Carestation 620/650/650c A1 recallZ-0813-2020 · GE Healthcare, LLC · initiated Nov 25, 2019

Class I Terminated All recalls posted in January 2020

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GE Healthcare has become aware that there is a potential for a loose cable connection inside specific manufactured anesthesia devices. This would cause a loss of mechanical ventilation and the system will provide high priority audio and visual alarms. Loss of mechanical ventilation could lead to hypoxia if the clinician does not intervene.

FDA root cause category: Process control

What the company is doing

The firm initiated their field notification by letter on November 25, 2019. The letter states the reason for recall and provides the following safety instructions: You can continue to use the anesthesia system. - If you observe the message Ventilate Manually! , change from mechanical to manual ventilation. At any time, the clinician may use a self-inflating bag to ventilate the patient and/or switch to another anesthesia device. Contact your GE Healthcare representative for repair of the device. - Perform the planned maintenance (PM) every 12-months at a minimum per the User s Reference Manual which includes inspection of the cable connection. Note: This inspection step is included in the annual PM described in the Technical Reference Manual. Performing this step in the PM would confirm the integrity of the cable connection. Contact Information: If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. The firm will be correcting the units. The firm issued a Press Release on January 13, 2020.

Recall details

Recall number
Z-0813-2020
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Nov 25, 2019
Posted by the FDA
Jan 23, 2020
Recalling firm
GE Healthcare, LLC
Quantity
3319 units
Product code
BSZ
FDA submissions
K151570
Distribution
Worldwide Distribution: US (Nationwide) and countries of: Albania, Algeria, Argentina, Australia, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Croatia and Hrvatska, Czech Republic, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Jamaica, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Maldives, Mexico, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, Oman, Paraguay, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Viet Nam.

Product description

Carestation 620/650/650c A1 Product Usage: anesthesia systems are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (pediatric and adult.

Code information (models, lots, serial numbers)

Show the full code information (180k characters)

CARESTATION 620 A1, Model Number 1012-9620-000 Mfg. Lot or Serial # UDI # SM619130003MA 010084068210398521SM619130003MA11190326 SM619120026MA 010084068210398521SM619120026MA11190321 SM619030025MA 010084068210398521SM619030025MA11190117 SM619020017MA 010084068210398521SM619020017MA11190108 SM619020018MA 010084068210398521SM619020018MA11190108 SM619180019MA 010084068210398521SM619180019MA11190502 SM619020063MA 010084068210398521SM619020063MA11190111 SM619020064MA 010084068210398521SM619020064MA11190111 SM619020065MA 010084068210398521SM619020065MA11190111 SM619110001MA 010084068210398521SM619110001MA11190311 SM618380015MA 010084068210398521SM618380015MA11180918 SM619060074MA 010084068210398521SM619060074MA11190208 SM618460073MA 010084068210398521SM618460073MA11181116 SM619170031MA 010084068210398521SM619170031MA11190422 SM619170032MA 010084068210398521SM619170032MA11190422 SM619170033MA 010084068210398521SM619170033MA11190422 SM618440026MA 010084068210398521SM618440026MA11181101 SM619020021MA 010084068210398521SM619020021MA11190110 SM619020022MA 010084068210398521SM619020022MA11190110 SM619140004MA 010084068210398521SM619140004MA11190402 SM618440027MA 010084068210398521SM618440027MA11181101 SM618440028MA 010084068210398521SM618440028MA11181101 SM619020014MA 010084068210398521SM619020014MA11190108 SM619020015MA 010084068210398521SM619020015MA11190108 SM619020016MA 010084068210398521SM619020016MA11190108 SM618500025MA 010084068210398521SM618500025MA11181214 SM618500026MA 010084068210398521SM618500026MA11181214 SM618500027MA 010084068210398521SM618500027MA11181214 SM619070010MA 010084068210398521SM619070010MA11190211 SM618470011MA 010084068210398521SM618470011MA11181120 SM618470012MA 010084068210398521SM618470012MA11181120 SM618400007MA 010084068210398521SM618400007MA11181004 SM618400008MA 010084068210398521SM618400008MA11181004 SM619090008MA 010084068210398521SM619090008MA11190225 SM619090009MA 010084068210398521SM619090009MA11190225 SM619230002MA 010084068210398521SM619230002MA11190603 SM618380016MA 010084068210398521SM618380016MA11180918 SM618430002MA 010084068210398521SM618430002MA11181022 SM618380041MA 010084068210398521SM618380041MA11180921 SM618380042MA 010084068210398521SM618380042MA11180921 SM618500024MA 010084068210398521SM618500024MA11181214 SM619140031MA 010084068210398521SM619140031MA11190405 SM619140032MA 010084068210398521SM619140032MA11190405 SM619140033MA 010084068210398521SM619140033MA11190405 SM619140034MA 010084068210398521SM619140034MA11190405 SM619140035MA 010084068210398521SM619140035MA11190405 SM618440033MA 010084068210398521SM618440033MA11181101 SM618490022MA 010084068210398521SM618490022MA11181206 SM618490023MA 010084068210398521SM618490023MA11181206 SM619030026MA 010084068210398521SM619030026MA11190117 SM619180005MA 010084068210398521SM619180005MA11190429 SM619180006MA 010084068210398521SM619180006MA11190429 SM619180007MA 010084068210398521SM619180007MA11190429 SM619180008MA 010084068210398521SM619180008MA11190429 SM618500006MA 010084068210398521SM618500006MA11181212 SM618500007MA 010084068210398521SM618500007MA11181212 SM618500008MA 010084068210398521SM618500008MA11181212 SM618500009MA 010084068210398521SM618500009MA11181212 SM618500010MA 010084068210398521SM618500010MA11181212 SM619070011MA 010084068210398521SM619070011MA11190211 SM619070012MA 010084068210398521SM619070012MA11190211 SM619070013MA 010084068210398521SM619070013MA11190211 SM619070014MA 010084068210398521SM619070014MA11190211 SM618430015MA 010084068210398521SM618430015MA11181025 SM618390017MA 010084068210398521SM618390017MA11180926 SM618390018MA 010084068210398521SM618390018MA11180926 SM618390019MA 010084068210398521SM618390019MA11180926 SM618390020MA 010084068210398521SM618390020MA11180926 SM619130009MA 010084068210398521SM619130009MA11190329 SM619130010MA 010084068210398521SM619130010MA11190329 SM619130011MA 010084068210398521SM619130011MA11190329 SM619130012MA 010084068210398521SM619130012MA11190329 SM619100010MA 010084068210398521SM619100010MA11190305 SM619190036MA 010084068210398521SM619190036MA11190508 SM618440029MA 010084068210398521SM618440029MA11181101 SM618440030MA 010084068210398521SM618440030MA11181101 SM618440032MA 010084068210398521SM618440032MA11181101 SM619050021MA 010084068210398521SM619050021MA11190201 SM619060029MA 010084068210398521SM619060029MA11190206 SM618500012MA 010084068210398521SM618500012MA11181213 SM618500013MA 010084068210398521SM618500013MA11181213 SM619040003MA 010084068210398521SM619040003MA11190122 SM619040004MA 010084068210398521SM619040004MA11190122 SM619060068MA 010084068210398521SM619060068MA11190206 SM619080001MA 010084068210398521SM619080001MA11190218 SM619050004MA 010084068210398521SM619050004MA11190129 SM619050005MA 010084068210398521SM619050005MA11190129 SM619050006MA 010084068210398521SM619050006MA11190129 SM619050007MA 010084068210398521SM619050007MA11190129 SM619050008MA 010084068210398521SM619050008MA11190129 SM619050009MA 010084068210398521SM619050009MA11190129 SM619050010MA 010084068210398521SM619050010MA11190129 SM618440031MA 010084068210398521SM618440031MA11181101 SM619160017MA 010084068210398521SM619160017MA11190418 SM619170083MA 010084068210398521SM619170083MA11190426 SM619170084MA 010084068210398521SM619170084MA11190426 SM619170085MA 010084068210398521SM619170085MA11190426 SM619140024MA 010084068210398521SM619140024MA11190403 SM619140025MA 010084068210398521SM619140025MA11190403 SM619210014MA 010084068210398521SM619210014MA11190522 SM619230012MA 010084068210398521SM619230012MA11190605 SM619060075MA 010084068210398521SM619060075MA11190208 SM619060076MA 010084068210398521SM619060076MA11190208 SM619060077MA 010084068210398521SM619060077MA11190208 SM619060078MA 010084068210398521SM619060078MA11190208 SM618440025MA 010084068210398521SM618440025MA11181101 SM619060073MA 010084068210398521SM619060073MA11190208 SM619140005MA 010084068210398521SM619140005MA11190402 SM619050012MA 010084068210398521SM619050012MA11190131 SM619050013MA 010084068210398521SM619050013MA11190131 SM619050014MA 010084068210398521SM619050014MA11190131 SM619050015MA 010084068210398521SM619050015MA11190131 SM619050016MA 010084068210398521SM619050016MA11190131 SM619270005MA 010084068210398521SM619270005MA11190702 SM619180011MA 010084068210398521SM619180011MA11190429 SM619180012MA 010084068210398521SM619180012MA11190429 SM619220059MA 010084068210398521SM619220059MA11190530 SM619030027MA 010084068210398521SM619030027MA11190117 SM618380043MA 010084068210398521SM618380043MA11180921 SM619030008MA 010084068210398521SM619030008MA11190114 SM619030009MA 010084068210398521SM619030009MA11190114 SM618430023MA 010084068210398521SM618430023MA11181026 SM618440024MA 010084068210398521SM618440024MA11181101 SM619180017MA 010084068210398521SM619180017MA11190501 SM619180018MA 010084068210398521SM619180018MA11190501 SM619030028MA 010084068210398521SM619030028MA11190117 SM619030024MA 010084068210398521SM619030024MA11190117 SM619180009MA 010084068210398521SM619180009MA11190429 SM619180010MA 010084068210398521SM619180010MA11190429 SM618410023WA 010084068210398521SM618410023WA11181011 SM618410024WA 010084068210398521SM618410024WA11181011 SM618410025WA 010084068210398521SM618410025WA11181011 SM618410026WA 010084068210398521SM618410026WA11181011 SM618410027WA 010084068210398521SM618410027WA11181011 SM618460043WA 010084068210398521SM618460043WA11181114 SM618420013WA 010084068210398521SM618420013WA11181016 SM618420016WA 010084068210398521SM618420016WA11181016 SM618420017WA 010084068210398521SM618420017WA11181016 SM619070023WA 010084068210398521SM619070023WA11190212 SM618420007WA 010084068210398521SM618420007WA11181016 SM618420010WA 010084068210398521SM618420010WA11181016 SM618420012WA 010084068210398521SM618420012WA11181016 SM618420015WA 010084068210398521SM618420015WA11181016 SM618420018WA 010084068210398521SM618420018WA11181016 SM618420008WA 010084068210398521SM618420008WA11181016 SM618420019WA 010084068210398521SM618420019WA11181016 SM618460046WA 010084068210398521SM618460046WA11181114 SM619070024WA 010084068210398521SM619070024WA11190212 SM619150034WA 010084068210398521SM619150034WA11190411 SM618460041WA 010084068210398521SM618460041WA11181114 SM618460042WA 010084068210398521SM618460042WA11181114 SM618460044WA 010084068210398521SM618460044WA11181114 SM619070021WA 010084068210398521SM619070021WA11190212 SM619070026WA 010084068210398521SM619070026WA11190212 SM618420006WA 010084068210398521SM618420006WA11181016 SM618420014WA 010084068210398521SM618420014WA11181016 SM618460045WA 010084068210398521SM618460045WA11181114 SM618460047WA 010084068210398521SM618460047WA11181114 SM619070022WA 010084068210398521SM619070022WA11190212 SM619070033WA 010084068210398521SM619070033WA11190214 SM619070019WA 010084068210398521SM619070019WA11190212 SM619070025WA 010084068210398521SM619070025WA11190212 SM619070027WA 010084068210398521SM619070027WA11190212 SM619070034WA 010084068210398521SM619070034WA11190214 SM618420004WA 010084068210398521SM618420004WA11181015 SM618420009WA 010084068210398521SM618420009WA11181016 SM618420011WA 010084068210398521SM618420011WA11181016 SM618460040WA 010084068210398521SM618460040WA11181114 SM618460054MA 010084068210398521SM618460054MA11181114 SM618460055MA 010084068210398521SM618460055MA11181114 SM618460056MA 010084068210398521SM618460056MA11181114 SM619080013MA 010084068210398521SM619080013MA11190220 SM619080014MA 010084068210398521SM619080014MA11190220 SM619080015MA 010084068210398521SM619080015MA11190220 SM619080016MA 010084068210398521SM619080016MA11190220 SM619080017MA 010084068210398521SM619080017MA11190220 SM619080023MA 010084068210398521SM619080023MA11190221 SM619080024MA 010084068210398521SM619080024MA11190221 SM619250014MA 010084068210398521SM619250014MA11190618 SM619250015MA 010084068210398521SM619250015MA11190618 SM619050001WA 010084068210398521SM619050001WA11190128 SM618450018WA 010084068210398521SM618450018WA11181107 SM619120005WA 010084068210398521SM619120005WA11190318 SM619120006WA 010084068210398521SM619120006WA11190318 SM619120007WA 010084068210398521SM619120007WA11190318 SM619020010WA 010084068210398521SM619020010WA11190108 SM619020012WA 010084068210398521SM619020012WA11190108 SM619020009WA 010084068210398521SM619020009WA11190108 SM619020011WA 010084068210398521SM619020011WA11190108 SM618440001WA 010084068210398521SM618440001WA11181029 SM618440002WA 010084068210398521SM618440002WA11181029 SM618440003WA 010084068210398521SM618440003WA11181029 SM619050020WA 010084068210398521SM619050020WA11190201 SM618460069WA 010084068210398521SM618460069WA11181116 SM619060080MA 010084068210398521SM619060080MA11190208 SM619060081MA 010084068210398521SM619060081MA11190208 SM619270012MA 010084068210398521SM619270012MA11190703 SM619270018MA 010084068210398521SM619270018MA11190703 SM619270033MA 010084068210398521SM619270033MA11190703 SM619270034MA 010084068210398521SM619270034MA11190703 SM619270035MA 010084068210398521SM619270035MA11190703 SM619270036MA 010084068210398521SM619270036MA11190703 SM619270037MA 010084068210398521SM619270037MA11190703 SM619270038MA 010084068210398521SM619270038MA11190703 SM619270039MA 010084068210398521SM619270039MA11190703 SM619270040MA 010084068210398521SM619270040MA11190703 SM619270041MA 010084068210398521SM619270041MA11190703 SM619270043MA 010084068210398521SM619270043MA11190703 SM619270050MA 010084068210398521SM619270050MA11190703 SM618360071WA 010084068210398521SM618360071WA11180908 SM619270044MA 010084068210398521SM619270044MA11190703 SM619270045MA 010084068210398521SM619270045MA11190703 SM618360068WA 010084068210398521SM618360068WA11180908 SM618360064WA 010084068210398521SM618360064WA11180908 SM618360061WA 010084068210398521SM618360061WA11180908 SM618360070WA 010084068210398521SM618360070WA11180908 SM619190047WA 010084068210398521SM619190047WA11190510 SM619190048WA 010084068210398521SM619190048WA11190510 SM619200001WA 010084068210398521SM619200001WA11190513 SM619200002WA 010084068210398521SM619200002WA11190513 SM619200003WA 010084068210398521SM619200003WA11190513 SM619200011WA 010084068210398521SM619200011WA11190513 SM619200012WA 010084068210398521SM619200012WA11190514 SM619270013MA 010084068210398521SM619270013MA11190703 SM619270014MA 010084068210398521SM619270014MA11190703 SM619270015MA 010084068210398521SM619270015MA11190703 SM619270016MA 010084068210398521SM619270016MA11190703 SM619270017MA 010084068210398521SM619270017MA11190703 SM619270019MA 010084068210398521SM619270019MA11190703 SM619270020MA 010084068210398521SM619270020MA11190703 SM619270021MA 010084068210398521SM619270021MA11190703 SM619270022MA 010084068210398521SM619270022MA11190703 SM619270023MA 010084068210398521SM619270023MA11190703 SM619270024MA 010084068210398521SM619270024MA11190703 SM619270025MA 010084068210398521SM619270025MA11190703 SM619270026MA 010084068210398521SM619270026MA11190703 SM619270027MA 010084068210398521SM619270027MA11190703 SM619270028MA 010084068210398521SM619270028MA11190703 SM619270029MA 010084068210398521SM619270029MA11190703 SM619270030MA 010084068210398521SM619270030MA11190703 SM619270031MA 010084068210398521SM619270031MA11190703 SM619270032MA 010084068210398521SM619270032MA11190703 SM619270042MA 010084068210398521SM619270042MA11190703 SM619270046MA 010084068210398521SM619270046MA11190703 SM619270047MA 010084068210398521SM619270047MA11190703 SM619270048MA 010084068210398521SM619270048MA11190703 SM619270049MA 010084068210398521SM619270049MA11190703 SM619270051MA 010084068210398521SM619270051MA11190703 SM618360033WA 010084068210398521SM618360033WA11180907 SM619190046WA 010084068210398521SM619190046WA11190510 SM618360032WA 010084068210398521SM618360032WA11180907 SM618360037WA 010084068210398521SM618360037WA11180907 SM618360066WA 010084068210398521SM618360066WA11180908 SM619140001WA 010084068210398521SM619140001WA11190402 SM619020025WA 010084068210398521SM619020025WA11190111 SM619020026WA 010084068210398521SM619020026WA11190111 SM619020027WA 010084068210398521SM619020027WA11190111 SM619140018WA 010084068210398521SM619140018WA11190403 SM619140019WA 010084068210398521SM619140019WA11190403 SM619140020WA 010084068210398521SM619140020WA11190403 SM618360038WA 010084068210398521SM618360038WA11180907 SM618360063WA 010084068210398521SM618360063WA11180908 SM618360062WA 010084068210398521SM618360062WA11180908 SM618360065WA 010084068210398521SM618360065WA11180908 SM619200004WA 010084068210398521SM619200004WA11190513 SM619200005WA 010084068210398521SM619200005WA11190513 SM619200006WA 010084068210398521SM619200006WA11190513 SM619200007WA 010084068210398521SM619200007WA11190513 SM619200008WA 010084068210398521SM619200008WA11190513 SM619200009WA 010084068210398521SM619200009WA11190513 SM619200010WA 010084068210398521SM619200010WA11190513 SM618360031WA 010084068210398521SM618360031WA11180907 SM618360069WA 010084068210398521SM618360069WA11180908 SM618360036WA 010084068210398521SM618360036WA11180907 SM619020024WA 010084068210398521SM619020024WA11190111 SM619020028WA 010084068210398521SM619020028WA11190111 SM619160005WA 010084068210398521SM619160005WA11190418 SM619160006WA 010084068210398521SM619160006WA11190418 SM619160007WA 010084068210398521SM619160007WA11190418 SM619160008WA 010084068210398521SM619160008WA11190418 SM619160009WA 010084068210398521SM619160009WA11190418 SM619160010WA 010084068210398521SM619160010WA11190418 SM619160011WA 010084068210398521SM619160011WA11190418 SM619160019WA 010084068210398521SM619160019WA11190419 SM619160020WA 010084068210398521SM619160020WA11190419 SM619160021WA 010084068210398521SM619160021WA11190419 SM618410028WA 010084068210398521SM618410028WA11181011 SM618410030WA 010084068210398521SM618410030WA11181011 SM618410046WA 010084068210398521SM618410046WA11181012 SM618360034WA 010084068210398521SM618360034WA11180907 SM618410029WA 010084068210398521SM618410029WA11181011 SM618410031WA 010084068210398521SM618410031WA11181011 SM619100001WA 010084068210398521SM619100001WA11190304 SM619100002WA 010084068210398521SM619100002WA11190304 SM619100003WA 010084068210398521SM619100003WA11190304 SM618360067WA 010084068210398521SM618360067WA11180908 SM618360072WA 010084068210398521SM618360072WA11180908 SM618360035WA 010084068210398521SM618360035WA11180907 SM618460033MA 010084068210398521SM618460033MA11181112 SM618460058MA 010084068210398521SM618460058MA11181114 SM618460053MA 010084068210398521SM618460053MA11181114 SM618450035MA 010084068210398521SM618450035MA11181108 SM618370064MA 010084068210398521SM618370064MA11180913 SM618360039MA 010084068210398521SM618360039MA11180907 SM618360040MA 010084068210398521SM618360040MA11180907 SM618360041MA 010084068210398521SM618360041MA11180907 SM618360042MA 010084068210398521SM618360042MA11180907 SM618360043MA 010084068210398521SM618360043MA11180907 SM618360044MA 010084068210398521SM618360044MA11180907 SM618360045MA 010084068210398521SM618360045MA11180907 SM618360046MA 010084068210398521SM618360046MA11180907 SM618360047MA 010084068210398521SM618360047MA11180907 SM618360048MA 010084068210398521SM618360048MA11180907 SM618360049MA 010084068210398521SM618360049MA11180907 SM618360050MA 010084068210398521SM618360050MA11180907 SM618360051MA 010084068210398521SM618360051MA11180907 SM618360052MA 010084068210398521SM618360052MA11180907 SM618360053MA 010084068210398521SM618360053MA11180907 SM618360054MA 010084068210398521SM618360054MA11180907 SM618360055MA 010084068210398521SM618360055MA11180907 SM618360056MA 010084068210398521SM618360056MA11180907 SM618360057MA 010084068210398521SM618360057MA11180907 SM618360058MA 010084068210398521SM618360058MA11180907 SM619150035MA 010084068210398521SM619150035MA11190411 SM619150036MA 010084068210398521SM619150036MA11190411 SM619280053MA 010084068210398521SM619280053MA11190711 SM619280054MA 010084068210398521SM619280054MA11190711 SM619090027MA 010084068210398521SM619090027MA11190228 SM619090028MA 010084068210398521SM619090028MA11190228 SM619090029MA 010084068210398521SM619090029MA11190228 SM619090030MA 010084068210398521SM619090030MA11190228 SM618490005MA 010084068210398521SM618490005MA11181203 SM618490006MA 010084068210398521SM618490006MA11181203 SM618490007MA 010084068210398521SM618490007MA11181203 SM618490008MA 010084068210398521SM618490008MA11181203 SM618490009MA 010084068210398521SM618490009MA11181203 SM619020013MA 010084068210398521SM619020013MA11190108 SM618370040MA 010084068210398521SM618370040MA11180912 SM618370042MA 010084068210398521SM618370042MA11180912 SM618370043MA 010084068210398521SM618370043MA11180912 SM618380031MA 010084068210398521SM618380031MA11180920 SM618460052MA 010084068210398521SM618460052MA11181114 SM618470010MA 010084068210398521SM618470010MA11181119 SM618500011MA 010084068210398521SM618500011MA11181213 SM619030029MA 010084068210398521SM619030029MA11190117 SM619060039MA 010084068210398521SM619060039MA11190206 SM619080025MA 010084068210398521SM619080025MA11190222 SM619080026MA 010084068210398521SM619080026MA11190222 SM619080027MA 010084068210398521SM619080027MA11190222 SM619080028MA 010084068210398521SM619080028MA11190222 SM619080029MA 010084068210398521SM619080029MA11190222 SM619080030MA 010084068210398521SM619080030MA11190222 SM619080031MA 010084068210398521SM619080031MA11190222 SM619080032MA 010084068210398521SM619080032MA11190222 SM619080033MA 010084068210398521SM619080033MA11190222 SM619080034MA 010084068210398521SM619080034MA11190222 SM619080035MA 010084068210398521SM619080035MA11190222 SM619080036MA 010084068210398521SM619080036MA11190222 SM619080037MA 010084068210398521SM619080037MA11190222 SM619080038MA 010084068210398521SM619080038MA11190222 SM619080039MA 010084068210398521SM619080039MA11190222 SM619080040MA 010084068210398521SM619080040MA11190222 SM619190039MA 010084068210398521SM619190039MA11190509 SM618450026MA 010084068210398521SM618450026MA11181107 SM618450027MA 010084068210398521SM618450027MA11181107 SM618450028MA 010084068210398521SM618450028MA11181107 SM618450029MA 