Allura Xper FD Allura Xper FD10/10 (REF 722027)Recalls & reported problems

Philips Medical Systems Nederland B.V. · Biplane Interventional X-Ray System with floor-mounted frontal stand (10-inch flat detector) and ceiling-mounted lateral stand ( 10-inch flat detector). · Class 2

All models of Allura Xper FD (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

18 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Philips Allura Xper is a biplane interventional X-ray system. Of the 603 reports naming it, most describe failed activation or radiation output, and some describe display or boot problems. The reports include no injuries and no deaths, and reports fell from 224 in the previous 12 months to 49 in the latest 12 months. Of its 19 recalls, 18 are open, covering aging hard drives, software faults, power loss and finger entrapment on the patient table.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
603
Deaths reported
0
Injuries reported
0
Recalls naming it
19

Summary

The FDA's MAUDE database contains 603 adverse event reports that list this exact device (DI 00884838054196), received between Feb 2, 2018 and May 8, 2026 — classified as 603 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (334 reports).

19 FDA recall records name this device's identifier, 18 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 13 reportsDec 2023: 14 reportsJan 2024: 9 reportsFeb 2024: 14 reportsMar 2024: 14 reportsApr 2024: 11 reportsMay 2024: 7 reportsJun 2024: 10 reportsJul 2024: 9 reportsAug 2024: 10 reportsSep 2024: 11 reportsOct 2024: 13 reportsNov 2024: 10 reportsDec 2024: 3 reportsJan 2025: 12 reportsFeb 2025: 15 reportsMar 2025: 4 reportsApr 2025: 12 reportsMay 2025: 13 reportsJun 2025: 21 reportsJul 2025: 23 reportsAug 2025: 52 reportsSep 2025: 16 reportsOct 2025: 43 reportsNov 2025: 40 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 4 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202652
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 29, 2026Class IIOpen, Classified

    Firmware issue potentially leading to loss of geometry movements.

    Z-2928-2026 · Root cause: Software design · Recall details · FDA record ↗

  2. Jun 17, 2026Class IIOpen, Classified

    Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.

    Z-2891-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Apr 30, 2026Class IIOpen, Classified

    It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.

    Z-2259-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Apr 28, 2026Class IIOpen, Classified

    The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

    Z-2228-2026 · Root cause: Component design/selection · Recall details · FDA record ↗

  5. Mar 3, 2026Class IIOpen, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1662-2026 · Root cause: Device Design · Recall details · FDA record ↗

  6. Dec 15, 2025Class IIOpen, Classified

    Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

    Z-1074-2026 · Root cause: Other · Recall details · FDA record ↗

  7. Sep 3, 2025Class IIOpen, Classified

    Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

    Z-2624-2025 · Root cause: Device Design · Recall details · FDA record ↗

  8. Sep 3, 2025Class IIOpen, Classified

    The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

    Z-0003-2026 · Root cause: Device Design · Recall details · FDA record ↗

  9. Aug 1, 2025Class IIOpen, Classified

    Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.

    Z-2504-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  10. Aug 1, 2025Class IIOpen, Classified

    Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.

    Z-2503-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  11. Jun 30, 2025Class IIOpen, Classified

    Potential for temporary loss of imaging (X-ray) functionality due to software issue.

    Z-2189-2025 · Root cause: Software design · Recall details · FDA record ↗

  12. Jan 13, 2025Class IOpen, Classified

    Possibility of the patient falling from the table related to the mattress used on the patient table.

    Z-1143-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  13. Jan 13, 2025Class IIOpen, Classified

    Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

    Z-1078-2025 · Root cause: Device Design · Recall details · FDA record ↗

  14. Aug 22, 2024Class IIOpen, Classified

    Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.

    Z-3166-2024 · Root cause: Device Design · Recall details · FDA record ↗

  15. Jan 2, 2024Class IIOpen, Classified

    Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.

    Z-1151-2024 · Root cause: Process design · Recall details · FDA record ↗

  16. Jan 2, 2024Class IIOpen, Classified

    Dual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.

    Z-1156-2024 · Root cause: Process design · Recall details · FDA record ↗

  17. Jan 2, 2024Class IIOpen, Classified

    Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.

    Z-1144-2024 · Root cause: Process design · Recall details · FDA record ↗

  18. Feb 2, 2022Class IIOpen, Classified

    Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.

    Z-0874-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  19. Nov 19, 2021Class IITerminated

    Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Suspension structure and fall down, may lead to patient, user or service engineer harm that may require medical intervention.

    Z-0416-2022 · Root cause: Other · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation Failure334
  • Radiation Output Failure176
  • No Display/Image48
  • Positioning Failure30
  • Detachment of Device or Device Component5
  • Display or Visual Feedback Problem4
  • Activation, Positioning or Separation Problem2
  • Erratic or Intermittent Display2

Patient problems

  • No Clinical Signs, Symptoms or Conditions599
  • Insufficient Information6
  • Cardiac Arrest1

Recent adverse event reports

The 25 most recent of 603 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM BECAME UNRESPONSIVE AND COULD NOT PERFORM EXPOSURE. THE SYSTEM WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25120842 ↗
  • MalfunctionMay 4, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM GAVE AN ALERT DURING SYSTEM START-UP, WHICH COULD IMPACT BOOTING. THE SYSTEM WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25069119 ↗
  • MalfunctionMar 30, 2026Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO ACTIVATE X-RAYS. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24728703 ↗
  • MalfunctionMar 21, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM GAVE AN ALERT INDICATING IT WAS CONNECTING TO THE HOST, PREVENTING THE SYSTEM FROM BOOTING. THE DEVICE WAS NOT IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24656596 ↗
  • MalfunctionMar 10, 2026Activation Failure

