VOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software… recallZ-2064-2019 · Fenwal Inc · initiated Jun 24, 2019

Class I Terminated All recalls posted in August 2019

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Fenwal Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of Volumat MC Agilia Software and versions 1.0 and 1.1 of Vigilant DrugLib

FDA root cause category: Other

What the company is doing

The firm initiated the recall by letter on 06/24/2019. The letter describes the issues and provided supplemental instructions to be implemented until the new software is installed. The firm issued a press release on 06/28/2019.

Recall details

Recall number
Z-2064-2019
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Jun 24, 2019
Posted by the FDA
Aug 12, 2019
Recalling firm
Fenwal Inc
Quantity
9461 units
Product code
FRN
FDA submissions
K121613
Distribution
US Nationwide Distribution

Product description

VOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software versions 1.7 and 1.9a Product Usage: The Volumat MC Agilia is a piece of transportable medical equipment intended to be used in healthcare facilities environment by healthcare trained professionals according to hospital protocols on adults, pediatrics and neonatal human patient to administer via a single channel or mounted on a multiple channels rack accessory: Intermittent or continuous delivery of parenteral fluids (solutions, colloids, parenteral nutrition) and medications (including but limited to diluted drugs, chemotherapy) through clinically accepted IV routes of administration). Transfusion of blood and blood derivatives products. Note that for transfusion procedures, use of a special administration set is required. Volumat MC Agilia can be used with the Dose Error Reduction Software (DERS) Vigilant Drug Lib to define the default and upper hard limits for the Bolus Volume, or dose, by drug. The health care professional can then operate the pump for that drug within the upper hard limits defined in the Drug Library.

Code information (models, lots, serial numbers)

All serial numbers distributed in the US