Impella 0042-0040-USRecalls & reported problems

Abiomed, Inc. · Optical, AIC, Impella Connect, Pkgd, US · Class 3

All models of Impella Heart Pump (34)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

6 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Abiomed Impella Connect controller works with a temporary heart support pump. Many of the 1,366 reports describe software, battery and sensing problems. The reports include 45 deaths, 61 injuries and 1,260 malfunctions. Reports rose from 412 in the prior 12 months to 631 in the latest 12 months. 8 recalls name this device, 6 of them open, covering software and cybersecurity issues, a controller not detecting a connected pump, and purge pressure parts.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,305
Deaths reported
41
Injuries reported
62
Recalls naming it
8

Summary

The FDA's MAUDE database contains 1,305 adverse event reports that list this exact device (DI 00813502011401), received between Aug 4, 2023 and Sep 30, 2026 — classified as 41 deaths, 62 injuries, 1,202 malfunctions.

The most frequently reported specific device problem is “Application Program Problem” (277 reports).

8 FDA recall records name this device's identifier, 6 still open.

19 other recalls cover devices cleared under the same FDA submission (P140003, P170011); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 9 reportsDec 2023: 12 reportsJan 2024: 21 reportsFeb 2024: 5 reportsMar 2024: 33 reportsApr 2024: 23 reportsMay 2024: 13 reportsJun 2024: 24 reportsJul 2024: 12 reportsAug 2024: 44 reportsSep 2024: 20 reportsOct 2024: 70 reportsNov 2024: 151 reportsDec 2024: 18 reportsJan 2025: 17 reportsFeb 2025: 14 reportsMar 2025: 13 reportsApr 2025: 26 reportsMay 2025: 36 reportsJun 2025: 33 reportsJul 2025: 32 reportsAug 2025: 22 reportsSep 2025: 37 reportsOct 2025: 13 reportsNov 2025: 19 reportsDec 2025: 22 reportsJan 2026: 64 reportsFeb 2026: 29 reportsMar 2026: 37 reportsApr 2026: 42 reportsMay 2026: 33 reportsJun 2026: 211 reportsJul 2026: 53 reportsAug 2026: 22 reportsSep 2026: 38 reportsOct 2026: 0 reports202420252026211
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 12, 2026Class IOpen, Classified

    The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.

    Z-3150-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. May 14, 2026Class IOpen, Classified

    Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.

    Z-2360-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Oct 1, 2025Class IOpen, Classified

    Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).

    Z-0069-2026 · Root cause: Software design · Recall details · FDA record ↗

  4. Sep 16, 2025Class IOpen, Classified

    Potential for Automated Impella Controller (AIC) purge pressure issues due to purge retainer failures.

    Z-0136-2026 · Root cause: Device Design · Recall details · FDA record ↗

  5. Jun 23, 2025Class IOpen, Classified

    A potential issue with the Automated Impella Controller (AIC) not detecting an Impella pump when it is connected

    Z-2164-2025 · Root cause: Device Design · Recall details · FDA record ↗

  6. Oct 10, 2023Class IITerminated

    Pump not detected as connected to controller due to software issue.

    Z-0441-2024 · Root cause: Software design · Recall details · FDA record ↗

  7. Mar 1, 2021Class IITerminated

    Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.

    Z-1175-2024 · Root cause: Software design · Recall details · FDA record ↗

  8. Feb 19, 2019Class IIOpen, Classified

    Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.

