HVAD Pump Implant Kit recallZ-1110-2022 · Medtronic Inc · initiated Mar 30, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Errors and inconsistencies that were identified in the Instruction for Use (IFU), Emergency Responder Guide (ERG), and Patient Manual (PM) for the HeartWare Ventricular Assist Device (HVAD) system.
FDA root cause category: Labeling design
What the company is doing
Starting 30-Mar-2022, Medtronic issued an Urgent Medical Device Notice by letter to VAD Coordinators at customer sites in order to inform them of the errors and the corrections impacting the Instructions for Use, Patient Manual, and Emergency Responder Guide for the HVAD System. In addition to the Urgent Medical Device Notice a patient template was provided to facilitate the VAD Coordinators in notifying their patients of the issues to the Patient Manual. Customer Instructions: "Please ensure your patients on support are aware of the changes to the PM. "Please share this notice with everyone in your organization who needs to be aware. "Continue to monitor patients per your practice's standard follow-up procedures. "Please complete the enclosed Customer Confirmation Form and email to [email protected]. For questions regarding this material, customers should contact their Medtronic Field Representative, MCS Customer Service or (877) 367-4823. .
Recall details
- Recall number
- Z-1110-2022
- Status
- Open, Classified
- Initiated by the company
- Mar 30, 2022
- Recalling firm
- Medtronic Inc
- Quantity
- 21,118 pieces of product literature
- Product code
- DSQ
- FDA submissions
- P100047
- Distribution
- Worldwide Distribution: US (nationwide) and OUS to countries of: Australia, Austria, Belgium, Denmark, France, Germany, Greece, Hungary, INDIA, Italy, Japan, Luxembourg, Netherlands, New Zealand, Norway, Poland, Slovakia, South Africa, Spain, Switzerland, Turkey, United Kingdom
Product description
Instructions for Use, Patient Manuals, and Emergency Responder Guides for HeartWare Ventricular Assist Device (HVAD) System. HVAD System. IFUs are distributed as part of the Pump Implant Kits models: 1103, MCS1705PU, 1104, and 1104JP. Additionally, IFUs, ERGs, and PMs are distributed with unique Catalog / Model Numbers: IFU00358, IFU00375, IFU00376, GR00526, IFU00520, IFU00526, IFU00593, IFU00614, M017475C001
Code information (models, lots, serial numbers)
Show the full code information (39k characters)
Device Model numbers affected: 1104: GTIN 00888707002028, Serial Numbers: