00105);  2097675-001, Not Applicable, (5447948-1);  2097675-001, Not Applicable, (5447949-1);  2097675-001, Not Applicable, (5456012-4);  2097675-001, Not Applicable, (5456012-5);  2097675-001, Not Applicable, (5456012-6);  2097675-001, Not Applicable, (5482957-1);  2097675-001, Not Applicable, (5482957-10);  2097675-001, Not Applicable, (5482957-2);  2097675-001, Not Applicable, (5482957-3);  2097675-001, Not Applicable, (5482957-4);  2097675-001, Not Applicable, (5482957-5);  2097675-001, Not Applicable, (5482957-6);  2097675-001, Not Applicable, (5482957-7);  2097675-001, Not Applicable, (5482957-8);  2097675-001, Not Applicable, (5482957-9);  1011-9050-000, 010084068210229221APWX0175811180720, (APWX01758);  1011-9050-000, 010084068210229221APWX0187011180803, (APWX01870);  2097675-001, Not Applicable, (5456719-10);  2097675-001, Not Applicable, (5456719-11);  2097675-001, Not Applicable, (5456719-12);  2097675-001, Not Applicable, (5456719-13);  2097675-001, Not Applicable, (5456719-14);  2097675-001, Not Applicable, (5456719-15);  2097675-001, Not Applicable, (5456719-16);  2097675-001, Not Applicable, (5456719-17);  2097675-001, Not Applicable, (5456719-18);  2097675-001, Not Applicable, (5456719-19);  2097675-001, Not Applicable, (5456719-2);  2097675-001, Not Applicable, (5456719-20);  2097675-001, Not Applicable, (5456719-21);  2097675-001, Not Applicable, (5456719-22);  2097675-001, Not Applicable, (5456719-23);  2097675-001, Not Applicable, (5456719-24);  2097675-001, Not Applicable, (5456719-25);  2097675-001, Not Applicable, (5456719-26);  2097675-001, Not Applicable, (5456719-27);  2097675-001, Not Applicable, (5456719-28);  2097675-001, Not Applicable, (5456719-29);  2097675-001, Not Applicable, (5456719-3);  2097675-001, Not Applicable, (5456719-30);  2097675-001, Not Applicable, (5456719-31);  2097675-001, Not Applicable, (5456719-32);  2097675-001, Not Applicable, (5456719-33);  2097675-001, Not Applicable, (5456719-34);  2097675-001, Not Applicable, (5456719-35);  2097675-001, Not Applicable, (5456719-36);  2097675-001, Not Applicable, (5456719-37);  2097675-001, Not Applicable, (5456719-38);  2097675-001, Not Applicable, (5456719-39);  2097675-001, Not Applicable, (5456719-4);  2097675-001, Not Applicable, (5456719-40);  2097675-001, Not Applicable, (5456719-41);  2097675-001, Not Applicable, (5456719-42);  2097675-001, Not Applicable, (5456719-43);  2097675-001, Not Applicable, (5456719-44);  2097675-001, Not Applicable, (5456719-45);  2097675-001, Not Applicable, (5456719-46);  2097675-001, Not Applicable, (5456719-5);  2097675-001, Not Applicable, (5456719-6);  2097675-001, Not Applicable, (5456719-7);  2097675-001, Not Applicable, (5456719-8);  2097675-001, Not Applicable, (5456719-9);  2097675-001, Not Applicable, (5482418-1);  1011-9050-000, 010084068210229221APWX0154411180627, (APWX01544);  1011-9050-000, 010084068210229221APWX0157811180703, (APWX01578);  1011-9050-000, 010084068210229221APWX0157911180703, (APWX01579);  1011-9050-000, 010084068210229221APWX0158011180703, (APWX01580);  1011-9050-000, 010084068210229221APWX0158111180703, (APWX01581);  1011-9050-000, 010084068210229221APWX0158211180703, (APWX01582);  1011-9050-000, 010084068210229221APWX0158311180703, (APWX01583);  1011-9050-000, 010084068210229221APWX0158411180703, (APWX01584);  1011-9050-000, 010084068210229221APWX0158511180703, (APWX01585);  2097675-PT, Not Applicable, (5457220-1);  1011-9050-000, 010084068210229221APWX0127711180604, (APWX01277);  1011-9050-000, 010084068210229221APWW0121411170608, (APWW01214);  1011-9050-000, 010084068210229221APWW0121511170608, (APWW01215);  1011-9050-000, 010084068210229221APWW0121611170608, (APWW01216);  1011-9050-000, 010084068210229221APWW0121711170608, (APWW01217);  1011-9050-000, 010084068210229221APWW0121811170608, (APWW01218);  1011-9050-000, 010084068210229221APWW0121911170608, (APWW01219);  1011-9050-000, 010084068210229221APWW0122911170608, (APWW01229);  1011-9050-000, 010084068210229221APWW0123011170608, (APWW01230);  1011-9050-000, 010084068210229221APWW0123111170608, (APWW01231);  1011-9050-000, 010084068210229221APWW0123211170608, (APWW01232);  1011-9050-000, 010084068210229221APWW0123311170608, (APWW01233);  1011-9050-000, 010084068210229221APWW0123411170608, (APWW01234);  1011-9050-000, 010084068210229221APWV0089111160426, (APWV00891);  1011-9050-000, 010084068210229221APWW0015511170105, (APWW00155);  1011-9050-000, 010084068210229221APWX0103111180427, (APWX01031);  1011-9050-000, 010084068210229221APWX0085711180328, (APWX00857);  1011-9050-000, 010084068210229221APWW0095111170505, (APWW00951);  1011-9050-000, 010084068210229221APWW0095211170505, (APWW00952);  1011-9050-000, 010084068210229221APWW0095311170505, (APWW00953);  1011-9050-000, 010084068210229221APWW0095411170505, (APWW00954);  1011-9050-000, 010084068210229221APWW0095511170505, (APWW00955);  1011-9050-000, 010084068210229221APWW0095611170505, (APWW00956);  1011-9050-000, 010084068210229221APWW0095711170505, (APWW00957);  1011-9050-000, 010084068210229221APWW0096011170505, (APWW00960);  1011-9050-000, 010084068210229221APWW0096111170505, (APWW00961);  1011-9050-000, 010084068210229221APWW0096211170505, (APWW00962);  1011-9050-000, 010084068210229221APWW0096311170505, (APWW00963);  1011-9050-000, 010084068210229221APWW0096411170505, (APWW00964);  1011-9050-000, 010084068210229221APWW0096511170505, (APWW00965);  1011-9050-000, 010084068210229221APWW0096611170505, (APWW00966);  1011-9050-000, 010084068210229221APWW0096711170505, (APWW00967);  1011-9050-000, 010084068210229221APWW0096811170505, (APWW00968);  1011-9050-000, 010084068210229221APWW0096911170505, (APWW00969);  1011-9050-000, 010084068210229221APWW0123511170608, (APWW01235);  1011-9050-000, 010084068210229221APWW0190811171109, (APWW01908);  1011-9050-000, 010084068210229221APWW0196111171122, (APWW01961);  1011-9050-000, 010084068210229221APWW0182411171026, (APWW01824);  1011-9050-000, 010084068210229221APWW0181111171026, (APWW01811);  1011-9050-000, 010084068210229221APWX0122911180530, (APWX01229);  1011-9050-000, 010084068210229221APWX0123011180530, (APWX01230);  1011-9050-000, 010084068210229221APWX0123111180530, (APWX01231);  1011-9050-000, 010084068210229221APWX0123211180531, (APWX01232);  1011-9050-000, 010084068210229221APWX0127211180604, (APWX01272);  1011-9050-000, 010084068210229221APWX0127311180604, (APWX01273);  1011-9050-000, 010084068210229221APWW0175211171013, (APWW01752);  1011-9050-000, 010084068210229221APWW0175311171013, (APWW01753);  1011-9050-000, 010084068210229221APWW0175411171013, (APWW01754);  1011-9050-000, 010084068210229221APWW0175511171013, (APWW01755);  1011-9050-000, 010084068210229221APWW0176611171013, (APWW01766);  1011-9050-000, 010084068210229221APWW0186611171109, (APWW01866);  1011-9050-000, 010084068210229221APWW0186711171109, (APWW01867);  1011-9050-000, 010084068210229221APWX0122611180530, (APWX01226);  1011-9050-000, 010084068210229221APWX0122711180530, (APWX01227);  1011-9050-000, 010084068210229221APWX0183211180802, (APWX01832);  1011-9050-000, 010084068210229221APWX0142511180618, (APWX01425);  1011-9050-000, 010084068210229221APWX0142611180618, (APWX01426);  1011-9050-000, 010084068210229221APWX0162111180711, (APWX01621);  1011-9050-000, 010084068210229221APWX0162211180711, (APWX01622);  1011-9050-000, 010084068210229221APWX0162311180711, (APWX01623);  1011-9050-000, 010084068210229221APWX0162411180711, (APWX01624);  1011-9050-000, 010084068210229221APWX0162511180711, (APWX01625);  1011-9050-000, 010084068210229221APWX0162611180711, (APWX01626);  1011-9050-000, 010084068210229221APWX0162711180711, (APWX01627);  1011-9050-000, 010084068210229221APWX0166711180716, (APWX01667);  1011-9050-000, 010084068210229221APWX0166811180716, (APWX01668);  1011-9050-000, 010084068210229221APWX0166911180716, (APWX01669);  1011-9050-000, 010084068210229221APWX0167011180716, (APWX01670);  1011-9050-000, 010084068210229221APWX0167111180716, (APWX01671);  1011-9050-000, 010084068210229221APWX0167211180716, (APWX01672);  1011-9050-000, 010084068210229221APWX0167311180716, (APWX01673);  1011-9050-000, 010084068210229221APWX0167411180716, (APWX01674);  1011-9050-000, 010084068210229221APWX0167511180716, (APWX01675);  1011-9050-000, 010084068210229221APWX0167611180716, (APWX01676);  1011-9050-000, 010084068210229221APWX0170811180718, (APWX01708);  1011-9050-000, 010084068210229221APWX0170911180718, (APWX01709);  1011-9050-000, 010084068210229221APWX0171011180718, (APWX01710);  1011-9050-000, 010084068210229221APWX0171111180718, (APWX01711);  1011-9050-000, 010084068210229221APWX0171211180718, (APWX01712);  1011-9050-000, 010084068210229221APWX0171311180718, (APWX01713);  1011-9050-000, 010084068210229221APWX0171411180718, (APWX01714);  1011-9050-000, 010084068210229221APWX0171511180718, (APWX01715);  1011-9050-000, 010084068210229221APWX0171611180718, (APWX01716);  1011-9050-000, 010084068210229221APWX0171711180718, (APWX01717);  1011-9050-000, 010084068210229221APWX0171811180718, (APWX01718);  1011-9050-000, 010084068210229221APWX0171911180718, (APWX01719);  1011-9050-000, 010084068210229221APWX0172011180718, (APWX01720);  1011-9050-000, 010084068210229221APWX0172111180718, (APWX01721);  1011-9050-000, 010084068210229221APWX0172211180718, (APWX01722);  1011-9050-000, 010084068210229221APWX0172311180718, (APWX01723);  1011-9050-000, 010084068210229221APWX0172411180718, (APWX01724);  1011-9050-000, 010084068210229221APWX0172511180718, (APWX01725);  1011-9050-000, 010084068210229221APWX0172611180718, (APWX01726);  1011-9050-000, 010084068210229221APWX0172711180718, (APWX01727);  1011-9050-000, 010084068210229221APWX0178311180720, (APWX01783);  1011-9050-000, 010084068210229221APWW0178211171023, (APWW01782);  1011-9050-000, 010084068210229221APWW0178311171023, (APWW01783);  1011-9050-000, 010084068210229221APWW0178411171023, (APWW01784);  1011-9050-000, 010084068210229221APWW0178511171023, (APWW01785);  1011-9050-000, 010084068210229221APWW0178611171023, (APWW01786);  1011-9050-000, 010084068210229221APWW0178711171023, (APWW01787);  1011-9050-000, 010084068210229221APWW0178811171023, (APWW01788);  1011-9050-000, 010084068210229221APWW0178911171023, (APWW01789);  1011-9050-000, 010084068210229221APWW0179011171023, (APWW01790);  1011-9050-000, 010084068210229221APWW0179111171023, (APWW01791);  2097675-001, Not Applicable, (5450983-1);  2097675-001, Not Applicable, (5450983-2);  2097675-001, Not Applicable, (5466569-1);  1011-9050-000, 010084068210229221APWX0027711180118, (APWX00277);  1011-9050-000, 010084068210229221APWX0027811180118, (APWX00278);  1011-9050-000, 010084068210229221APWX0027911180118, (APWX00279);  1011-9050-000, 010084068210229221APWX0028011180118, (APWX00280);  1011-9050-000, 010084068210229221APWX0028111180118, (APWX00281);  1011-9050-000, 010084068210229221APWX0028211180118, (APWX00282);  1011-9050-000, 010084068210229221APWX0028311180118, (APWX00283);  1011-9050-000, 010084068210229221APWX0176211180720, (APWX01762);  1011-9050-000, 010084068210229221APWX0176311180720, (APWX01763);  1011-9050-000, 010084068210229221APWX0176411180720, (APWX01764);  1011-9050-000, 010084068210229221APWX0176511180720, (APWX01765);  1011-9050-000, 010084068210229221APWX0090611180328, (APWX00906);  1011-9050-000, 010084068210229221APWX0054011180222, (APWX00540);  1011-9050-000, 010084068210229221APWX0054311180301, (APWX00543);  1011-9050-000, 010084068210229221APWX0053111180222, (APWX00531);  1011-9050-000, 010084068210229221APWX0086311180328, (APWX00863);  1011-9050-000, 010084068210229221APWX0086411180328, (APWX00864);  1011-9050-000, 010084068210229221APWW0207011171206, (APWW02070);  1011-9050-000, 010084068210229221APWX0053011180222, (APWX00530);  1011-9050-000, 010084068210229221APWX0054211180301, (APWX00542);  1011-9050-000, 010084068210229221APWX0054111180301, (APWX00541);  1011-9050-000, 010084068210229221APWX0053611180222, (APWX00536);  1011-9050-000, 010084068210229221APWX0053911180222, (APWX00539);  1011-9050-000, 010084068210229221APWX0054411180301, (APWX00544);  1011-9050-000, Not Applicable, (APWT02872);  1011-9050-000, Not Applicable, (APWT02873);  1011-9050-000, Not Applicable, (APWT02876);  1011-9050-000, Not Applicable, (APWT02877);  1011-9050-000, Not Applicable, (APWT02878);  1011-9050-000, Not Applicable, (APWT02927);  1011-9050-000, Not Applicable, (APWU00128);  1011-9050-000, Not Applicable, (APWU00129);  1011-9050-000, Not Applicable, (APWU01145);  1011-9050-000, Not Applicable, (APWU01425);  2097675-001, Not Applicable, (5449421-1);  2097675-001, Not Applicable, (5476468-18);  2097675-001, Not Applicable, (5476468-19);  2097675-001, Not Applicable, (5476468-20);  2097675-001, Not Applicable, (5476468-21);  2097675-001, Not Applicable, (5476468-22);  2097675-001, Not Applicable, (5476468-23);  2097675-001, Not Applicable, (5476468-24);  2097675-001, Not Applicable, (5476468-25);  2097675-001, Not Applicable, (5476468-26);  2097675-001, Not Applicable, (5476468-27);  2097675-001, Not Applicable, (5476468-28);  2097675-001, Not Applicable, (5476468-29);  2097675-001, Not Applicable, (5476468-30);  2097675-001, Not Applicable, (5476468-31);  2097675-001, Not Applicable, (5476468-32);  2097675-001, Not Applicable, (5476468-33);  2097675-001, Not Applicable, (5476468-34);  2097675-001, Not Applicable, (5476468-35);  2097675-001, Not Applicable, (5476468-36);  2097675-001, Not Applicable, (5476468-37);  2097675-001, Not Applicable, (5476468-38);  2097675-001, Not Applicable, (5476468-39);  