    Z-3251-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Application Program Problem277
  • Electrical /Electronic Property Problem195
  • Battery Problem149
  • Device Sensing Problem129
  • Failure to Read Input Signal121
  • Material Integrity Problem118
  • Computer Operating System Problem47
  • No Pressure45

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,231
  • Hemodynamic instability63
  • Cardiovascular Dysfunction/Insufficiency16
  • Insufficient Information11
  • Arrhythmia2
  • Cardiac Arrest2
  • Tachycardia2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 1,305 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947866 ↗
  • MalfunctionSep 29, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937296 ↗
  • InjurySep 29, 2026Device Sensing Problem

    An Impella CP with SmartAssist (SA) was inserted via the right femoral artery in a 42-year-old male weighing 127 kg and measuring 188 cm in height. The patient presented with acute myocardial infarction complicated by cardiogenic shock. Comorbidities were not reported. The patient's clinical state prior to initiation of support was Society for Cardiovascular Angiography and Interventions (SCAI) Shock Stage D identified when there is evidence of shock that worsens or fails to improve despite escalation therapies. Prior to Impella support, the patient was receiving inotropes, vasopressors, and respiratory support.Device Course & Treatment:During implantation, a 0.035-inch guidewire was unable to make the turn into the aortic arch and a 0.018-inch guidewire was subsequently used for aortic valve assessment. While the guidewire was being advanced, it contacted the Impella motor, resulting in fraying and fracture of the wire. During removal, the guidewire became entangled within the mitral valve chordae and subsequently fractured. Multiple attempts were made to snare the fractured guidewire fragment, which was ultimately successfully retrieved and removed from the patient. The Impella C

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936087 ↗
  • MalfunctionSep 29, 2026Material Integrity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933718 ↗
  • MalfunctionSep 29, 2026Failure to Read Input Signal

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932302 ↗
  • MalfunctionSep 29, 2026Material Integrity Problem

    The complainant reported that during service and repair activities, an Automated Impella Controller (AIC) underwent preventive maintenance and functional testing. During the purge cassette portion of the preventive maintenance procedure, the console failed testing when a crack was identified in the Purge Disk Retainer. The Purge Disk Retainer was replaced. Following replacement, the preventive maintenance procedure was completed without further issues. The console subsequently passed testing and was returned to service.The issue was identified during service and repair activities and not during clinical use. No patient involvement was reported in association with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932095 ↗
  • MalfunctionSep 29, 2026Battery Problem; Unintended Application Program Shut Down

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931952 ↗
  • MalfunctionSep 28, 2026Device Sensing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50925630 ↗
  • MalfunctionSep 28, 2026Application Program Problem

    The complainant reported that, after providing patient care, the pump alarmed ¿controller error¿. The bedside nurse assessed the Impella cord. After applying pressure to Impella plug into console, alarm resolved. No adverse effect on the patient.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923989 ↗
  • MalfunctionSep 28, 2026Application Program Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917594 ↗
  • MalfunctionSep 28, 2026Material Integrity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916108 ↗
  • MalfunctionSep 28, 2026Material Integrity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914757 ↗
  • MalfunctionSep 25, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50899174 ↗
  • MalfunctionSep 25, 2026Electrical /Electronic Property Problem; Material Integrity Problem

    It was reported that, while a performing preventative maintenance procedure, the console displayed a batter failure alarm after power up. The console batteries (0042-3016) were replaced. During that same preventative maintenance procedure that a hairline crack was found in the purge retainer (b)(6). The purge retainer was replaced. The console was returned back to specification and was shipped back to customer. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50898466 ↗
  • MalfunctionSep 25, 2026Priming Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50897750 ↗
  • MalfunctionSep 25, 2026Material Integrity Problem

    It was reported that, while performing a functional check, a crack found in the purge disk retainer. The purge disk retainer (3001195) was replaced and passed all specifications. No patient harm has been reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50897477 ↗
  • MalfunctionSep 25, 2026Material Integrity Problem

    It was reported that, while performing a functional check, a crack was found in the purge retainer ((b)(6)). The purge retainer was replaced and the console performed to specification and was shipped back to the customer. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50897163 ↗
  • MalfunctionSep 24, 2026Material Integrity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889340 ↗
  • MalfunctionSep 24, 2026Application Program Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889055 ↗
  • MalfunctionSep 24, 2026Battery Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889043 ↗
  • MalfunctionSep 24, 2026Electrical /Electronic Property Problem