HW28848, HW29011, HW29154, HW29199, HW29211, HW29212, HW29219, HW29229, HW29246, HW29248, HW29249, HW29252, HW29265, HW29266, HW29268, HW29271, HW29273, HW29274, HW29276, HW29279, HW29282, HW29283, HW29284, HW29285, HW29286, HW29287, HW29290, HW29293, HW29302, HW29309, HW29310, HW29329, HW29333, HW29387, HW29426, HW29465, HW29472, HW29485, HW29502, HW29503, HW29522, HW29526, HW29532, HW29533, HW29534, HW29535, HW29536, HW29538, HW29539, HW29540, HW29541, HW29542, HW29543, HW29546, HW29547, HW29548, HW29549, HW29551, HW29552, HW29562, HW29567, HW29579, HW29583, HW29586, HW29587, HW29588, HW29589, HW29590, HW29592, HW29593, HW29594, HW29595, HW29600, HW29602, HW29603, HW29604, HW29612, HW29629, HW29635, HW29650, HW29660, HW29661, HW29664, HW29665, HW29666, HW29667, HW29668, HW29669, HW29670, HW29671, HW29672, HW29673, HW29674, HW29675, HW29676, HW29685, HW29686, HW29687, HW29700, HW29701, HW29702, HW29717, HW29727, HW29741, HW29754, HW29755, HW29764, HW29776, HW29780, HW29781, HW29782, HW29784, HW29786, HW29787, HW29788, HW29790, HW29791, HW29798, HW29815, HW29820, HW29825, HW29832, HW29836, HW29850, HW29866, HW29867, HW29873, HW29877, HW29887, HW29894, HW29895, HW29896, HW29900, HW29903, HW29904, HW29909, HW29911, HW29915, HW29922, HW29923, HW29927, HW29928, HW29932, HW29949, HW29965, HW29976, HW29983, HW29990, HW29995, HW30038, HW30039, HW30040, HW30079, HW30080, HW30081, HW30082, HW30083, HW30084, HW30085, HW30086, HW30087, HW30088, HW30089, HW30102, HW30103, HW30104, HW30105, HW30106, HW30107, HW30108, HW30109, HW30110, HW30111, HW30112, HW30113, HW30115, HW30116, HW30117, HW30118, HW30147, HW30148, HW30149, HW30150, HW30151, HW30152, HW30153, HW30154, HW30155, HW30156, HW30157, HW30178, HW30193, HW30194, HW30195, HW30196, HW30197, HW30198, HW30199, HW30200, HW30201, HW30202, HW30204, HW30205, HW30207, HW30208, HW30209, HW30210, HW30211, HW30212, HW30213, HW30214, HW30224, HW30225, HW30226, HW30227, HW30231, HW30267, HW30319, HW30320, HW30342, HW30358, HW30360, HW30363, HW30364, HW30365, HW30366, HW30367, HW30369, HW30370, HW30372, HW30376, HW30378, HW30380, HW30382, HW30383, HW30384, HW30385, HW30387, HW30390, HW30392, HW30393, HW30397, HW30398, HW30399, HW30400, HW30402, HW30404, HW30406, HW30407, HW30408, HW30409, HW30410, HW30430, HW30449, HW30455, HW30461, HW30462, HW30463, HW30471, HW30472, HW30473, HW30475, HW30483, HW30484, HW30485, HW30486, HW30491, HW30494, HW30498, HW30595, HW30596, HW30597, HW30598, HW30599, HW30600, HW30601, HW30602, HW30603, HW30604, HW30606, HW30607, HW30608, HW30651, HW30652, HW30654, HW30656, HW30768, HW30769, HW30770, HW30771, HW30808, HW30809, HW30810, HW30811, HW30813, HW30814, HW30815, HW30816, HW30818, HW30820, HW30821, HW30822, HW30823, HW30824, HW30825, HW30829, HW30830, HW30863, HW30864, HW30865, HW30866, HW30867, HW30868, HW30869, HW30870, HW30871, HW30953, HW30954, HW30955, HW30956, HW30957, HW30958, HW30959, HW30960, HW30961, HW30962, HW30963, HW30964, HW30966, HW30967, HW30968, HW30970, HW30971, HW30972, HW30973, HW30974, HW30976, HW31015, HW31020, HW31072, HW31076, HW31087, HW31101, HW31103, HW31104, HW31105, HW31107, HW31111, HW31113, HW31114, HW31115, HW31117, HW31119, HW31128, HW31132, HW31134, HW31192, HW31219, HW31223, HW31226, HW31228, HW31232, HW31235, HW31242, HW31244, HW31249, HW31251, HW31254, HW31255, HW31258, HW31260, HW31261, HW31262, HW31267, HW31383