010084068210398521SM618450029MA11181107 SM618450030MA 010084068210398521SM618450030MA11181107 SM618460007MA 010084068210398521SM618460007MA11181112 SM618460006MA 010084068210398521SM618460006MA11181112 SM619070005MA 010084068210398521SM619070005MA11190211 SM619070006MA 010084068210398521SM619070006MA11190211 SM619070007MA 010084068210398521SM619070007MA11190211 SM619070008MA 010084068210398521SM619070008MA11190211 SM619070009MA 010084068210398521SM619070009MA11190211 SM619230018MA 010084068210398521SM619230018MA11190606 SM618390003MA 010084068210398521SM618390003MA11180925 SM618420036MA 010084068210398521SM618420036MA11181019 SM618440022MA 010084068210398521SM618440022MA11181101 SM618440023MA 010084068210398521SM618440023MA11181101 SM618440009MA 010084068210398521SM618440009MA11181029 SM618480053MA 010084068210398521SM618480053MA11181128 SM618480054MA 010084068210398521SM618480054MA11181128 SM618480055MA 010084068210398521SM618480055MA11181128 SM619010105MA 010084068210398521SM619010105MA11190103 SM619070054MA 010084068210398521SM619070054MA11190215 SM619250020MA 010084068210398521SM619250020MA11190618 SM619180004MA 010084068210398521SM619180004MA11190429 SM618390004MA 010084068210398521SM618390004MA11180925 SM618390005MA 010084068210398521SM618390005MA11180925 SM618450036MA 010084068210398521SM618450036MA11181108 SM618510002MA 010084068210398521SM618510002MA11181218 SM618460004MA 010084068210398521SM618460004MA11181112 SM618460005MA 010084068210398521SM618460005MA11181112 SM618370025MA 010084068210398521SM618370025MA11180912 SM618370026MA 010084068210398521SM618370026MA11180912 SM618370027MA 010084068210398521SM618370027MA11180912 SM618370028MA 010084068210398521SM618370028MA11180912 SM618370029MA 010084068210398521SM618370029MA11180912 SM618370030MA 010084068210398521SM618370030MA11180912 SM618370031MA 010084068210398521SM618370031MA11180912 SM618370032MA 010084068210398521SM618370032MA11180912 SM618370033MA 010084068210398521SM618370033MA11180912 SM618370034MA 010084068210398521SM618370034MA11180912 SM618370036MA 010084068210398521SM618370036MA11180912 SM618370037MA 010084068210398521SM618370037MA11180912 SM618370038MA 010084068210398521SM618370038MA11180912 SM618370039MA 010084068210398521SM618370039MA11180912 SM618450039MA 010084068210398521SM618450039MA11181108 SM619060037MA 010084068210398521SM619060037MA11190206 SM619190034MA 010084068210398521SM619190034MA11190508 SM619190035MA 010084068210398521SM619190035MA11190508 SM618450043MA 010084068210398521SM618450043MA11181109 SM618450044MA 010084068210398521SM618450044MA11181109 SM619060067MA 010084068210398521SM619060067MA11190206 SM619200025MA 010084068210398521SM619200025MA11190515 SM618450037MA 010084068210398521SM618450037MA11181108 SM618450038MA 010084068210398521SM618450038MA11181108 SM618450040MA 010084068210398521SM618450040MA11181108 SM618370035MA 010084068210398521SM618370035MA11180912 SM618480008MA 010084068210398521SM618480008MA11181126 SM618480009MA 010084068210398521SM618480009MA11181126 SM618480010MA 010084068210398521SM618480010MA11181126 SM618480011MA 010084068210398521SM618480011MA11181126 SM618480012MA 010084068210398521SM618480012MA11181126 SM618480013MA 010084068210398521SM618480013MA11181126 SM618480014MA 010084068210398521SM618480014MA11181126 SM618480015MA 010084068210398521SM618480015MA11181126 SM618480016MA 010084068210398521SM618480016MA11181126 SM618480017MA 010084068210398521SM618480017MA11181126 SM618480018MA 010084068210398521SM618480018MA11181126 SM618480019MA 010084068210398521SM618480019MA11181126 SM618480020MA 010084068210398521SM618480020MA11181126 SM618480021MA 010084068210398521SM618480021MA11181126 SM618480022MA 010084068210398521SM618480022MA11181126 SM618480023MA 010084068210398521SM618480023MA11181126 SM619070055MA 010084068210398521SM619070055MA11190215 SM619070056MA 010084068210398521SM619070056MA11190215 SM619070057MA 010084068210398521SM619070057MA11190215 SM619070058MA 010084068210398521SM619070058MA11190215 SM619070045MA 010084068210398521SM619070045MA11190215 SM619070041MA 010084068210398521SM619070041MA11190215 SM619070042MA 010084068210398521SM619070042MA11190215 SM619070044MA 010084068210398521SM619070044MA11190215 SM619070046MA 010084068210398521SM619070046MA11190215 SM619230019MA 010084068210398521SM619230019MA11190606 SM619060027MA 010084068210398521SM619060027MA11190206 SM619100020MA 010084068210398521SM619100020MA11190308 SM619180002MA 010084068210398521SM619180002MA11190429 SM619180003MA 010084068210398521SM619180003MA11190429 SM618360024MA 010084068210398521SM618360024MA11180906 SM619070032MA 010084068210398521SM619070032MA11190213 SM618340083MA 010084068210398521SM618340083MA11180824 SM618480004MA 010084068210398521SM618480004MA11181126 SM619200035MA 010084068210398521SM619200035MA11190515 SM619060023MA 010084068210398521SM619060023MA11190206 SM619060024MA 010084068210398521SM619060024MA11190206 SM619060025MA 010084068210398521SM619060025MA11190206 SM619150010MA 010084068210398521SM619150010MA11190409 SM619150039MA 010084068210398521SM619150039MA11190412 SM618480052MA 010084068210398521SM618480052MA11181128 SM618480026MA 010084068210398521SM618480026MA11181126 SM619060022MA 010084068210398521SM619060022MA11190206 SM619190022MA 010084068210398521SM619190022MA11190507 SM619190025MA 010084068210398521SM619190025MA11190507 SM619190027MA 010084068210398521SM619190027MA11190507 SM619010106MA 010084068210398521SM619010106MA11190103 SM619010107MA 010084068210398521SM619010107MA11190103 SM619010108MA 010084068210398521SM619010108MA11190103 SM619210011MA 010084068210398521SM619210011MA11190521 SM619210012MA 010084068210398521SM619210012MA11190521 SM618370066MA 010084068210398521SM618370066MA11180913 SM618370067MA 010084068210398521SM618370067MA11180913 SM618370048MA 010084068210398521SM618370048MA11180912 SM619060016MA 010084068210398521SM619060016MA11190206 SM619060026MA 010084068210398521SM619060026MA11190206 SM619060030MA 010084068210398521SM619060030MA11190206 SM619060031MA 010084068210398521SM619060031MA11190206 SM619210009MA 010084068210398521SM619210009MA11190521 SM619210010MA 010084068210398521SM619210010MA11190521 SM618460011MA 010084068210398521SM618460011MA11181112 SM618460012MA 010084068210398521SM618460012MA11181112 SM618460013MA 010084068210398521SM618460013MA11181112 SM618460014MA 010084068210398521SM618460014MA11181112 SM618370044MA 010084068210398521SM618370044MA11180912 SM618370045MA 010084068210398521SM618370045MA11180912 SM618370046MA 010084068210398521SM618370046MA11180912 SM618370047MA 010084068210398521SM618370047MA11180912 SM619130004MA 010084068210398521SM619130004MA11190327 SM619210006MA 010084068210398521SM619210006MA11190521 SM619210007MA 010084068210398521SM619210007MA11190521 SM619210008MA 010084068210398521SM619210008MA11190521 SM618340074MA 010084068210398521SM618340074MA11180823 SM618340075MA 010084068210398521SM618340075MA11180823 SM618340076MA 010084068210398521SM618340076MA11180823 SM618340077MA 010084068210398521SM618340077MA11180823 SM618480006MA 010084068210398521SM618480006MA11181126 SM618480007MA 010084068210398521SM618480007MA11181126 SM618480039MA 010084068210398521SM618480039MA11181128 SM618480040MA 010084068210398521SM618480040MA11181128 SM619060028MA 010084068210398521SM619060028MA11190206 SM619200036MA 010084068210398521SM619200036MA11190515 SM619200037MA 010084068210398521SM619200037MA11190515 SM618410044MA 010084068210398521SM618410044MA11181011 SM619180014MA 010084068210398521SM619180014MA11190501 SM619180015MA 010084068210398521SM619180015MA11190501 SM619210013MA 010084068210398521SM619210013MA11190522 SM618510001MA 010084068210398521SM618510001MA11181218 SM618370049MA 010084068210398521SM618370049MA11180912 SM618370050MA 010084068210398521SM618370050MA11180912 SM618480049MA 010084068210398521SM618480049MA11181128 SM618480050MA 010084068210398521SM618480050MA11181128 SM618480051MA 010084068210398521SM618480051MA11181128 SM619250016MA 010084068210398521SM619250016MA11190618 SM619250017MA 010084068210398521SM619250017MA11190618 SM619160012MA 010084068210398521SM619160012MA11190418 SM618480056MA 010084068210398521SM618480056MA11181128 SM618480057MA 010084068210398521SM618480057MA11181128 SM619060064MA 010084068210398521SM619060064MA11190206 SM619060065MA 010084068210398521SM619060065MA11190206 SM619090031MA 010084068210398521SM619090031MA11190301 SM619090032MA 010084068210398521SM619090032MA11190301 SM619090033MA 010084068210398521SM619090033MA11190301 SM619090034MA 010084068210398521SM619090034MA11190301 SM619090035MA 010084068210398521SM619090035MA11190301 SM619090036MA 010084068210398521SM619090036MA11190301 SM619090037MA 010084068210398521SM619090037MA11190301 SM619090038MA 010084068210398521SM619090038MA11190301 SM619220028MA 010084068210398521SM619220028MA11190527 SM619220029MA 010084068210398521SM619220029MA11190527 SM619220030MA 010084068210398521SM619220030MA11190527 SM619220031MA 010084068210398521SM619220031MA11190527 SM619220032MA 010084068210398521SM619220032MA11190527 SM619220033MA 010084068210398521SM619220033MA11190527 SM619220034MA 010084068210398521SM619220034MA11190527 SM619220035MA 010084068210398521SM619220035MA11190527 SM618380044MA 010084068210398521SM618380044MA11180921 SM619060062MA 010084068210398521SM619060062MA11190206 SM619060063MA 010084068210398521SM619060063MA11190206 SM619060066MA 010084068210398521SM619060066MA11190206 SM619180001MA 010084068210398521SM619180001MA11190429 SM618450015MA 010084068210398521SM618450015MA11181106 SM618460015MA 010084068210398521SM618460015MA11181112 SM618460016MA 010084068210398521SM618460016MA11181112 SM619070043MA 010084068210398521SM619070043MA11190215 SM619070047MA 010084068210398521SM619070047MA11190215 SM619070048MA 010084068210398521SM619070048MA11190215 SM619070049MA 010084068210398521SM619070049MA11190215 SM619070050MA 010084068210398521SM619070050MA11190215 SM619070051MA 010084068210398521SM619070051MA11190215 SM619070052MA 010084068210398521SM619070052MA11190215 SM619070053MA 010084068210398521SM619070053MA11190215 SM619090004MA 010084068210398521SM619090004MA11190225 SM619090005MA 010084068210398521SM619090005MA11190225 SM619090006MA 010084068210398521SM619090006MA11190225 SM619220026MA 010084068210398521SM619220026MA11190527 SM619220027MA 010084068210398521SM619220027MA11190527 SM619230003MA 010084068210398521SM619230003MA11190604 SM619230005MA 010084068210398521SM619230005MA11190604 SM618390006MA 010084068210398521SM618390006MA11180925 SM619060033MA 010084068210398521SM619060033MA11190206 SM619060035MA 010084068210398521SM619060035MA11190206 SM619060036MA 010084068210398521SM619060036MA11190206 SM619060041MA 010084068210398521SM619060041MA11190206 SM619250035MA 010084068210398521SM619250035MA11190619 SM618480041MA 010084068210398521SM618480041MA11181128 SM618480042MA 010084068210398521SM618480042MA11181128 SM618480043MA 010084068210398521SM618480043MA11181128 SM618480044MA 010084068210398521SM618480044MA11181128 SM618480045MA 010084068210398521SM618480045MA11181128 SM618480046MA 010084068210398521SM618480046MA11181128 SM618480047MA 010084068210398521SM618480047MA11181128 SM618480048MA 010084068210398521SM618480048MA11181128 SM618480002MA 010084068210398521SM618480002MA11181126 SM618480003MA 010084068210398521SM618480003MA11181126 SM619200024MA 010084068210398521SM619200024MA11190515 SM618480027MA 010084068210398521SM618480027MA11181126 SM619200030MA 010084068210398521SM619200030MA11190515 SM619200031MA 010084068210398521SM619200031MA11190515 SM619200032MA 010084068210398521SM619200032MA11190515 SM619200033MA 010084068210398521SM619200033MA11190515 SM619200034MA 010084068210398521SM619200034MA11190515 SM618450016MA 010084068210398521SM618450016MA11181106 SM618450017MA 010084068210398521SM618450017MA11181106 SM619190023MA 010084068210398521SM619190023MA11190507 SM619190024MA 010084068210398521SM619190024MA11190507 SM619190026MA 010084068210398521SM619190026MA11190507 SM619190028MA 010084068210398521SM619190028MA11190507 SM619190029MA 010084068210398521SM619190029MA11190507 SM619060013MA 010084068210398521SM619060013MA11190206 SM619200046MA 010084068210398521SM619200046MA11190516 SM619200047MA 010084068210398521SM619200047MA11190516 SM619200048MA 010084068210398521SM619200048MA11190516 SM619200049MA 010084068210398521SM619200049MA11190516 SM619200050MA 010084068210398521SM619200050MA11190516 SM619200051MA 010084068210398521SM619200051MA11190516 SM619060032MA 010084068210398521SM619060032MA11190206 SM619060034MA 010084068210398521SM619060034MA11190206 SM619060014MA 010084068210398521SM619060014MA11190206 SM619060015MA 010084068210398521SM619060015MA11190206 SM619060038MA 010084068210398521SM619060038MA11190206 SM619060040MA 010084068210398521SM619060040MA11190206 SM618370007MA 010084068210398521SM618370007MA11180910 SM618450023MA 010084068210398521SM618450023MA11181107 SM618480062MA 010084068210398521SM618480062MA11181129 SM618490028MA 010084068210398521SM618490028MA11181207 SM618360026MA 010084068210398521SM618360026MA11180906 SM618370008MA 010084068210398521SM618370008MA11180910 SM618370009MA 010084068210398521SM618370009MA11180910 SM619060001MA 010084068210398521SM619060001MA11190205 SM619060002MA 010084068210398521SM619060002MA11190205 SM619060003MA 010084068210398521SM619060003MA11190205 SM619060004MA 010084068210398521SM619060004MA11190205 SM619060005MA 010084068210398521SM619060005MA11190205 SM619060006MA 010084068210398521SM619060006MA11190205 SM619060007MA 010084068210398521SM619060007MA11190205 SM619060008MA 010084068210398521SM619060008MA11190205 SM619060009MA 010084068210398521SM619060009MA11190205 SM619060010MA 010084068210398521SM619060010MA11190205 SM619060011MA 010084068210398521SM619060011MA11190205 SM619060012MA 010084068210398521SM619060012MA11190205 SM619200026MA 010084068210398521SM619200026MA11190515 SM619200027MA 010084068210398521SM619200027MA11190515 SM619200028MA 010084068210398521SM619200028MA11190515 SM619200029MA 010084068210398521SM619200029MA11190515 SM618360025MA 010084068210398521SM618360025MA11180906 SM618450024MA 010084068210398521SM618450024MA11181107 SM618450025MA 010084068210398521SM618450025MA11181107 SM619230004MA 010084068210398521SM619230004MA11190604 SM618450014MA 010084068210398521SM618450014MA11181106 SM619060017MA 010084068210398521SM619060017MA11190206 SM619060018MA 010084068210398521SM619060018MA11190206 SM619060019MA 010084068210398521SM619060019MA11190206 SM619060020MA 010084068210398521SM619060020MA11190206 SM619060021MA 010084068210398521SM619060021MA11190206 SM619180016MA 010084068210398521SM619180016MA11190501 SM618390012WA 010084068210398521SM618390012WA11180926 SM619250021MA 010084068210398521SM619250021MA11190618 SM619250022MA 010084068210398521SM619250022MA11190618 SM619250023MA 010084068210398521SM619250023MA11190618 SM619250024MA 010084068210398521SM619250024MA11190618 SM619250025MA 010084068210398521SM619250025MA11190618 SM618360014MA 010084068210398521SM618360014MA11180905 SM618360015MA 010084068210398521SM618360015MA11180905 SM618360016MA 010084068210398521SM618360016MA11180905 SM619170051MA 010084068210398521SM619170051MA11190425 SM619170052MA 010084068210398521SM619170052MA11190425 SM619170054MA 010084068210398521SM619170054MA11190425 SM619170055MA 010084068210398521SM619170055MA11190425 SM619170057MA 010084068210398521SM619170057MA11190425 SM619170058MA 010084068210398521SM619170058MA11190425 SM619170059MA 010084068210398521SM619170059MA11190425 SM619170060MA 010084068210398521SM619170060MA11190425 SM619190011MA 010084068210398521SM619190011MA11190506 SM619190012MA 010084068210398521SM619190012MA11190506 SM619190013MA 010084068210398521SM619190013MA11190506 SM619190014MA 010084068210398521SM619190014MA11190506 SM619190015MA 010084068210398521SM619190015MA11190506 SM619190016MA 010084068210398521SM619190016MA11190506 SM619190017MA 010084068210398521SM619190017MA11190506 SM619220001MA 010084068210398521SM619220001MA11190527 SM619220002MA 010084068210398521SM619220002MA11190527 SM619220003MA 010084068210398521SM619220003MA11190527 SM619220004MA 010084068210398521SM619220004MA11190527 SM619220005MA 010084068210398521SM619220005MA11190527 SM619220006MA 010084068210398521SM619220006MA11190527 SM619220007MA 010084068210398521SM619220007MA11190527 SM619220008MA 010084068210398521SM619220008MA11190527 SM619220009MA 010084068210398521SM619220009MA11190527 SM619220010MA 010084068210398521SM619220010MA11190527 SM619220011MA 010084068210398521SM619220011MA11190527 SM619220012MA 010084068210398521SM619220012MA11190527 SM619220013MA 010084068210398521SM619220013MA11190527 SM619220014MA 010084068210398521SM619220014MA11190527 SM619220015MA 010084068210398521SM619220015MA11190527 SM619220016MA 010084068210398521SM619220016MA11190527 SM619220017MA 010084068210398521SM619220017MA11190527 SM619220018MA 010084068210398521SM619220018MA11190527 SM619220019MA 010084068210398521SM619220019MA11190527 SM619220020MA 010084068210398521SM619220020MA11190527 SM619220021MA 010084068210398521SM619220021MA11190527 SM619220022MA 010084068210398521SM619220022MA11190527 SM619220023MA 010084068210398521SM619220023MA11190527 SM619220024MA 010084068210398521SM619220024MA11190527 SM619220025MA 010084068210398521SM619220025MA11190527 SM619170053MA 010084068210398521SM619170053MA11190425 SM619190018MA 010084068210398521SM619190018MA11190506 SM619190009MA 010084068210398521SM619190009MA11190506 SM618380024WA 010084068210398521SM618380024WA11180920 SM618380025WA 010084068210398521SM618380025WA11180920 SM618380026WA 010084068210398521SM618380026WA11180920 SM618380027WA 010084068210398521SM618380027WA11180920 SM618380028WA 010084068210398521SM618380028WA11180920 SM618380029WA 010084068210398521SM618380029WA11180920 SM618380030WA 010084068210398521SM618380030WA11180920 SM618380032WA 010084068210398521SM618380032WA11180921 SM618380033WA 010084068210398521SM618380033WA11180921 SM618380034WA 010084068210398521SM618380034WA11180921 SM618380035WA 010084068210398521SM618380035WA11180921 SM618380036WA 010084068210398521SM618380036WA11180921 SM618380037WA 010084068210398521SM618380037WA11180921 SM618380038WA 010084068210398521SM618380038WA11180921 SM618390007WA 010084068210398521SM618390007WA11180926 SM618390008WA 010084068210398521SM618390008WA11180926 SM618390011WA 010084068210398521SM618390011WA11180926 SM618350017WA 010084068210398521SM618350017WA11180830 SM618360001WA 010084068210398521SM618360001WA11180903 SM618370011WA 010084068210398521SM618370011WA11180911 SM618410003WA 010084068210398521SM618410003WA11181008 SM618410004WA 010084068210398521SM618410004WA11181008 SM618410005WA 010084068210398521SM618410005WA11181008 SM618410006WA 010084068210398521SM618410006WA11181008 SM618410016WA 010084068210398521SM618410016WA11181009 SM618410017WA 010084068210398521SM618410017WA11181009 SM618410018WA 010084068210398521SM618410018WA11181009 SM618410019WA 010084068210398521SM618410019WA11181010 SM618420001WA 010084068210398521SM618420001WA11181015 SM618430010WA 010084068210398521SM618430010WA11181025 SM618430011WA 010084068210398521SM618430011WA11181025 SM618430012WA 010084068210398521SM618430012WA11181025 SM618430013WA 010084068210398521SM618430013WA11181025 SM618430014WA 010084068210398521SM618430014WA11181025 SM618450001WA 010084068210398521SM618450001WA11181105 SM618450002WA 010084068210398521SM618450002WA11181105 SM618450003WA 010084068210398521SM618450003WA11181105 SM618450004WA 010084068210398521SM618450004WA11181105 SM618460001WA 010084068210398521SM618460001WA11181112 SM618460002WA 010084068210398521SM618460002WA11181112 SM618460037WA 010084068210398521SM618460037WA11181113 SM618460038WA 010084068210398521SM618460038WA11181113 SM618460060WA 010084068210398521SM618460060WA11181115 SM618460061WA 010084068210398521SM618460061WA11181115 SM618500022WA 010084068210398521SM618500022WA11181214 SM618500023WA 010084068210398521SM618500023WA11181214 SM619020001WA 010084068210398521SM619020001WA11190107 SM619020019WA 010084068210398521SM619020019WA11190109 SM619070061WA 010084068210398521SM619070061WA11190216 SM619080003WA 010084068210398521SM619080003WA11190219 SM619080004WA 010084068210398521SM619080004WA11190219 SM619080005WA 010084068210398521SM619080005WA11190219 SM619080006WA 010084068210398521SM619080006WA11190219 SM619080007WA 010084068210398521SM619080007WA11190219 SM619090010WA 010084068210398521SM619090010WA11190226 SM619110025WA 010084068210398521SM619110025WA11190314 SM618360002WA 010084068210398521SM618360002WA11180903 SM618400011MA 010084068210398521SM618400011MA11181004 SM618400012MA 010084068210398521SM618400012MA11181004 SM618400013MA 010084068210398521SM618400013MA11181004 SM618400014MA 010084068210398521SM618400014MA11181004 SM618400015MA 010084068210398521SM618400015MA11181004 SM618520013MA 010084068210398521SM618520013MA11181227 SM619050023MA 010084068210398521SM619050023MA11190201 SM619050024MA 010084068210398521SM619050024MA11190201 SM619090024MA 010084068210398521SM619090024MA11190228 SM619090025MA 010084068210398521SM619090025MA11190228 SM619090026MA 010084068210398521SM619090026MA11190228 SM619170056MA 010084068210398521SM619170056MA11190425 SM619170061MA 010084068210398521SM619170061MA11190425 SM619170062MA 010084068210398521SM619170062MA11190425 SM619170063MA 010084068210398521SM619170063MA11190425 SM619170064MA 010084068210398521SM619170064MA11190425 SM619170065MA 010084068210398521SM619170065MA11190425 SM619170066MA 010084068210398521SM619170066MA11190425 SM619170067MA 010084068210398521SM619170067MA11190425 SM619170068MA 010084068210398521SM619170068MA11190425 SM619170069MA 010084068210398521SM619170069MA11190425 SM619170070MA 010084068210398521SM619170070MA11190425 SM619170071MA 010084068210398521SM619170071MA11190425 SM619170072MA 010084068210398521SM619170072MA11190425 SM619170073MA 010084068210398521SM619170073MA11190425 SM619170074MA 010084068210398521SM619170074MA11190425 SM619170075MA 010084068210398521SM619170075MA11190425 SM619170076MA 010084068210398521SM619170076MA11190425 SM619170077MA 010084068210398521SM619170077MA11190425 SM619170078MA 010084068210398521SM619170078MA11190425 SM619170079MA 010084068210398521SM619170079MA11190425 SM619190010MA 010084068210398521SM619190010MA11190506 SM618380023WA 010084068210398521SM618380023WA11180919 SM618450007WA 010084068210398521SM618450007WA11181106 SM618450008WA 010084068210398521SM618450008WA11181106 SM618470013WA 010084068210398521SM618470013WA11181121 SM618470014WA 010084068210398521SM618470014WA11181121 