2097675-001, Not Applicable, (5476468-40);  2097675-001, Not Applicable, (5476468-41);  2097675-001, Not Applicable, (5476468-42);  2097675-001, Not Applicable, (5476468-43);  2097675-001, Not Applicable, (5476468-44);  2097675-001, Not Applicable, (5476468-45);  2097675-001, Not Applicable, (5476468-46);  2097675-001, Not Applicable, (5476468-47);  2097675-001, Not Applicable, (5476468-48);  2097675-001, Not Applicable, (5476468-49);  2097675-001, Not Applicable, (5476468-50);  2097675-001, Not Applicable, (5476468-51);  2097675-001, Not Applicable, (5476468-52);  2097675-001, Not Applicable, (5476468-53);  2097675-001, Not Applicable, (5476468-54);  2097675-001, Not Applicable, (5476468-55);  2097675-001, Not Applicable, (5476468-56);  2097675-001, Not Applicable, (5476468-57);  2097675-001, Not Applicable, (5476468-58);  2097675-001, Not Applicable, (5476468-59);  2097675-001, Not Applicable, (5476468-60);  2097675-001, Not Applicable, (5476468-61);  2097675-001, Not Applicable, (5476468-62);  2097675-001, Not Applicable, (5476468-63);  2097675-001, Not Applicable, (5476468-64);  2097675-001, Not Applicable, (5476468-65);  2097675-001, Not Applicable, (5476468-66);  2097675-001, Not Applicable, (5476468-67);  2097675-001, Not Applicable, (5476468-68);  2097675-001, Not Applicable, (5476468-69);  1011-9050-000, 010084068210229221APWX0086011180328, (APWX00860);  1011-9050-000, 010084068210229221APWX0088011180328, (APWX00880);  1011-9050-000, 010084068210229221APWX0089811180328, (APWX00898);  1011-9050-000, 010084068210229221APWX0090711180328, (APWX00907);  1011-9050-000, 010084068210229221APWX0126011180604, (APWX01260);  1011-9050-000, 010084068210229221APWX0126111180604, (APWX01261);  1011-9050-000, 010084068210229221APWX0182811180802, (APWX01828);  1011-9050-000, 010084068210229221APWX0182911180802, (APWX01829);  1011-9050-000, 010084068210229221APWX0192411180815, (APWX01924);  1011-9050-000, 010084068210229221APWX0053811180222, (APWX00538);  1011-9050-000, 010084068210229221APWX0053711180222, (APWX00537);  1011-9050-000, 010084068210229221APWV0187911160831, (APWV01879);  1011-9050-000, 010084068210229221APWW0012111170103, (APWW00121);  1011-9050-000, 010084068210229221APWW0013111170103, (APWW00131);  1011-9050-000, 010084068210229221APWW0052011170215, (APWW00520);  1011-9050-000, 010084068210229221APWW0204611171130, (APWW02046);  1011-9050-000, 010084068210229221APWW0204711171130, (APWW02047);  1011-9050-000, 010084068210229221APWW0204811171130, (APWW02048);  1011-9050-000, 010084068210229221APWW0204911171130, (APWW02049);  1011-9050-000, 010084068210229221APWW0205011171130, (APWW02050);  1011-9050-000, 010084068210229221APWX0017411180105, (APWX00174);  1011-9050-000, 010084068210229221APWX0024211180116, (APWX00242);  1011-9050-000, 010084068210229221APWX0025311180116, (APWX00253);  1011-9050-000, 010084068210229221APWX0025411180116, (APWX00254);  1011-9050-000, 010084068210229221APWX0025511180116, (APWX00255);  1011-9050-000, 010084068210229221APWX0025611180116, (APWX00256);  1011-9050-000, 010084068210229221APWX0135711180611, (APWX01357);  1011-9050-000, 010084068210229221APWW0199811171127, (APWW01998);  1011-9050-000, 010084068210229221APWW0143011170809, (APWW01430);  1011-9050-000, 010084068210229221APWW0143111170822, (APWW01431);  1011-9050-000, 010084068210229221APWW0143211170822, (APWW01432);  1011-9050-000, 010084068210229221APWW0143311170822, (APWW01433);  1011-9050-000, 010084068210229221APWW0143411170822, (APWW01434);  1011-9050-000, 010084068210229221APWW0143511170822, (APWW01435);  1011-9050-000, 010084068210229221APWW0143611170822, (APWW01436);  1011-9050-000, 010084068210229221APWX0033611180123, (APWX00336);  1011-9050-000, 010084068210229221APWX0033711180123, (APWX00337);  1011-9050-000, 010084068210229221APWX0033811180123, (APWX00338);  1011-9050-000, 010084068210229221APWX0033911180123, (APWX00339);  1011-9050-000, 010084068210229221APWW0173511171013, (APWW01735);  1011-9050-000, 010084068210229221APWW0173611171013, (APWW01736);  1011-9050-000, 010084068210229221APWW0173711171013, (APWW01737);  1011-9050-000, 010084068210229221APWW0173811171013, (APWW01738);  2097675-SV, Not Applicable, (5483715-1);  1011-9050-000, 010084068210229221APWV0060811160310, (APWV00608);  2097675-SV, Not Applicable, (5452850-1);  1011-9050-000, 010084068210229221APWW0134511170724, (APWW01345);  1011-9050-000, Not Applicable, (APWS00143);  1011-9050-000, 010084068210229221APWX0109411180502, (APWX01094);  1011-9050-000, 010084068210229221APWX0109511180502, (APWX01095);  1011-9050-000, 010084068210229221APWX0121211180530, (APWX01212);  1011-9050-000, 010084068210229221APWX0151811180625, (APWX01518);  1011-9050-000, 010084068210229221APWX0151911180625, (APWX01519);  1011-9050-000, 010084068210229221APWX0165211180716, (APWX01652);  1011-9050-000, 010084068210229221APWX0168711180716, (APWX01687);  1011-9050-000, 010084068210229221APWX0188211180803, (APWX01882);  1011-9050-000, 010084068210229221APWX0188311180803, (APWX01883);  1011-9050-000, 010084068210229221APWX0186611180803, (APWX01866);  1011-9050-000, 010084068210229221APWX0186711180803, (APWX01867);  1011-9050-000, 010084068210229221APWX0167711180716, (APWX01677);  1011-9050-000, 010084068210229221APWX0167811180716, (APWX01678);  1011-9050-000, 010084068210229221APWX0167911180716, (APWX01679);  1011-9050-000, 010084068210229221APWX0168011180716, (APWX01680);  1011-9050-000, 010084068210229221APWX0016311180105, (APWX00163);  1011-9050-000, 010084068210229221APWX0016411180105, (APWX00164);  1011-9050-000, 010084068210229221APWX0016511180105, (APWX00165);  1011-9050-000, 010084068210229221APWX0173311180720, (APWX01733);  1011-9050-000, 010084068210229221APWX0173411180720, (APWX01734);  1011-9050-000, 010084068210229221APWX0175111180720, (APWX01751);  1011-9050-000, 010084068210229221APWX0175211180720, (APWX01752);  1011-9050-000, 010084068210229221APWX0175311180720, (APWX01753);  1011-9050-000, 010084068210229221APWX0175411180720, (APWX01754);  1011-9050-000, 010084068210229221APWX0175511180720, (APWX01755);  1011-9050-000, 010084068210229221APWX0175611180720, (APWX01756);  1011-9050-000, 010084068210229221APWX0155811180627, (APWX01558);  1011-9050-000, 010084068210229221APWX0155911180627, (APWX01559);  1011-9050-000, 010084068210229221APWX0156011180627, (APWX01560);  1011-9050-000, 010084068210229221APWX0156111180703, (APWX01561);  1011-9050-000, 010084068210229221APWX0156211180703, (APWX01562);  1011-9050-000, 010084068210229221APWX0156311180703, (APWX01563);  1011-9050-000, 010084068210229221APWX0116711180517, (APWX01167);  1011-9050-000, 010084068210229221APWX0116811180517, (APWX01168);  1011-9050-000, 010084068210229221APWX0116911180517, (APWX01169);  1011-9050-000, 