Skip to content

Revolution CT recallZ-2004-2021 · GE Healthcare, LLC · initiated Mar 19, 2021

Class II Completed All recalls posted in July 2021

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GE Healthcare has become aware of a potential issue on the Revolution CT system where a collimator blade does not find its appropriate home position. This issue could result in the delivered dose of x-ray radiation being either larger or smaller than its prescribed value.

FDA root cause category: Radiation Control for Health and Safety Act

What the company is doing

GE Healthcare will perform an initial site assessment and replace the collimator, if needed. This will be followed by bi-weekly remote monitoring of the system s log file. Finally, GE Healthcare will install a currently under development software patch, added to the daily fast calibration to automatically and reliably detect this failure should it occur. All of these actions will be performed at no cost to the customer. For questions call 262-544-3217.

Recall details

Recall number
Z-2004-2021
Recall class
Class II
Status
Completed
Initiated by the company
Mar 19, 2021
Posted by the FDA
Jul 8, 2021
Recalling firm
GE Healthcare, LLC
Quantity
197 (US = 40; OUS = 157)
Product code
JAK
FDA submissions
K133705, K163213, K191777
Distribution
Worldwide distribution - US Nationwide

Product description

Revolution CT

Code information (models, lots, serial numbers)

5590000 5590000-6 5590000-2

Other devices under the same FDA submission

5 devices cleared under K133705, K163213, K191777 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates