HVAD Pump Implant Kit recallZ-2188-2021 · Heartware, Inc. · initiated Jun 3, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Heartware, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when compared to other commercially available left ventricular assist devices, as well as complaints that the internal pump may delay or fail to restart.
FDA root cause category: Unknown/Undetermined by firm
What the company is doing
In the United States, beginning 03-Jun-2021, this notice was delivered to Risk Managers via UPS 2-day delivery in the United States as well as through verbal and in person follow ups from the Medtronic field staff. In addition, a courtesy copy was sent to all implanting and following physicians via UPS 2-day delivery. Finally, an e-mail notice was sent as a courtesy copy to all consignees who had opted into communication from Medtronic. Medtronic began an additional communication on 7 July 2021 to all consignees who, according to Medtronic records had unexpired product (see Attachment D Product List) at the time of the initial communication on 03-JUN-2021. The consignees are being asked to confirm receipt of the notification and return all affected product within their possession. This communication also stresses the importance of not returning peripherals (batteries, cables, patient packs, etc.) and surgical tool kits if they still have patients on support. In the US, this communication was delivered to VAD Coordinators via UPS 2-day delivery. Outside of the United States, beginning 07-JUL-2021 this communication will be delivered to consignees.
Recall details
- Recall number
- Z-2188-2021
- Status
- Open, Classified
- Initiated by the company
- Jun 3, 2021
- Recalling firm
- Heartware, Inc.
- Quantity
- 647 units
- Product code
- DSQ
- FDA submissions
- P100047
- Distribution
- Worldwide distribution. US nationwide and Armenia, AUSTRIA, Bahrain, Belgium, CROATIA, CZECH REPUBLIC, Denmark, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, Hungary, Israel, ITALY, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, LITHUANIA, LUXEMBOURG, Macedonia, NETHERLANDS, Norway, POLAND, Qatar, Romania, SAUDI ARABIA, Serbia, Slovakia, SOUTH AFRICA, Spain, Sweden, SWITZERLAND, TAIWAN, Turkey, Ukraine, UNITED ARAB EMIRATES, and United Kingdom.
Product description
HeartWare HVAD Pump Implant Kit, REF 1103
Code information (models, lots, serial numbers)
GTIN: 00888707003261 All serial numbers
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
Product lines
Other devices under the same FDA submission
32 devices cleared under P100047 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HeartWare® Controller 1403USHeartware Inc. · Ventricular (Assisst) Bypass3 recalls10 deaths1,495 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assisst) Bypass1 recall1,126 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass2 recalls606 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall615 reports
- HVAD® Pump Implant Kit MCS1705PUHeartware Inc. · Ventricular (Assist) Bypass7 recalls36 deaths203 reports
- HVAD® Controller 1403USHeartware Inc. · Ventricular (Assist) Bypass548 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall487 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assist) Bypass1 recall281 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall159 reports
- HVAD® Pump Implant Kit MCS1705PUHeartware Inc. · Ventricular (Assist) Bypass4 recalls1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 88197).
- Z-2189-2021 · Open, ClassifiedHeartWare HVAD Pump Implant Kit, REF 1104
- Z-2190-2021 · Open, ClassifiedHeartWare HVAD Pump Implant Kit, REF 1125
- Z-2191-2021 · Open, ClassifiedHeartWare HVAD Pump Implant Kit, REF 1153
- Z-2192-2021 · Open, ClassifiedHeartWare HVAD Driveline Extension Cable, REF 100
- Z-2193-2021 · Open, ClassifiedHeartWare HVAD Driveline Extension Cable, REF 100US
- Z-2194-2021 · Open, ClassifiedHeartWare HVAD Implant Kit, REF 1104JP
- Z-2195-2021 · Open, ClassifiedHeartWare HVAD Implant Kit, REF MCS1705PU
- Z-2196-2021 · Open, ClassifiedHeartWare HVAD Outflow Graft, REF MCS1725OG
- Z-2197-2021 · Open, ClassifiedHeartWare HVAD Pump Accessories, REF MCS1753AK
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page