010084068210229221APWX0117011180517, (APWX01170);  1011-9050-000, 010084068210229221APWX0117111180517, (APWX01171);  1011-9050-000, 010084068210229221APWX0117211180517, (APWX01172);  1011-9050-000, 010084068210229221APWX0010111180104, (APWX00101);  1011-9050-000, 010084068210229221APWX0192211180815, (APWX01922);  2097675-001, Not Applicable, (5465604-1);  1011-9050-000, 010084068210229221APWX0047211180208, (APWX00472);  1011-9050-000, 010084068210229221APWX0047311180208, (APWX00473);  1011-9050-000, 010084068210229221APWX0047411180208, (APWX00474);  1011-9050-000, 010084068210229221APWX0047511180208, (APWX00475);  1011-9050-000, 010084068210229221APWX0090011180328, (APWX00900);  1011-9050-000, 010084068210229221APWX0090111180328, (APWX00901);  1011-9050-000, 010084068210229221APWX0173511180720, (APWX01735);  1011-9050-000, 010084068210229221APWX0173611180720, (APWX01736);  1011-9050-000, 010084068210229221APWX0173711180720, (APWX01737);  1011-9050-000, 010084068210229221APWX0173811180720, (APWX01738);  1011-9050-000, 010084068210229221APWX0173911180720, (APWX01739);  1011-9050-000, 010084068210229221APWX0174011180720, (APWX01740);  1011-9050-000, 010084068210229221APWX0174111180720, (APWX01741);  1011-9050-000, 010084068210229221APWX0174211180720, (APWX01742);  1011-9050-000, 010084068210229221APWX0174311180720, (APWX01743);  1011-9050-000, 010084068210229221APWX0103211180427, (APWX01032);  1011-9050-000, 010084068210229221APWX0103311180427, (APWX01033);  1011-9050-000, 010084068210229221APWX0103411180427, (APWX01034);  1011-9050-000, 010084068210229221APWX0103511180427, (APWX01035);  1011-9050-000, 010084068210229221APWX0103611180427, (APWX01036);  1011-9050-000, 010084068210229221APWX0103711180427, (APWX01037);  1011-9050-000, 010084068210229221APWX0103811180427, (APWX01038);  1011-9050-000, 010084068210229221APWX0111711180507, (APWX01117);  1011-9050-000, 010084068210229221APWV0054111160302, (APWV00541);  1011-9050-000, 010084068210229221APWV0054211160302, (APWV00542);  1011-9050-000, 010084068210229221APWV0054311160302, (APWV00543);  1011-9050-000, 010084068210229221APWV0054411160302, (APWV00544);  1011-9050-000, 010084068210229221APWV0054511160302, (APWV00545);  1011-9050-000, 010084068210229221APWV0054611160302, (APWV00546);  1011-9050-000, 010084068210229221APWV0235711161108, (APWV02357);  1011-9050-000, 010084068210229221APWV0235811161108, (APWV02358);  1011-9050-000, 010084068210229221APWV0235911161108, (APWV02359);  1011-9050-000, 010084068210229221APWV0236011161108, (APWV02360);  1011-9050-000, 010084068210229221APWV0236111161108, (APWV02361);  1011-9050-000, 010084068210229221APWV0236211161108, (APWV02362);  1011-9050-000, 010084068210229221APWX0020211180115, (APWX00202);  1011-9050-000, 010084068210229221APWX0020311180115, (APWX00203);  1011-9050-000, 010084068210229221APWX0020411180115, (APWX00204);  2097675-001, Not Applicable, (5448868-1);  2097675-001, Not Applicable, (5448868-2);  2097675-001, Not Applicable, (5454532-1);  2097675-001, Not Applicable, (5454532-2);  1011-9050-000, 010084068210229221APWW0171511171013, (APWW01715);  1011-9050-000, 010084068210229221APWX0111211180507, (APWX01112);  1011-9050-000, 010084068210229221APWX0111311180507, (APWX01113);  1011-9050-000, 010084068210229221APWX0111411180507, (APWX01114);  1011-9050-000, 010084068210229221APWX0152411180627, (APWX01524);  1011-9050-000, 010084068210229221APWX0152511180627, (APWX01525);  1011-9050-000, 010084068210229221APWX0152611180627, (APWX01526);  1011-9050-000, 010084068210229221APWX0152711180627, (APWX01527);  1011-9050-000, 010084068210229221APWX0025711180116, (APWX00257);  1011-9050-000, Not Applicable, (APWT02277);  1011-9050-000, 010084068210229221APWX0186411180803, (APWX01864);  1011-9050-000, 010084068210229221APWX0186511180803, (APWX01865);  1011-9050-000, 010084068210229221APWX0123411180531, (APWX01234);  1011-9050-000, 010084068210229221APWX0123511180531, (APWX01235);  1011-9050-000, 010084068210229221APWX0123611180531, (APWX01236);  1011-9050-000, 010084068210229221APWX0123711180531, (APWX01237);  1011-9050-000, 010084068210229221APWV0092411160427, (APWV00924);  1011-9050-000, 010084068210229221APWV0246411161117, (APWV02464);  1011-9050-000, 010084068210229221APWV0246511161117, (APWV02465);  2097675-001, Not Applicable, (5477529-1);  1011-9050-000, 010084068210229221APWW0149611170830, (APWW01496);  1011-9050-000, 010084068210229221APWX0049911180208, (APWX00499);  1011-9050-000, 010084068210229221APWX0050011180208, (APWX00500);  1011-9050-000, 010084068210229221APWX0050111180222, (APWX00501);  1011-9050-000, 010084068210229221APWX0050211180222, (APWX00502);  1011-9050-000, 010084068210229221APWX0050311180222, (APWX00503);  1011-9050-000, 010084068210229221APWX0050411180222, (APWX00504);  1011-9050-000, 010084068210229221APWX0050511180222, (APWX00505);  1011-9050-000, 010084068210229221APWX0050611180222, (APWX00506);  1011-9050-000, 010084068210229221APWX0050711180222, (APWX00507);  1011-9050-000, 010084068210229221APWX0050811180222, (APWX00508);  1011-9050-000, 010084068210229221APWX0050911180222, (APWX00509);  1011-9050-000, 010084068210229221APWX0051011180222, (APWX00510);  1011-9050-000, 010084068210229221APWX0051111180222, (APWX00511);  1011-9050-000, 010084068210229221APWX0051211180222, (APWX00512);  1011-9050-000, 010084068210229221APWX0051311180222, (APWX00513);  1011-9050-000, 010084068210229221APWX0051411180222, (APWX00514);  1011-9050-000, 010084068210229221APWX0051511180222, (APWX00515);  1011-9050-000, 010084068210229221APWX0051611180222, (APWX00516);  1011-9050-000, 010084068210229221APWX0051711180222, (APWX00517);  1011-9050-000, 010084068210229221APWX0051811180222, (APWX00518);  1011-9050-000, 010084068210229221APWX0051911180222, (APWX00519);  1011-9050-000, 010084068210229221APWX0052011180222, (APWX00520);  1011-9050-000, 010084068210229221APWX0052111180222, (APWX00521);  1011-9050-000, 010084068210229221APWX0052211180222, (APWX00522);  1011-9050-000, 010084068210229221APWX0052311180222, (APWX00523);  1011-9050-000, 010084068210229221APWX0052411180222, (APWX00524);  1011-9050-000, 010084068210229221APWX0052511180222, (APWX00525);  1011-9050-000, 010084068210229221APWX0052611180222, (APWX00526);  1011-9050-000, 010084068210229221APWX0110411180502, (APWX01104);  1011-9050-000, 010084068210229221APWX0110511180502, (APWX01105);  1011-9050-000, 010084068210229221APWX0110611180502, (APWX01106);  1011-9050-000, 010084068210229221APWX0110711180502, (APWX01107);  1011-9050-000, 010084068210229221APWX0110811180502, (APWX01108);  1011-9050-000, 010084068210229221APWX0110911180502, (APWX01109);  1011-9050-000, 010084068210229221APWX0111011180502, (APWX01110);  1011-9050-000, 010084068210229221APWX0111111180507, (APWX01111);  1011-9050-000, 