CoaguChek XS PST Meters- IVD (Patient Self-Testing)… recallZ-0804-2021 · Roche Diagnostics Operations, Inc. · initiated Dec 2, 2020
Class II Terminated All recalls posted in January 2021
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Roche Diagnostics Operations, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Possible Misinterpretation of Displayed Measuring Units from INR to %Quick or Seconds; providing instructions to users for ensuring the meter is displaying INR results
FDA root cause category: Device Design
What the company is doing
Roche issued on 12/2/2020, UMDC letter notifications via UPS Ground (signature required upon receipt) to Healthcare Professionals, Distributors and Patients provides instructions to users for ensuring the meter is displaying INR results as follows: For the CoaguChek XS Pro system: UMDC for Healthcare Professionals (TP-01147); UMDC TP-01150 (Distributors), including an enclosed UMDC for Healthcare Professionals (TP-01147). For the CoaguChek XS PST system: UMDC for IDTFs (TP-01148) Including an enclosed UMDC for Patients (TP-01142);" UMDC for Clinics (TP-01149);o Including an enclosed UMDC for Patients (TP-01142). The UMDCs posted to the diagnostics.roche.com website. Distributors of CoaguChek Products are instructed as follows: " Read the enclosed Healthcare Professionals UMDC TP-01147 " Within 14 days from receipt of this UMDC distribute a copy of the enclosed Healthcare Professionals UMDC TP-01147 to all customers to whom you have shipped test strips to use with the Coaguchek XS Professional meter within the last 2 years. " If your organization is contractually authorized by Roche to ship test strips to use with the CoaguChek XS Professional meter to redistributors, then your organization must immediately send a copy of this Distributor UMDC TP-01150 and a copy of the enclosed Healthcare Professional UMDC TP-01147 to those re-distributor customers upon your receipt of this UMDC. Such redistributors should take each of the steps included in this Actions Required section of this Distributor UMDC with respect to their own healthcare professional customers. " Complete all sections of the enclosed fax back form (TP-01153); fax or email it according to the instructions on the form. " File this Urgent Medical Device Correction (UMDC) and the enclosed Healthcare Professional UMDC (TP-01147) for future " Healthcare Professionals (HCPs) considered direct government customers (i.e. VA, other government accounts) " If testing a patients INR at your f
Recall details
- Recall number
- Z-0804-2021
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 2, 2020
- Posted by the FDA
- Jan 7, 2021
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Quantity
- 716,880 units
- Product code
- GJS
- FDA submissions
- K062925
- Distribution
- US Nationwide
Product description
CoaguChek XS PST Meters- IVD (Patient Self-Testing) measures blood-clotting time for people who are taking anticoagulation medications such as Coumadin or warfarin. Catalog 04837738001
Code information (models, lots, serial numbers)
All serial numbers
Other devices under the same FDA submission
3 devices cleared under K062925 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- COAGUCHEK XS PT TEST 6 STRIP 4625374160Roche Diagnostics GmbH · TEST, TIME, PROTHROMBIN3 deaths1,681 reports
- CoaguChek XS Patient Self Testing Care Kit 03666794001Roche Diagnostics GmbH · Test, Time, Prothrombin212 reports
- CoaguChek XS Patient Self Testing Care Kit 4837738001Roche Diagnostics GmbH · TEST, TIME, PROTHROMBIN134 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 86951).
- Z-0803-2021 · TerminatedCoaguChek XS Prof (Professional) Meters- IVD professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. Catalog: 048379750…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page