Skip to content

Ultrasound Navigation Software 1.0 and Ultrasound… recallZ-0056-2021 · Brainlab AG · initiated Aug 31, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Brainlab AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Brainlab Ultrasound Navigation Software does not support the modification of the probe's image width.

FDA root cause category: Software design

What the company is doing

The intraoperative Brainlab Ultrasound Navigation Software does not support the modification of the probes image width, an optional feature of the BK Medical Ultrasound System, resulting in an incorrect display of the depth of the ultrasound overlay by the navigation for non-linear ultrasound probes, with the deviation depending on the probe and the difference between the calibrated and used image width.

Recall details

Recall number
Z-0056-2021
Status
Terminated
Initiated by the company
Aug 31, 2020
Recalling firm
Brainlab AG
Quantity
271 units
Product code
HAW
FDA submissions
K092467
Distribution
Worldwide distribution - US Nationwide distribution including in the states of Illinois, Massachusetts, Texas, California, Idaho, Missouri, Oregon, Connecticut and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Curacao, Czech Republic, Egypt, Finland, France, Germany, Switzerland, Netherlands, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kuwait, Malaysia, Myanmar, Norway, Pakistan, Romania, Slovakia, South Africa, Spain, Sweden, Thailand, Turkey and United Kingdom.

Product description

Ultrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x and 3.1.x - Product Usage: intended to be an intra-operative image guided localization system to enable minimally invasive surgery.

Code information (models, lots, serial numbers)

Model/catalogue numbers: 22582 IGSONIC BK MEDICAL SOFTWARE 22582A ULTRASOUND INTEGRATION SW (BK) 22582B ULTRASOUND INTEGRATION SW 22583 LICENSE UPDATE TO IGSONIC SW FOR BK MED 26506 ULTRASOUND SW BK800 SUBSCRIPTION 26506A ULTRASOUND SW BK SUBSCRIPTION 26508A ULTRASOUND SW BK UPDATE  UDI: The GTIN for Ultrasound Navigation 1.0 is: 04056481140137 The GTIN for Cranial/ENT Navigation 3.1 is: 04056481132439

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

11 devices cleared under K092467 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page