, HW31386, HW31387, HW31388, HW31390, HW31398, HW31399, HW31402, HW31409, HW31410, HW31411, HW31412, HW31414, HW31445, HW31446, HW31447, HW31448, HW31449, HW31451, HW31452, HW31453, HW31454, HW31455, HW31456, HW31457, HW31458, HW31541, HW31543, HW31544, HW31545, HW31547, HW31548, HW31549, HW31550, HW31552, HW31559, HW31564, HW31571, HW31573, HW31578, HW31579, HW31686, HW31694, HW31697, HW31731, HW31749, HW31750, HW31777, HW31778, HW31810, HW31819, HW31830, HW31831, HW31832, HW31834, HW31835, HW31836, HW31838, HW31839, HW31840, HW31842, HW31843, HW31844, HW31845, HW31846, HW31847, HW31849, HW31855, HW31858, HW31876, HW31883, HW31944, HW31947, HW31979, HW31981, HW31986, HW31987, HW31988, HW31989, HW31991, HW32068, HW32116, HW32121, HW32141, HW32143, HW32156, HW32159, HW32161, HW32162, HW32164, HW32167, HW32172, HW32179, HW32189, HW32190, HW32192, HW32193, HW32194, HW32197, HW32199, HW32201, HW32204, HW32206, HW32209, HW32210, HW32214, HW32215, HW32216, HW32218, HW32220, HW32222, HW32223, HW32224, HW32228, HW32229, HW32230, HW32231, HW32248, HW32258, HW32259, HW32260, HW32261, HW32273, HW32279, HW32281, HW32285, HW32286, HW32290, HW32292, HW32315, HW32335, HW32357, HW32386, HW32393, HW32400, HW32419, HW32423, HW32426, HW32427, HW32457, HW32458, HW32459, HW32464, HW32465, HW32468, HW32470, HW32471, HW32472, HW32479, HW32499, HW32504, HW32508, HW32509, HW32510, HW32511, HW32516, HW32534, HW34001, HW34002, HW34025, HW34034, HW34035, HW34036, HW34037, HW34038, HW34040, HW34041, HW34042, HW34043, HW34044, HW34045, HW34046, HW34047, HW34048, HW34051, HW34066, HW34096, HW34100, HW34104, HW34106, HW34109, HW34110, HW34111, HW34113, HW34117, HW34121, HW34126, HW34151, HW34169, HW34179, HW34181, HW34186, HW34187, HW34192, HW34193, HW34196, HW34197, HW34202, HW34204, HW34205, HW34207, HW34209, HW34211, HW34226, HW34317, HW34319, HW34326, HW34328, HW34329, HW34330, HW34331, HW34336, HW34342, HW34344, HW34347, HW34368, HW34369, HW34371, HW34372, HW34373, HW34375, HW34376, HW34377, HW34378, HW34379, HW34425, HW34427, HW34428, HW34429, HW34432, HW34446, HW34495, HW34519, HW34520, HW34521, HW34522, HW34523, HW34546, HW34547, HW34550, HW34551, HW34555, HW34556, HW34588, HW34590, HW34596, HW34603, HW34605, HW34609, HW34610, HW34643, HW34673, HW34752, HW34768, HW34805, HW34806, HW34808, HW34810, HW34824, HW34840, HW34842, HW34845, HW34866, HW34868, HW34877, HW34902, HW34903, HW34904, HW34908, HW34910, HW34913, HW34914, HW34923, HW34924, HW34925, HW34928, HW34933, HW34934, HW34938, HW34939, HW34944, HW34946, HW34947, HW34948, HW34949, HW34957, HW34979, HW34990, HW34997, HW35012, HW35032, HW35099, HW35122, HW35123, HW35124, HW35125, HW35128, HW35131, HW35133, HW35134, HW35136, HW35138, HW35141, HW35143, HW35161, HW35189, HW35191, HW35204, HW35205, HW35209, HW35211, HW35212, HW35213, HW35214, HW35215, HW35220, HW35284, HW35305, HW35317, HW35319, HW35321, HW35324, HW35325, HW35326, HW35332, HW35334, HW35337