SM618470015WA 010084068210398521SM618470015WA11181121 SM618470016WA 010084068210398521SM618470016WA11181121 SM618470017WA 010084068210398521SM618470017WA11181121 SM618470018WA 010084068210398521SM618470018WA11181121 SM618470019WA 010084068210398521SM618470019WA11181121 SM618470020WA 010084068210398521SM618470020WA11181121 SM618470021WA 010084068210398521SM618470021WA11181121 SM619030001WA 010084068210398521SM619030001WA11190114 SM619030002WA 010084068210398521SM619030002WA11190114 SM619030003WA 010084068210398521SM619030003WA11190114 SM619030004WA 010084068210398521SM619030004WA11190114 SM619030005WA 010084068210398521SM619030005WA11190114 SM619030006WA 010084068210398521SM619030006WA11190114 SM619050017WA 010084068210398521SM619050017WA11190201 SM619050018WA 010084068210398521SM619050018WA11190201 SM619050019WA 010084068210398521SM619050019WA11190201 SM619070015WA 010084068210398521SM619070015WA11190212 SM619070016WA 010084068210398521SM619070016WA11190212 SM619070017WA 010084068210398521SM619070017WA11190212 SM619070018WA 010084068210398521SM619070018WA11190212 SM619070020WA 010084068210398521SM619070020WA11190212 SM619110028WA 010084068210398521SM619110028WA11190314 SM619110029WA 010084068210398521SM619110029WA11190314 SM619110030WA 010084068210398521SM619110030WA11190314 SM619110031WA 010084068210398521SM619110031WA11190314 SM619110032WA 010084068210398521SM619110032WA11190314 SM619110033WA 010084068210398521SM619110033WA11190314 SM619110034WA 010084068210398521SM619110034WA11190314 SM619110036WA 010084068210398521SM619110036WA11190315 SM619110037WA 010084068210398521SM619110037WA11190315 SM619110038WA 010084068210398521SM619110038WA11190315 SM619110039WA 010084068210398521SM619110039WA11190315 SM619110040WA 010084068210398521SM619110040WA11190315 SM619110041WA 010084068210398521SM619110041WA11190315 SM619110042WA 010084068210398521SM619110042WA11190315 SM619110043WA 010084068210398521SM619110043WA11190315 SM619130007MA 010084068210398521SM619130007MA11190328 SM619140023WA 010084068210398521SM619140023WA11190403 SM619150002WA 010084068210398521SM619150002WA11190408 SM619150003WA 010084068210398521SM619150003WA11190408 SM619150004WA 010084068210398521SM619150004WA11190408 SM619150005WA 010084068210398521SM619150005WA11190408 SM619150006WA 010084068210398521SM619150006WA11190408 SM619150019WA 010084068210398521SM619150019WA11190410 SM619150020WA 010084068210398521SM619150020WA11190410 SM619150021WA 010084068210398521SM619150021WA11190410 SM619150022WA 010084068210398521SM619150022WA11190410 SM619160023WA 010084068210398521SM619160023WA11190420 SM619160025WA 010084068210398521SM619160025WA11190420 SM619160026WA 010084068210398521SM619160026WA11190420 SM619170081MA 010084068210398521SM619170081MA11190425 SM619190019WA 010084068210398521SM619190019WA11190507 SM619190020WA 010084068210398521SM619190020WA11190507 SM619210020WA 010084068210398521SM619210020WA11190524 SM619220051WA 010084068210398521SM619220051WA11190529 SM619220052WA 010084068210398521SM619220052WA11190529 SM619230006WA 010084068210398521SM619230006WA11190605 SM619230007WA 010084068210398521SM619230007WA11190605 SM619230008WA 010084068210398521SM619230008WA11190605 SM619230015WA 010084068210398521SM619230015WA11190606 SM619230016WA 010084068210398521SM619230016WA11190606 SM619230017WA 010084068210398521SM619230017WA11190606 SM618440013MA 010084068210398521SM618440013MA11181029 SM618440014MA 010084068210398521SM618440014MA11181029 SM619200013WA 010084068210398521SM619200013WA11190514 SM619200014WA 010084068210398521SM619200014WA11190514 SM619200015WA 010084068210398521SM619200015WA11190514 SM619200016WA 010084068210398521SM619200016WA11190514 SM619200017WA 010084068210398521SM619200017WA11190514 SM619200018WA 010084068210398521SM619200018WA11190514 SM619200019WA 010084068210398521SM619200019WA11190514 SM619200020WA 010084068210398521SM619200020WA11190514 SM619200021WA 010084068210398521SM619200021WA11190514 SM619200022WA 010084068210398521SM619200022WA11190514 SM619220056MA 010084068210398521SM619220056MA11190530 SM619220057MA 010084068210398521SM619220057MA11190530 SM619220058MA 010084068210398521SM619220058MA11190530 SM619090018MA 010084068210398521SM619090018MA11190227 SM619090019MA 010084068210398521SM619090019MA11190227 SM619220044MA 010084068210398521SM619220044MA11190527 SM618450006WA 010084068210398521SM618450006WA11181105 SM618450031WA 010084068210398521SM618450031WA11181108 SM619150017WA 010084068210398521SM619150017WA11190410 SM619150018WA 010084068210398521SM619150018WA11190410 SM618340003WA 010084068210398521SM618340003WA11180820 SM619200055WA 010084068210398521SM619200055WA11190517 SM619170044WA 010084068210398521SM619170044WA11190424 SM619170046WA 010084068210398521SM619170046WA11190424 SM619040009WA 010084068210398521SM619040009WA11190125 SM619170043WA 010084068210398521SM619170043WA11190424 SM619170045WA 010084068210398521SM619170045WA11190424 SM619190021WA 010084068210398521SM619190021WA11190507 SM619200053WA 010084068210398521SM619200053WA11190517 SM619040010WA 010084068210398521SM619040010WA11190125 SM619100019WA 010084068210398521SM619100019WA11190308 SM618450009WA 010084068210398521SM618450009WA11181106 SM618450010WA 010084068210398521SM618450010WA11181106 SM618450011WA 010084068210398521SM618450011WA11181106 SM619200038WA 010084068210398521SM619200038WA11190516 SM619200039WA 010084068210398521SM619200039WA11190516 SM619200040WA 010084068210398521SM619200040WA11190516 SM619200041WA 010084068210398521SM619200041WA11190516 SM619200054WA 010084068210398521SM619200054WA11190517 SM619200052WA 010084068210398521SM619200052WA11190517 SM618340004WA 010084068210398521SM618340004WA11180820 SM618340005WA 010084068210398521SM618340005WA11180820 SM619040011WA 010084068210398521SM619040011WA11190125 SM618350015WA 010084068210398521SM618350015WA11180830 SM618340001WA 010084068210398521SM618340001WA11180820 SM618340002WA 010084068210398521SM618340002WA11180820 SM619200045WA 010084068210398521SM619200045WA11190516 SM619100016WA 010084068210398521SM619100016WA11190308 SM619100017WA 010084068210398521SM619100017WA11190308 SM618350040WA 010084068210398521SM618350040WA11180831 SM619150023WA 010084068210398521SM619150023WA11190410 SM619150024WA 010084068210398521SM619150024WA11190410 SM619150025WA 010084068210398521SM619150025WA11190410 SM619150026WA 010084068210398521SM619150026WA11190410 SM619150028WA 010084068210398521SM619150028WA11190411 SM619150029WA 010084068210398521SM619150029WA11190411 SM619150030WA 010084068210398521SM619150030WA11190411 SM619150031WA 010084068210398521SM619150031WA11190411 SM619150032WA 010084068210398521SM619150032WA11190411 SM619150033WA 010084068210398521SM619150033WA11190411 SM618410015WA 010084068210398521SM618410015WA11181009 SM618350041WA 010084068210398521SM618350041WA11180831 SM619040007WA 010084068210398521SM619040007WA11190125 SM619040008WA 010084068210398521SM619040008WA11190125 SM619100018WA 010084068210398521SM619100018WA11190308 SM619200042WA 010084068210398521SM619200042WA11190516 SM619200058WA 010084068210398521SM619200058WA11190517 SM618340011WA 010084068210398521SM618340011WA11180821 SM618340012WA 010084068210398521SM618340012WA11180821 SM618340014WA 010084068210398521SM618340014WA11180821 SM618340006WA 010084068210398521SM618340006WA11180821 SM618340007WA 010084068210398521SM618340007WA11180821 SM618340008WA 010084068210398521SM618340008WA11180821 SM618340010WA 010084068210398521SM618340010WA11180821 SM618340013WA 010084068210398521SM618340013WA11180821 SM618340015WA 010084068210398521SM618340015WA11180821 SM618340016WA 010084068210398521SM618340016WA11180821 SM619050011WA 010084068210398521SM619050011WA11190131 SM619070036WA 010084068210398521SM619070036WA11190215 SM619070037WA 010084068210398521SM619070037WA11190215 SM619070038WA 010084068210398521SM619070038WA11190215 SM619070039WA 010084068210398521SM619070039WA11190215 SM619070040WA 010084068210398521SM619070040WA11190215 SM618340009WA 010084068210398521SM618340009WA11180821 SM618390037MA 010084068210398521SM618390037MA11180927 SM618390038MA 010084068210398521SM618390038MA11180927 SM618390039MA 010084068210398521SM618390039MA11180927 SM618390040MA 010084068210398521SM618390040MA11180927 SM618400003MA 010084068210398521SM618400003MA11181001 SM618420025MA 010084068210398521SM618420025MA11181018 SM618420026MA 010084068210398521SM618420026MA11181018 SM618420027MA 010084068210398521SM618420027MA11181018 SM618420028MA 010084068210398521SM618420028MA11181018 SM618420029MA 010084068210398521SM618420029MA11181018 SM618420030MA 010084068210398521SM618420030MA11181018 SM618420031MA 010084068210398521SM618420031MA11181018 SM618420032MA 010084068210398521SM618420032MA11181018 SM618420033MA 010084068210398521SM618420033MA11181018 SM618420034MA 010084068210398521SM618420034MA11181018 SM619220045MA 010084068210398521SM619220045MA11190527 SM619220046MA 010084068210398521SM619220046MA11190527 SM619220047MA 010084068210398521SM619220047MA11190527 SM619220048MA 010084068210398521SM619220048MA11190527 SM619220049MA 010084068210398521SM619220049MA11190527 SM618500033MA 010084068210398521SM618500033MA11181214 SM618520023MA 010084068210398521SM618520023MA11181228 SM618520024MA 010084068210398521SM618520024MA11181228 SM619130008MA 010084068210398521SM619130008MA11190328 SM618380022WA 010084068210398521SM618380022WA11180919 SM618440015WA 010084068210398521SM618440015WA11181030 SM618470027WA 010084068210398521SM618470027WA11181123 SM618470028WA 010084068210398521SM618470028WA11181123 SM618480058WA 010084068210398521SM618480058WA11181129 SM618480059WA 010084068210398521SM618480059WA11181129 SM618480060WA 010084068210398521SM618480060WA11181129 SM619040002WA 010084068210398521SM619040002WA11190122 SM619050002WA 010084068210398521SM619050002WA11190128 SM619050003WA 010084068210398521SM619050003WA11190128 SM619080008WA 010084068210398521SM619080008WA11190219 SM619080009WA 010084068210398521SM619080009WA11190219 SM619080010WA 010084068210398521SM619080010WA11190220 SM619090021WA 010084068210398521SM619090021WA11190228 SM619140021WA 010084068210398521SM619140021WA11190403 SM619140022WA 010084068210398521SM619140022WA11190403 SM619210021WA 010084068210398521SM619210021WA11190524 SM619220050WA 010084068210398521SM619220050WA11190528 SM618400005MA 010084068210398521SM618400005MA11181003 SM619250026MA 010084068210398521SM619250026MA11190618 SM619250027MA 010084068210398521SM619250027MA11190618 SM619250028MA 010084068210398521SM619250028MA11190618 SM619250029MA 010084068210398521SM619250029MA11190618 SM619090023WA 010084068210398521SM619090023WA11190228 SM619210003WA 010084068210398521SM619210003WA11190521 SM619210004WA 010084068210398521SM619210004WA11190521 SM619190008WA 010084068210398521SM619190008WA11190506 SM619210005WA 010084068210398521SM619210005WA11190521 SM618520002WA 010084068210398521SM618520002WA11181225 SM618520003WA 010084068210398521SM618520003WA11181225 SM619090015WA 010084068210398521SM619090015WA11190226 SM618360017WA 010084068210398521SM618360017WA11180906 SM619020004WA 010084068210398521SM619020004WA11190108 SM619090017WA 010084068210398521SM619090017WA11190227 SM618360019WA 010084068210398521SM618360019WA11180906 SM619090020WA 010084068210398521SM619090020WA11190228 SM618480066WA 010084068210398521SM618480066WA11181130 SM618440034WA 010084068210398521SM618440034WA11181102 SM618470032WA 010084068210398521SM618470032WA11181124 SM619090016WA 010084068210398521SM619090016WA11190227 SM618360005WA 010084068210398521SM618360005WA11180903 SM619240004WA 010084068210398521SM619240004WA11190612 SM618430006WA 010084068210398521SM618430006WA11181024 SM619070004WA 010084068210398521SM619070004WA11190211 SM619140027WA 010084068210398521SM619140027WA11190404 SM619140028WA 010084068210398521SM619140028WA11190404 SM618490003WA 010084068210398521SM618490003WA11181203 SM618360018WA 010084068210398521SM618360018WA11180906 SM618420003WA 010084068210398521SM618420003WA11181015 SM619220060WA 010084068210398521SM619220060WA11190531 SM618470034WA 010084068210398521SM618470034WA11181124 SM618360020WA 010084068210398521SM618360020WA11180906 SM618360021WA 010084068210398521SM618360021WA11180906 SM618480031WA 010084068210398521SM618480031WA11181127 SM619090022WA 010084068210398521SM619090022WA11190228 SM618410021WA 010084068210398521SM618410021WA11181010 SM619020002WA 010084068210398521SM619020002WA11190107 SM619090014WA 010084068210398521SM619090014WA11190226 SM618360004WA 010084068210398521SM618360004WA11180903 SM619150038WA 010084068210398521SM619150038WA11190412 SM619160002WA 010084068210398521SM619160002WA11190415 SM619160003WA 010084068210398521SM619160003WA11190415 SM619160004WA 010084068210398521SM619160004WA11190415 SM619150037WA 010084068210398521SM619150037WA11190412 SM618490002WA 010084068210398521SM618490002WA11181203 SM619040005WA 010084068210398521SM619040005WA11190123 SM619180026WA 010084068210398521SM619180026WA11190505 SM618450005WA 010084068210398521SM618450005WA11181105 SM618410020WA 010084068210398521SM618410020WA11181010 SM618480001WA 010084068210398521SM618480001WA11181126 SM618470033WA 010084068210398521SM618470033WA11181124 SM619100008WA 010084068210398521SM619100008WA11190305 SM619100009WA 010084068210398521SM619100009WA11190305 SM619110010WA 010084068210398521SM619110010WA11190312 SM619140026WA 010084068210398521SM619140026WA11190404 SM619160001WA 010084068210398521SM619160001WA11190415 SM618520005WA 010084068210398521SM618520005WA11181225 SM618410047WA 010084068210398521SM618410047WA11181012 SM618520004WA 010084068210398521SM618520004WA11181225 SM618480032WA 010084068210398521SM618480032WA11181127 SM618410022WA 010084068210398521SM618410022WA11181010 SM618460003WA 010084068210398521SM618460003WA11181112 SM618480033WA 010084068210398521SM618480033WA11181127 SM618520006WA 010084068210398521SM618520006WA11181225 SM619130006WA 010084068210398521SM619130006WA11190328 SM619150008WA 010084068210398521SM619150008WA11190409 SM619020005WA 010084068210398521SM619020005WA11190108 SM619020006WA 010084068210398521SM619020006WA11190108 SM619020007WA 010084068210398521SM619020007WA11190108 SM618520001WA 010084068210398521SM618520001WA11181225 SM619010101MA 010084068210398521SM619010101MA11190102 SM619010102MA 010084068210398521SM619010102MA11190102 SM619060079MA 010084068210398521SM619060079MA11190208 SM619120023MA 010084068210398521SM619120023MA11190319 SM619260009MA 010084068210398521SM619260009MA11190628 SM619260010MA 010084068210398521SM619260010MA11190628 SM619260011MA 010084068210398521SM619260011MA11190628 SM619260012MA 010084068210398521SM619260012MA11190628 SM619260014MA 010084068210398521SM619260014MA11190628 SM619170080MA 010084068210398521SM619170080MA11190425 SM618390035MA 010084068210398521SM618390035MA11180927 SM618390036MA 010084068210398521SM618390036MA11180927 SM618460071MA 010084068210398521SM618460071MA11181116 SM618460072MA 010084068210398521SM618460072MA11181116 SM618520021MA 010084068210398521SM618520021MA11181228 SM619020029MA 010084068210398521SM619020029MA11190111 SM619020030MA 010084068210398521SM619020030MA11190111 SM619020031MA 010084068210398521SM619020031MA11190111 SM619020032MA 010084068210398521SM619020032MA11190111 SM619020033MA 010084068210398521SM619020033MA11190111 SM619020034MA 010084068210398521SM619020034MA11190111 SM619020035MA 010084068210398521SM619020035MA11190111 SM619020036MA 010084068210398521SM619020036MA11190111 SM619020037MA 010084068210398521SM619020037MA11190111 SM619020038MA 010084068210398521SM619020038MA11190111 SM619020039MA 010084068210398521SM619020039MA11190111 SM619020040MA 010084068210398521SM619020040MA11190111 SM619020041MA 010084068210398521SM619020041MA11190111 SM619020042MA 010084068210398521SM619020042MA11190111 SM619020043MA 010084068210398521SM619020043MA11190111 SM619020044MA 010084068210398521SM619020044MA11190111 SM619020045MA 010084068210398521SM619020045MA11190111 SM619020046MA 010084068210398521SM619020046MA11190111 SM619020047MA 010084068210398521SM619020047MA11190111 SM619020048MA 010084068210398521SM619020048MA11190111 SM619020049MA 010084068210398521SM619020049MA11190111 SM619020050MA 010084068210398521SM619020050MA11190111 SM619020051MA 010084068210398521SM619020051MA11190111 SM619020052MA 010084068210398521SM619020052MA11190111 SM619020053MA 010084068210398521SM619020053MA11190111 SM619020054MA 010084068210398521SM619020054MA11190111 SM619020055MA 010084068210398521SM619020055MA11190111 SM619020056MA 010084068210398521SM619020056MA11190111 SM619020057MA 010084068210398521SM619020057MA11190111 SM619020058MA 010084068210398521SM619020058MA11190111 SM619020059MA 010084068210398521SM619020059MA11190111 SM619020060MA 010084068210398521SM619020060MA11190111 SM619020061MA 010084068210398521SM619020061MA11190111 SM619020062MA 010084068210398521SM619020062MA11190111 SM619030017MA 010084068210398521SM619030017MA11190115 SM619030018MA 010084068210398521SM619030018MA11190115 SM619030019MA 010084068210398521SM619030019MA11190115 SM619030020MA 010084068210398521SM619030020MA11190115 SM619030021MA 010084068210398521SM619030021MA11190115 SM619030022MA 010084068210398521SM619030022MA11190115 SM619030023MA 010084068210398521SM619030023MA11190115 SM618350018MA 010084068210398521SM618350018MA11180830 SM618370065MA 010084068210398521SM618370065MA11180913 SM618400006MA 010084068210398521SM618400006MA11181003 SM618400010MA 010084068210398521SM618400010MA11181004 SM618500030MA 010084068210398521SM618500030MA11181214 SM619160016MA 010084068210398521SM619160016MA11190418 SM619200043WA 010084068210398521SM619200043WA11190516 SM619200044WA 010084068210398521SM619200044WA11190516 SM618350001MA 010084068210398521SM618350001MA11180828 SM618350002MA 010084068210398521SM618350002MA11180828 SM618350003MA 010084068210398521SM618350003MA11180828 SM619220036MA 010084068210398521SM619220036MA11190527 SM619220037MA 010084068210398521SM619220037MA11190527 SM619220038MA 010084068210398521SM619220038MA11190527 SM619220039MA 010084068210398521SM619220039MA11190527 SM619220040MA 010084068210398521SM619220040MA11190527 SM619220041MA 010084068210398521SM619220041MA11190527 SM619220042MA 010084068210398521SM619220042MA11190527 SM619220043MA 010084068210398521SM619220043MA11190527 SM618350004MA 010084068210398521SM618350004MA11180828 SM618350005MA 010084068210398521SM618350005MA11180828 SM618350006MA 010084068210398521SM618350006MA11180828 SM618350007MA 010084068210398521SM618350007MA11180828 SM618350008MA 010084068210398521SM618350008MA11180828 SM618350009MA 010084068210398521SM618350009MA11180828 SM618350010MA 010084068210398521SM618350010MA11180828 SM618350011MA 010084068210398521SM618350011MA11180828 SM618370061MA 010084068210398521SM618370061MA11180913 SM619090007MA 010084068210398521SM619090007MA11190225 SM619260002MA 010084068210398521SM619260002MA11190626 SM619210016MA 010084068210398521SM619210016MA11190523 SM619210017MA 010084068210398521SM619210017MA11190523 SM619210018MA 010084068210398521SM619210018MA11190523 SM619210019MA 010084068210398521SM619210019MA11190523 SM618370058MA 010084068210398521SM618370058MA11180913 SM618370059MA 010084068210398521SM618370059MA11180913 SM618420035MA 010084068210398521SM618420035MA11181018 SM618370057MA 010084068210398521SM618370057MA11180913 SM618470004MA 010084068210398521SM618470004MA11181119 SM618470005MA 010084068210398521SM618470005MA11181119 SM618470006MA 010084068210398521SM618470006MA11181119 SM618470007MA 010084068210398521SM618470007MA11181119 SM618410011MA 010084068210398521SM618410011MA11181008 SM618410012MA 010084068210398521SM618410012MA11181008 SM618410013MA 010084068210398521SM618410013MA11181008 SM618410014MA 010084068210398521SM618410014MA11181008 SM619260001MA 010084068210398521SM619260001MA11190625 SM618370062MA 010084068210398521SM618370062MA11180913 SM619270052MA 010084068210398521SM619270052MA11190703 SM619150027MA 010084068210398521SM619150027MA11190410 SM618340082MA 010084068210398521SM618340082MA11180824 SM618360059MA 010084068210398521SM618360059MA11180907 SM618360060MA 010084068210398521SM618360060MA11180907 SM618410010MA 010084068210398521SM618410010MA11181008 SM618410008MA 010084068210398521SM618410008MA11181008 SM618410009MA 010084068210398521SM618410009MA11181008 SM618370060MA 010084068210398521SM618370060MA11180913 SM618370063MA 010084068210398521SM618370063MA11180913 SM618340079WA 010084068210398521SM618340079WA11180824 SM618340080WA 010084068210398521SM618340080WA11180824 SM618340081WA 010084068210398521SM618340081WA11180824 SM619160022WA 010084068210398521SM619160022WA11190420 SM619080021WA 010084068210398521SM619080021WA11190221 SM619100005WA 010084068210398521SM619100005WA11190305 SM619100006WA 010084068210398521SM619100006WA11190305 SM619100007WA 010084068210398521SM619100007WA11190305 SM619080020WA 010084068210398521SM619080020WA11190221 SM619160024WA 010084068210398521SM619160024WA11190420 SM619080018WA 010084068210398521SM619080018WA11190221 SM619080019WA 010084068210398521SM619080019WA11190221 SM619150011MA 010084068210398521SM619150011MA11190409 SM619150012MA 010084068210398521SM619150012MA11190409 SM618430016MA 010084068210398521SM618430016MA11181025 SM618460034MA 010084068210398521SM618460034MA11181112 SM618490021MA 010084068210398521SM618490021MA11181206 SM618350012WA 010084068210398521SM618350012WA11180830 SM618350014WA 010084068210398521SM618350014WA11180830 SM618350013WA 010084068210398521SM618350013WA11180830 SM618410048WA 010084068210398521SM618410048WA11181012 SM619070059WA 010084068210398521SM619070059WA11190216 SM619070060WA 010084068210398521SM619070060WA11190216 SM619140002WA 010084068210398521SM619140002WA11190402 SM619100013WA 010084068210398521SM619100013WA11190308 SM619100014WA 010084068210398521SM619100014WA11190308 SM619100015WA 010084068210398521SM619100015WA11190308 SM618340058WA 010084068210398521SM618340058WA11180822 SM618340059WA 010084068210398521SM618340059WA11180822 SM618340060WA 010084068210398521SM618340060WA11180822 SM618430017WA 010084068210398521SM618430017WA11181026 SM618430018WA 010084068210398521SM618430018WA11181026 SM618340055WA 010084068210398521SM618340055WA11180822 SM618340056WA 010084068210398521SM618340056WA11180822 SM618340057WA 010084068210398521SM618340057WA11180822 SM618430021WA 010084068210398521SM618430021WA11181026 SM618430022WA 010084068210398521SM618430022WA11181026 SM618450041WA 010084068210398521SM618450041WA11181109 SM618450042WA 010084068210398521SM618450042WA11181109 SM619080011WA 