Touchscreen display kit. Sold under the following product… recallZ-0413-2019 · GE Healthcare, LLC · initiated Sep 27, 2018
Class II Terminated All recalls posted in November 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.
FDA root cause category: Software design
What the company is doing
GE Healthcare notified customers on about 09/27/2018 via "URGENT MEDICAL DEVICE CORRECTION" letter. The letter informed customers that they could continue to use the anesthesia device. The issue can only occur if the patient stops spontaneous breathing in PSVPro ventilation mode after completion of lung recruitment breaths. Stopping the Cycling Procedure resolves the issue. Continue to monitor patient as is normal for Apnea. A GE Healthcare representative will contact each customer to arrange for the correction. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-0413-2019
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 27, 2018
- Posted by the FDA
- Nov 9, 2018
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 4,207 units total
- Product code
- BSZ
- FDA submissions
- K110213, K170872
- Distribution
- Worldwide distribution - US nationwide in the states of CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, ,MA, MD, MI, MN, MO, MT, NC, NE, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. Countries of Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea (Republic of), Kuwait, Macedonia, Mexico, Netherlands, New Zealand, Nigeria, Norway, Peru, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
Product description
Touchscreen display kit. Sold under the following product names: TOUCHSCREEN DISPLAY KIT 11.X ENGLISH FOR USA, TOUCHSCREEN DISPLAY KIT 11.X ENGLISH, TOUCHSCREEN DISPLAY KIT 11.X GERMAN, TOUCHSCREEN DISPLAY KIT 11.X DUTCH, TOUCHSCREEN DISPLAY KIT 11.X FRENCH, TOUCHSCREEN DISPLAY KIT 11.X DANISH, TOUCHSCREEN DISPLAY KIT 11.X ESTONIAN, TOUCHSCREEN DISPLAY KIT 11.X ITALIAN, TOUCHSCREEN DISPLAY KIT 11.X NORWEGIAN. Product Usage: The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.
Code information (models, lots, serial numbers)
Show the full code information (8k characters)
Model Number, UDI Number, (Serial Number): 2097673-001-US (4651698-1); 2097673-001-US (4651698-3); 2097673-001-US (4651698-4); 2097673-001-US (4651726-1); 2097673-001-US (4652206-1); 2097673-001-US (4652206-2); 2097673-001-US (4649618-1); 2097673-001-US (4649618-2); 2097673-001-US (4649618-3); 2097673-001-US (4649618-4); 2097673-001-US (4649618-5); 2097673-001-US (4653227-1); 2097673-001-US (4653227-2); 2097673-001-US (4653227-3); 2097673-001-US (5474296-1); 2097673-001-US (4652587-1); 2097673-001-US (4652587-2); 2097673-001-US (4652587-3); 2097673-001-US (4652587-4); 2097673-001-US (4652587-5); 2097673-001-US (4652587-6); 2097673-001-US (4632861-1); 2097673-001-US (4632861-2); 2097673-001-US (4632861-3); 2097673-001-US (4644953-1); 2097673-001 (5463757-1); 2097673-001 (5467309-16); 2097673-001 (5471699-1); 2097673-001 (5471699-2); 2097673-001 (5471699-3); 2097673-001 (5475515-1); 2097673-001 (5475515-2); 2097673-001 (5475516-1); 2097673-001 (5475516-10); 2097673-001 (5475516-2); 2097673-001 (5475516-3); 2097673-001 (5475516-4); 2097673-001 (5475516-5); 2097673-001 (5475516-6); 2097673-001 (5475516-7); 2097673-001 (5475516-8); 2097673-001 (5475516-9); 2097673-001 (5475528-1); 2097673-001 (5475528-10); 2097673-001 (5475528-11); 2097673-001 (5475528-12); 2097673-001 (5475528-13); 2097673-001 (5475528-14); 2097673-001 (5475528-15); 2097673-001 (5475528-16); 2097673-001 (5475528-17); 2097673-001 (5475528-18); 097673-001 (5475528-19); 