010084068210229221APWX0011011180104, (APWX00110);  1011-9050-000, 010084068210229221APWX0058711180307, (APWX00587);  1011-9050-000, 010084068210229221APWX0058911180307, (APWX00589);  1011-9050-000, 010084068210229221APWX0010911180104, (APWX00109);  1011-9050-000, 010084068210229221APWW0142611170809, (APWW01426);  1011-9050-000, 010084068210229221APWW0171711171013, (APWW01717);  1011-9050-000, 010084068210229221APWW0171911171013, (APWW01719);  1011-9050-000, 010084068210229221APWW0172011171013, (APWW01720);  1011-9050-000, 010084068210229221APWW0174111171013, (APWW01741);  1011-9050-000, 010084068210229221APWW0174211171013, (APWW01742);  1011-9050-000, 010084068210229221APWW0174311171013, (APWW01743);  1011-9050-000, 010084068210229221APWX0043011180206, (APWX00430);  1011-9050-000, 010084068210229221APWX0043111180206, (APWX00431);  1011-9050-000, 010084068210229221APWW0224511171219, (APWW02245);  1011-9050-000, 010084068210229221APWW0224611171219, (APWW02246);  1011-9050-000, 010084068210229221APWW0224711171219, (APWW02247);  1011-9050-000, 010084068210229221APWW0224811171219, (APWW02248);  1011-9050-000, 010084068210229221APWX0087511180328, (APWX00875);  1011-9050-000, 010084068210229221APWX0087611180328, (APWX00876);  1011-9050-000, 010084068210229221APWX0144011180618, (APWX01440);  1011-9050-000, 010084068210229221APWX0144111180619, (APWX01441);  1011-9050-000, 010084068210229221APWX0144211180619, (APWX01442);  1011-9050-000, 010084068210229221APWX0144311180619, (APWX01443);  1011-9050-000, 010084068210229221APWX0144411180619, (APWX01444);  1011-9050-000, 010084068210229221APWX0144511180619, (APWX01445);  1011-9050-000, 010084068210229221APWX0144611180619, (APWX01446);  1011-9050-000, 010084068210229221APWX0144711180619, (APWX01447);  1011-9050-000, 010084068210229221APWX0144811180619, (APWX01448);  1011-9050-000, 010084068210229221APWX0144911180619, (APWX01449);  1011-9050-000, 010084068210229221APWX0145011180619, (APWX01450);  1011-9050-000, 010084068210229221APWX0145111180619, (APWX01451);  1011-9050-000, 010084068210229221APWX0145211180619, (APWX01452);  1011-9050-000, 010084068210229221APWX0145311180619, (APWX01453);  1011-9050-000, 010084068210229221APWX0142311180618, (APWX01423);  1011-9050-000, 010084068210229221APWW0207111171206, (APWW02071);  1011-9050-000, 010084068210229221APWW0207211171206, (APWW02072);  1011-9050-000, 010084068210229221APWW0207311171206, (APWW02073);  1011-9050-000, 010084068210229221APWW0207411171206, (APWW02074);  1011-9050-000, 010084068210229221APWW0207511171206, (APWW02075);  1011-9050-000, 010084068210229221APWW0207611171206, (APWW02076);  1011-9050-000, 010084068210229221APWW0207711171206, (APWW02077);  1011-9050-000, 010084068210229221APWX0058811180307, (APWX00588);  1011-9050-000, 010084068210229221APWW0085111170412, (APWW00851);  1011-9050-000, 010084068210229221APWW0142511170809, (APWW01425);  1011-9050-000, 010084068210229221APWW0142711170809, (APWW01427);  1011-9050-000, 010084068210229221APWW0142811170809, (APWW01428);  1011-9050-000, 010084068210229221APWW0149211170830, (APWW01492);  1011-9050-000, 010084068210229221APWW0149411170830, (APWW01494);  1011-9050-000, 010084068210229221APWW0149511170830, (APWW01495);  1011-9050-000, 010084068210229221APWW0149311170830, (APWW01493);  1011-9050-000, 010084068210229221APWW0171611171013, (APWW01716);  1011-9050-000, 010084068210229221APWW0171811171013, (APWW01718);  1011-9050-000, 010084068210229221APWX0014011180104, (APWX00140);  1011-9050-000, 010084068210229221APWX0014111180105, (APWX00141);  1011-9050-000, 010084068210229221APWX0014211180105, (APWX00142);  1011-9050-000, 010084068210229221APWX0014311180105, (APWX00143);  1011-9050-000, 010084068210229221APWX0043211180206, (APWX00432);  1011-9050-000, 010084068210229221APWX0090811180328, (APWX00908);  1011-9050-000, 010084068210229221APWX0090911180328, (APWX00909);  1011-9050-000, 010084068210229221APWX0091011180328, (APWX00910);  1011-9050-000, 010084068210229221APWX0091111180410, (APWX00911);  1011-9050-000, 010084068210229221APWX0091211180410, (APWX00912);  1011-9050-000, 010084068210229221APWX0091311180410, (APWX00913);  1011-9050-000, 010084068210229221APWX0091411180410, (APWX00914);  1011-9050-000, 010084068210229221APWX0091511180410, (APWX00915);  1011-9050-000, 010084068210229221APWX0091611180410, (APWX00916);  1011-9050-000, 010084068210229221APWX0091711180410, (APWX00917);  1011-9050-000, 010084068210229221APWX0091811180410, (APWX00918);  1011-9050-000, 010084068210229221APWX0091911180410, (APWX00919);  1011-9050-000, 010084068210229221APWX0092011180410, (APWX00920);  1011-9050-000, 010084068210229221APWX0092111180410, (APWX00921);  1011-9050-000, 010084068210229221APWX0092211180410, (APWX00922);  1011-9050-000, 010084068210229221APWX0092311180410, (APWX00923);  1011-9050-000, 010084068210229221APWX0092411180410, (APWX00924);  1011-9050-000, 010084068210229221APWX0092511180410, (APWX00925);  1011-9050-000, 010084068210229221APWX0092611180410, (APWX00926);  1011-9050-000, 010084068210229221APWX0189711180803, (APWX01897);  2097675-001, Not Applicable, (5477193-1);  2097675-001, Not Applicable, (5477193-10);  2097675-001, Not Applicable, (5477193-11);  2097675-001, Not Applicable, (5477193-12);  2097675-001, Not Applicable, (5477193-2);  2097675-001, Not Applicable, (5477193-3);  2097675-001, Not Applicable, (5477193-4);  2097675-001, Not Applicable, (5477193-5);  2097675-001, Not Applicable, (5477193-6);  2097675-001, Not Applicable, (5477193-7);  2097675-001, Not Applicable, (5477193-8);  2097675-001, Not Applicable, (5477193-9);  1011-9050-000, 010084068210229221APWX0126311180604, (APWX01263);  1011-9050-000, 010084068210229221APWX0126411180604, (APWX01264);  1011-9050-000, 010084068210229221APWX0126511180604, (APWX01265);  1011-9050-000, 010084068210229221APWX0126611180604, (APWX01266);  1011-9050-000, 010084068210229221APWX0126711180604, (APWX01267);  1011-9050-000, 010084068210229221APWX0126811180604, (APWX01268);  1011-9050-000, 010084068210229221APWX0126911180604, (APWX01269);  1011-9050-000, 010084068210229221APWX0127011180604, (APWX01270);  1011-9050-000, 010084068210229221APWW0167611171004, (APWW01676);  1011-9050-000, 010084068210229221APWX0015311180105, (APWX00153);  1011-9050-000, 010084068210229221APWX0017911180105, (APWX00179);  1011-9050-000, 010084068210229221APWX0018011180105, (APWX00180);  1011-9050-000, 010084068210229221APWX0018111180115, (APWX00181);  1011-9050-000, 010084068210229221APWX0018211180115, (APWX00182);  1011-9050-000, 010084068210229221APWX0018311180115, (APWX00183);  1011-9050-000, 010084068210229221APWX0018411180115, (APWX00184);  1011-9050-000, 010084068210229221APWX0018511180115, (APWX00185);  1011-9050-000, 010084068210229221APWX0018611180115, (APWX00186);  