010084068210398521SM619080011WA11190220 SM619080012WA 010084068210398521SM619080012WA11190220 SM618410007WA 010084068210398521SM618410007WA11181008 SM619190032WA 010084068210398521SM619190032WA11190508 SM619190033WA 010084068210398521SM619190033WA11190508 SM618430019WA 010084068210398521SM618430019WA11181026 SM618430020WA 010084068210398521SM618430020WA11181026 SM618470022WA 010084068210398521SM618470022WA11181122 SM618470023WA 010084068210398521SM618470023WA11181122 SM619170047WA 010084068210398521SM619170047WA11190425 SM619090001WA 010084068210398521SM619090001WA11190225 SM619090002WA 010084068210398521SM619090002WA11190225 SM619090003WA 010084068210398521SM619090003WA11190225 SM619220053WA 010084068210398521SM619220053WA11190530 SM618480038WA 010084068210398521SM618480038WA11181128 SM618460035WA 010084068210394721SM618460035WA11181115 SM618460036WA 010084068210394721SM618460036WA11181115 SM618510007WA 010084068210398521SM618510007WA11181221 SM619020008WA 010084068210398521SM619020008WA11190108 SM618380040WA 010084068210398521SM618380040WA11180921 SM618480034WA 010084068210398521SM618480034WA11181127 SM618450032WA 010084068210398521SM618450032WA11181108 SM618490011WA 010084068210398521SM618490011WA11181205 SM618450019MA 010084068210398521SM618450019MA11181107 SM618450020MA 010084068210398521SM618450020MA11181107 SM618450021MA 010084068210398521SM618450021MA11181107 SM618450022MA 010084068210398521SM618450022MA11181107 SM618490001WA 010084068210398521SM618490001WA11181203 SM618480065WA 010084068210398521SM618480065WA11181130 SM619020023WA 010084068210398521SM619020023WA11190111 SM619220055WA 010084068210398521SM619220055WA11190530 SM618490012WA 010084068210398521SM618490012WA11181205 SM618510008WA 010084068210398521SM618510008WA11181221 SM618460070WA 010084068210398521SM618460070WA11181116 SM618430009WA 010084068210398521SM618430009WA11181025 SM618360006WA 010084068210398521SM618360006WA11180903 SM618360003WA 010084068210398521SM618360003WA11180903 SM618440018WA 010084068210398521SM618440018WA11181031 SM619190037WA 010084068210398521SM619190037WA11190509 SM619190038WA 010084068210398521SM619190038WA11190509 SM618420002WA 010084068210398521SM618420002WA11181015 SM619020003WA 010084068210398521SM619020003WA11190107 SM619100012WA 010084068210398521SM619100012WA11190308 SM618480037WA 010084068210398521SM618480037WA11181128 SM619170048WA 010084068210398521SM619170048WA11190425 SM618340053WA 010084068210398521SM618340053WA11180822 SM618340054WA 010084068210398521SM618340054WA11180822 SM619140029MA 010084068210398521SM619140029MA11190404 SM619140030MA 010084068210398521SM619140030MA11190404 SM618340017WA 010084068210398521SM618340017WA11180821 SM619240001MA 010084068210398521SM619240001MA11190610 SM619240002MA 010084068210398521SM619240002MA11190610 SM618520015MA 010084068210398521SM618520015MA11181228 SM619170049MA 010084068210398521SM619170049MA11190425 SM619260013MA 010084068210398521SM619260013MA11190628 SM619120008MA 010084068210398521SM619120008MA11190319 SM619120009MA 010084068210398521SM619120009MA11190319 SM619160014MA 010084068210398521SM619160014MA11190418 SM619160018MA 010084068210398521SM619160018MA11190418 SM619190031MA 010084068210398521SM619190031MA11190507 SM619170050MA 010084068210398521SM619170050MA11190425 SM618410045MA 010084068210398521SM618410045MA11181011 SM618520020MA 010084068210398521SM618520020MA11181228 SM618520018MA 010084068210398521SM618520018MA11181228 SM618520019MA 010084068210398521SM618520019MA11181228 SM619120010MA 010084068210398521SM619120010MA11190319 SM619120011MA 010084068210398521SM619120011MA11190319 SM619120012MA 010084068210398521SM619120012MA11190319 SM618390001MA 010084068210398521SM618390001MA11180925 SM618390002MA 010084068210398521SM618390002MA11180925 SM618410043MA 010084068210398521SM618410043MA11181011 SM619160013MA 010084068210398521SM619160013MA11190418 SM619260004MA 010084068210398521SM619260004MA11190627 SM619260005MA 010084068210398521SM619260005MA11190627 SM619260006MA 010084068210398521SM619260006MA11190627 SM619260007MA 010084068210398521SM619260007MA11190627 SM619260008MA 010084068210398521SM619260008MA11190627 SM618390032MA 010084068210398521SM618390032MA11180927 SM618390033MA 010084068210398521SM618390033MA11180927 SM618410041MA 010084068210398521SM618410041MA11181011 SM618520017MA 010084068210398521SM618520017MA11181228 SM619060070MA 010084068210398521SM619060070MA11190206 SM619130002MA 010084068210398521SM619130002MA11190326 SM619170082MA 010084068210398521SM619170082MA11190425 SM619230013MA 010084068210398521SM619230013MA11190605 SM619230014MA 010084068210398521SM619230014MA11190605 SM619240003MA 010084068210398521SM619240003MA11190610 SM619250030MA 010084068210398521SM619250030MA11190618 SM619250032MA 010084068210398521SM619250032MA11190618 SM619250033MA 010084068210398521SM619250033MA11190618 SM619250034MA 010084068210398521SM619250034MA11190618 SM619060069MA 010084068210398521SM619060069MA11190206 SM618520022MA 010084068210398521SM618520022MA11181228 SM619110002MA 010084068210398521SM619110002MA11190311 SM618410039MA 010084068210398521SM618410039MA11181011 SM618410040MA 010084068210398521SM618410040MA11181011 SM618520010MA 010084068210398521SM618520010MA11181226 SM618520011MA 010084068210398521SM618520011MA11181226 SM618520012MA 010084068210398521SM618520012MA11181226 SM618410049MA 010084068210398521SM618410049MA11181012 SM619030010MA 010084068210398521SM619030010MA11190114 SM619030011MA 010084068210398521SM619030011MA11190114 SM618410042MA 010084068210398521SM618410042MA11181011 SM618520016MA 010084068210398521SM618520016MA11181228 SM619110003MA 010084068210398521SM619110003MA11190311 SM619120027MA 010084068210398521SM619120027MA11190322 SM618390034MA 010084068210398521SM618390034MA11180927 SM618500028MA 010084068210398521SM618500028MA11181214 SM618500029MA 010084068210398521SM618500029MA11181214 SM619180023WA 010084068210398521SM619180023WA11190505 SM619180024WA 010084068210398521SM619180024WA11190505 SM619180025WA 010084068210398521SM619180025WA11190505 SM619190001WA 010084068210398521SM619190001WA11190506 SM619190002WA 010084068210398521SM619190002WA11190506 SM619190003WA 010084068210398521SM619190003WA11190506 SM619190004WA 010084068210398521SM619190004WA11190506 SM619190005WA 010084068210398521SM619190005WA11190506 SM619200023WA 010084068210398521SM619200023WA11190515 SM619190042WA 010084068210398521SM619190042WA11190510 SM619190043WA 010084068210398521SM619190043WA11190510 SM619190044WA 010084068210398521SM619190044WA11190510 SM619190045WA 010084068210398521SM619190045WA11190510 SM619200059WA 010084068210398521SM619200059WA11190517 SM619190006WA 010084068210398521SM619190006WA11190506 SM619220054WA 010084068210398521SM619220054WA11190530 SM619160027WA 010084068210398521SM619160027WA11190420 SM619160028WA 010084068210398521SM619160028WA11190420 SM619170004WA 010084068210398521SM619170004WA11190422 SM619170005WA 010084068210398521SM619170005WA11190422 SM619170006WA 010084068210398521SM619170006WA11190422 SM619170007WA 010084068210398521SM619170007WA11190422 SM619170008WA 010084068210398521SM619170008WA11190422 SM619170009WA 010084068210398521SM619170009WA11190422 SM619170010WA 010084068210398521SM619170010WA11190422 SM619170034WA 010084068210398521SM619170034WA11190423 SM619170035WA 010084068210398521SM619170035WA11190423 SM619170036WA 010084068210398521SM619170036WA11190423 SM619170037WA 010084068210398521SM619170037WA11190423 SM619170038WA 010084068210398521SM619170038WA11190423 SM618460059WA 010084068210398521SM618460059WA11181115 SM619180020WA 010084068210398521SM619180020WA11190505 SM619190007WA 010084068210398521SM619190007WA11190506 SM618460063WA 010084068210398521SM618460063WA11181116 SM618460064WA 010084068210398521SM618460064WA11181116 SM618460065WA 010084068210398521SM618460065WA11181116 SM618460066WA 010084068210398521SM618460066WA11181116 SM618460067WA 010084068210398521SM618460067WA11181116 SM618460068WA 010084068210398521SM618460068WA11181116 SM619180027WA 010084068210398521SM619180027WA11190505 SM618470029WA 010084068210398521SM618470029WA11181123 SM618470030WA 010084068210398521SM618470030WA11181123 SM618470031WA 010084068210398521SM618470031WA11181123 SM619180021WA 010084068210398521SM619180021WA11190505 SM619180022WA 010084068210398521SM619180022WA11190505 SM618400001MA 010084068210398521SM618400001MA11181001 SM618400002MA 010084068210398521SM618400002MA11181001 SM618400009MA 010084068210398521SM618400009MA11181004 SM618430003MA 010084068210398521SM618430003MA11181023 SM618430004MA 010084068210398521SM618430004MA11181023 SM618500031MA 010084068210398521SM618500031MA11181214 SM618500032MA 010084068210398521SM618500032MA11181214 SM618400004MA 010084068210398521SM618400004MA11181003 SM618410032MA 010084068210398521SM618410032MA11181011 SM618430001MA 010084068210398521SM618430001MA11181022 SM619070028MA 010084068210398521SM619070028MA11190212 SM619070029MA 010084068210398521SM619070029MA11190212 SM619070030MA 010084068210398521SM619070030MA11190212 SM619070031MA 010084068210398521SM619070031MA11190212 SM619190030MA 010084068210398521SM619190030MA11190507 SM618470008MA 010084068210398521SM618470008MA11181119 SM618470009MA 010084068210398521SM618470009MA11181119 SM619210015MA 010084068210398521SM619210015MA11190522 SM619120024MA 010084068210398521SM619120024MA11190320 SM618460008MA 010084068210398521SM618460008MA11181112 SM618460009MA 010084068210398521SM618460009MA11181112 SM618390014MA 010084068210398521SM618390014MA11180926 SM618390015MA 010084068210398521SM618390015MA11180926 SM618390016MA 010084068210398521SM618390016MA11180926 SM618410038MA 010084068210398521SM618410038MA11181011 SM618430024MA 010084068210398521SM618430024MA11181026 SM618430025MA 010084068210398521SM618430025MA11181026 SM618440012MA 010084068210398521SM618440012MA11181029 SM618410035MA 010084068210398521SM618410035MA11181011 SM618410036MA 010084068210398521SM618410036MA11181011 SM619120025MA 010084068210398521SM619120025MA11190320 SM618460024MA 010084068210398521SM618460024MA11181112 SM618460025MA 010084068210398521SM618460025MA11181112 SM618390013MA 010084068210398521SM618390013MA11180926 SM618460023MA 010084068210398521SM618460023MA11181112 SM618480030MA 010084068210398521SM618480030MA11181126 SM618340067MA 010084068210398521SM618340067MA11180822 SM618340061MA 010084068210398521SM618340061MA11180822 SM618340063MA 010084068210398521SM618340063MA11180822 SM618340068MA 010084068210398521SM618340068MA11180822 SM619120014MA 010084068210398521SM619120014MA11190319 SM618340072MA 010084068210398521SM618340072MA11180822 SM618460027MA 010084068210398521SM618460027MA11181112 SM618460028MA 010084068210398521SM618460028MA11181112 SM618430026MA 010084068210398521SM618430026MA11181026 SM618340062MA 010084068210398521SM618340062MA11180822 SM618340065MA 010084068210398521SM618340065MA11180822 SM618340071MA 010084068210398521SM618340071MA11180822 SM618340069MA 010084068210398521SM618340069MA11180822 SM618340064MA 010084068210398521SM618340064MA11180822 SM618340019MA 010084068210398521SM618340019MA11180821 SM618340022MA 010084068210398521SM618340022MA11180821 SM618410034MA 010084068210398521SM618410034MA11181011 SM619060071MA 010084068210398521SM619060071MA11190207 SM619060072MA 010084068210398521SM619060072MA11190207 SM619120015MA 010084068210398521SM619120015MA11190319 SM619230001MA 010084068210398521SM619230001MA11190603 SM618340018MA 010084068210398521SM618340018MA11180821 SM618340021MA 010084068210398521SM618340021MA11180821 SM618340023MA 010084068210398521SM618340023MA11180821 SM618340026MA 010084068210398521SM618340026MA11180821 SM618340027MA 010084068210398521SM618340027MA11180821 SM618340028MA 010084068210398521SM618340028MA11180821 SM618440006MA 010084068210398521SM618440006MA11181029 SM618450050MA 010084068210398521SM618450050MA11181109 SM618480024MA 010084068210398521SM618480024MA11181126 SM618480025MA 010084068210398521SM618480025MA11181126 SM619050022MA 010084068210398521SM619050022MA11190201 SM619120016MA 010084068210398521SM619120016MA11190319 SM619120017MA 010084068210398521SM619120017MA11190319 SM619180013MA 010084068210398521SM619180013MA11190429 SM619250031MA 010084068210398521SM619250031MA11190618 SM618340020MA 010084068210398521SM618340020MA11180821 SM618410033MA 010084068210398521SM618410033MA11181011 SM618440008MA 010084068210398521SM618440008MA11181029 SM618450049MA 010084068210398521SM618450049MA11181109 SM618460030MA 010084068210398521SM618460030MA11181112 SM618460031MA 010084068210398521SM618460031MA11181112 SM618460032MA 010084068210398521SM618460032MA11181112 SM619120013MA 010084068210398521SM619120013MA11190319 SM619120018MA 010084068210398521SM619120018MA11190319 SM619120019MA 010084068210398521SM619120019MA11190319 SM619120020MA 010084068210398521SM619120020MA11190319 SM619120021MA 010084068210398521SM619120021MA11190319 SM619120022MA 010084068210398521SM619120022MA11190319 SM618410037MA 010084068210398521SM618410037MA11181011 SM618450045MA 010084068210398521SM618450045MA11181109 SM618450046MA 010084068210398521SM618450046MA11181109 SM618450047MA 010084068210398521SM618450047MA11181109 SM618450048MA 010084068210398521SM618450048MA11181109 SM618460026MA 010084068210398521SM618460026MA11181112 SM619010104MA 010084068210398521SM619010104MA11190103 SM619150015MA 010084068210398521SM619150015MA11190409 SM619150016MA 010084068210398521SM619150016MA11190409 SM619160015MA 010084068210398521SM619160015MA11190418 SM618440007MA 010084068210398521SM618440007MA11181029 SM618340066MA 010084068210398521SM618340066MA11180822 SM618340070MA 010084068210398521SM618340070MA11180822 SM618440011MA 010084068210398521SM618440011MA11181029 SM618340024MA 010084068210398521SM618340024MA11180821 SM618340025MA 010084068210398521SM618340025MA11180821 SM618460029MA 010084068210398521SM618460029MA11181112 SM619010103MA 010084068210398521SM619010103MA11190103 SM618340073MA 010084068210398521SM618340073MA11180822 SM618360023MA 010084068210398521SM618360023MA11180906 SM618390021MA 010084068210398521SM618390021MA11180926 SM618390022MA 010084068210398521SM618390022MA11180926 SM618390023MA 010084068210398521SM618390023MA11180926 SM618390024MA 010084068210398521SM618390024MA11180926 SM618440010MA 010084068210398521SM618440010MA11181029 SM618440020MA 010084068210398521SM618440020MA11181031 SM618440021MA 010084068210398521SM618440021MA11181031 SM618480028MA 010084068210398521SM618480028MA11181126 SM618440005MA 010084068210398521SM618440005MA11181029 SM618480029MA 010084068210398521SM618480029MA11181126 CARESTATION 650 A1, Model Number 1012-9650-000 Mfg. Lot or Serial # UDI # SM718390004MA 010084068210394721SM718390004MA11180924 SM718390005MA 010084068210394721SM718390005MA11180924 SM718390006MA 010084068210394721SM718390006MA11180924 SM718390007MA 010084068210394721SM718390007MA11180924 SM718390008MA 010084068210394721SM718390008MA11180924 SM718390009MA 010084068210394721SM718390009MA11180924 SM718390010MA 010084068210394721SM718390010MA11180924 SM718390011MA 010084068210394721SM718390011MA11180924 SM718390012MA 010084068210394721SM718390012MA11180924 SM718390013MA 010084068210394721SM718390013MA11180924 SM718390014MA 010084068210394721SM718390014MA11180924 SM718390015MA 010084068210394721SM718390015MA11180924 SM718390016MA 010084068210394721SM718390016MA11180924 SM718390017MA 010084068210394721SM718390017MA11180924 SM718390018MA 010084068210394721SM718390018MA11180924 SM718390019MA 010084068210394721SM718390019MA11180924 SM718390020MA 010084068210394721SM718390020MA11180924 SM718390021MA 010084068210394721SM718390021MA11180924 SM718390022MA 010084068210394721SM718390022MA11180924 SM718390023MA 010084068210394721SM718390023MA11180924 SM718390024MA 010084068210394721SM718390024MA11180924 SM718390025MA 010084068210394721SM718390025MA11180924 SM718390026MA 010084068210394721SM718390026MA11180924 SM718390027MA 010084068210394721SM718390027MA11180924 SM718510007MA 010084068210394721SM718510007MA11181219 SM719060030MA 010084068210394721SM719060030MA11190208 SM719220050MA 010084068210394721SM719220050MA11190529 SM719060016MA 010084068210394721SM719060016MA11190206 SM719060017MA 010084068210394721SM719060017MA11190206 SM719060018MA 010084068210394721SM719060018MA11190206 SM719060019MA 010084068210394721SM719060019MA11190206 SM719060020MA 010084068210394721SM719060020MA11190206 SM719060021MA 010084068210394721SM719060021MA11190206 SM719060022MA 010084068210394721SM719060022MA11190206 SM719060023MA 010084068210394721SM719060023MA11190206 SM719060024MA 010084068210394721SM719060024MA11190206 SM719060025MA 010084068210394721SM719060025MA11190206 SM718420056MA 010084068210394721SM718420056MA11181018 SM719010114MA 010084068210394721SM719010114MA11190103 SM719010115MA 010084068210394721SM719010115MA11190103 SM719010116MA 010084068210394721SM719010116MA11190103 SM719010117MA 010084068210394721SM719010117MA11190103 SM719090043MA 010084068210394721SM719090043MA11190301 SM719090044MA 010084068210394721SM719090044MA11190301 SM719010104MA 010084068210394721SM719010104MA11190103 SM719010105MA 010084068210394721SM719010105MA11190103 SM719010106MA 010084068210394721SM719010106MA11190103 SM719010107MA 010084068210394721SM719010107MA11190103 SM719010108MA 010084068210394721SM719010108MA11190103 SM719010109MA 010084068210394721SM719010109MA11190103 SM719010110MA 010084068210394721SM719010110MA11190103 SM719010111MA 010084068210394721SM719010111MA11190103 SM719010112MA 010084068210394721SM719010112MA11190103 SM719010113MA 010084068210394721SM719010113MA11190103 SM719050013MA 010084068210394721SM719050013MA11190128 SM719050014MA 010084068210394721SM719050014MA11190128 SM719050015MA 010084068210394721SM719050015MA11190128 SM719050016MA 010084068210394721SM719050016MA11190128 SM718390051MA 010084068210394721SM718390051MA11180927 SM719230035MA 010084068210394721SM719230035MA11190607 SM718400017MA 010084068210394721SM718400017MA11181003 SM718400018MA 010084068210394721SM718400018MA11181003 SM719230017MA 010084068210394721SM719230017MA11190603 SM719230018MA 010084068210394721SM719230018MA11190603 SM719170020MA 010084068210394721SM719170020MA11190425 SM719170021MA 010084068210394721SM719170021MA11190425 SM718370072MA 010084068210394721SM718370072MA11180913 SM719170027MA 010084068210394721SM719170027MA11190426 SM719170028MA 010084068210394721SM719170028MA11190426 SM719170029MA 010084068210394721SM719170029MA11190426 SM719050010MA 010084068210394721SM719050010MA11190128 SM719050011MA 010084068210394721SM719050011MA11190128 SM719050012MA 010084068210394721SM719050012MA11190128 SM719240013MA 010084068210394721SM719240013MA11190611 SM719110028MA 010084068210394721SM719110028MA11190314 SM718500032MA 010084068210394721SM718500032MA11181213 SM718500033MA 010084068210394721SM718500033MA11181213 SM718500034MA 010084068210394721SM718500034MA11181213 SM718500035MA 010084068210394721SM718500035MA11181213 SM719030043MA 010084068210394721SM719030043MA11190117 SM719030044MA 010084068210394721SM719030044MA11190117 SM719030045MA 010084068210394721SM719030045MA11190117 SM719030046MA 010084068210394721SM719030046MA11190117 SM719130029MA 010084068210394721SM719130029MA11190328 SM719130030MA 010084068210394721SM719130030MA11190328 SM718340012MA 010084068210394721SM718340012MA11180822 SM719130027MA 010084068210394721SM719130027MA11190328 SM719130028MA 010084068210394721SM719130028MA11190328 SM719280016MA 010084068210394721SM719280016MA11190712 SM718470012MA 010084068210394721SM718470012MA11181119 SM718410034MA 010084068210394721SM718410034MA11181011 SM718500017MA 010084068210394721SM718500017MA11181211 SM718500018MA 010084068210394721SM718500018MA11181211 SM719140041MA 010084068210394721SM719140041MA11190404 SM718440019MA 010084068210394721SM718440019MA11181101 SM718440020MA 010084068210394721SM718440020MA11181101 SM718440021MA 010084068210394721SM718440021MA11181101 SM718420057MA 010084068210394721SM718420057MA11181018 SM718460041MA 010084068210394721SM718460041MA11181116 SM718460042MA 010084068210394721SM718460042MA11181116 SM718430045MA 010084068210394721SM718430045MA11181024 SM718430046MA 010084068210394721SM718430046MA11181024 SM718390059MA 010084068210394721SM718390059MA11180928 SM718390060MA 010084068210394721SM718390060MA11180928 SM718430002MA 010084068210394721SM718430002MA11181022 SM718430003MA 010084068210394721SM718430003MA11181022 SM719200020MA 010084068210394721SM719200020MA11190516 SM718370026MA 010084068210394721SM718370026MA11180912 SM718370027MA 010084068210394721SM718370027MA11180912 SM718370028MA 010084068210394721SM718370028MA11180912 SM718370029MA 010084068210394721SM718370029MA11180912 SM719020011MA 010084068210394721SM719020011MA11190108 SM718470013MA 010084068210394721SM718470013MA11181119 SM718470014MA 010084068210394721SM718470014MA11181119 SM719030019MA 010084068210394721SM719030019MA11190115 SM719030020MA 010084068210394721SM719030020MA11190115 SM719030021MA 010084068210394721SM719030021MA11190115 SM719030022MA 010084068210394721SM719030022MA11190115 SM719030023MA 010084068210394721SM719030023MA11190115 SM719030024MA 010084068210394721SM719030024MA11190115 SM719030025MA 010084068210394721SM719030025MA11190115 SM719030026MA 010084068210394721SM719030026MA11190115 SM719230012MA 010084068210394721SM719230012MA11190603 SM719230013MA 010084068210394721SM719230013MA11190603 SM719010128MA 010084068210394721SM719010128MA11190103 SM719220011MA 010084068210394721SM719220011MA11190527 SM719220012MA 010084068210394721SM719220012MA11190527 SM719100012MA 010084068210394721SM719100012MA11190305 SM719100013MA 010084068210394721SM719100013MA11190305 SM718470034MA 010084068210394721SM718470034MA11181120 SM718470035MA 010084068210394721SM718470035MA11181120 SM718470036MA 010084068210394721SM718470036MA11181120 SM718380042MA 010084068210394721SM718380042MA11180921 SM718390054MA 010084068210394721SM718390054MA11180927 SM718390055MA 010084068210394721SM718390055MA11180927 SM718390056MA 010084068210394721SM718390056MA11180927 SM718350008MA 010084068210394721SM718350008MA11180830 SM718350009MA 010084068210394721SM718350009MA11180830 SM718370044MA 010084068210394721SM718370044MA11180913 SM718370045MA 010084068210394721SM718370045MA11180913 SM718370046MA 010084068210394721SM718370046MA11180913 SM718370047MA 010084068210394721SM718370047MA11180913 SM718370048MA 010084068210394721SM718370048MA11180913 SM718370049MA 010084068210394721SM718370049MA11180913 SM718370050MA 010084068210394721SM718370050MA11180913 SM718370051MA 010084068210394721SM718370051MA11180913 