, HW35339, HW35340, HW35341, HW35357, HW35360, HW35361, HW35367, HW35385, HW35464, HW35477, HW35478, HW35479, HW35480, HW35481, HW35482, HW35483, HW35487, HW35488, HW35489, HW35492, HW35493, HW35494, HW35495, HW35499, HW35500, HW35501, HW35502, HW35504, HW35506, HW35508, HW35534, HW35535, HW35539, HW35540, HW35541, HW35542, HW35543, HW35545, HW35548, HW35563, HW35577, HW35578, HW35580, HW35582, HW35583, HW35584, HW35586, HW35587, HW35588, HW35624, HW35625, HW35626, HW35627, HW35628, HW35629, HW35630, HW35631, HW35632, HW35633, HW35634, HW35635, HW35636, HW35637, HW35639, HW35641, HW35644, HW35646, HW35647, HW35668, HW35669, HW35670, HW35671, HW35673, HW35674, HW35676, HW35677, HW35678, HW35680, HW35681, HW35682, HW35683, HW35684, HW35686, HW35687, HW35688, HW35689, HW35690, HW35691, HW35692, HW35693, HW35695, HW35696, HW35697, HW35698, HW35701, HW35702, HW35703, HW35704, HW35705, HW35707, HW35708, HW35730, HW35731, HW29208R01, HW29209R01, HW29225R01, HW29226R01, HW29241R01, HW29242R01, HW29245R01, HW29247R01; GTIN 00888707003063, Serial Numbers: HW41903, HW41911, HW41924, HW41925, HW41926, HW41927, HW41928, HW41929, HW41950, HW41953, HW42004, HW42006, HW42007, HW42008, HW42018, HW42019, HW42089, HW42158, HW42160, HW42161, HW42219, HW42230, HW42235, HW42237, HW42238, HW42650, HW42651, HW42678, HW42720, HW42724, HW42790, HW42793, HW42797, HW42829, HW42831, HW43009; GTIN 00888707003070, Serial Numbers: HW41892, HW41894, HW41895, HW41897, HW41899, HW41900, HW41901, HW41932, HW41934, HW41936, HW41937, HW41938, HW41939, HW41940, HW41941, HW41971, HW41972, HW41973, HW41974, HW41975, HW41977, HW41986, HW41991, HW41993, HW41994, HW41997, HW42000, HW42001, HW42002, HW42005, HW42009, HW42021, HW42022, HW42026, HW42099, HW42101, HW42103, HW42104, HW42110, HW42111, HW42114, HW42116, HW42148, HW42153, HW42155, HW42157, HW42223, HW42225, HW42228, HW42229, HW42232, HW42233, HW42252, HW42291, HW42307, HW42326, HW42328, HW42329, HW42330, HW42331, HW42332, HW42333, HW42500, HW42666, HW42915; GTIN 00888707003100, Serial Numbers: HW41961, HW41962, HW41964, HW41965, HW41966, HW41967, HW41968, HW42313, HW42660, HW42663, HW42671, HW42672, HW42901; GTIN 00888707003117, Serial Numbers: HW41955, HW41957, HW41958, HW41959, HW41960, HW42023, HW42024, HW42025, HW42061, HW42091, HW42130, HW42292, HW42335, HW42337, HW42339, HW42647, HW42665; GTIN 00888707003209, Serial Numbers: HW41915, HW41916, HW41918, HW41919, HW41945, HW41946, HW41947, HW41948, HW41978, HW41979, HW41981, HW41982, HW41983, HW41984, HW41988, HW41990; GTIN 00888707003254, Serial Numbers: HW41943, HW41944, HW42213; GTIN 00888707005968, Serial Numbers: HW35801, HW35802, HW35803, HW35804, HW35805, HW35806, HW35807, HW35808, HW35809, HW35810, HW35811, HW35812, HW35813, HW35814, HW35815, HW35816, HW35817, HW35818, HW35820, HW35847, HW35848, HW35850, HW35851, HW35852, HW35853, HW35854, HW35855, HW35856, HW35857, HW35858, HW35859, HW35860, HW35861, HW35862, HW35863, HW35864, HW35867, HW35870, HW35875, HW35923, HW35924, HW35927, HW35929, HW35931, HW35935, HW35936, HW35937, HW35938, HW35939, HW35966, HW35967, HW35968, HW35969, HW35970, HW35971, HW35972, HW35975, HW35976, HW35978, HW35988, HW35989, HW35990, HW35991, HW35993, HW35994, HW35995, HW35996, HW35997; GTIN 00888707006330, Serial Numbers: HW31419, HW31699, HW31733, HW32505, HW35393, HW35396, HW35562, HW35571, HW35726, HW35749, HW35766, HW35769, HW35784, HW35829, HW35872, HW36005, HW36006, HW36007, HW36008, HW36009, HW36010, HW36013, HW36018, HW36019, HW36020, HW36021, HW36069, HW36070, HW36071, HW36131, HW36132, HW36138, HW36174, HW36175, HW36176, HW36177, HW36178, HW36179, HW36180, HW36184, HW36185, HW36186, HW36192, HW36193, HW36194, HW36195, HW36197, HW36198, HW36199, HW36200, HW36201, HW36224, HW36225, HW36226, HW36227, HW36228, HW36229, HW36230, HW36231, HW36232, HW36233, HW36248, HW36251, HW36252, HW36253, HW36254, HW36255, HW36256, HW36258, HW36259, HW36260, HW36276, HW36298, HW36299, HW36300, HW36303, HW36305, HW36307, HW36309, HW36310, HW36313, HW36314, HW36316, HW36317, HW36318, HW36319, HW36320, HW36321, HW36323, HW36324, HW36325, HW36326, HW36328, HW36332, HW36333, HW36334, HW36335, HW36354, HW36355, HW36356, HW36357, HW36358, HW36359, HW36360, HW36361, HW36362, HW36363, HW36396, HW36397, HW36398, HW36399, HW36400, HW36401, HW36403, HW36404, HW36405, HW36406, HW36407, HW36408, HW36409, HW36410, HW36411, HW36412, HW36419, HW36428, HW36429, HW36430, HW36432, HW36433, HW36434, HW36435, HW36436, HW36437, HW36438, HW36500, HW36501, HW36502, HW36503, HW36504, HW36505, HW36506, HW36507, HW36508, HW36509, HW36510, HW36511, HW36513, HW36514, HW36515, HW36530, HW36553, HW36554, HW36555, HW36556, HW36557, HW36559, HW36560, HW36561, HW36562, HW36563, HW36564, HW36595, HW36596, HW36597, HW36598, HW36599, HW36600, HW36601, HW36602, HW36603, HW36624, HW36625, HW36626, HW36628, HW36630, HW36631, HW36633, HW36636, HW36637, HW36654, HW36655, HW36659, HW36660, HW36662, HW36664, HW36669, HW36670, HW36671, HW36672, HW36673, HW36677, HW36678, HW36680, HW36681, HW36683, HW36685, HW36702, HW36703, HW36704, HW36705, HW36708, HW36709, HW36710, HW36711, HW36712, HW36713, HW36718, HW36719, HW36720, HW36722, HW36723, HW36724, HW36725, HW36726, HW36727, HW36728, HW36750, HW36752, HW36753, HW36754, HW36755, HW36785, HW36786, HW36788, HW36789, HW36812, HW36813, HW36815, HW36816, HW36817, HW36818, HW36823, HW36827, HW36829, HW36830, HW36831, HW36832, HW36833, HW36834, HW36835, HW36836, HW36837, HW36839, HW36840, HW36853, HW36854, HW36855, HW36877, HW36881, HW36882, HW36883, HW36891, HW36897, HW36898, HW36899, HW36900, HW36901, HW36902, HW36919, HW36936, HW36937, HW36940, HW36942, HW36944, HW36945, HW36964, HW36966, HW36969, HW36982, HW36988, HW36996, HW36997, HW36999, HW37000, HW37002, HW37005, HW37184, HW37185, HW37186, HW37187, HW37188, HW37189, HW37190, HW37193, HW37194, HW37195, HW37196, HW37199, HW37200, HW37210, HW37212, HW37213, HW37214, HW37215, HW37216, HW37218, HW37222, HW37223, HW37225, HW37226, HW37228, HW37230, HW37231, HW37232, HW37233, HW37234, HW37241, HW37242, HW37243, HW37245, HW37246, HW37273, HW37276, HW37350, HW37362, HW37365, HW37366, HW37367, HW37369; 