    IT HAS BEEN REPORTED TO PHILIPS THAT THE SYSTEM DID NOT BOOT UP SUCCESSFULLY. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24559103 ↗
  • MalfunctionMar 9, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S GEOMETRY COMPUTER WAS UNABLE TO BOOT, PREVENTING GEOMETRY MOVEMENTS. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24546675 ↗
  • MalfunctionJan 13, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS OUTSIDE CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24056748 ↗
  • MalfunctionDec 24, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM GAVE AN ALERT INDICATING THERE WAS A GEOMETRY ERROR. UPON REBOOTING, THE SYSTEM WAS UNABLE TO PERFORM GEOMETRY MOVEMENTS. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23886187 ↗
  • MalfunctionDec 12, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT PROPERLY. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23790513 ↗
  • MalfunctionNov 26, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S FLEXVISION COMPUTER WAS UNABLE TO REACH TO THE HOST COMPUTER, PREVENTING THE SYSTEM FROM BOOTING. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23655393 ↗
  • MalfunctionNov 25, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS OUTSIDE CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23652566 ↗
  • MalfunctionNov 25, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. NO HARM WAS REPORTED TO PHILIPS. THE DEVICE WAS NOT IN CLINICAL USE AT THE TIME OF REPORTED EVENT. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23644433 ↗
  • MalfunctionNov 24, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE FLEXVISION PC FAILED TO BOOT WITH 00:00 ERROR ON AN ALLURA XPER FD10/10. THE CLINICAL USE OF THE SYSTEM WAS OUTSIDE OF USE. THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23630337 ↗
  • MalfunctionNov 24, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE WORKSTATION FAILED TO BOOT UP DURING THE SETUP PHASE ON AN ALLURA XPER FD10/10. THE CLINICAL USE OF THE SYSTEM WAS OUTSIDE OF USE. THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23629662 ↗
  • MalfunctionNov 21, 2025Radiation Output Failure; Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT 'WIRELESS FOOTSWITCH CANNOT BE USED'. THIS IS CONNECTED TO LOSS OF IMAGE RELATED TO A ALLURA XPER FD10/10. THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23615019 ↗
  • MalfunctionNov 20, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE X-RAY PC FAILED TO BOOT DURING CLINICAL USE ON AN ALLURA XPER FD10/10. THE CLINICAL USE OF THE SYSTEM WAS IN USE. THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23603691 ↗
  • MalfunctionNov 19, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM STUCK AT ZERO COUNTDOWN DURING BOOT-UP ON A ALLURA XPER FD10/10. THE CLINICAL USE OF THE SYSTEM WAS IN USE. THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23597271 ↗
  • MalfunctionNov 19, 2025No Display/Image; No Display/Image

    IT WAS REPORTED TO PHILIPS THAT STORAGE NOT AVAILABLE ERROR PREVENTED IMAGING ON A ALLURA XPER FD10/10. THE CLINICAL USE OF THE SYSTEM WAS UNKNOWN. THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23595202 ↗
  • MalfunctionNov 19, 2025Radiation Output Failure; Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT EXPOSURE WAS NOT POSSIBLE DUE TO AN IMAGE PROCESSING COMMUNICATION ERROR ON AN ALLURA XPER FD10/10. THE CLINICAL USE OF THE SYSTEM WAS IN USE. THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23595193 ↗
  • MalfunctionNov 19, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THERE WAS AN INTERMITTENT IPPC BOOT FAILURE; ACQUISITION DISK ERROR ON A ALLURA XPER FD10/10. THE CLINICAL USE OF THE SYSTEM WAS IN USE. THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23595190 ↗
  • MalfunctionNov 19, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT CONNECTION TO ALLURA FAILED; CORRUPT STARTUP REPORTED ON A ALLURA XPER FD10/10. THE CLINICAL USE OF THE SYSTEM WAS IN USE. THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23592152 ↗
  • MalfunctionNov 19, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT STARTUP FAILURE OF THE IMAGE STORAGE DEVICE OCCURRED, MAKING EXPOSURE NOT POSSIBLE ON A ALLURA XPER FD10/10. THE CLINICAL USE OF THE SYSTEM WAS IN USE. THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23592149 ↗
  • MalfunctionNov 19, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM FAILED TO BOOT, DISPLAYING A BLUE SCREEN DURING STARTUP ON A ALLURA XPER FD10/10. THE CLINICAL USE OF THE SYSTEM WAS IN USE. THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23592145 ↗
  • MalfunctionNov 18, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT FLEXVISION NOT TURNING ON. THIS IS CONNECTED TO STARTUP ISSUES RELATED TO A ALLURA XPER FD10/10. THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23582651 ↗
  • MalfunctionNov 18, 2025No Display/Image; No Display/Image

    IT WAS REPORTED TO PHILIPS THAT THERE WAS A LOSS OF IMAGE MID-CASE, AND THE SYSTEM REQUIRED A REBOOT ON AN ALLURA XPER FD10/10. THE CLINICAL USE OF THE SYSTEM WAS IN USE. THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23581001 ↗

Other Interventional Fluoroscopic X-Ray System devices

All Interventional Fluoroscopic X-Ray System devices →

Sources and updates