SM718370052MA 010084068210394721SM718370052MA11180913 SM718370053MA 010084068210394721SM718370053MA11180913 SM718370054MA 010084068210394721SM718370054MA11180913 SM718370055MA 010084068210394721SM718370055MA11180913 SM718370056MA 010084068210394721SM718370056MA11180913 SM718370057MA 010084068210394721SM718370057MA11180913 SM718370058MA 010084068210394721SM718370058MA11180913 SM718370059MA 010084068210394721SM718370059MA11180913 SM718370060MA 010084068210394721SM718370060MA11180913 SM718370061MA 010084068210394721SM718370061MA11180913 SM718370062MA 010084068210394721SM718370062MA11180913 SM718370063MA 010084068210394721SM718370063MA11180913 SM718370064MA 010084068210394721SM718370064MA11180913 SM718370065MA 010084068210394721SM718370065MA11180913 SM718370066MA 010084068210394721SM718370066MA11180913 SM718370067MA 010084068210394721SM718370067MA11180913 SM718370068MA 010084068210394721SM718370068MA11180913 SM718370069MA 010084068210394721SM718370069MA11180913 SM718370070MA 010084068210394721SM718370070MA11180913 SM718370071MA 010084068210394721SM718370071MA11180913 SM718420011MA 010084068210394721SM718420011MA11181015 SM718420044MA 010084068210394721SM718420044MA11181018 SM718420045MA 010084068210394721SM718420045MA11181018 SM718420046MA 010084068210394721SM718420046MA11181018 SM718420047MA 010084068210394721SM718420047MA11181018 SM718420048MA 010084068210394721SM718420048MA11181018 SM718420049MA 010084068210394721SM718420049MA11181018 SM718420050MA 010084068210394721SM718420050MA11181018 SM718420051MA 010084068210394721SM718420051MA11181018 SM719100025MA 010084068210394721SM719100025MA11190306 SM718390001MA 010084068210394721SM718390001MA11180924 SM718390002MA 010084068210394721SM718390002MA11180924 SM718390003MA 010084068210394721SM718390003MA11180924 SM718480001MA 010084068210394721SM718480001MA11181126 SM719120013MA 010084068210394721SM719120013MA11190321 SM719120014MA 010084068210394721SM719120014MA11190321 SM719120015MA 010084068210394721SM719120015MA11190321 SM719120016MA 010084068210394721SM719120016MA11190321 SM719120017MA 010084068210394721SM719120017MA11190321 SM719120018MA 010084068210394721SM719120018MA11190321 SM719120019MA 010084068210394721SM719120019MA11190321 SM719120020MA 010084068210394721SM719120020MA11190321 SM719120021MA 010084068210394721SM719120021MA11190321 SM719170026MA 010084068210394721SM719170026MA11190426 SM718430010MA 010084068210394721SM718430010MA11181022 SM718430011MA 010084068210394721SM718430011MA11181022 SM719120035MA 010084068210394721SM719120035MA11190322 SM719180017MA 010084068210394721SM719180017MA11190501 SM719180018MA 010084068210394721SM719180018MA11190501 SM718390037MA 010084068210394721SM718390037MA11180924 SM718480062MA 010084068210394721SM718480062MA11181129 SM718480063MA 010084068210394721SM718480063MA11181129 SM718480064MA 010084068210394721SM718480064MA11181129 SM719260014MA 010084068210394721SM719260014MA11190627 SM719260015MA 010084068210394721SM719260015MA11190627 SM719260016MA 010084068210394721SM719260016MA11190627 SM718370036MA 010084068210394721SM718370036MA11180913 SM718380034MA 010084068210394721SM718380034MA11180920 SM718380035MA 010084068210394721SM718380035MA11180920 SM718380036MA 010084068210394721SM718380036MA11180920 SM718380037MA 010084068210394721SM718380037MA11180920 SM718380038MA 010084068210394721SM718380038MA11180920 SM718380039MA 010084068210394721SM718380039MA11180920 SM718380040MA 010084068210394721SM718380040MA11180920 SM718380043MA 010084068210394721SM718380043MA11180921 SM718380044MA 010084068210394721SM718380044MA11180921 SM718380045MA 010084068210394721SM718380045MA11180921 SM718380046MA 010084068210394721SM718380046MA11180921 SM718380047MA 010084068210394721SM718380047MA11180921 SM718380048MA 010084068210394721SM718380048MA11180921 SM718380049MA 010084068210394721SM718380049MA11180921 SM718380050MA 010084068210394721SM718380050MA11180921 SM718390028MA 010084068210394721SM718390028MA11180924 SM718390029MA 010084068210394721SM718390029MA11180924 SM718390030MA 010084068210394721SM718390030MA11180924 SM718390031MA 010084068210394721SM718390031MA11180924 SM718390032MA 010084068210394721SM718390032MA11180924 SM718390033MA 010084068210394721SM718390033MA11180924 SM718390034MA 010084068210394721SM718390034MA11180924 SM718440017MA 010084068210394721SM718440017MA11181101 SM718440018MA 010084068210394721SM718440018MA11181101 SM719150082MA 010084068210394721SM719150082MA11190412 SM719150083MA 010084068210394721SM719150083MA11190412 SM719150084MA 010084068210394721SM719150084MA11190412 SM719150085MA 010084068210394721SM719150085MA11190412 SM719160014MA 010084068210394721SM719160014MA11190418 SM719160015MA 010084068210394721SM719160015MA11190418 SM719160016MA 010084068210394721SM719160016MA11190418 SM719250008MA 010084068210394721SM719250008MA11190619 SM718390039MA 010084068210394721SM718390039MA11180925 SM718390040MA 010084068210394721SM718390040MA11180925 SM718400025MA 010084068210394721SM718400025MA11181005 SM718480043MA 010084068210394721SM718480043MA11181128 SM718480044MA 010084068210394721SM718480044MA11181128 SM718480045MA 010084068210394721SM718480045MA11181128 SM718480046MA 010084068210394721SM718480046MA11181128 SM719240011MA 010084068210394721SM719240011MA11190611 SM719240012MA 010084068210394721SM719240012MA11190611 SM719180004MA 010084068210394721SM719180004MA11190429 SM719180005MA 010084068210394721SM719180005MA11190429 SM719190004MA 010084068210394721SM719190004MA11190506 SM719200021MA 010084068210394721SM719200021MA11190516 SM719200022MA 010084068210394721SM719200022MA11190516 SM719200023MA 010084068210394721SM719200023MA11190516 SM719200024MA 010084068210394721SM719200024MA11190516 SM719060026MA 010084068210394721SM719060026MA11190206 SM719160012MA 010084068210394721SM719160012MA11190418 SM719160013MA 010084068210394721SM719160013MA11190418 SM719020032MA 010084068210394721SM719020032MA11190110 SM719020033MA 010084068210394721SM719020033MA11190110 SM719180001MA 010084068210394721SM719180001MA11190429 SM719180002MA 010084068210394721SM719180002MA11190429 SM719180003MA 010084068210394721SM719180003MA11190429 SM719170025MA 010084068210394721SM719170025MA11190425 SM718410008MA 010084068210394721SM718410008MA11181008 SM719150034MA 010084068210394721SM719150034MA11190409 SM718490010MA 010084068210394721SM718490010MA11181203 SM718490011MA 010084068210394721SM718490011MA11181203 SM719170030MA 010084068210394721SM719170030MA11190426 SM719170031MA 010084068210394721SM719170031MA11190426 SM718480017MA 010084068210394721SM718480017MA11181127 SM719110013MA 010084068210394721SM719110013MA11190312 SM719110014MA 010084068210394721SM719110014MA11190312 SM719250009MA 010084068210394721SM719250009MA11190619 SM718500022MA 010084068210394721SM718500022MA11181212 SM718500023MA 010084068210394721SM718500023MA11181212 SM718500024MA 010084068210394721SM718500024MA11181212 SM718500025MA 010084068210394721SM718500025MA11181212 SM718500026MA 010084068210394721SM718500026MA11181212 SM718500027MA 010084068210394721SM718500027MA11181212 SM718500028MA 010084068210394721SM718500028MA11181212 SM718500029MA 010084068210394721SM718500029MA11181212 SM718500030MA 010084068210394721SM718500030MA11181212 SM718500031MA 010084068210394721SM718500031MA11181212 SM718510018MA 010084068210394721SM718510018MA11181220 SM718510019MA 010084068210394721SM718510019MA11181220 SM718370034MA 010084068210394721SM718370034MA11180913 SM719250013MA 010084068210394721SM719250013MA11190621 SM719250014MA 010084068210394721SM719250014MA11190621 SM719250015MA 010084068210394721SM719250015MA11190621 SM719250016MA 010084068210394721SM719250016MA11190621 SM719250017MA 010084068210394721SM719250017MA11190621 SM718420054MA 010084068210394721SM718420054MA11181018 SM718480037MA 010084068210394721SM718480037MA11181128 SM718480038MA 010084068210394721SM718480038MA11181128 SM718480039MA 010084068210394721SM718480039MA11181128 SM718480040MA 010084068210394721SM718480040MA11181128 SM718480041MA 010084068210394721SM718480041MA11181128 SM718480042MA 010084068210394721SM718480042MA11181128 SM719260007MA 010084068210394721SM719260007MA11190625 SM719050029MA 010084068210394721SM719050029MA11190129 SM719050030MA 010084068210394721SM719050030MA11190129 SM719050031MA 010084068210394721SM719050031MA11190129 SM719100023MA 010084068210394721SM719100023MA11190306 SM719100024MA 010084068210394721SM719100024MA11190306 SM719260011MA 010084068210394721SM719260011MA11190627 SM719260012MA 010084068210394721SM719260012MA11190627 SM719260013MA 010084068210394721SM719260013MA11190627 SM719020030MA 010084068210394721SM719020030MA11190110 SM719020031MA 010084068210394721SM719020031MA11190110 SM719120005MA 010084068210394721SM719120005MA11190319 SM719020038MA 010084068210394721SM719020038MA11190111 SM719020039MA 010084068210394721SM719020039MA11190111 SM719020040MA 010084068210394721SM719020040MA11190111 SM718500019MA 010084068210394721SM718500019MA11181211 SM719050028MA 010084068210394721SM719050028MA11190129 SM719020037MA 010084068210394721SM719020037MA11190111 SM719230014MA 010084068210394721SM719230014MA11190603 SM719230015MA 010084068210394721SM719230015MA11190603 SM718430044MA 010084068210394721SM718430044MA11181024 SM719230016MA 010084068210394721SM719230016MA11190603 SM718510035MA 010084068210394721SM718510035MA11181221 SM718510036MA 010084068210394721SM718510036MA11181221 SM718510037MA 010084068210394721SM718510037MA11181221 SM718510038MA 010084068210394721SM718510038MA11181221 SM718510039MA 010084068210394721SM718510039MA11181221 SM718510040MA 010084068210394721SM718510040MA11181221 SM718510041MA 010084068210394721SM718510041MA11181221 SM718510042MA 010084068210394721SM718510042MA11181221 SM718510043MA 010084068210394721SM718510043MA11181221 SM718510044MA 010084068210394721SM718510044MA11181221 SM718510045MA 010084068210394721SM718510045MA11181221 SM718510046MA 010084068210394721SM718510046MA11181221 SM719150063MA 010084068210394721SM719150063MA11190411 SM719240001MA 010084068210394721SM719240001MA11190610 SM719240002MA 010084068210394721SM719240002MA11190610 SM719260017MA 010084068210394721SM719260017MA11190628 SM718390057MA 010084068210394721SM718390057MA11180927 SM718390058MA 010084068210394721SM718390058MA11180927 SM718500020MA 010084068210394721SM718500020MA11181211 SM718500021MA 010084068210394721SM718500021MA11181211 SM718390036MA 010084068210394721SM718390036MA11180924 SM718390038MA 010084068210394721SM718390038MA11180924 SM718430012MA 010084068210394721SM718430012MA11181022 SM718430013MA 010084068210394721SM718430013MA11181022 SM718430014MA 010084068210394721SM718430014MA11181022 SM718370039MA 010084068210394721SM718370039MA11180913 SM718410035MA 010084068210394721SM718410035MA11181011 SM718400002MA 010084068210394721SM718400002MA11181002 SM718400003MA 010084068210394721SM718400003MA11181002 SM719110018MA 010084068210394721SM719110018MA11190313 SM719110019MA 010084068210394721SM719110019MA11190313 SM719110020MA 010084068210394721SM719110020MA11190313 SM719110021MA 010084068210394721SM719110021MA11190313 SM719110022MA 010084068210394721SM719110022MA11190313 SM719110023MA 010084068210394721SM719110023MA11190313 SM719110024MA 010084068210394721SM719110024MA11190313 SM719110025MA 010084068210394721SM719110025MA11190313 SM719110026MA 010084068210394721SM719110026MA11190313 SM719150009MA 010084068210394721SM719150009MA11190408 SM719150010MA 010084068210394721SM719150010MA11190408 SM719150011MA 010084068210394721SM719150011MA11190408 SM719150012MA 010084068210394721SM719150012MA11190408 SM719150013MA 010084068210394721SM719150013MA11190408 SM719150014MA 010084068210394721SM719150014MA11190408 SM719150015MA 010084068210394721SM719150015MA11190408 SM719020021MA 010084068210394721SM719020021MA11190109 SM719020022MA 010084068210394721SM719020022MA11190109 SM719040016MA 010084068210394721SM719040016MA11190122 SM719100007MA 010084068210394721SM719100007MA11190304 SM719060027MA 010084068210394721SM719060027MA11190206 SM719060028MA 010084068210394721SM719060028MA11190206 SM719060029MA 010084068210394721SM719060029MA11190206 SM718370025MA 010084068210394721SM718370025MA11180912 SM718420053MA 010084068210394721SM718420053MA11181018 SM718420055MA 010084068210394721SM718420055MA11181018 SM718440022MA 010084068210394721SM718440022MA11181101 SM718440023MA 010084068210394721SM718440023MA11181101 SM718440024MA 010084068210394721SM718440024MA11181101 SM718440025MA 010084068210394721SM718440025MA11181101 SM718440026MA 010084068210394721SM718440026MA11181101 SM718480010MA 010084068210394721SM718480010MA11181127 SM718480011MA 010084068210394721SM718480011MA11181127 SM718480012MA 010084068210394721SM718480012MA11181127 SM718480013MA 010084068210394721SM718480013MA11181127 SM718480014MA 010084068210394721SM718480014MA11181127 SM718480015MA 010084068210394721SM718480015MA11181127 SM718480016MA 010084068210394721SM718480016MA11181127 SM718480018MA 010084068210394721SM718480018MA11181127 SM718480019MA 010084068210394721SM718480019MA11181127 SM719030018MA 010084068210394721SM719030018MA11190115 SM719060001MA 010084068210394721SM719060001MA11190204 SM719060002MA 010084068210394721SM719060002MA11190204 SM719060003MA 010084068210394721SM719060003MA11190204 SM719060004MA 010084068210394721SM719060004MA11190204 SM719060005MA 010084068210394721SM719060005MA11190204 SM719150035MA 010084068210394721SM719150035MA11190409 SM719150036MA 010084068210394721SM719150036MA11190409 SM719150037MA 010084068210394721SM719150037MA11190409 SM719150038MA 010084068210394721SM719150038MA11190409 SM718510020MA 010084068210394721SM718510020MA11181220 SM718510021MA 010084068210394721SM718510021MA11181220 SM719210018MA 010084068210394721SM719210018MA11190521 SM718510022MA 010084068210394721SM718510022MA11181220 SM718510023MA 010084068210394721SM718510023MA11181220 SM718510024MA 010084068210394721SM718510024MA11181220 SM718510025MA 010084068210394721SM718510025MA11181220 SM718510026MA 010084068210394721SM718510026MA11181220 SM718510027MA 010084068210394721SM718510027MA11181220 SM718370035MA 010084068210394721SM718370035MA11180913 SM719130002WA 010084068210394721SM719130002WA11190327 SM719130003WA 010084068210394721SM719130003WA11190327 SM718430008MA 010084068210394721SM718430008MA11181022 SM718430009MA 010084068210394721SM718430009MA11181022 SM718430004MA 010084068210394721SM718430004MA11181022 SM718430005MA 010084068210394721SM718430005MA11181022 SM718430006MA 010084068210394721SM718430006MA11181022 SM718430007MA 010084068210394721SM718430007MA11181022 SM719110003MA 010084068210394721SM719110003MA11190311 SM719110004MA 010084068210394721SM719110004MA11190311 SM718430048WA 010084068210394721SM718430048WA11181025 SM718430049WA 010084068210394721SM718430049WA11181025 SM718430050WA 010084068210394721SM718430050WA11181025 SM718430051WA 010084068210394721SM718430051WA11181025 SM718440013WA 010084068210394721SM718440013WA11181030 SM719190008WA 010084068210394721SM719190008WA11190507 SM719190009WA 010084068210394721SM719190009WA11190507 SM719190010WA 010084068210394721SM719190010WA11190507 SM719090034WA 010084068210394721SM719090034WA11190228 SM719210037WA 010084068210394721SM719210037WA11190524 SM718460004WA 010084068210394721SM718460004WA11181112 SM719090030WA 010084068210394721SM719090030WA11190228 SM719190033WA 010084068210394721SM719190033WA11190510 SM718350003WA 010084068210394721SM718350003WA11180830 SM718460002WA 010084068210394721SM718460002WA11181112 SM718460003WA 010084068210394721SM718460003WA11181112 SM718410025WA 010084068210394721SM718410025WA11181010 SM718340022WA 010084068210394721SM718340022WA11180823 SM718340023WA 010084068210394721SM718340023WA11180823 SM718340024WA 010084068210394721SM718340024WA11180823 SM719090029WA 010084068210394721SM719090029WA11190228 SM719210036WA 010084068210394721SM719210036WA11190524 SM719210038WA 010084068210394721SM719210038WA11190524 SM719090031WA 010084068210394721SM719090031WA11190228 SM719200012WA 010084068210394721SM719200012WA11190516 SM718430054WA 010084068210394721SM718430054WA11181026 SM719090032WA 010084068210394721SM719090032WA11190228 SM719090033WA 010084068210394721SM719090033WA11190228 SM718410031WA 010084068210394721SM718410031WA11181011 SM718410032WA 010084068210394721SM718410032WA11181011 SM718410033WA 010084068210394721SM718410033WA11181011 SM718400016MA 010084068210394721SM718400016MA11181002 SM718410029WA 010084068210394721SM718410029WA11181011 SM718410030WA 010084068210394721SM718410030WA11181011 SM718480006WA 010084068210394721SM718480006WA11181127 SM718480007WA 010084068210394721SM718480007WA11181127 SM718480008WA 010084068210394721SM718480008WA11181127 SM719090001WA 010084068210394721SM719090001WA11190225 SM719150046WA 010084068210394721SM719150046WA11190410 SM719160019WA 010084068210394721SM719160019WA11190420 SM719190005WA 010084068210394721SM719190005WA11190507 SM719190006WA 010084068210394721SM719190006WA11190507 SM719230029WA 010084068210394721SM719230029WA11190605 SM718360040MA 010084068210394721SM718360040MA11180905 SM718480002MA 010084068210394721SM718480002MA11181126 SM718480003MA 010084068210394721SM718480003MA11181126 SM719010101MA 010084068210394721SM719010101MA11190103 SM719010102MA 010084068210394721SM719010102MA11190103 SM718440014MA 010084068210394721SM718440014MA11181031 SM718440015MA 010084068210394721SM718440015MA11181031 SM719040035WA 010084068210394721SM719040035WA11190124 SM719100010WA 010084068210394721SM719100010WA11190305 SM718480028WA 010084068210394721SM718480028WA11181128 SM718480029WA 010084068210394721SM718480029WA11181128 SM718480030WA 010084068210394721SM718480030WA11181128 SM719040038WA 010084068210394721SM719040038WA11190124 SM718420021WA 010084068210394721SM718420021WA11181017 SM718420022WA 010084068210394721SM718420022WA11181017 SM718420023WA 010084068210394721SM718420023WA11181017 SM718420024WA 010084068210394721SM718420024WA11181017 SM718420025WA 010084068210394721SM718420025WA11181017 SM718420026WA 010084068210394721SM718420026WA11181017 SM718420027WA 010084068210394721SM718420027WA11181017 SM718420028WA 010084068210394721SM718420028WA11181017 SM718420029WA 010084068210394721SM718420029WA11181017 SM719040036WA 010084068210394721SM719040036WA11190124 SM719040037WA 010084068210394721SM719040037WA11190124 SM719190023WA 010084068210394721SM719190023WA11190509 SM719190024WA 010084068210394721SM719190024WA11190509 SM719190025WA 010084068210394721SM719190025WA11190509 SM719190026WA 010084068210394721SM719190026WA11190509 SM719190027WA 010084068210394721SM719190027WA11190509 SM719190028WA 010084068210394721SM719190028WA11190509 SM719190029WA 010084068210394721SM719190029WA11190509 SM719190030WA 010084068210394721SM719190030WA11190510 SM719190031WA 010084068210394721SM719190031WA11190510 SM719110015MA 010084068210394721SM719110015MA11190312 SM719110016MA 010084068210394721SM719110016MA11190312 SM718350010MA 010084068210394721SM718350010MA11180830 SM718350011MA 010084068210394721SM718350011MA11180830 SM718350012MA 010084068210394721SM718350012MA11180830 SM718350013MA 010084068210394721SM718350013MA11180830 SM718350014MA 010084068210394721SM718350014MA11180830 SM718350015MA 010084068210394721SM718350015MA11180830 SM718350016MA 010084068210394721SM718350016MA11180830 SM718350017MA 010084068210394721SM718350017MA11180830 SM718350018MA 010084068210394721SM718350018MA11180830 SM718350019MA 010084068210394721SM718350019MA11180830 SM718350020MA 010084068210394721SM718350020MA11180830 SM718350021MA 010084068210394721SM718350021MA11180830 SM718350022MA 010084068210394721SM718350022MA11180830 SM718350023MA 010084068210394721SM718350023MA11180830 SM718350024MA 010084068210394721SM718350024MA11180830 SM718350025MA 010084068210394721SM718350025MA11180830 SM718350026MA 010084068210394721SM718350026MA11180830 SM718350027MA 010084068210394721SM718350027MA11180830 SM718350028MA 010084068210394721SM718350028MA11180830 SM718460043MA 010084068210394721SM718460043MA11181116 SM718460044MA 010084068210394721SM718460044MA11181116 SM718460045MA 010084068210394721SM718460045MA11181116 SM718460046MA 010084068210394721SM718460046MA11181116 SM718460047MA 010084068210394721SM718460047MA11181116 SM718520021MA 010084068210394721SM718520021MA11181228 SM718520022MA 010084068210394721SM718520022MA11181228 SM719070039MA 010084068210394721SM719070039MA11190213 SM719150040MA 010084068210394721SM719150040MA11190409 SM719150041MA 010084068210394721SM719150041MA11190409 SM719150042MA 010084068210394721SM719150042MA11190409 SM719150043MA 010084068210394721SM719150043MA11190409 SM719150044MA 010084068210394721SM719150044MA11190409 SM718350029MA 010084068210394721SM718350029MA11180830 SM718340033MA 010084068210394721SM718340033MA11180824 SM718340034MA 010084068210394721SM718340034MA11180824 SM718340035MA 010084068210394721SM718340035MA11180824 SM718340036MA 010084068210394721SM718340036MA11180824 SM718340037MA 010084068210394721SM718340037MA11180824 SM718340038MA 010084068210394721SM718340038MA11180824 SM718340039MA 010084068210394721SM718340039MA11180824 SM718340040MA 010084068210394721SM718340040MA11180824 SM718340041MA 010084068210394721SM718340041MA11180824 SM718340042MA 010084068210394721SM718340042MA11180824 SM718370022MA 010084068210394721SM718370022MA11180912 SM718370023MA 010084068210394721SM718370023MA11180912 SM719220010MA 010084068210394721SM719220010MA11190527 SM719270023MA 010084068210394721SM719270023MA11190702 SM719270024MA 010084068210394721SM719270024MA11190702 SM719270025MA 010084068210394721SM719270025MA11190702 SM719270026MA 010084068210394721SM719270026MA11190702 SM719270027MA 010084068210394721SM719270027MA11190702 SM719270028MA 010084068210394721SM719270028MA11190702 SM719270029MA 010084068210394721SM719270029MA11190702 SM719270030MA 010084068210394721SM719270030MA11190702 SM719120006MA 010084068210394721SM719120006MA11190319 SM719120007MA 010084068210394721SM719120007MA11190319 SM719120008MA 010084068210394721SM719120008MA11190319 SM719120009MA 010084068210394721SM719120009MA11190319 SM719120010MA 010084068210394721SM719120010MA11190319 SM718450065MA 010084068210394721SM718450065MA11181109 SM718450066MA 010084068210394721SM718450066MA11181109 SM719200029MA 010084068210394721SM719200029MA11190516 SM718370016MA 010084068210394721SM718370016MA11180912 SM718370017MA 010084068210394721SM718370017MA11180912 SM718370018MA 010084068210394721SM718370018MA11180912 SM718370019MA 010084068210394721SM718370019MA11180912 SM718370020MA 010084068210394721SM718370020MA11180912 SM718370021MA 