1103: GTIN 00888707002639: Serial Numbers: HW28807R01, HW28965, HW29296, HW29337R01, HW29448, HW29498R01, HW29530, HW29642R01, HW29738, HW29744, HW29865R02, HW29905R01, HW29906, HW29969, HW30008, HW30122, HW30174, HW30233, HW30235, HW30268R01, HW30269R01, HW30312, HW30441, HW30630, HW30673, HW30925, HW31182, HW31259, HW31324, HW31336, HW30685, HW30757, HW31127, hw31193, HW31463, HW29890, HW30760, HW30773, HW30010, HW30565, HW30755, HW31191, HW31298, hw31024, HW30023R01, HW30070, HW30071, HW30295, HW30315, HW30629, HW30747, HW31404, HW30620, HW29632, HW29907, HW30050, HW30064, HW30132, HW30220, HW30250, HW30259R01, HW30492, HW30617, HW30745, hw30947, hw31078, HW31462, HW28956, HW29301, HW29499R01, HW29984R02, HW30442, HW31459, HW31460, HW30677, HW30703, hw31013, HW31038, HW31053, HW31004, HW31046, HW31060, HW31149, hw31217, HW31010, HW30581, HW30605, HW30613, HW30628, HW30884, HW30439, HW31065, HW31321, HW29720R01, HW30007, HW30009R01, HW30474, HW30493, HW30458, HW30611, HW29326R01, HW29334R01, hw29828R01, HW30724, HW30750, HW31623, HW34506, HW34654, HW34798, HW31407, HW31537, HW31765, HW31914, HW32339, HW32424, HW34127, HW34298, HW34299, HW34865, HW34874, HW29910R01, HW30177, HW30337, HW30349R01, HW30541, HW30557, HW30751, hw30936, HW31220, HW31575, HW31591, HW31649, HW29739, HW29777, HW29889, HW30059, HW30061, HW30077, HW32485, HW32545, HW34831, HW34832, HW29626, HW29960, HW30129, HW30159, HW30302, HW30457, HW31730, HW32003, HW32451, HW34162, HW34634, HW34679, HW34736, HW34984, HW35152, HW35299, HW29710, HW29730, HW29732, HW29733, HW29813, HW29855R01, HW30065, HW30246, HW30265R01, HW30334, HW30340, HW30377, HW30424, HW30426, HW30440, HW30551, HW30681, HW30772, HW30789, HW30911, hw30942, hw30943, HW30977, HW30997, HW31002, HW31029, hw31057, HW31198, HW31213, hw31224, HW34137, HW32005, HW32267, HW34741, HW30238, HW30631, HW30649, HW30782, HW30786, hw31292, HW31326, HW31099, HW31177, HW31504, HW31738, HW32045, HW32418, HW34572, HW34790, HW35434, HW30063, HW30317, HW30344, HW30723, HW30779, HW31272, HW31401, HW31609, HW31613, HW32327, HW34486, HW35014, HW32019, HW34635, HW29640R01, HW29938, HW30004, HW30011, HW30076, HW30173, HW30175, HW30622, HW30759, HW30780, HW30907, HW30908, HW30921, hw31011, HW31050, HW31325, HW31368, HW31363, HW31595, HW31636, HW31638, HW31671, HW31681, HW31690, HW31962, HW31963, HW31975, HW32037, HW32069, HW32111, HW32352, HW32359, HW32380, HW32398, HW32438, HW32439, HW34135, HW34394, HW34396, HW34397, HW34671, HW34709, HW34917, HW34937, HW35110, HW35111, HW35166, HW35168, HW35431, HW35449, HW31720, HW34074, HW34323, HW34451, HW34657, HW34662, HW29962R02, HW30665, HW30701, HW30712