    It was reported that, while performing preventative maintenance, a failed measurement at 128 mmHg spec <125 was noted. The purge pressure sensor (b)(6) was replaced and the device was returned to specification. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50888973 ↗
  • MalfunctionSep 24, 2026Electrical /Electronic Property Problem; Material Integrity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50888129 ↗
  • MalfunctionSep 24, 2026Electrical /Electronic Property Problem; Material Integrity Problem

    The complainant reported that while performing preventative maintenance a crack purge disk retainer was found and the purge film was damaged. The purge disk retainer and purge disk film were replaced. The retainer was torqued to specification and failed measurement at three mmHg spec 0 mmHg +/- 1. After replacing the purge pressure sensor, all specifications were passed. No patient was involved.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883107 ↗
  • MalfunctionSep 24, 2026Material Integrity Problem

    During service repair, a crack was identified in the purge retainer. Purge retainer was replaced. No patient involvement has been reported.Purge pressure sensor kit was also replaced due to the volume of dextrose.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883011 ↗
  • MalfunctionSep 24, 2026Material Integrity Problem

    The complainant reported that the console failed preventative maintenance (PM) due to a crack in the Purge Disk Retainer. The Purge Disk Retainer was replaced. After replacing the retainer, the PM was completed with no further issues. No patient was involved.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50882915 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140003, P170011. One submission often covers many models, so check the product and lot details in each record.

  • Z-3149-2026 · Aug 12, 2026 · Open, Classified
    Impella Optical Controller with the below product descriptions and corresponding Product Codes. 1. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU. 2. Impella
  • Z-3148-2026 · Aug 12, 2026 · Open, Classified
    Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella
  • Z-2457-2026 · May 18, 2026 · Open, Classified
    Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.
  • Z-2211-2026 · Apr 20, 2026 · Open, Classified
    Automated Impella Controller (AIC); Product Code: 0042-0000-US;
  • Z-1589-2026 · Feb 18, 2026 · Open, Classified
    Purge Cassette, sold within Impella Pump Sets and Individually Packaged. Labeled and distributed in the below configurations and corresponding product codes. 1. Purge Cassette, 5
  • Z-1591-2026 · Feb 16, 2026 · Open, Classified
    Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Contr
  • Z-1469-2026 · Jan 27, 2026 · Open, Classified
    Impella RP with SmartAssist. Product Code: 0046-0035.
  • Z-1470-2026 · Jan 27, 2026 · Open, Classified
    Impella RP Flex with SmartAssist. Product Code: 1000323.
  • Z-1471-2026 · Jan 27, 2026 · Open, Classified
    Impella RP. Product Code: 0046-0011.
  • Z-1030-2026 · Dec 4, 2025 · Open, Classified
    Impella 5.5 with SmartAssist S2, Impella 5.5 with SmartAssist S2 Set, AU.
  • Z-0829-2025 · Dec 12, 2024 · Open, Classified
    Impella RP with SmartAssist; Product Number: 0046-0035;
  • Z-0830-2025 · Dec 12, 2024 · Open, Classified
    Impella RP Flex with SmartAssist; Product Number: 1000323;
  • Z-2319-2024 · May 31, 2024 · Open, Classified
    Impella CP with SmartAssist, Product No.: 1000080 (REF 0048-0003). left heart support blood pump
  • Z-0980-2024 · Dec 27, 2023 · Open, Classified
    Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Code 005042 - Impella 2.5 (2) Product Code 005062 - Impella 5
  • Z-0636-2024 · Dec 6, 2023 · Open, Classified
    Impella Connect, software portal for online device management that allows clinicians and Impella support staff to remotely view Impella data.

…and 4 more.

Other Temporary Non-Roller Type Right Heart Support Blood Pump devices

All Temporary Non-Roller Type Right Heart Support Blood Pump devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page