010084068210394721SM718370021MA11180912 SM718480047MA 010084068210394721SM718480047MA11181128 SM718480048MA 010084068210394721SM718480048MA11181128 SM718480049MA 010084068210394721SM718480049MA11181128 SM718480050MA 010084068210394721SM718480050MA11181128 SM718480051MA 010084068210394721SM718480051MA11181128 SM718480052MA 010084068210394721SM718480052MA11181128 SM718480053MA 010084068210394721SM718480053MA11181128 SM718480054MA 010084068210394721SM718480054MA11181128 SM719060006MA 010084068210394721SM719060006MA11190204 SM719060007MA 010084068210394721SM719060007MA11190204 SM719060008MA 010084068210394721SM719060008MA11190204 SM719060009MA 010084068210394721SM719060009MA11190204 SM719060010MA 010084068210394721SM719060010MA11190204 SM719060011MA 010084068210394721SM719060011MA11190204 SM718450060MA 010084068210394721SM718450060MA11181109 SM718450061MA 010084068210394721SM718450061MA11181109 SM718480036MA 010084068210394721SM718480036MA11181128 SM718370008MA 010084068210394721SM718370008MA11180910 SM718370010MA 010084068210394721SM718370010MA11180910 SM718480035MA 010084068210394721SM718480035MA11181128 SM719100038MA 010084068210394721SM719100038MA11190308 SM719170022MA 010084068210394721SM719170022MA11190425 SM719170023MA 010084068210394721SM719170023MA11190425 SM719170024MA 010084068210394721SM719170024MA11190425 SM718520019MA 010084068210394721SM718520019MA11181227 SM718340025MA 010084068210394721SM718340025MA11180823 SM718400001MA 010084068210394721SM718400001MA11181001 SM718480032MA 010084068210394721SM718480032MA11181128 SM718480033MA 010084068210394721SM718480033MA11181128 SM719240016MA 010084068210394721SM719240016MA11190612 SM719240017MA 010084068210394721SM719240017MA11190612 SM718480034MA 010084068210394721SM718480034MA11181128 SM718460028MA 010084068210394721SM718460028MA11181114 SM719060014MA 010084068210394721SM719060014MA11190206 SM719060015MA 010084068210394721SM719060015MA11190206 SM718370009MA 010084068210394721SM718370009MA11180910 SM719230036MA 010084068210394721SM719230036MA11190607 SM719230037MA 010084068210394721SM719230037MA11190607 SM718370024MA 010084068210394721SM718370024MA11180912 SM718380014MA 010084068210394721SM718380014MA11180920 SM718380015MA 010084068210394721SM718380015MA11180920 SM718380016MA 010084068210394721SM718380016MA11180920 SM718380017MA 010084068210394721SM718380017MA11180920 SM718380018MA 010084068210394721SM718380018MA11180920 SM718380019MA 010084068210394721SM718380019MA11180920 SM718380020MA 010084068210394721SM718380020MA11180920 SM718380021MA 010084068210394721SM718380021MA11180920 SM718380022MA 010084068210394721SM718380022MA11180920 SM718380023MA 010084068210394721SM718380023MA11180920 SM719210019MA 010084068210394721SM719210019MA11190521 SM719210020MA 010084068210394721SM719210020MA11190522 SM719210021MA 010084068210394721SM719210021MA11190522 SM719210022MA 010084068210394721SM719210022MA11190522 SM719160010MA 010084068210394721SM719160010MA11190418 SM719160011MA 010084068210394721SM719160011MA11190418 SM719200025MA 010084068210394721SM719200025MA11190516 SM719200026MA 010084068210394721SM719200026MA11190516 SM719200027MA 010084068210394721SM719200027MA11190516 SM719200028MA 010084068210394721SM719200028MA11190516 SM719060012MA 010084068210394721SM719060012MA11190206 SM719060013MA 010084068210394721SM719060013MA11190206 SM718370007MA 010084068210394721SM718370007MA11180910 SM718450025MA 010084068210394721SM718450025MA11181107 SM718450042WA 010084068210394721SM718450042WA11181109 SM718450043WA 010084068210394721SM718450043WA11181109 SM718450044WA 010084068210394721SM718450044WA11181109 SM718450045WA 010084068210394721SM718450045WA11181109 SM718450046WA 010084068210394721SM718450046WA11181109 SM718450047WA 010084068210394721SM718450047WA11181109 SM718450048WA 010084068210394721SM718450048WA11181109 SM718450049WA 010084068210394721SM718450049WA11181109 SM718450050WA 010084068210394721SM718450050WA11181109 SM718450051WA 010084068210394721SM718450051WA11181109 SM719050022WA 010084068210394721SM719050022WA11190129 SM719050023WA 010084068210394721SM719050023WA11190129 SM719050024WA 010084068210394721SM719050024WA11190129 SM719050025WA 010084068210394721SM719050025WA11190129 SM719050026WA 010084068210394721SM719050026WA11190129 SM719050027WA 010084068210394721SM719050027WA11190129 SM719050032WA 010084068210394721SM719050032WA11190130 SM719050033WA 010084068210394721SM719050033WA11190130 SM719050034WA 010084068210394721SM719050034WA11190130 SM719050035WA 010084068210394721SM719050035WA11190130 SM719190011WA 010084068210394721SM719190011WA11190507 SM719190012WA 010084068210394721SM719190012WA11190507 SM719190013WA 010084068210394721SM719190013WA11190508 SM719190014WA 010084068210394721SM719190014WA11190508 SM719190015WA 010084068210394721SM719190015WA11190508 SM719190016WA 010084068210394721SM719190016WA11190508 SM719190017WA 010084068210394721SM719190017WA11190508 SM719190018WA 010084068210394721SM719190018WA11190508 SM719190019WA 010084068210394721SM719190019WA11190508 SM719190020WA 010084068210394721SM719190020WA11190508 SM719040017MA 010084068210394721SM719040017MA11190122 SM719040018MA 010084068210394721SM719040018MA11190122 SM719040019MA 010084068210394721SM719040019MA11190122 SM719040020MA 010084068210394721SM719040020MA11190122 SM719140020MA 010084068210394721SM719140020MA11190403 SM719130010MA 010084068210394721SM719130010MA11190327 SM719130013MA 010084068210394721SM719130013MA11190327 SM719150058MA 010084068210394721SM719150058MA11190411 SM719150059MA 010084068210394721SM719150059MA11190411 SM719150060MA 010084068210394721SM719150060MA11190411 SM719150061MA 010084068210394721SM719150061MA11190411 SM719130011MA 010084068210394721SM719130011MA11190327 SM719130012MA 010084068210394721SM719130012MA11190327 SM719140022MA 010084068210394721SM719140022MA11190403 SM719140023MA 010084068210394721SM719140023MA11190403 SM719140024MA 010084068210394721SM719140024MA11190403 SM719140025MA 010084068210394721SM719140025MA11190403 SM719140026MA 010084068210394721SM719140026MA11190403 SM719140027MA 010084068210394721SM719140027MA11190403 SM719140028MA 010084068210394721SM719140028MA11190403 SM719140029MA 010084068210394721SM719140029MA11190403 SM719140030MA 010084068210394721SM719140030MA11190403 SM719140031MA 010084068210394721SM719140031MA11190403 SM719140032MA 010084068210394721SM719140032MA11190403 SM719140033MA 010084068210394721SM719140033MA11190403 SM719150062MA 010084068210394721SM719150062MA11190411 SM719150064MA 010084068210394721SM719150064MA11190411 SM719150065MA 010084068210394721SM719150065MA11190411 SM719150066MA 010084068210394721SM719150066MA11190411 SM719150067MA 010084068210394721SM719150067MA11190411 SM719160001MA 010084068210394721SM719160001MA11190415 SM719160002MA 010084068210394721SM719160002MA11190415 SM719160003MA 010084068210394721SM719160003MA11190415 SM719160004MA 010084068210394721SM719160004MA11190415 SM719160005MA 010084068210394721SM719160005MA11190415 SM719160006MA 010084068210394721SM719160006MA11190415 SM719160007MA 010084068210394721SM719160007MA11190415 SM719160008MA 010084068210394721SM719160008MA11190415 SM719160009MA 010084068210394721SM719160009MA11190415 SM719220001MA 010084068210394721SM719220001MA11190527 SM719220002MA 010084068210394721SM719220002MA11190527 SM719220003MA 010084068210394721SM719220003MA11190527 SM719220004MA 010084068210394721SM719220004MA11190527 SM719220005MA 010084068210394721SM719220005MA11190527 SM719220006MA 010084068210394721SM719220006MA11190527 SM719220007MA 010084068210394721SM719220007MA11190527 SM719220008MA 010084068210394721SM719220008MA11190527 SM719220009MA 010084068210394721SM719220009MA11190527 SM719280017MA 010084068210394721SM719280017MA11190712 SM719280018MA 010084068210394721SM719280018MA11190712 SM719280019MA 010084068210394721SM719280019MA11190712 SM719280020MA 010084068210394721SM719280020MA11190712 SM719280021MA 010084068210394721SM719280021MA11190712 SM719140019MA 010084068210394721SM719140019MA11190403 SM719140021MA 010084068210394721SM719140021MA11190403 SM718440030WA 010084068210394721SM718440030WA11181102 SM719110017WA 010084068210394721SM719110017WA11190313 SM718390041WA 010084068210394721SM718390041WA11180926 SM718420003WA 010084068210394721SM718420003WA11181015 SM718510033WA 010084068210394721SM718510033WA11181221 SM719150025WA 010084068210394721SM719150025WA11190409 SM719190032WA 010084068210394721SM719190032WA11190510 SM718350006WA 010084068210394721SM718350006WA11180830 SM718410009WA 010084068210394721SM718410009WA11181009 SM718430021WA 010084068210394721SM718430021WA11181023 SM718430022WA 010084068210394721SM718430022WA11181023 SM718430023WA 010084068210394721SM718430023WA11181023 SM718430024WA 010084068210394721SM718430024WA11181023 SM718430034WA 010084068210394721SM718430034WA11181024 SM718430035WA 010084068210394721SM718430035WA11181024 SM718430036WA 010084068210394721SM718430036WA11181024 SM718430037WA 010084068210394721SM718430037WA11181024 SM718430038WA 010084068210394721SM718430038WA11181024 SM718430042WA 010084068210394721SM718430042WA11181024 SM718430043WA 010084068210394721SM718430043WA11181024 SM718440031WA 010084068210394721SM718440031WA11181102 SM718440032WA 010084068210394721SM718440032WA11181102 SM718450010WA 010084068210394721SM718450010WA11181106 SM718450012WA 010084068210394721SM718450012WA11181107 SM718450013WA 010084068210394721SM718450013WA11181107 SM718450014WA 010084068210394721SM718450014WA11181107 SM718450015WA 010084068210394721SM718450015WA11181107 SM718450016WA 010084068210394721SM718450016WA11181107 SM718450017WA 010084068210394721SM718450017WA11181107 SM718450018WA 010084068210394721SM718450018WA11181107 SM718470048WA 010084068210394721SM718470048WA11181122 SM718370004WA 010084068210394721SM718370004WA11180910 SM719040001WA 010084068210394721SM719040001WA11190121 SM719040002WA 010084068210394721SM719040002WA11190121 SM718480061WA 010084068210394721SM718480061WA11181129 SM719020035WA 010084068210394721SM719020035WA11190111 SM719020036WA 010084068210394721SM719020036WA11190111 SM719020018WA 010084068210394721SM719020018WA11190109 SM719020019WA 010084068210394721SM719020019WA11190109 SM719140015WA 010084068210394721SM719140015WA11190403 SM718470058WA 010084068210394721SM718470058WA11181122 SM719220027WA 010084068210394721SM719220027WA11190528 SM719020005WA 010084068210394721SM719020005WA11190107 SM719020006WA 010084068210394721SM719020006WA11190107 SM719220057WA 010084068210394721SM719220057WA11190531 SM719170018WA 010084068210394721SM719170018WA11190425 SM719170019WA 010084068210394721SM719170019WA11190425 SM719170017WA 010084068210394721SM719170017WA11190425 SM719090002WA 010084068210394721SM719090002WA11190225 SM718470059WA 010084068210394721SM718470059WA11181122 SM718480005WA 010084068210394721SM718480005WA11181127 SM719100009WA 010084068210394721SM719100009WA11190305 SM718370030WA 010084068210394721SM718370030WA11180913 SM719150026WA 010084068210394721SM719150026WA11190409 SM719090003WA 010084068210394721SM719090003WA11190225 SM719090004WA 010084068210394721SM719090004WA11190225 SM719090005WA 010084068210394721SM719090005WA11190225 SM719120001WA 010084068210394721SM719120001WA11190318 SM719120002WA 010084068210394721SM719120002WA11190318 SM719080047WA 010084068210394721SM719080047WA11190222 SM719080048WA 010084068210394721SM719080048WA11190222 SM719080049WA 010084068210394721SM719080049WA11190222 SM719080050WA 010084068210394721SM719080050WA11190222 SM719080051WA 010084068210394721SM719080051WA11190222 SM719040003WA 010084068210394721SM719040003WA11190121 SM719040004WA 010084068210394721SM719040004WA11190121 SM718450019WA 010084068210394721SM718450019WA11181107 SM719160020WA 010084068210394721SM719160020WA11190420 SM718360050WA 010084068210394721SM718360050WA11180906 SM718360051WA 010084068210394721SM718360051WA11180906 SM718450008WA 010084068210394721SM718450008WA11181106 SM718360005WA 010084068210394721SM718360005WA11180903 SM718360006WA 010084068210394721SM718360006WA11180904 SM718360007WA 010084068210394721SM718360007WA11180904 SM718360008WA 010084068210394721SM718360008WA11180904 SM718360009WA 010084068210394721SM718360009WA11180904 SM718360010WA 010084068210394721SM718360010WA11180904 SM718360011WA 010084068210394721SM718360011WA11180904 SM718360012WA 010084068210394721SM718360012WA11180904 SM718360013WA 010084068210394721SM718360013WA11180904 SM718360014WA 010084068210394721SM718360014WA11180904 SM718360015WA 010084068210394721SM718360015WA11180904 SM718360016WA 010084068210394721SM718360016WA11180904 SM718360029WA 010084068210394721SM718360029WA11180905 SM718370001WA 010084068210394721SM718370001WA11180910 SM718460021WA 010084068210394721SM718460021WA11181113 SM718460022WA 010084068210394721SM718460022WA11181113 SM718480009WA 010084068210394721SM718480009WA11181127 SM718480022WA 010084068210394721SM718480022WA11181128 SM718490038WA 010084068210394721SM718490038WA11181207 SM718490039WA 010084068210394721SM718490039WA11181207 SM718490040WA 010084068210394721SM718490040WA11181207 SM719140017WA 010084068210394721SM719140017WA11190403 SM719150027WA 010084068210394721SM719150027WA11190409 SM719080044WA 010084068210394721SM719080044WA11190222 SM719080045WA 010084068210394721SM719080045WA11190222 SM719080046WA 010084068210394721SM719080046WA11190222 SM718460038WA 010084068210394721SM718460038WA11181116 SM718460039WA 010084068210394721SM718460039WA11181116 SM719080017WA 010084068210394721SM719080017WA11190220 SM719080018WA 010084068210394721SM719080018WA11190220 SM719220058WA 010084068210394721SM719220058WA11190531 SM719220059WA 010084068210394721SM719220059WA11190531 SM719230001WA 010084068210394721SM719230001WA11190603 SM719230002WA 010084068210394721SM719230002WA11190603 SM719230003WA 010084068210394721SM719230003WA11190603 SM719230004WA 010084068210394721SM719230004WA11190603 SM719230005WA 010084068210394721SM719230005WA11190603 SM719230006WA 010084068210394721SM719230006WA11190603 SM719230007WA 010084068210394721SM719230007WA11190603 SM719230008WA 010084068210394721SM719230008WA11190603 SM719230009WA 010084068210394721SM719230009WA11190603 SM719170015WA 010084068210394721SM719170015WA11190425 SM719170016WA 010084068210394721SM719170016WA11190425 SM719080027WA 010084068210394721SM719080027WA11190220 SM719080028WA 010084068210394721SM719080028WA11190220 SM719080029WA 010084068210394721SM719080029WA11190220 SM719080030WA 010084068210394721SM719080030WA11190220 SM719080031WA 010084068210394721SM719080031WA11190220 SM719080032WA 010084068210394721SM719080032WA11190220 SM719080033WA 010084068210394721SM719080033WA11190220 SM719080034WA 010084068210394721SM719080034WA11190220 SM718470046WA 010084068210394721SM718470046WA11181122 SM719010118WA 010084068210394721SM719010118WA11190104 SM719230026WA 010084068210394721SM719230026WA11190604 SM718510034WA 010084068210394721SM718510034WA11181221 SM718460040WA 010084068210394721SM718460040WA11181116 SM718410019WA 010084068210394721SM718410019WA11181010 SM718410020WA 010084068210394721SM718410020WA11181010 SM718410021WA 010084068210394721SM718410021WA11181010 SM718410022WA 010084068210394721SM718410022WA11181010 SM718410023WA 010084068210394721SM718410023WA11181010 SM719030012MA 010084068210394721SM719030012MA11190114 SM719050049WA 010084068210394721SM719050049WA11190131 SM719050050WA 010084068210394721SM719050050WA11190131 SM719050051WA 010084068210394721SM719050051WA11190201 SM719080038WA 010084068210394721SM719080038WA11190221 SM719080039WA 010084068210394721SM719080039WA11190221 SM719120011MA 010084068210394721SM719120011MA11190319 SM719120012MA 010084068210394721SM719120012MA11190319 SM719130004WA 010084068210394721SM719130004WA11190327 SM719130005WA 010084068210394721SM719130005WA11190327 SM719130006WA 010084068210394721SM719130006WA11190327 SM719130007WA 010084068210394721SM719130007WA11190327 SM719140042MA 010084068210394721SM719140042MA11190404 SM719200015MA 010084068210394721SM719200015MA11190516 SM719200016MA 010084068210394721SM719200016MA11190516 SM719200017MA 010084068210394721SM719200017MA11190516 SM719200018MA 010084068210394721SM719200018MA11190516 SM719200019MA 010084068210394721SM719200019MA11190516 SM719200030MA 010084068210394721SM719200030MA11190516 SM719200031MA 010084068210394721SM719200031MA11190516 SM719200032MA 010084068210394721SM719200032MA11190516 SM719200033MA 010084068210394721SM719200033MA11190516 SM719200034MA 010084068210394721SM719200034MA11190516 SM719200035MA 010084068210394721SM719200035MA11190516 SM719220013MA 010084068210394721SM719220013MA11190527 SM719220014MA 010084068210394721SM719220014MA11190527 SM719230010WA 010084068210394721SM719230010WA11190603 SM719230011WA 010084068210394721SM719230011WA11190603 SM718440003WA 010084068210394721SM718440003WA11181029 SM718440004WA 010084068210394721SM718440004WA11181029 SM718360039MA 010084068210394721SM718360039MA11180905 SM718440016MA 010084068210394721SM718440016MA11181101 SM718470029MA 010084068210394721SM718470029MA11181119 SM718470030MA 010084068210394721SM718470030MA11181119 SM719150068MA 010084068210394721SM719150068MA11190411 SM719150069MA 010084068210394721SM719150069MA11190411 SM719150070MA 010084068210394721SM719150070MA11190411 SM719150071MA 010084068210394721SM719150071MA11190411 SM719150072MA 010084068210394721SM719150072MA11190411 SM718510031WA 010084068210394721SM718510031WA11181221 SM719030004WA 010084068210394721SM719030004WA11190114 SM718500011WA 010084068210394721SM718500011WA11181211 SM719230030WA 010084068210394721SM719230030WA11190605 SM718340005WA 010084068210394721SM718340005WA11180820 SM718340006WA 010084068210394721SM718340006WA11180820 SM718410005WA 010084068210394721SM718410005WA11181008 SM719230031WA 010084068210394721SM719230031WA11190605 SM719230032WA 010084068210394721SM719230032WA11190605 SM719230033WA 010084068210394721SM719230033WA11190605 SM718500013WA 010084068210394721SM718500013WA11181211 SM719220052WA 010084068210394721SM719220052WA11190530 SM719220053WA 010084068210394721SM719220053WA11190530 SM718340001WA 010084068210398521SM718340001WA11180820 SM718340002WA 010084068210398521SM718340002WA11180820 SM718340003WA 010084068210398521SM718340003WA11180820 SM718340004WA 010084068210398521SM718340004WA11180820 SM718470060WA 010084068210394721SM718470060WA11181123 SM718470061WA 010084068210394721SM718470061WA11181123 SM718470062WA 010084068210394721SM718470062WA11181123 SM718470063WA 010084068210394721SM718470063WA11181123 SM718470064WA 010084068210394721SM718470064WA11181123 SM718470065WA 010084068210394721SM718470065WA11181123 SM718470066WA 010084068210394721SM718470066WA11181123 SM718470067WA 010084068210394721SM718470067WA11181123 SM718500007WA 010084068210394721SM718500007WA11181210 SM719110009WA 010084068210394721SM719110009WA11190312 SM718370005WA 010084068210394721SM718370005WA11180910 SM718500012WA 010084068210394721SM718500012WA11181211 SM718440035WA 010084068210394721SM718440035WA11181102 SM718440036WA 010084068210394721SM718440036WA11181102 SM718500006WA 010084068210394721SM718500006WA11181210 SM719070004WA 010084068210394721SM719070004WA11190211 SM719100037WA 010084068210394721SM719100037WA11190308 SM719050054WA 010084068210394721SM719050054WA11190201 SM719110007WA 010084068210394721SM719110007WA11190312 SM719110008WA 010084068210394721SM719110008WA11190312 SM719220051WA 010084068210394721SM719220051WA11190530 SM719200010WA 010084068210394721SM719200010WA11190515 SM718500008WA 010084068210394721SM718500008WA11181210 SM718500009WA 010084068210394721SM718500009WA11181211 SM718500010WA 010084068210394721SM718500010WA11181211 SM719010123WA 010084068210394721SM719010123WA11190104 SM719010124WA 010084068210394721SM719010124WA11190104 SM719010125WA 010084068210394721SM719010125WA11190104 SM719010126WA 010084068210394721SM719010126WA11190104 SM719010127WA 010084068210394721SM719010127WA11190104 SM719200001WA 010084068210394721SM719200001WA11190515 SM719200002WA 010084068210394721SM719200002WA11190515 SM719200003WA 010084068210394721SM719200003WA11190515 SM719200004WA 010084068210394721SM719200004WA11190515 SM719200005WA 010084068210394721SM719200005WA11190515 SM719200006WA 010084068210394721SM719200006WA11190515 SM719200007WA 010084068210394721SM719200007WA11190515 SM719200008WA 010084068210394721SM719200008WA11190515 SM718410003WA 010084068210394721SM718410003WA11181008 SM718410004WA 010084068210394721SM718410004WA11181008 SM719210008WA 010084068210394721SM719210008WA11190520 SM719210009WA 010084068210394721SM719210009WA11190520 SM719210010WA 010084068210394721SM719210010WA11190521 SM719220040WA 010084068210394721SM719220040WA11190529 SM719220041WA 010084068210394721SM719220041WA11190529 SM719220042WA 010084068210394721SM719220042WA11190529 SM719220043WA 010084068210394721SM719220043WA11190529 SM718410006WA 010084068210394721SM718410006WA11181008 SM718410007WA 010084068210394721SM718410007WA11181008 SM718370006WA 010084068210394721SM718370006WA11180910 SM718490016WA 010084068210394721SM718490016WA11181204 SM718490017WA 010084068210394721SM718490017WA11181204 SM718490018WA 010084068210394721SM718490018WA11181204 SM718490019WA 010084068210394721SM718490019WA11181204 SM718490020WA 010084068210394721SM718490020WA11181204 SM718490021WA 010084068210394721SM718490021WA11181204 SM719070001WA 010084068210394721SM719070001WA11190211 SM719070002WA 010084068210394721SM719070002WA11190211 SM719070003WA 010084068210394721SM719070003WA11190211 SM718380041WA 010084068210394721SM718380041WA11180921 SM719210024MA 010084068210394721SM719210024MA11190523 SM719210025MA 010084068210394721SM719210025MA11190523 SM718340007WA 010084068210394721SM718340007WA11180820 SM718340008WA 010084068210394721SM718340008WA11180820 SM718340009WA 010084068210394721SM718340009WA11180820 SM719200009WA 010084068210394721SM719200009WA11190515 SM719160021WA 010084068210394721SM719160021WA11190420 SM719160018WA 010084068210394721SM719160018WA11190420 SM719220018WA 010084068210394721SM719220018WA11190528 SM719070005WA 010084068210394721SM719070005WA11190211 SM719070006WA 010084068210394721SM719070006WA11190211 SM719210026MA 010084068210394721SM719210026MA11190523 SM719210027MA 010084068210394721SM719210027MA11190523 SM719210028MA 010084068210394721SM719210028MA11190523 SM719210029MA 010084068210394721SM719210029MA11190523 SM719210030MA 010084068210394721SM719210030MA11190523 SM719210031MA 010084068210394721SM719210031MA11190523 SM719210032MA 010084068210394721SM719210032MA11190523 SM719210033MA 010084068210394721SM719210033MA11190523 