1011-9050-000, 010084068210229221APWX0018711180115, (APWX00187);  1011-9050-000, 010084068210229221APWX0018811180115, (APWX00188);  1011-9050-000, 010084068210229221APWX0018911180115, (APWX00189);  1011-9050-000, 010084068210229221APWX0019011180115, (APWX00190);  1011-9050-000, 010084068210229221APWX0019111180115, (APWX00191);  2097675-001, Not Applicable, (5477192-1);  2097675-001, Not Applicable, (5477192-10);  2097675-001, Not Applicable, (5477192-11);  2097675-001, Not Applicable, (5477192-12);  2097675-001, Not Applicable, (5477192-2);  2097675-001, Not Applicable, (5477192-3);  2097675-001, Not Applicable, (5477192-4);  2097675-001, Not Applicable, (5477192-5);  2097675-001, Not Applicable, (5477192-6);  2097675-001, Not Applicable, (5477192-7);  2097675-001, Not Applicable, (5477192-8);  2097675-001, Not Applicable, (5477192-9);  1011-9050-000, 010084068210229221APWX0126211180604, (APWX01262);  1011-9050-000, 010084068210229221APWW0187511171109, (APWW01875);  1011-9050-000, 010084068210229221APWX0164611180716, (APWX01646);  1011-9050-000, 010084068210229221APWX0164711180716, (APWX01647);  1011-9050-000, 010084068210229221APWX0173011180718, (APWX01730);  1011-9050-000, 010084068210229221APWX0173111180720, (APWX01731);   Additional Serial Numbers - Model Number, UDI Number, (Serial Number): 2097675-001-US, Not Applicable, (APWV01637);  2097675-001-US, Not Applicable, (APWV01638);  2097675-001-US, Not Applicable, (APWU01054);  2097675-001-US, Not Applicable, (APWU01055);  2097675-001-US, Not Applicable, (APWU01056);  2097675-001-US, Not Applicable, (APWV01635);  2097675-001-US, Not Applicable, (APWV01631);  2097675-001-US, Not Applicable, (APWV01632);  2097675-001-US, Not Applicable, (APWV01633);  2097675-001-US, Not Applicable, (APWV01634);  2097675-001-US, Not Applicable, (APWV01636);  2097673-001-US, Not Applicable, (ANAM00306);  2097673-001-US, Not Applicable, (ANAP00354);  2097675-001-US, Not Applicable, (ANAP00691);  2097673-001-US, Not Applicable, (APWU00297);  2097673-001-US, Not Applicable, (ANAP00715);  2097673-001-US, Not Applicable, (ANAR00602);  2097673-001-US, Not Applicable, (ANAR00603);  1011-9050-000, Not Applicable, (APWU01813);  2097675-001-US, Not Applicable, (APWU01910);  2097675-001-US, Not Applicable, (ANAS01096);  2097673-001-US, Not Applicable, (ANAN00169);  2097673-001-US, Not Applicable, (ANAP00574);  2097673-001-US, Not Applicable, (ANAP00575);  2097673-001-US, Not Applicable, (ANAP00576);  2097673-001-US, Not Applicable, (ANAP00577);  2097673-001-US, Not Applicable, (ANAQ01175);  2097673-001-US, Not Applicable, (ANAP00172);  2097673-001-US, Not Applicable, (ANAP00861);  2097673-001-US, Not Applicable, (ANAP00870);  2097673-001-US, Not Applicable, (ANAP00872);  2097675-001-US, Not Applicable, (APWW00907);  2097675-001-US, Not Applicable, (APWW00908);  2097675-001-US, Not Applicable, (APWW00917);  2097675-001-US, Not Applicable, (APWW00920);  2097675-001-US, Not Applicable, (APWW01364);  2097675-001-US, Not Applicable, (APWW01365);  2097675-001-US, Not Applicable, (APWW01366);  2097675-001-US, Not Applicable, (APWW01480);  2097675-001-US, Not Applicable, (APWW01481);  2097675-001-US, Not Applicable, (APWW01497);  2097675-001-US, Not Applicable, (APWW01498);  2097675-001-US, Not Applicable, (APWW01499);  2097675-001-US, Not Applicable, (APWW01500);  2097675-001-US, Not Applicable, (APWW01501);  2097675-001-US, Not Applicable, (APWW01502);  2097675-001-US, Not Applicable, (APWW01503);  2097675-001-US, Not Applicable, (APWW01504);  2097675-001-US, Not Applicable, (APWW01505);  2097675-001-US, Not Applicable, (APWW01506);  2097675-001-US, Not Applicable, (APWW01519);  2097675-001-US, Not Applicable, (ANAS00396);  2097675-001-US, Not Applicable, (ANAS00393);  2097675-001-US, Not Applicable, (ANAS00394);  2097675-001-US, Not Applicable, (ANAS00395);  2097675-001-US, Not Applicable, (ANAS00397);  2097673-001-US, Not Applicable, (ANAS00398 );  2097675-001-US, Not Applicable, (ANAS00399);  2097675-001-US, Not Applicable, (ANAS00400);  2097675-001-US, Not Applicable, (ANAS00401);  2097675-001-US, Not Applicable, (ANAS00402);  2097675-001-US, Not Applicable, (ANAS00403);  2097675-001-US, Not Applicable, (ANAS00404);  2097675-001-US, Not Applicable, (ANAS00405);  2097675-001-US, Not Applicable, (ANAS00406);  2097675-001-US, Not Applicable, (ANAS00407);  2097675-001-US, Not Applicable, (ANAS00408);  2097675-001-US, Not Applicable, (ANAS00409);  2097675-001-US, Not Applicable, (ANAS00410);  2097675-001-US, Not Applicable, (ANAS00483);  2097675-001-US, Not Applicable, (APWU00274);  2097675-001-US, Not Applicable, (APWU00275);  2097675-001-US, Not Applicable, (APWU00276);  2097675-001-US, Not Applicable, (APWU00277);  2097675-001-US, Not Applicable, (APWU00278);  2097675-001-US, Not Applicable, (APWU00279);  2097675-001-US, Not Applicable, (APWU00280);  2097675-001-US, Not Applicable, (APWU00281);  2097675-001-US, Not Applicable, (APWU00282);  2097675-001-US, Not Applicable, (APWU00283);  2097675-001-US, Not Applicable, (APWU00284);  2097675-001-US, Not Applicable, (APWU00285);  2097675-001-US, Not Applicable, (APWU00286);  2097675-001-US, Not Applicable, (APWU00287);  2097675-001-US, Not Applicable, (APWU00288);  2097675-001-US, Not Applicable, (APWU00289);  2097675-001-US, Not Applicable, (APWU00290);  2097675-001-US, Not Applicable, (APWU00291);  2097675-001-US, Not Applicable, (APWU00292);  2097673-001-US, Not Applicable, (ANAL00105);  2097673-001-US, Not Applicable, (ANAL00106);  2097673-001-US, Not Applicable, (ANAL00107);  2097673-001-US, Not Applicable, (ANAL00108);  2097673-001-US, Not Applicable, (ANAL00109);  2097673-001-US, Not Applicable, (ANAL00116);  2097673-001-US, Not Applicable, (ANAM00439);  2097673-001-US, Not Applicable.