2097673-001 (5475528-2); 2097673-001 (5475528-3); 2097673-001 (5475528-4); 2097673-001 (5475528-5); 2097673-001 (5475528-6); 2097673-001 (5475528-7); 2097673-001 (5475528-8); 2097673-001 (5475528-9); 2097673-001 (5481650-1); 2097673-001 (5481650-10); 2097673-001 (5481650-2); 2097673-001 (5481650-3); 097673-001 (5481650-4); 2097673-001 (5481650-5); 2097673-001 (5481650-6); 2097673-001 (5481650-7); 2097673-001 (5481650-8); 2097673-001 (5481650-9); 2097673-001 (5471701-1); 2097673-001 (5471701-2); 2097673-001 (5471702-1); 2097673-001 (5472013-1); 2097673-001 (5472013-10); 097673-001 (5472013-12); 2097673-001 (5472013-14); 2097673-001 (5472013-15); 2097673-001 (5472013-16); 2097673-001 (5472013-17); 2097673-001 (5472013-18); 2097673-001 (5472013-3); 2097673-001 (5472013-4); 2097673-001 (5472013-5); 2097673-001 (5472013-6); 2097673-001 (5472013-7); 2097673-001 (5472013-8); 2097673-001 (5463631-1); 2097673-001 (5463631-2); 2097673-001 (5466892-1); 2097673-001 (5466892-10); 2097673-001 (5466892-2); 2097673-001 (5466892-3); 2097673-001 (5466892-4); 2097673-001 (5466892-5); 2097673-001 (5466892-6); 2097673-001 (5466892-7); 2097673-001 (5466892-8); 2097673-001 (5466892-9); 2097673-DE (5480919-1); 2097673-NL (5473301-1); 2097673-NL (5473301-2); 2097673-NL (5473301-3); 2097673-NL (5473301-4); 2097673-NL (5473301-5); 2097673-NL (5473301-6); 2097673-FR (5481162-1); 2097673-NL (5481369-1); 2097673-NL (5481369-2); 2097673-NL (5481369-3); 2097673-NL (5481369-4); 2097673-NL (5481369-5); 2097673-NL (5481369-6); 2097673-NL (5481369-7); 2097673-NL (5481369-8); 2097673-NL (5481369-9); 2097673-001 (5462738-1); 2097673-001 (5462738-2); 2097673-FR (5475371-1); 2097673-FR (5475371-2); 2097673-FR (5475371-3); 2097673-FR (5475371-4); 2097673-FR (5475371-5); 2097673-FR (5479900-1); 2097673-DA (5453900-1); 2097673-ET (5461749-1); 2097673-FR (5481462-1); 2097673-FR (5481462-2); 2097673-FR (5481462-3); 2097673-FR (5481462-4); 2097673-FR (5481462-5); 2097673-FR (5481462-6); 2097673-FR (5481462-7); 2097673-FR (5481462-8); 2097673-FR (5481462-9); 2097673-FR (5473743-1); 2097673-FR (5473743-2); 2097673-FR (5473743-3); 2097673-FR (5473743-4); 2097673-FR (5473743-5); 2097673-FR (5473744-1); 2097673-FR (5473744-2); 2097673-FR (5473744-3); 2097673-FR (5473744-4); 2097673-FR (5473744-5); 2097673-FR (5473744-6); 2097673-FR (5458277-1); 2097673-FR (5458277-2); 2097673-001 (5455478-1); 2097673-IT (5476554-1); 2097673-IT (5476554-2); 2097673-IT (5476554-3); 2097673-NL (5450447-1); 2097673-NL (5450447-3); 2097673-NL (5450447-5); 2097673-NL (5450447-6); 2097673-NO (5474251-1); 2097673-001 (5476468-70); 2097673-001 (5476468-71); 2097673-001 (5476468-72); 2097673-001 (5476468-73); 2097673-001 (5476468-74); 2097673-001 (5476468-75); 2097673-001 (5476468-76); 2097673-001 (5476468-77); 2097673-001 (5476468-78); 2097673-001 (5476468-79); 2097673-001 (5476468-80); 2097673-001 (5476468-81); 2097673-001 (5476468-82); 2097673-001 (5476468-83); 2097673-001 (5476468-84); 2097673-001 (5476468-85); 2097673-001 (5476468-86); 2097673-001-US (4673852-1); 2097673-001-US (4673852-2); 2097673-001-US (4673852-3); Additional Serial Numbers - Model Number, UDI Number, (Serial Number): 2097673-001-US, Not Applicable, (ANAM00306); 2097673-001-US, Not Applicable, (ANAP00354); 2097673-001-US, Not Applicable, (APWU00297); 2097673-001-US, Not Applicable, (ANAP00715); 2097673-001-US, Not Applicable, (ANAR00602); 2097673-001-US, Not Applicable, (ANAR00603); 2097673-001-US, Not Applicable, (ANAN00169); 2097673-001-US, Not Applicable, (ANAP00574); 2097673-001-US, Not Applicable, (ANAP00575); 2097673-001-US, Not