Why it was recalled

This one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period. The pump may have migrated out of position and possibly caused injury. The positioning and suction alarms were also disabled.

FDA root cause category: Under Investigation by firm

What the company is doing

The affected device was retrieved from the only customer on 9/4/19. An Urgent Medical Device Recall notification letter dated 9/12/19 was sent to that customer as well.

Recall details

Recall number
Z-0345-2020
Status
Terminated
Initiated by the company
Sep 4, 2019
Recalling firm
Abiomed, Inc.
Quantity
1
Product code
OZD
FDA submissions
P140003
Distribution
The product was distributed to GA

Product description

lmpella CP with Smart Assist, Model Number: 0048-0003. Left heart support blood pump, for Use During Cardiogenic Shock and High-Risk PCI

Code information (models, lots, serial numbers)

UDI: (01)00813502011371(10)1430951(17)210331(21)192148A;  Serial Number: 192148A

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

11 devices cleared under P140003 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page