SM719210034MA 010084068210394721SM719210034MA11190523 SM719210035MA 010084068210394721SM719210035MA11190523 SM719050001WA 010084068210394721SM719050001WA11190128 SM719050002WA 010084068210394721SM719050002WA11190128 SM719050003WA 010084068210394721SM719050003WA11190128 SM719050042WA 010084068210394721SM719050042WA11190131 SM719170005WA 010084068210394721SM719170005WA11190423 SM719170006WA 010084068210394721SM719170006WA11190423 SM719170007WA 010084068210394721SM719170007WA11190423 SM719170008WA 010084068210394721SM719170008WA11190424 SM719170009WA 010084068210394721SM719170009WA11190424 SM719170010WA 010084068210394721SM719170010WA11190424 SM719170011WA 010084068210394721SM719170011WA11190424 SM719170012WA 010084068210394721SM719170012WA11190424 SM719170013WA 010084068210394721SM719170013WA11190424 SM719170014WA 010084068210394721SM719170014WA11190424 SM718340016WA 010084068210394721SM718340016WA11180823 SM718410024WA 010084068210394721SM718410024WA11181010 SM718420043WA 010084068210394721SM718420043WA11181018 SM718430039WA 010084068210394721SM718430039WA11181024 SM718430040WA 010084068210394721SM718430040WA11181024 SM718430041WA 010084068210394721SM718430041WA11181024 SM718440008WA 010084068210394721SM718440008WA11181029 SM718480066WA 010084068210394721SM718480066WA11181130 SM718480067WA 010084068210394721SM718480067WA11181130 SM718480068WA 010084068210394721SM718480068WA11181130 SM718480069WA 010084068210394721SM718480069WA11181130 SM718480070WA 010084068210394721SM718480070WA11181130 SM718480071WA 010084068210394721SM718480071WA11181130 SM718480072WA 010084068210394721SM718480072WA11181130 SM719010121WA 010084068210394721SM719010121WA11190104 SM719010122WA 010084068210394721SM719010122WA11190104 SM719040005WA 010084068210394721SM719040005WA11190121 SM719040006WA 010084068210394721SM719040006WA11190121 SM719040008WA 010084068210394721SM719040008WA11190122 SM719050052WA 010084068210394721SM719050052WA11190201 SM719050053WA 010084068210394721SM719050053WA11190201 SM719090006WA 010084068210394721SM719090006WA11190225 SM719090007WA 010084068210394721SM719090007WA11190225 SM719100033WA 010084068210394721SM719100033WA11190307 SM719140016WA 010084068210394721SM719140016WA11190403 SM719140037WA 010084068210394721SM719140037WA11190404 SM719150006WA 010084068210394721SM719150006WA11190409 SM719190022WA 010084068210394721SM719190022WA11190509 SM719210014WA 010084068210394721SM719210014WA11190521 SM719210015WA 010084068210394721SM719210015WA11190521 SM719210016WA 010084068210394721SM719210016WA11190521 SM719210017WA 010084068210394721SM719210017WA11190521 SM719210013WA 010084068210394721SM719210013WA11190521 SM719240014WA 010084068210394721SM719240014WA11190612 SM719030013WA 010084068210394721SM719030013WA11190115 SM719030014WA 010084068210394721SM719030014WA11190115 SM719030015WA 010084068210394721SM719030015WA11190115 SM719030016WA 010084068210394721SM719030016WA11190115 SM719030017WA 010084068210394721SM719030017WA11190115 SM718430026WA 010084068210394721SM718430026WA11181023 SM718430027WA 010084068210394721SM718430027WA11181023 SM719150080WA 010084068210394721SM719150080WA11190412 SM719150081WA 010084068210394721SM719150081WA11190412 SM718470093WA 010084068210394721SM718470093WA11181124 SM718490008WA 010084068210394721SM718490008WA11181203 SM718490002WA 010084068210394721SM718490002WA11181203 SM719190021WA 010084068210394721SM719190021WA11190509 SM719230034WA 010084068210394721SM719230034WA11190606 SM719040009WA 010084068210394721SM719040009WA11190122 SM719040010WA 010084068210394721SM719040010WA11190122 SM719040011WA 010084068210394721SM719040011WA11190122 SM719040012WA 010084068210394721SM719040012WA11190122 SM719040013WA 010084068210394721SM719040013WA11190122 SM718350048WA 010084068210394721SM718350048WA11180901 SM718350050WA 010084068210394721SM718350050WA11180901 SM718350051WA 010084068210394721SM718350051WA11180901 SM718410027WA 010084068210394721SM718410027WA11181010 SM719150074WA 010084068210394721SM719150074WA11190412 SM718490014WA 010084068210394721SM718490014WA11181204 SM719070017WA 010084068210394721SM719070017WA11190212 SM718350038WA 010084068210394721SM718350038WA11180831 SM718520010WA 010084068210394721SM718520010WA11181224 SM719210005WA 010084068210394721SM719210005WA11190520 SM719040023WA 010084068210394721SM719040023WA11190123 SM719130008WA 010084068210394721SM719130008WA11190327 SM719040025WA 010084068210394721SM719040025WA11190123 SM719040026WA 010084068210394721SM719040026WA11190123 SM718410038WA 010084068210394721SM718410038WA11181012 SM718410039WA 010084068210394721SM718410039WA11181012 SM718410040WA 010084068210394721SM718410040WA11181012 SM718410041WA 010084068210394721SM718410041WA11181012 SM719090010WA 010084068210394721SM719090010WA11190226 SM718440027WA 010084068210394721SM718440027WA11181102 SM718470075WA 010084068210394721SM718470075WA11181124 SM718470073WA 010084068210394721SM718470073WA11181124 SM718480024WA 010084068210394721SM718480024WA11181128 SM719130019WA 010084068210394721SM719130019WA11190328 SM719130020WA 010084068210394721SM719130020WA11190328 SM718480023WA 010084068210394721SM718480023WA11181128 SM719210006WA 010084068210394721SM719210006WA11190520 SM719210007WA 010084068210394721SM719210007WA11190520 SM719040021WA 010084068210394721SM719040021WA11190123 SM718470080WA 010084068210394721SM718470080WA11181124 SM719180019WA 010084068210394721SM719180019WA11190505 SM718350039WA 010084068210394721SM718350039WA11180831 SM719150073WA 010084068210394721SM719150073WA11190412 SM718430047WA 010084068210394721SM718430047WA11181025 SM718520008WA 010084068210394721SM718520008WA11181224 SM719080012WA 010084068210394721SM719080012WA11190218 SM719080013WA 010084068210394721SM719080013WA11190218 SM719140018WA 010084068210394721SM719140018WA11190403 SM719150047WA 010084068210394721SM719150047WA11190410 SM719150077WA 010084068210394721SM719150077WA11190412 SM719150078WA 010084068210394721SM719150078WA11190412 SM719150079WA 010084068210394721SM719150079WA11190412 SM718490006WA 010084068210394721SM718490006WA11181203 SM718490007WA 010084068210394721SM718490007WA11181203 SM718440006WA 010084068210394721SM718440006WA11181029 SM718470078WA 010084068210394721SM718470078WA11181124 SM719210004WA 010084068210394721SM719210004WA11190520 SM718350049WA 010084068210394721SM718350049WA11180901 SM718350040WA 010084068210394721SM718350040WA11180901 SM718350044WA 010084068210394721SM718350044WA11180901 SM718350045WA 010084068210394721SM718350045WA11180901 SM719210012WA 010084068210394721SM719210012WA11190521 SM718520006WA 010084068210394721SM718520006WA11181224 SM718350036WA 010084068210394721SM718350036WA11180831 SM719160017WA 010084068210394721SM719160017WA11190419 SM719080042WA 010084068210394721SM719080042WA11190222 SM719080043WA 010084068210394721SM719080043WA11190222 SM719020009WA 010084068210394721SM719020009WA11190108 SM718430053WA 010084068210394721SM718430053WA11181026 SM718470079WA 010084068210394721SM718470079WA11181124 SM718410012WA 010084068210394721SM718410012WA11181009 SM718410013WA 010084068210394721SM718410013WA11181009 SM718410014WA 010084068210394721SM718410014WA11181009 SM718410015WA 010084068210394721SM718410015WA11181009 SM718410016WA 010084068210394721SM718410016WA11181009 SM718410017WA 010084068210394721SM718410017WA11181009 SM719040015WA 010084068210394721SM719040015WA11190122 SM719040014WA 010084068210394721SM719040014WA11190122 SM719040024WA 010084068210394721SM719040024WA11190123 SM719080011WA 010084068210394721SM719080011WA11190218 SM719180010WA 010084068210394721SM719180010WA11190430 SM719180011WA 010084068210394721SM719180011WA11190430 SM719180012WA 010084068210394721SM719180012WA11190430 SM719180013WA 010084068210394721SM719180013WA11190430 SM719180014WA 010084068210394721SM719180014WA11190430 SM719180015WA 010084068210394721SM719180015WA11190430 SM719180016WA 010084068210394721SM719180016WA11190430 SM719070007WA 010084068210394721SM719070007WA11190211 SM719210003WA 010084068210394721SM719210003WA11190520 SM719090028WA 010084068210394721SM719090028WA11190228 SM718450021WA 010084068210394721SM718450021WA11181107 SM718450022WA 010084068210394721SM718450022WA11181107 SM718490004WA 010084068210394721SM718490004WA11181203 SM719090025WA 010084068210394721SM719090025WA11190228 SM718520009WA 010084068210394721SM718520009WA11181224 SM719070009WA 010084068210394721SM719070009WA11190211 SM719130026WA 010084068210394721SM719130026WA11190328 SM719130022WA 010084068210394721SM719130022WA11190328 SM718490001WA 010084068210394721SM718490001WA11181203 SM718520007WA 010084068210394721SM718520007WA11181224 SM718360001WA 010084068210394721SM718360001WA11180903 SM718360002WA 010084068210394721SM718360002WA11180903 SM718360003WA 010084068210394721SM718360003WA11180903 SM719070018WA 010084068210394721SM719070018WA11190212 SM719090026WA 010084068210394721SM719090026WA11190228 SM719090027WA 010084068210394721SM719090027WA11190228 SM719220031WA 010084068210394721SM719220031WA11190528 SM719220032WA 010084068210394721SM719220032WA11190528 SM718470074WA 010084068210394721SM718470074WA11181124 SM718490005WA 010084068210394721SM718490005WA11181203 SM719080037WA 010084068210394721SM719080037WA11190221 SM719100034WA 010084068210394721SM719100034WA11190307 SM719070008WA 010084068210394721SM719070008WA11190211 SM718430025WA 010084068210394721SM718430025WA11181023 SM718470076WA 010084068210394721SM718470076WA11181124 SM718520005WA 010084068210394721SM718520005WA11181224 SM718450001WA 010084068210394721SM718450001WA11181105 SM718520004WA 010084068210394721SM718520004WA11181224 SM719140001WA 010084068210394721SM719140001WA11190402 SM718350041WA 010084068210394721SM718350041WA11180901 SM718350042WA 010084068210394721SM718350042WA11180901 SM718350043WA 010084068210394721SM718350043WA11180901 SM718520002WA 010084068210394721SM718520002WA11181224 SM719040022WA 010084068210394721SM719040022WA11190123 SM719210001WA 010084068210394721SM719210001WA11190520 SM719210002WA 010084068210394721SM719210002WA11190520 SM718350047WA 010084068210394721SM718350047WA11180901 SM718360004WA 010084068210394721SM718360004WA11180903 SM719130023WA 010084068210394721SM719130023WA11190328 SM719130024WA 010084068210394721SM719130024WA11190328 SM719150057WA 010084068210394721SM719150057WA11190411 SM719140039WA 010084068210394721SM719140039WA11190404 SM719150075WA 010084068210394721SM719150075WA11190412 SM719050006WA 010084068210394721SM719050006WA11190128 SM718410028WA 010084068210394721SM718410028WA11181010 SM719050005WA 010084068210394721SM719050005WA11190128 SM718350037WA 010084068210394721SM718350037WA11180831 SM718410026WA 010084068210394721SM718410026WA11181010 SM718440007WA 010084068210394721SM718440007WA11181029 SM718440010WA 010084068210394721SM718440010WA11181030 SM718440011WA 010084068210394721SM718440011WA11181030 SM718440012WA 010084068210394721SM718440012WA11181030 SM718470069WA 010084068210394721SM718470069WA11181124 SM718470072WA 010084068210394721SM718470072WA11181124 SM718470084WA 010084068210394721SM718470084WA11181124 SM718470092WA 010084068210394721SM718470092WA11181124 SM718490003WA 010084068210394721SM718490003WA11181203 SM718490012WA 010084068210394721SM718490012WA11181204 SM718490015WA 010084068210394721SM718490015WA11181204 SM718520001WA 010084068210394721SM718520001WA11181224 SM718520003WA 010084068210394721SM718520003WA11181224 SM719100003WA 010084068210394721SM719100003WA11190304 SM719100004WA 010084068210394721SM719100004WA11190304 SM719100008WA 010084068210394721SM719100008WA11190305 SM719130021WA 010084068210394721SM719130021WA11190328 SM719150076WA 010084068210394721SM719150076WA11190412 SM718480026WA 010084068210394721SM718480026WA11181128 SM718490009WA 010084068210394721SM718490009WA11181203 SM718490013WA 010084068210394721SM718490013WA11181204 SM718470068WA 010084068210394721SM718470068WA11181123 SM718350046WA 010084068210394721SM718350046WA11180901 SM718380013WA 010084068210394721SM718380013WA11180920 SM718450009WA 010084068210394721SM718450009WA11181106 SM718470070WA 010084068210394721SM718470070WA11181124 SM718470071WA 010084068210394721SM718470071WA11181124 SM718470077WA 010084068210394721SM718470077WA11181124 SM718470081WA 010084068210394721SM718470081WA11181124 SM718470082WA 010084068210394721SM718470082WA11181124 SM718470083WA 010084068210394721SM718470083WA11181124 SM718470085WA 010084068210394721SM718470085WA11181124 SM718470086WA 010084068210394721SM718470086WA11181124 SM718470087WA 010084068210394721SM718470087WA11181124 SM718470088WA 010084068210394721SM718470088WA11181124 SM718470089WA 010084068210394721SM718470089WA11181124 SM718470090WA 010084068210394721SM718470090WA11181124 SM718470091WA 010084068210394721SM718470091WA11181124 SM718480025WA 010084068210394721SM718480025WA11181128 SM719080040WA 010084068210394721SM719080040WA11190221 SM719130025WA 010084068210394721SM719130025WA11190328 SM719140002WA 010084068210394721SM719140002WA11190402 SM719140003WA 010084068210394721SM719140003WA11190402 SM719140004WA 010084068210394721SM719140004WA11190402 SM719140005WA 010084068210394721SM719140005WA11190402 SM719140006WA 010084068210394721SM719140006WA11190402 SM719140007WA 010084068210394721SM719140007WA11190402 SM719140008WA 010084068210394721SM719140008WA11190403 SM719140009WA 010084068210394721SM719140009WA11190403 SM719140010WA 010084068210394721SM719140010WA11190403 SM719220033WA 010084068210394721SM719220033WA11190528 SM719220035WA 010084068210394721SM719220035WA11190529 SM719220036WA 010084068210394721SM719220036WA11190529 SM719220037WA 010084068210394721SM719220037WA11190529 SM719220038WA 010084068210394721SM719220038WA11190529 SM718380012WA 010084068210394721SM718380012WA11180920 SM718350035WA 010084068210394721SM718350035WA11180831 SM718480065MA 010084068210394721SM718480065MA11181129 SM718450002WA 010084068210394721SM718450002WA11181105 SM718450003WA 010084068210394721SM718450003WA11181105 SM718460001WA 010084068210394721SM718460001WA11181112 SM719050045WA 010084068210394721SM719050045WA11190131 SM719050046WA 010084068210394721SM719050046WA11190131 SM719050047WA 010084068210394721SM719050047WA11190131 SM719170002WA 010084068210394721SM719170002WA11190423 SM719170003WA 010084068210394721SM719170003WA11190423 SM719170004WA 010084068210394721SM719170004WA11190423 SM719240008WA 010084068210394721SM719240008WA11190611 SM719240009WA 010084068210394721SM719240009WA11190611 SM719240010WA 010084068210394721SM719240010WA11190611 SM718410042WA 010084068210394721SM718410042WA11181012 SM718410043WA 010084068210394721SM718410043WA11181012 SM718410044WA 010084068210394721SM718410044WA11181012 SM718410045WA 010084068210394721SM718410045WA11181012 SM718410046WA 010084068210394721SM718410046WA11181012 SM719150001WA 010084068210394721SM719150001WA11190408 SM719150002WA 010084068210394721SM719150002WA11190408 SM719150003WA 010084068210394721SM719150003WA11190408 SM719150004WA 010084068210394721SM719150004WA11190408 SM718340026WA 010084068210394721SM718340026WA11180824 SM718340027WA 010084068210394721SM718340027WA11180824 SM718420037WA 010084068210394721SM718420037WA11181018 SM718420038WA 010084068210394721SM718420038WA11181018 SM718420039WA 010084068210394721SM718420039WA11181018 SM718410010WA 010084068210394721SM718410010WA11181009 SM719040042WA 010084068210394721SM719040042WA11190125 SM718460024WA 010084068210394721SM718460024WA11181113 SM718410011WA 010084068210394721SM718410011WA11181009 SM718420009WA 010084068210394721SM718420009WA11181015 SM718420010WA 010084068210394721SM718420010WA11181015 SM718350002WA 010084068210394721SM718350002WA11180830 SM719020002WA 010084068210394721SM719020002WA11190107 SM719020003WA 010084068210394721SM719020003WA11190107 SM719020004WA 010084068210394721SM719020004WA11190107 SM719070078WA 010084068210394721SM719070078WA11190216 SM719070079WA 010084068210394721SM719070079WA11190216 SM719140038WA 010084068210394721SM719140038WA11190404 SM719180006WA 010084068210394721SM719180006WA11190430 SM719180007WA 010084068210394721SM719180007WA11190430 SM719180008WA 010084068210394721SM719180008WA11190430 SM719180009WA 010084068210394721SM719180009WA11190430 SM719190001WA 010084068210394721SM719190001WA11190506 SM719190002WA 010084068210394721SM719190002WA11190506 SM719190003WA 010084068210394721SM719190003WA11190506 SM719190007WA 010084068210394721SM719190007WA11190507 SM719220044MA 010084068210394721SM719220044MA11190529 SM719220045MA 010084068210394721SM719220045MA11190529 SM719220046MA 010084068210394721SM719220046MA11190529 SM719220047MA 010084068210394721SM719220047MA11190529 SM719220048MA 010084068210394721SM719220048MA11190529 SM719220049MA 010084068210394721SM719220049MA11190529 SM718490031MA 010084068210394721SM718490031MA11181206 SM719090009MA 010084068210394721SM719090009MA11190225 SM719140043MA 010084068210394721SM719140043MA11190405 SM719270041MA 010084068210394721SM719270041MA11190703 SM719270042MA 010084068210394721SM719270042MA11190703 SM719150048MA 010084068210394721SM719150048MA11190410 SM719150049MA 010084068210394721SM719150049MA11190410 SM719030027MA 010084068210394721SM719030027MA11190117 SM719020027WA 010084068210394721SM719020027WA11190110 SM719100014WA 010084068210394721SM719100014WA11190306 SM719230019WA 010084068210394721SM719230019WA11190604 SM719230020WA 010084068210394721SM719230020WA11190604 SM719230021WA 010084068210394721SM719230021WA11190604 SM718470045WA 010084068210394721SM718470045WA11181121 SM719020020WA 010084068210394721SM719020020WA11190109 SM719020023WA 010084068210394721SM719020023WA11190110 SM718430028WA 010084068210394721SM718430028WA11181023 SM718440005WA 010084068210394721SM718440005WA11181029 SM719030005MA 010084068210394721SM719030005MA11190114 SM719030006MA 010084068210394721SM719030006MA11190114 SM719030007MA 010084068210394721SM719030007MA11190114 SM719030008MA 010084068210394721SM719030008MA11190114 SM719030009MA 010084068210394721SM719030009MA11190114 SM719030010MA 010084068210394721SM719030010MA11190114 SM719030011MA 010084068210394721SM719030011MA11190114 SM719150039MA 010084068210394721SM719150039MA11190409 SM718360041MA 010084068210394721SM718360041MA11180905 SM718400019MA 010084068210394721SM718400019MA11181003 SM718450062MA 010084068210394721SM718450062MA11181109 SM718450063MA 010084068210394721SM718450063MA11181109 SM718450064MA 010084068210394721SM718450064MA11181109 SM719020029MA 010084068210394721SM719020029MA11190110 SM718420001WA 010084068210394721SM718420001WA11181015 SM718420002WA 010084068210394721SM718420002WA11181015 SM718460005WA 010084068210394721SM718460005WA11181112 SM718430055WA 010084068210394721SM718430055WA11181026 SM718430056WA 010084068210394721SM718430056WA11181026 SM718430057WA 010084068210394721SM718430057WA11181026 SM718370011WA 010084068210394721SM718370011WA11180911 SM719210040WA 010084068210394721SM719210040WA11190524 SM719210041WA 010084068210394721SM719210041WA11190524 SM718350032WA 010084068210394721SM718350032WA11180831 SM718350033WA 010084068210394721SM718350033WA11180831 SM718350034WA 010084068210394721SM718350034WA11180831 SM719210039WA 010084068210394721SM719210039WA11190524 SM718430059WA 010084068210394721SM718430059WA11181026 SM718440001WA 010084068210394721SM718440001WA11181029 SM719040039WA 010084068210394721SM719040039WA11190125 SM718470047WA 010084068210394721SM718470047WA11181122 SM718360030WA 010084068210394721SM718360030WA11180905 SM718360031WA 010084068210394721SM718360031WA11180905 SM718360032WA 010084068210394721SM718360032WA11180905 SM718350001WA 010084068210394721SM718350001WA11180830 SM718350030WA 010084068210394721SM718350030WA11180831 SM718350031WA 010084068210394721SM718350031WA11180831 SM719100035WA 010084068210394721SM719100035WA11190307 SM718430058WA 010084068210394721SM718430058WA11181026 SM718450004WA 010084068210394721SM718450004WA11181105 SM718460023WA 010084068210394721SM718460023WA11181113 SM719050021WA 010084068210394721SM719050021WA11190129 SM719080041WA 010084068210394721SM719080041WA11190221 SM718460006WA 010084068210394721SM718460006WA11181112 SM718480056WA 010084068210394721SM718480056WA11181129 SM719130001WA 010084068210394721SM719130001WA11190327 SM719050017WA 010084068210394721SM719050017WA11190129 SM719050018WA 010084068210394721SM719050018WA11190129 SM719050019WA 010084068210394721SM719050019WA11190129 SM719050020WA 010084068210394721SM719050020WA11190129 SM718470054WA 010084068210394721SM718470054WA11181122 SM719070016WA 010084068210394721SM719070016WA11190212 SM719220028WA 010084068210394721SM719220028WA11190528 SM719220029WA 010084068210394721SM719220029WA11190528 SM719220030WA 010084068210394721SM719220030WA11190528 SM718410001WA 010084068210394721SM718410001WA11181008 SM718410002WA 010084068210394721SM718410002WA11181008 SM718450005WA 010084068210394721SM718450005WA11181105 SM718450006WA 010084068210394721SM718450006WA11181105 SM718450007WA 010084068210394721SM718450007WA11181106 SM719210011WA 010084068210394721SM719210011WA11190521 SM719250002MA 010084068210394721SM719250002MA11190618 SM719250003MA 010084068210394721SM719250003MA11190618 SM719250004MA 010084068210394721SM719250004MA11190618 SM719250005MA 010084068210394721SM719250005MA11190618 SM719250006MA 010084068210394721SM719250006MA11190618 SM719250007MA 010084068210394721SM719250007MA11190618 SM718400005MA 010084068210394721SM718400005MA11181002 SM718400008MA 010084068210394721SM718400008MA11181002 SM718400009MA 010084068210394721SM718400009MA11181002 SM718400010MA 010084068210394721SM718400010MA11181002 SM718400012MA 010084068210394721SM718400012MA11181002 SM718400013MA 010084068210394721SM718400013MA11181002 SM718400004MA 010084068210394721SM718400004MA11181002 SM718400006MA 010084068210394721SM718400006MA11181002 SM718400007MA 010084068210394721SM718400007MA11181002 SM718400011MA 010084068210394721SM718400011MA11181002 SM718400014MA 010084068210394721SM718400014MA11181002 SM718400015MA 010084068210394721SM718400015MA11181002 SM718500036MA 010084068210394721SM718500036MA11181213 SM718510001MA 010084068210394721SM718510001MA11181218 SM718510002MA 010084068210394721SM718510002MA11181218 SM718510003MA 010084068210394721SM718510003MA11181218 SM718510004MA 010084068210394721SM718510004MA11181218 SM718510005MA 010084068210394721SM718510005MA11181218 SM718510006MA 010084068210394721SM718510006MA11181218 SM718450024MA 010084068210394721SM718450024MA11181107 SM719120036MA 010084068210394721SM719120036MA11190322 SM719120037MA 010084068210394721SM719120037MA11190322 SM719150050MA 