(ANAM00440);  2097673-001-US, Not Applicable, (ANAM00441);  2097675-001-US, Not Applicable, (APWW02034);  2097675-001-US, Not Applicable, (APWW02024);  2097675-001-US, Not Applicable, (APWW02025);  2097675-001-US, Not Applicable, (APWW02026);  2097675-001-US, Not Applicable, (APWW02030);  2097675-001-US, Not Applicable, (APWW02031);  2097675-001-US, Not Applicable, (APWW02032);  2097675-001-US, Not Applicable, (APWW02033);  2097675-001-US, Not Applicable, (APWW02035);  2097675-001-US, Not Applicable, (APWW02027);  2097675-001-US, Not Applicable, (APWW02028);  2097675-001-US, Not Applicable, (APWW02029);  2097675-001-US, Not Applicable, (APWW02212);  2097675-001-US, Not Applicable, (APWW02213);  1011-9050-000, Not Applicable, (APWW01679);  2097673-001-US, Not Applicable, (ANAP01088);  2097673-001-US, Not Applicable, (ANAP01090);  2097673-001-US, Not Applicable, (ANAP01091);  2097673-001, Not Applicable, (ANAN00604);  2097673-001, Not Applicable, (ANAM01165);  2097673-001, Not Applicable, (ANAM01158);  2097675-001, Not Applicable, (APWW01491);  2097675-001, Not Applicable, (APWU00251);  2097675-001, Not Applicable, (APWU00172);  2097675-001, Not Applicable, (APWU00173);  2097675-001, Not Applicable, (APWU00174);  2097675-001, Not Applicable, (APWU00823);  2097675-001, Not Applicable, (APWU00824);  2097675-001, Not Applicable, (APWU01779);  2097675-001, Not Applicable, (APWU01780);  2097675-001, Not Applicable, (APWV01441);  2097675-001, Not Applicable, (APWV01442);  2097675-001, Not Applicable, (APWV01443);  2097675-001, Not Applicable, (APWV01444);  2097675-001, Not Applicable, (APWV01445);  2097675-001, Not Applicable, (APWV02328);  2097675-001, Not Applicable, (APWW00161);  2097675-001, Not Applicable, (APWW00162);  2097675-001, Not Applicable, (APWW00163);  2097675-001, Not Applicable, (APWW00164);  2097675-001, Not Applicable, (APWW00165);  2097675-001, Not Applicable, (APWW00166);  2097675-001, Not Applicable, (APWW00167);  2097675-001, Not Applicable, (APWW00637);  2097675-001, Not Applicable, (APWW01088);  2097675-001, Not Applicable, (APWW01116);  2097675-001, Not Applicable, (APWW01117);  2097675-001, Not Applicable, (APWW01118);  2097675-001, Not Applicable, (APWW01119);  2097675-001, Not Applicable, (APWW01120);  2097675-001, Not Applicable, (APWW01121);  2097675-001, Not Applicable, (APWW01122);  2097675-001, Not Applicable, (APWW01123);  2097675-001, Not Applicable, (APWW01124);  2097675-001, Not Applicable, (APWW01125);  2097675-001, Not Applicable, (APWW01884);  2097675-001, Not Applicable, (APWW00792);  2097675-001, Not Applicable, (APWW00782);  2097675-001, Not Applicable, (APWW00783);  2097675-001, Not Applicable, (APWW00789);  2097675-001, Not Applicable, (APWW00790);  2097675-001, Not Applicable, (APWW00791);  2097675-001, Not Applicable, (APWW00793);  2097675-001, Not Applicable, (APWW01297);  2097675-001, Not Applicable, (APWW01298);  2097675-001, Not Applicable, (APWW01299);  2097675-001, Not Applicable, (APWV00685);  2097675-001, Not Applicable, (APWV00686);  2097675-001, Not Applicable, (APWV00687);  2097675-001, Not Applicable, (APWV00688);  2097675-001, Not Applicable, (APWV00689);  2097675-001, Not Applicable, (APWV00690);  2097675-001, Not Applicable, (APWV00691);  2097675-001, Not Applicable, (APWV00692);  2097675-001, Not Applicable, (APWV00693);  2097673-001, Not Applicable, (ANAM00815);  2097673-001, Not Applicable, (ANAM00816);  2097673-001, Not Applicable, (ANAM00817);  2097673-001, Not Applicable, (ANAM00818);  2097673-001, Not Applicable, (ANAM00819);  2097673-001, Not Applicable, (ANAM00820);  2097673-001, Not Applicable, (ANAM00821);  2097673-001, Not Applicable, (ANAM00822);  2097673-001, Not Applicable, (ANAM00823);  2097673-001, Not Applicable, (ANAM00824);  1011-9050-000, Not Applicable, (APWU01809);  1011-9050-000, Not Applicable, (APWU01810);  2097673-DE, Not Applicable, (ANAP01156);  2097675-DE, Not Applicable, (ANAP01234);  2097675-DE, Not Applicable, (APWS00156);  2097675-DE, Not Applicable, (APWT00183);  2097675-DE, Not Applicable, (APWT02507);  2097675-DE, Not Applicable, (APWT02509);  2097675-DE, Not Applicable, (APWU01889);  2097675-DE, Not Applicable, (APWV01053);  2097675-DE, Not Applicable, (APWV01498);  2097675-DE, Not Applicable, (APWV01499);  2097675-DE, Not Applicable, (APWV01709);  2097675-DE, Not Applicable, (APWV01710);  2097675-DE, Not Applicable, (APWV01728);  2097675-DE, Not Applicable, (APWV02233);  2097675-DE, Not Applicable, (APWV02428);  2097675-DE, Not Applicable, (APWV02429);  2097675-DE, Not Applicable, (APWV02430);  2097675-DE, Not Applicable, (APWW00827);  2097675-DE, Not Applicable, (APWW00828);  2097675-DE, Not Applicable, (APWW00829);  1011-9050-000-561097, Not Applicable, (APWX00896);  1011-9050-000-561097, Not Applicable, (APWX00897);  1011-9050-000-540673, 010084068210229221APWW0199211171127, (APWX00534);  1011-9050-000-540673, 010084068210229221APWW0199211171127, (APWX00535);  1011-9050-000-560349, 010084068210229221APWX0102411180417, (APWX01065);  1011-9050-000-560349, 010084068210229221APWX0102411180417, (APWX01066);  1011-9050-000-560349, 010084068210229221APWX0102411180417, (APWX01067);  1011-9050-000-560349, 010084068210229221APWX0102411180417, (APWX01068);  1011-9050-000-560349, 010084068210229221APWX0102411180417, (APWX01069);  2097673-001, Not Applicable, (ANAR00318);  2097673-001, Not Applicable, (ANAR00319);  2097675-001, Not Applicable, (ANAS00379);  2097675-001, Not Applicable, (ANAS00380);  2097675-001, Not Applicable, (ANAS00381);  2097675-001, Not Applicable, (ANAS00382);  2097675-001, Not Applicable, (ANAS00383);  2097675-001, Not Applicable, (ANAS00384);  2097675-001, Not Applicable, (APWW00613);  2097675-001, Not Applicable, (APWW00614);  2097675-001, Not Applicable, (APWW00615);  2097675-001, Not Applicable, (APWW00616);  2097675-001, Not Applicable, (APWW00617);  2097675-001, Not Applicable, (APWW00618);  2097675-001, Not Applicable, (APWW00619);  2097675-001, Not Applicable, (APWU01851);  2097675-001, Not Applicable, (APWU01852);  2097675-001, Not Applicable, (APWU01853);  2097675-001, Not Applicable, (APWU01854);  2097675-001, Not Applicable, (APWU01855);  2097675-001, Not Applicable, (APWU01856);  2097675-001, Not Applicable, (APWU01857);  2097675-001, Not Applicable, (APWU01858);  2097675-001, Not Applicable, (APWU01859);  2097675-001, Not Applicable, (APWU01860);  2097675-001, Not Applicable, (APWU01861);  2097675-001, Not Applicable, (APWU01862);  2097675-001, Not Applicable, (APWU01863);  2097675-001, Not Applicable, (APWU01864);  2097675-001, Not Applicable, (APWU01865);  2097675-001, Not Applicable, (APWU01866);  2097675-001, Not Applicable, (APWU01867);  2097675-001, Not Applicable, (APWU01868);  2097675-001, Not Applicable, (APWU01869);  2097675-001, Not Applicable, (APWU01870);  2097675-001, Not Applicable, (APWU01871);  2097675-001, Not Applicable, (APWU01872);  2097675-001, Not Applicable, (APWU01873);  2097675-001, Not Applicable, (APWU01874);  2097675-001, Not Applicable, (APWU01875);  1011-9050-000, Not Applicable, (APWU00216);  2097675-001, Not Applicable, (APWV01175);  2097675-001, Not Applicable, (APWV01184);  2097675-001, Not Applicable, (APWV01185);  2097675-001, Not Applicable, (APWV01186);  2097675-001, Not Applicable, (APWV01187);  2097675-001, Not Applicable, (APWV01188);  2097675-001, Not Applicable, (APWV01189);  2097675-001, Not Applicable, (APWV01190);  2097675-001, Not Applicable, (APWV01191);  2097675-001, Not Applicable, (APWV01192);  2097675-001, Not Applicable, (APWV01517);  2097675-001, Not Applicable, (APWV01518);  2097675-001, Not Applicable, (APWV01519);  2097675-001, Not Applicable, (APWV01520);  2097675-001, Not Applicable, (APWV01521);  2097675-001, Not Applicable, (APWV01522);  