Allura Xper FD recallZ-0416-2022 · Philips North America Llc · initiated Nov 19, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Suspension structure and fall down, may lead to patient, user or service engineer harm that may require medical intervention.

FDA root cause category: Other

What the company is doing

Philips issued Medical Device Correction letter to US consignees on 19-Nov-2021. Letters outside of the U.S.A. will be distributed through the Philips Markets Organization. Letter states reason for recall, health risk and action to take: Move the MCS with a 7th or 8th monitor only before or after a procedure. " Do not position the MCS above the patient/sterile field during the procedure. " Place this Field Safety Notice with the system documentation until your system has been fixed. " Circulate this notice to all users of this device so they are aware of the product issues. " Return the attached reply form to Philips to confirm that the users of the system have reviewed and understood this Field Safety Notice. A Philips representative will secure the set of screws of all affected systems. You will be contacted by your local Philips representative to schedule the service. If you need any further information, please contact your local Philips representative (reference to FC072200488).

Recall details

Recall number
Z-0416-2022
Status
Terminated
Initiated by the company
Nov 19, 2021
Recalling firm
Philips North America Llc
Quantity
USA: 5 systems R.O.W.: 15 systems
Product code
OWB
FDA submissions
K141979, K130842, K161563
Distribution
Worldwide distribution - US Nationwide distribution in the states of DC, GA, MA, TX, VA and the countries of Dominican Republic, INDIA, Japan, Poland, Spain, Taiwan.

Product description

Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. Model Number: 722005, 722006, 722008, 722010, 722012, 722013, 722025, 722026, 722027, 722028.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

US Serial Number/UDI: 1 (01)00884838054202(21)1; 527 (01)00884838059054(21)527; 408 (01)00884838054202(21)408; 306 (01)00884838054202(21)306; 61 (01)00884838059030(21)61 OUS: Model Product S/N UDI: 722006 Allura Xper FD20 1997 (01)00884838059191(10)1997; 722010 Allura Xper FD10 1177 (01)00884838059030(21)1177; 722010 Allura Xper FD10 1178 (01)00884838059030(21)1178; 722013 Allura Xper FD20 Biplane 544;; 722010 Allura Xper FD10 514 (01)00884838059030(21)514; 722013 Allura Xper FD20 Biplane 271; 722025 Allura Xper FD20 Biplane OR Table 6; 722005 Allura Xper FD10/10 519; 722008 Allura Xper FD20 Biplane 531; 722012 Allura Xper FD20 1901 (01)00884838059054(21)1901; 722012 Allura Xper FD20 2505 (01)00884838059054(21)2505; 722013 Allura Xper FD20 Biplane 137; 722026 Allura Xper FD10 439 (01)00884838054189(21)439; 722134 Field ext. Xper vascular systems R7.6 98736; 722027 Allura Xper FD10/10 172 (01)00884838054196(21)172;