010084068210394721SM719150050MA11190410 SM718520012MA 010084068210394721SM718520012MA11181227 SM718520013MA 010084068210394721SM718520013MA11181227 SM718520014MA 010084068210394721SM718520014MA11181227 SM718520015MA 010084068210394721SM718520015MA11181227 SM718520016MA 010084068210394721SM718520016MA11181227 SM718520017MA 010084068210394721SM718520017MA11181227 SM718520018MA 010084068210394721SM718520018MA11181227 SM718520020MA 010084068210394721SM718520020MA11181227 SM719020010MA 010084068210394721SM719020010MA11190108 SM719140044MA 010084068210394721SM719140044MA11190405 SM719140045MA 010084068210394721SM719140045MA11190405 SM719140046MA 010084068210394721SM719140046MA11190405 SM719140047MA 010084068210394721SM719140047MA11190405 SM719140048MA 010084068210394721SM719140048MA11190405 SM719140049MA 010084068210394721SM719140049MA11190405 SM719270011MA 010084068210394721SM719270011MA11190701 SM719270012MA 010084068210394721SM719270012MA11190701 SM719280015MA 010084068210394721SM719280015MA11190711 SM718410048MA 010084068210394721SM718410048MA11181012 SM718410049MA 010084068210394721SM718410049MA11181012 SM718410050MA 010084068210394721SM718410050MA11181012 SM718410051MA 010084068210394721SM718410051MA11181012 SM718450027WA 010084068210394721SM718450027WA11181108 SM718480027WA 010084068210394721SM718480027WA11181128 SM718480060WA 010084068210394721SM718480060WA11181129 SM718500037MA 010084068210394721SM718500037MA11181214 SM719090054WA 010084068210394721SM719090054WA11190302 SM718360044WA 010084068210394721SM718360044WA11180906 SM718360045WA 010084068210394721SM718360045WA11180906 SM718360046WA 010084068210394721SM718360046WA11180906 SM718360047WA 010084068210394721SM718360047WA11180906 SM718360048WA 010084068210394721SM718360048WA11180906 SM718360049WA 010084068210394721SM718360049WA11180906 SM718380007WA 010084068210394721SM718380007WA11180919 SM718380008WA 010084068210394721SM718380008WA11180919 SM718380009WA 010084068210394721SM718380009WA11180919 SM718420005WA 010084068210394721SM718420005WA11181015 SM718420006WA 010084068210394721SM718420006WA11181015 SM718420007WA 010084068210394721SM718420007WA11181015 SM718420008WA 010084068210394721SM718420008WA11181015 SM718480031WA 010084068210394721SM718480031WA11181128 SM718480055WA 010084068210394721SM718480055WA11181129 SM718480057WA 010084068210394721SM718480057WA11181129 SM718480058WA 010084068210394721SM718480058WA11181129 SM718480059WA 010084068210394721SM718480059WA11181129 SM719100030WA 010084068210394721SM719100030WA11190307 SM719100031WA 010084068210394721SM719100031WA11190307 SM719100032WA 010084068210394721SM719100032WA11190307 SM719240004WA 010084068210394721SM719240004WA11190611 SM719240005WA 010084068210394721SM719240005WA11190611 SM719240006WA 010084068210394721SM719240006WA11190611 SM719240007WA 010084068210394721SM719240007WA11190611 SM718340010WA 010084068210394721SM718340010WA11180822 SM718340011WA 010084068210394721SM718340011WA11180822 SM718430052WA 010084068210394721SM718430052WA11181025 SM718440002WA 010084068210394721SM718440002WA11181029 SM719080014WA 010084068210394721SM719080014WA11190219 SM719080015WA 010084068210394721SM719080015WA11190219 SM718460007WA 010084068210394721SM718460007WA11181112 SM718460008WA 010084068210394721SM718460008WA11181112 SM718460009WA 010084068210394721SM718460009WA11181112 SM718500014WA 010084068210394721SM718500014WA11181211 SM718500015WA 010084068210394721SM718500015WA11181211 SM719030001WA 010084068210394721SM719030001WA11190114 SM719030002WA 010084068210394721SM719030002WA11190114 SM719030003WA 010084068210394721SM719030003WA11190114 SM719040007WA 010084068210394721SM719040007WA11190122 SM719050007WA 010084068210394721SM719050007WA11190128 SM719050008WA 010084068210394721SM719050008WA11190128 SM719070020WA 010084068210394721SM719070020WA11190213 SM719070042WA 010084068210394721SM719070042WA11190214 SM719070043WA 010084068210394721SM719070043WA11190214 SM719070044WA 010084068210394721SM719070044WA11190214 SM719070045WA 010084068210394721SM719070045WA11190214 SM719070046WA 010084068210394721SM719070046WA11190214 SM719070047WA 010084068210394721SM719070047WA11190214 SM719070048WA 010084068210394721SM719070048WA11190214 SM719070049WA 010084068210394721SM719070049WA11190214 SM719070050WA 010084068210394721SM719070050WA11190214 SM719070051WA 010084068210394721SM719070051WA11190214 SM719070052WA 010084068210394721SM719070052WA11190214 SM719070053WA 010084068210394721SM719070053WA11190214 SM719070068WA 010084068210394721SM719070068WA11190215 SM719070069WA 010084068210394721SM719070069WA11190215 SM719070070WA 010084068210394721SM719070070WA11190215 SM719070071WA 010084068210394721SM719070071WA11190215 SM719110010WA 010084068210394721SM719110010WA11190312 SM719110011WA 010084068210394721SM719110011WA11190312 SM719110012WA 010084068210394721SM719110012WA11190312 SM719220055WA 010084068210394721SM719220055WA11190531 SM719090008MA 010084068210394721SM719090008MA11190225 SM718490022WA 010084068210394721SM718490022WA11181204 SM718410047MA 010084068210394721SM718410047MA11181012 SM718440009MA 010084068210394721SM718440009MA11181029 SM718450023MA 010084068210394721SM718450023MA11181107 SM719210023MA 010084068210394721SM719210023MA11190522 SM718340030MA 010084068210394721SM718340030MA11180824 SM718340031MA 010084068210394721SM718340031MA11180824 SM718340032MA 010084068210394721SM718340032MA11180824 SM719010103MA 010084068210394721SM719010103MA11190103 SM719040040WA 010084068210394721SM719040040WA11190125 SM719040041WA 010084068210394721SM719040041WA11190125 SM719150005WA 010084068210394721SM719150005WA11190408 SM718450020WA 010084068210394721SM718450020WA11181107 SM719050004WA 010084068210394721SM719050004WA11190128 SM718500003WA 010084068210394721SM718500003WA11181210 SM719020024WA 010084068210394721SM719020024WA11190110 SM719020025WA 010084068210394721SM719020025WA11190110 SM719020026WA 010084068210394721SM719020026WA11190110 SM718500001WA 010084068210394721SM718500001WA11181210 SM718500002WA 010084068210394721SM718500002WA11181210 SM719220026WA 010084068210394721SM719220026WA11190528 SM719220054WA 010084068210394721SM719220054WA11190530 SM718420040WA 010084068210394721SM718420040WA11181018 SM718460032WA 010084068210394721SM718460032WA11181115 SM719100011WA 010084068210394721SM719100011WA11190305 SM719110002WA 010084068210394721SM719110002WA11190311 SM719220039WA 010084068210394721SM719220039WA11190529 SM719220056WA 010084068210394721SM719220056WA11190531 SM718470057WA 010084068210394721SM718470057WA11181122 SM719200036WA 010084068210394721SM719200036WA11190517 SM718440028WA 010084068210394721SM718440028WA11181102 SM719110006WA 010084068210394721SM719110006WA11190312 SM718380001WA 010084068210394721SM718380001WA11180917 SM718430015WA 010084068210394721SM718430015WA11181023 SM718430016WA 010084068210394721SM718430016WA11181023 SM718500004WA 010084068210394721SM718500004WA11181210 SM718500005WA 010084068210394721SM718500005WA11181210 SM719010119WA 010084068210394721SM719010119WA11190104 SM719010120WA 010084068210394721SM719010120WA11190104 SM719150007WA 010084068210394721SM719150007WA11190409 SM719150008WA 010084068210394721SM719150008WA11190409 SM719150023WA 010084068210394721SM719150023WA11190409 SM719150024WA 010084068210394721SM719150024WA11190409 SM718510032WA 010084068210394721SM718510032WA11181221 SM719020016WA 010084068210394721SM719020016WA11190109 SM719020017WA 010084068210394721SM719020017WA11190109 SM719020007WA 010084068210394721SM719020007WA11190107 SM719020008WA 010084068210394721SM719020008WA11190107 SM719020012WA 010084068210394721SM719020012WA11190109 SM718420004WA 010084068210394721SM718420004WA11181015 SM719020014WA 010084068210394721SM719020014WA11190109 SM719020015WA 010084068210394721SM719020015WA11190109 SM719080016WA 010084068210394721SM719080016WA11190219 SM719020013WA 010084068210394721SM719020013WA11190109 SM719200013WA 010084068210394721SM719200013WA11180516 SM718390052MA 010084068210394721SM718390052MA11180927 SM718390053MA 010084068210394721SM718390053MA11180927 SM719240003MA 010084068210394721SM719240003MA11190610 SM718450057MA 010084068210394721SM718450057MA11181109 SM718450058MA 010084068210394721SM718450058MA11181109 SM718450059MA 010084068210394721SM718450059MA11181109 SM719070028MA 010084068210394721SM719070028MA11190213 SM719070029MA 010084068210394721SM719070029MA11190213 SM719070030MA 010084068210394721SM719070030MA11190213 SM719070031MA 010084068210394721SM719070031MA11190213 SM719070032MA 010084068210394721SM719070032MA11190213 SM719070033MA 010084068210394721SM719070033MA11190213 SM719070034MA 010084068210394721SM719070034MA11190213 SM719070035MA 010084068210394721SM719070035MA11190213 SM719070036MA 010084068210394721SM719070036MA11190213 SM719070037MA 010084068210394721SM719070037MA11190213 SM719070066MA 010084068210394721SM719070066MA11190214 SM719070067MA 010084068210394721SM719070067MA11190214 SM718400020MA 010084068210394721SM718400020MA11181003 SM718420052MA 010084068210394721SM718420052MA11181018 SM718400021MA 010084068210394721SM718400021MA11181003 SM718400022MA 010084068210394721SM718400022MA11181003 SM718400023MA 010084068210394721SM718400023MA11181003 SM718400024MA 010084068210394721SM718400024MA11181003 SM718450052MA 010084068210394721SM718450052MA11181109 SM718450053MA 010084068210394721SM718450053MA11181109 SM718450054MA 010084068210394721SM718450054MA11181109 SM718450055MA 010084068210394721SM718450055MA11181109 SM718450056MA 010084068210394721SM718450056MA11181109 SM718450011WA 010084068210394721SM718450011WA11181106 SM719020001WA 010084068210394721SM719020001WA11190107 SM718340017WA 010084068210394721SM718340017WA11180823 SM718470022WA 010084068210394721SM718470022WA11181120 SM718470023WA 010084068210394721SM718470023WA11181120 SM718470024WA 010084068210394721SM718470024WA11181120 SM718470025WA 010084068210394721SM718470025WA11181120 SM718470026WA 010084068210394721SM718470026WA11181120 SM718470044WA 010084068210394721SM718470044WA11181121 SM719170001WA 010084068210394721SM719170001WA11190422 SM718450026WA 010084068210394721SM718450026WA11181108 SM719050048WA 010084068210394721SM719050048WA11190131 SM718350007MA 010084068210394721SM718350007MA11180830 SM719150028MA 010084068210394721SM719150028MA11190409 SM719150029MA 010084068210394721SM719150029MA11190409 SM719150030MA 010084068210394721SM719150030MA11190409 SM719150031MA 010084068210394721SM719150031MA11190409 SM719150032MA 010084068210394721SM719150032MA11190409 SM719150033MA 010084068210394721SM719150033MA11190409 SM718470004WA 010084068210394721SM718470004WA11181119 SM718470005WA 010084068210394721SM718470005WA11181119 SM718470006WA 010084068210394721SM718470006WA11181119 SM718470007WA 010084068210394721SM718470007WA11181119 SM719020034WA 010084068210394721SM719020034WA11190111 SM719050036WA 010084068210394721SM719050036WA11190130 SM719050037WA 010084068210394721SM719050037WA11190130 SM719050038WA 010084068210394721SM719050038WA11190130 SM719050039WA 010084068210394721SM719050039WA11190130 SM719050040WA 010084068210394721SM719050040WA11190130 SM719050041WA 010084068210394721SM719050041WA11190130 SM719050043WA 010084068210394721SM719050043WA11190131 SM719050044WA 010084068210394721SM719050044WA11190131 SM719070072WA 010084068210394721SM719070072WA11190215 SM719070073WA 010084068210394721SM719070073WA11190216 SM719070074WA 010084068210394721SM719070074WA11190216 SM719070075WA 010084068210394721SM719070075WA11190216 SM719070076WA 010084068210394721SM719070076WA11190216 SM719070077WA 010084068210394721SM719070077WA11190216 SM719080001WA 010084068210394721SM719080001WA11190218 SM719080002WA 010084068210394721SM719080002WA11190218 SM719080003WA 010084068210394721SM719080003WA11190218 SM719080004WA 010084068210394721SM719080004WA11190218 SM719080005WA 010084068210394721SM719080005WA11190218 SM719080006WA 010084068210394721SM719080006WA11190218 SM719080007WA 010084068210394721SM719080007WA11190218 SM719080008WA 010084068210394721SM719080008WA11190218 SM719080009WA 010084068210394721SM719080009WA11190218 SM719080010WA 010084068210394721SM719080010WA11190218 SM718420041WA 010084068210394721SM718420041WA11181018 SM718420042WA 010084068210394721SM718420042WA11181018 SM718490023WA 010084068210394721SM718490023WA11181205 SM719120022WA 010084068210394721SM719120022WA11190322 SM719120023WA 010084068210394721SM719120023WA11190322 SM719120024WA 010084068210394721SM719120024WA11190322 SM719120025WA 010084068210394721SM719120025WA11190322 SM719120026WA 010084068210394721SM719120026WA11190322 SM719120027WA 010084068210394721SM719120027WA11190322 SM719120028WA 010084068210394721SM719120028WA11190322 SM719120029WA 010084068210394721SM719120029WA11190322 SM719120030WA 010084068210394721SM719120030WA11190322 SM719120031WA 010084068210394721SM719120031WA11190322 SM719120032WA 010084068210394721SM719120032WA11190322 SM719120033WA 010084068210394721SM719120033WA11190322 SM719120034WA 010084068210394721SM719120034WA11190322 SM719120038WA 010084068210394721SM719120038WA11190323 SM719120039WA 010084068210394721SM719120039WA11190323 SM719120040WA 010084068210394721SM719120040WA11190323 SM719120041WA 010084068210394721SM719120041WA11190323 SM719120042WA 010084068210394721SM719120042WA11190323 SM719120043WA 010084068210394721SM719120043WA11190323 SM719120044WA 010084068210394721SM719120044WA11190323 SM719120045WA 010084068210394721SM719120045WA11190323 SM719120046WA 010084068210394721SM719120046WA11190323 SM719120047WA 010084068210394721SM719120047WA11190323 SM719120048WA 010084068210394721SM719120048WA11190323 SM719120049WA 010084068210394721SM719120049WA11190323 SM719120050WA 010084068210394721SM719120050WA11190323 SM719130009WA 010084068210394721SM719130009WA11190327 SM719130015WA 010084068210394721SM719130015WA11190328 SM719130016WA 010084068210394721SM719130016WA11190328 SM719130017WA 010084068210394721SM719130017WA11190328 SM719130018WA 010084068210394721SM719130018WA11190328 SM719150054WA 010084068210394721SM719150054WA11190411 SM718460029WA 010084068210394721SM718460029WA11181115 SM718460030WA 010084068210394721SM718460030WA11181115 SM718460031WA 010084068210394721SM718460031WA11181115 SM719100015WA 010084068210394721SM719100015WA11190306 SM719100016WA 010084068210394721SM719100016WA11190306 SM719100017WA 010084068210394721SM719100017WA11190306 SM719100018WA 010084068210394721SM719100018WA11190306 SM719100019WA 010084068210394721SM719100019WA11190306 SM719100020WA 010084068210394721SM719100020WA11190306 SM719100021WA 010084068210394721SM719100021WA11190306 SM719100022WA 010084068210394721SM719100022WA11190306 SM719100026WA 010084068210394721SM719100026WA11190307 SM719100027WA 010084068210394721SM719100027WA11190307 SM719100028WA 010084068210394721SM719100028WA11190307 SM719100029WA 010084068210394721SM719100029WA11190307 SM719120003WA 010084068210394721SM719120003WA11190318 SM718340021WA 010084068210394721SM718340021WA11180823 SM719150052WA 010084068210394721SM719150052WA11190411 SM719150053WA 010084068210394721SM719150053WA11190411 SM719090023WA 010084068210394721SM719090023WA11190227 SM719120004WA 010084068210394721SM719120004WA11190318 SM719140034WA 010084068210394721SM719140034WA11190404 SM719140035WA 010084068210394721SM719140035WA11190404 SM719140036WA 010084068210394721SM719140036WA11190404 SM719150055WA 010084068210394721SM719150055WA11190411 SM719150056WA 010084068210394721SM719150056WA11190411 SM718340028WA 010084068210394721SM718340028WA11180824 SM718340029WA 010084068210394721SM718340029WA11180824 SM719100006WA 010084068210394721SM719100006WA11190304 SM718470001WA 010084068210394721SM718470001WA11181119 SM718470002WA 010084068210394721SM718470002WA11181119 SM718470003WA 010084068210394721SM718470003WA11181119 SM719150051WA 010084068210394721SM719150051WA11190411 SM718430017WA 010084068210394721SM718430017WA11181023 SM718430018WA 010084068210394721SM718430018WA11181023 SM718430019WA 010084068210394721SM718430019WA11181023 SM718430020WA 010084068210394721SM718430020WA11181023 SM718370002WA 010084068210394721SM718370002WA11180910 SM718370003WA 010084068210394721SM718370003WA11180910 SM718370014WA 010084068210394721SM718370014WA11180912 SM718460025WA 010084068210394721SM718460025WA11181113 SM718460026WA 010084068210394721SM718460026WA11181113 SM718460027WA 010084068210394721SM718460027WA11181113 SM718460033WA 010084068210394721SM718460033WA11181115 SM718460034WA 010084068210394721SM718460034WA11181115 SM718460035WA 010084068210394721SM718460035WA11181115 SM718460036WA 010084068210394721SM718460036WA11181115 SM718460037WA 010084068210394721SM718460037WA11181115 SM718480004WA 010084068210394721SM718480004WA11181127 SM718480020WA 010084068210394721SM718480020WA11181128 SM718480021WA 010084068210394721SM718480021WA11181128 SM719090011WA 010084068210394721SM719090011WA11190226 SM719090012WA 010084068210394721SM719090012WA11190226 SM719090013WA 010084068210394721SM719090013WA11190226 SM719090035WA 010084068210394721SM719090035WA11190228 SM719090055WA 010084068210394721SM719090055WA11190302 SM719100001WA 010084068210394721SM719100001WA11190304 SM719100002WA 010084068210394721SM719100002WA11190304 SM719110005WA 010084068210394721SM719110005WA11190312 SM719040027WA 010084068210394721SM719040027WA11190123 SM719040028WA 010084068210394721SM719040028WA11190123 SM719040029WA 010084068210394721SM719040029WA11190124 SM719040030WA 010084068210394721SM719040030WA11190124 SM719040031WA 010084068210394721SM719040031WA11190124 SM719040032WA 010084068210394721SM719040032WA11190124 SM719040033WA 010084068210394721SM719040033WA11190124 SM719040034WA 010084068210394721SM719040034WA11190124 SM718420030WA 010084068210394721SM718420030WA11181017 SM718420031WA 010084068210394721SM718420031WA11181018 SM718420032WA 010084068210394721SM718420032WA11181018 SM718420033WA 010084068210394721SM718420033WA11181018 SM718420034WA 010084068210394721SM718420034WA11181018 SM718420035WA 010084068210394721SM718420035WA11181018 SM718420036WA 010084068210394721SM718420036WA11181018 SM718470008WA 010084068210394721SM718470008WA11181119 SM718470027WA 010084068210394721SM718470027WA11181120 SM718470037WA 010084068210394721SM718470037WA11181121 SM718470038WA 010084068210394721SM718470038WA11181121 SM718470039WA 010084068210394721SM718470039WA11181121 SM718470040WA 010084068210394721SM718470040WA11181121 SM718470041WA 010084068210394721SM718470041WA11181121 SM718470042WA 010084068210394721SM718470042WA11181121 SM718470043WA 010084068210394721SM718470043WA11181121 SM719100005WA 010084068210394721SM719100005WA11190304 SM718470009WA 010084068210394721SM718470009WA11181119 SM718470010WA 010084068210394721SM718470010WA11181119 SM718470015WA 010084068210394721SM718470015WA11181120 SM718470016WA 010084068210394721SM718470016WA11181120 SM718470017WA 010084068210394721SM718470017WA11181120 SM718470018WA 010084068210394721SM718470018WA11181120 SM718470019WA 010084068210394721SM718470019WA11181120 SM718470020WA 010084068210394721SM718470020WA11181120 SM718410036WA 010084068210394721SM718410036WA11181012 SM718410037WA 010084068210394721SM718410037WA11181012 SM718470021WA 010084068210394721SM718470021WA11181120 SM719080053WA 010084068210394721SM719080053WA11190223 SM719080054WA 010084068210394721SM719080054WA11190223 SM719080055WA 010084068210394721SM719080055WA11190223 SM719080056WA 010084068210394721SM719080056WA11190223 SM719080057WA 010084068210394721SM719080057WA11190223 SM719080058WA 010084068210394721SM719080058WA11190223 SM719080059WA 010084068210394721SM719080059WA11190223 SM719080060WA 010084068210394721SM719080060WA11190223 SM719080061WA 010084068210394721SM719080061WA11190223 SM719080062WA 010084068210394721SM719080062WA11190223 SM719090014WA 010084068210394721SM719090014WA11190226 SM719090015WA 010084068210394721SM719090015WA11190227 SM719090016WA 010084068210394721SM719090016WA11190227 SM719090017WA 010084068210394721SM719090017WA11190227 SM719090018WA 010084068210394721SM719090018WA11190227 SM719090019WA 010084068210394721SM719090019WA11190227 SM719090020WA 010084068210394721SM719090020WA11190227 SM719090021WA 010084068210394721SM719090021WA11190227 SM719090022WA 010084068210394721SM719090022WA11190227 SM719090024WA 010084068210394721SM719090024WA11190227 SM719090036WA 010084068210394721SM719090036WA11190301 SM719090037WA 010084068210394721SM719090037WA11190301 SM719090038WA 010084068210394721SM719090038WA11190301 SM719090039WA 010084068210394721SM719090039WA11190301 SM719090040WA 010084068210394721SM719090040WA11190301 SM719090041WA 010084068210394721SM719090041WA11190301 SM719090042WA 010084068210394721SM719090042WA11190301 SM719090045WA 010084068210394721SM719090045WA11190302 SM719090046WA 010084068210394721SM719090046WA11190302 SM719090047WA 010084068210394721SM719090047WA11190302 SM719090048WA 010084068210394721SM719090048WA11190302 SM719090049WA 010084068210394721SM719090049WA11190302 SM719090050WA 010084068210394721SM719090050WA11190302 SM719090051WA 010084068210394721SM719090051WA11190302 SM719090052WA 010084068210394721SM719090052WA11190302 SM719090053WA 010084068210394721SM719090053WA11190302 SM718350004WA 010084068210394721SM718350004WA11180830 SM718350005WA 010084068210394721SM718350005WA11180830 SM718470049WA 010084068210394721SM718470049WA11181122 SM718470050WA 010084068210394721SM718470050WA11181122 SM718470051WA 010084068210394721SM718470051WA11181122 SM718470052WA 010084068210394721SM718470052WA11181122 SM718470053WA 010084068210394721SM718470053WA11181122 SM719020028MA 010084068210394721SM719020028MA11190110 SM718460018MA 010084068210394721SM718460018MA11181112 SM718460019MA 010084068210394721SM718460019MA11181112 SM718460020MA 010084068210394721SM718460020MA11181112 SM718390035MA 010084068210394721SM718390035MA11180924 SM718390042MA 010084068210394721SM718390042MA11180926 SM718390043MA 010084068210394721SM718390043MA11180926 SM718390044MA 010084068210394721SM718390044MA11180926 SM718390045MA 010084068210394721SM718390045MA11180926 SM718390046MA 010084068210394721SM718390046MA11180926 SM718390047MA 010084068210394721SM718390047MA11180926 SM718390048MA 010084068210394721SM718390048MA11180926 SM718390049MA 010084068210394721SM718390049MA11180926 SM718390050MA 010084068210394721SM718390050MA11180926 SM719150045MA 010084068210394721SM719150045MA11190409 SM719070038MA 010084068210394721SM719070038MA11190213 SM718040037MA 010084068210394721SM718040037MA11180123 CARESTATION 650C A1, Model Number 1012-9655-000 Mfg. Lot or Serial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