Applicable, (ANAP00576); 2097673-001-US, Not Applicable, (ANAP00577); 2097673-001-US, Not Applicable, (ANAQ01175); 2097673-001-US, Not Applicable, (ANAP00172); 2097673-001-US, Not Applicable, (ANAP00861); 2097673-001-US, Not Applicable, (ANAP00870); 2097673-001-US, Not Applicable, (ANAP00872); 2097673-001-US, Not Applicable, (ANAS00398 ); 2097673-001-US, Not Applicable, (ANAL00105); 2097673-001-US, Not Applicable, (ANAL00106); 2097673-001-US, Not Applicable, (ANAL00107); 2097673-001-US, Not Applicable, (ANAL00108); 2097673-001-US, Not Applicable, (ANAL00109); 2097673-001-US, Not Applicable, (ANAL00116); 2097673-001-US, Not Applicable, (ANAM00439); 2097673-001-US, Not Applicable, (ANAM00440); 2097673-001-US, Not Applicable, (ANAM00441); 2097673-001-US, Not Applicable, (ANAP01088); 2097673-001-US, Not Applicable, (ANAP01090); 2097673-001-US, Not Applicable, (ANAP01091); 2097673-001, Not Applicable, (ANAN00604); 2097673-001, Not Applicable, (ANAM01165); 2097673-001, Not Applicable, (ANAM01158); 2097673-001, Not Applicable, (ANAM00815); 2097673-001, Not Applicable, (ANAM00816); 2097673-001, Not Applicable, (ANAM00817); 2097673-001, Not Applicable, (ANAM00818); 2097673-001, Not Applicable, (ANAM00819); 2097673-001, Not Applicable, (ANAM00820); 2097673-001, Not Applicable, (ANAM00821); 2097673-001, Not Applicable, (ANAM00822); 2097673-001, Not Applicable, (ANAM00823); 2097673-001, Not Applicable, (ANAM00824); 2097673-DE, Not Applicable, (ANAP01156); 2097673-001, Not Applicable, (ANAR00318); 2097673-001, Not Applicable, (ANAR00319); 2097673-FR, Not Applicable, (ANAR05232); 2097673-FR, Not Applicable, (ANAR05233); 2097673-FR, Not Applicable, (ANAR05234); 2097673-FR, Not Applicable, (ANAR05235); 2097673-FR, Not Applicable, (ANAR05236); 2097673-FR, Not Applicable, (ANAP01419); 1011-9000-000, Not Applicable, (APWT00661); 1011-9000-000, Not Applicable, (APWT00662); 1011-9000-000, Not Applicable, (APWT00664); 1011-9000-000, Not Applicable, (ANAN00664); 1011-9000-000, Not Applicable, (ANAN00666); 1011-9000-000, Not Applicable, (ANAN00667); 1011-9000-000, Not Applicable, (ANAN00669); 1011-9000-000, Not Applicable, (ANAN00670); 1011-9000-000, Not Applicable, (ANAN00671); 1011-9000-000, Not Applicable, (ANAN00890); 1011-9000-000, Not Applicable, (ANAN00891); 1011-9000-000, Not Applicable, (ANAR00396); 1011-9000-000, Not Applicable, (APWT02012); 1011-9000-000, Not Applicable, (APWT02013); 1011-9000-000, Not Applicable, (ANAN01101); 1011-9000-000, Not Applicable, (ANAP00605); 2097673-FR, Not Applicable, (ANAQ00803); 2097673-FR, Not Applicable, (ANAQ00806); 2097673-FR, Not Applicable.(ANAQ00948); 2097673-FR, Not Applicable, (ANAR05145); 2097673-FR, Not Applicable, (ANAR05150); 2097673-FR, Not Applicable, (ANAR05513); 2097673-FR, Not Applicable, (ANAR05517); 2097673-FR, Not Applicable, (ANAS00665); 2097673-FR, Not Applicable, (ANAS00666); 2097673-FR, Not Applicable, (ANAS00668); 2097673-FR, Not Applicable, (ANAS00971); 2097673-NO, Not Applicable, (ANAQ01098);
Other devices under the same FDA submission
2 devices cleared under K110213, K170872 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Aisys CS2Datex-Ohmeda Inc · Gas-machine, anesthesia5 recalls2,048 reports
- Aisys CS2 WITH ET CONTROLDatex-Ohmeda Inc · Gas-machine, anesthesia2 recalls98 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 81182).
- Z-0411-2019 · TerminatedAisys Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRADE K…
- Z-0412-2019 · TerminatedAisys CS2 Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRA…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.