2097675-001, Not Applicable, (APWV01523);  2097675-001, Not Applicable, (APWV01524);  2097675-001, Not Applicable, (APWU00263);  2097675-001, Not Applicable, (APWV02603);  2097675-001, Not Applicable, (ANAS00471);  2097675-001, Not Applicable, (ANAS00474);  2097675-001, Not Applicable, (APWU01596);  2097675-001, Not Applicable, (APWU01597);  2097675-001, Not Applicable, (ANAS00470);  2097675-001, Not Applicable, (ANAS00472);  2097675-001, Not Applicable, (ANAS00473);  2097675-001, Not Applicable, (ANAS00469);  2097675-001, Not Applicable, (APWU02599);  2097675-001, Not Applicable, (APWU02600);  2097675-001, Not Applicable, (APWU02601);  2097675-001, Not Applicable, (APWU02602);  2097675-001, Not Applicable, (APWU02603);  2097675-001, Not Applicable, (APWV00237);  2097675-001, Not Applicable, (APWV00238);  2097675-001, Not Applicable, (APWV00239);  2097675-001, Not Applicable, (APWV00240);  2097675-001, Not Applicable, (APWV00241);  2097675-001, Not Applicable, (APWV00242);  2097675-001, Not Applicable, (APWV00243);  2097675-001, Not Applicable, (APWV00244);  2097675-001, Not Applicable, (APWV00245);  2097675-001, Not Applicable, (APWV00246);  2097675-001, Not Applicable, (APWV00247);  2097675-001, Not Applicable, (APWV00248);  2097675-001, Not Applicable, (APWV00249);  2097675-001, Not Applicable, (APWV00250);  2097675-001, Not Applicable, (APWV00251);  2097675-001, Not Applicable, (APWV00252);  2097675-001, Not Applicable, (APWV00253);  2097675-001, Not Applicable, (APWV00254);  2097675-001, Not Applicable, (APWV00255);  2097675-001, Not Applicable, (APWV00256);  2097675-001, Not Applicable, (APWV00257);  2097675-001, Not Applicable, (APWV00258);  2097675-001, Not Applicable, (APWV00259);  2097675-001, Not Applicable, (APWV00260);  2097675-001, Not Applicable, (APWV00261);  2097675-001, Not Applicable, (APWV00262);  2097675-001, Not Applicable, (APWV00263);  2097675-001, Not Applicable, (APWV00264);  2097675-001, Not Applicable, (APWV00265);  2097675-001, Not Applicable, (APWV00297);  2097675-001, Not Applicable, (APWV00298);  2097675-001, Not Applicable, (APWV00299);  2097675-001, Not Applicable, (APWV00300);  2097675-001, Not Applicable, (APWV00301);  2097675-001, Not Applicable, (APWW00521);  1011-9050-000, Not Applicable, (APWU00217);  2097673-FR, Not Applicable, (ANAR05232);  2097673-FR, Not Applicable, (ANAR05233);  2097673-FR, Not Applicable, (ANAR05234);  2097673-FR, Not Applicable, (ANAR05235);  2097673-FR, Not Applicable, (ANAR05236);  2097673-FR, Not Applicable, (ANAP01419);  2097675-001, Not Applicable, (APWU01928);  2097675-001, Not Applicable, (APWW00869);  2097675-001, Not Applicable, (APWW00870);  2097675-FI, Not Applicable, (APWW01104);  2097675-FI, Not Applicable, (APWT02991);  2097675-FI, Not Applicable, (APWT02992);  2097675-FI, Not Applicable, (APWT02994);  2097675-FI, Not Applicable, (APWU00325);  2097675-FI, Not Applicable, (APWU00921);  2097675-FI, Not Applicable, (APWU00922);  2097675-FI, Not Applicable, (APWU00923);  2097675-FI, Not Applicable, (APWU00969);  2097675-FI, Not Applicable, (APWU00970);  2097675-FI, Not Applicable, (APWU00971);  2097675-FI, Not Applicable, (APWW01047);  2097675-FI, Not Applicable, (APWW01048);  2097675-FI, Not Applicable, (APWW01094);  2097675-FI, Not Applicable, (APWW01095);  2097675-FI, Not Applicable, (APWW01730);  2097675-FI, Not Applicable, (APWU00402);  2097675-FI, Not Applicable, (APWT02993);  2097675-FI, Not Applicable, (ANAP01426);  2097675-FI, Not Applicable, (ANAP01427);  2097675-FI, Not Applicable, (APWT02990);  1011-9000-000, Not Applicable, (APWT00661);  1011-9000-000, Not Applicable, (APWT00662);  1011-9000-000, Not Applicable, (APWT00664);  1011-9000-000, Not Applicable, (ANAN00664);  1011-9000-000, Not Applicable, (ANAN00666);  1011-9000-000, Not Applicable, (ANAN00667);  1011-9000-000, Not Applicable, (ANAN00669);  1011-9000-000, Not Applicable, (ANAN00670);  1011-9000-000, Not Applicable, (ANAN00671);  1011-9000-000, Not Applicable, (ANAN00890);  1011-9000-000, Not Applicable, (ANAN00891);  1011-9000-000, Not Applicable, (ANAR00396);  1011-9000-000, Not Applicable, (APWT02012);  1011-9000-000, Not Applicable, (APWT02013);  1011-9000-000, Not Applicable, (ANAN01101);  1011-9000-000, Not Applicable, (ANAP00605);  2097675-FR, Not Applicable, (APWS00272);  1011-9050-000, Not Applicable, (APWT00233);  1011-9050-000-xxxx, Not Applicable, (APWW01257);  1011-9050-000, Not Applicable, (APWS00487);  1011-9050-000, Not Applicable, (APWS00488);  1011-9050-000-xxxx, Not Applicable, (APWW01258);  1011-9050-000-xxxx, Not Applicable, (APWW01259);  1011-9050-000-xxxx, Not Applicable, (APWW01260);  1011-9050-000-xxxx, Not Applicable, (APWW01261);  2097673-FR, Not Applicable, (ANAQ00803);  2097673-FR, Not Applicable, (ANAQ00806);  2097673-FR, Not Applicable, (ANAQ00948);  2097673-FR, Not Applicable, (ANAR05145);  2097673-FR, Not Applicable, (ANAR05150);  2097673-FR, Not Applicable, (ANAR05513);  2097673-FR, Not Applicable, (ANAR05517);  2097673-FR, Not Applicable, (ANAS00665);  2097673-FR, Not Applicable, (ANAS00666);  2097673-FR, Not Applicable, (ANAS00668);  2097673-FR, Not Applicable, (ANAS00971);  1011-9050-000, Not Applicable, (APWT00670);  1011-9050-000-548347, Not Applicable, (APWX00823);  1011-9050-000-548347, Not Applicable, (APWX00824);  1011-9050-000, 010084068210229221APWW0138111170802, (APWW01380);  1011-9050-000, 010084068210229221APWW0138111170802, (APWW01382);  1011-9050-000, Not Applicable, (APWV01462);  1011-9050-000, Not Applicable, (APWU00215);  2097675-NL, Not Applicable, (APWV00486);  2097675-NL, Not Applicable, (APWV00488);  2097675-NL, Not Applicable, (APWV00489);  2097675-NL, Not Applicable, (APWV00490);  2097675-NL, Not Applicable, (APWV00491);  2097675-NL, Not Applicable, (APWV00492);  2097675-NL, Not Applicable, (APWV00493);  2097675-NL, Not Applicable, (APWV00494);  2097675-NL, Not Applicable, (APWV00495);  2097675-NL, Not Applicable, (APWV00497);  2097675-NL, Not Applicable, (APWV00498);  2097675-NL, Not Applicable, (APWV00501);  2097675-NL, Not Applicable, (APWV00503);  2097675-NL, Not Applicable, (APWV00504);  2097675-NL, Not Applicable, (APWV00505);  2097675-NL, Not Applicable, (APWV00487);  2097675-NL, Not Applicable, (APWV00496);  2097675-NL, Not Applicable, (APWV00499);  2097675-NL, Not Applicable, (APWV00500);  2097675-NL, Not Applicable, (APWV00502);  2097673-NO, Not Applicable, (ANAQ01098);  1011-9050-000, Not Applicable, (APWU00218);  1011-9050-000, Not Applicable, (APWU00219);  1011-9050-000, Not Applicable, (APWU01811);  2097675-SV, Not Applicable, (APWW00601);  2097675-001, Not Applicable, (APWW00850);

Other devices under the same FDA submission

2 devices cleared under K110213, K170872 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 81182).