, HW31043, HW31064, HW31241, HW31344, HW32284, HW34242, HW34264, HW34602, HW30051, HW30066, HW30887, HW31890, HW32176, HW32361, HW34189, HW34714, HW30553, HW30727, HW31862, HW34952, HW30183, HW30191, HW30882, HW30893, HW31061, HW34167, HW34532, HW30781, HW31776, HW34194, HW34763, HW35058, HW30330, HW32405, HW34416, HW34585, HW34872, HW35015, HW35081, HW35422, HW29528, HW29807, HW29860, HW29902, HW29933, HW29953, HW29977R02, HW29987R02, HW29998, HW30281, HW30318, HW30403, HW30554, HW30555, HW30567, HW30583, HW30627, HW30897, hw31102, HW31476, HW31484, HW29623, HW30248, HW30573, HW30700, HW35003, hw30932, HW31122, HW31131, HW34525, HW34676, HW34931, HW31648, HW32145, HW32409, HW34854, HW35257, HW30636, HW30640, HW30641, HW30918, hw30941, hw31073R01, HW31142, HW31186, HW31214, hw31000, HW31643, HW31747, HW32130, HW35008, HW29919R01, HW30582, HW30886, HW31126, HW31479, HW32117, HW30610, HW31719, HW32425, HW32469, HW32498, HW34505, HW35450, HW29823R02, HW30241, HW30257, HW30258, HW30427, HW30460, HW30898, HW30975, HW34386, HW34793, HW29891, HW30448, HW30587, HW30687, HW31159, HW31203, HW31478, HW34699, HW35069, HW29792, HW29804, HW29814, HW29816, HW29948, hw30929, hw31237, HW31256, HW31490, HW31189, HW34067, HW34230, HW35193, HW29805, HW29930, HW30261R01, HW30282, HW30285, HW30571, HW30586, HW30758, HW30787, HW30788, HW30790, HW31339, HW31349, HW31361, HW31389, HW34263, HW34265, HW34314, HW34483, HW34777, HW35176, HW35183, hw29709R01, HW29742, HW29929, HW29939, HW30171, HW30270R01, HW30339, HW30350, HW30696, hw30990, hw31059, HW31184, HW31215, HW34055, HW34155, HW34156, HW34978, hw29643R01, HW30013, HW30014, HW30616, HW31152, HW31529, HW31540, HW31359, HW31570, HW31951, HW34188, HW35026, HW35107, HW29897, HW29925, HW30180, HW30181, HW30243, HW30274R01, HW30294, HW30298, HW31783, HW32456, HW34118, HW30578, hw31008, HW31047, HW31678, HW30128, HW30131, HW30185, HW30417, HW30418, HW30563, HW30883, HW30889, HW31888, HW32395, HW34469, HW30015, HW31174, HW31181, HW31265, HW31869, HW34645, HW34970, HW35452, HW31480, HW31599, HW32375, HW29636, HW30311, HW30531, HW30756, HW31330, HW35206, HW29638, HW29663, HW29678, HW29751, HW29778, HW29829R02, HW30056R01, HW30062, HW30127, HW30203, HW30206, HW30273, HW30332, HW30356, HW30357, HW30450, HW30754, HW30891, hw30945, hw31005, HW31135, HW29874, HW31608, HW31797, HW34064, HW34689, HW29766, HW31942, HW34282, HW34492, HW34653, HW35415, HW29818, HW29819, HW29862, HW29899R01, HW30215, HW30244, HW30280R01, HW30283, HW30303, HW30327, HW30336, HW30416, HW30527, HW30576, HW30625, HW30626, HW30693, HW30704, HW30713, HW30721, HW30730, HW30742, HW30749, HW30775, HW30903, HW30906, HW30909, HW30922, hw30935, hw30994, hw30995, HW31030, hw31031, HW31034, HW31069, HW31209, HW31212, HW31327, HW31465, HW31466, HW34315, HW34715, HW35180, HW35182, HW35264, HW29857, HW30672, HW31289, HW32271, HW32280, HW34016, HW34566, HW35115, HW28961R01, HW29338R01, HW29398R01, HW29443, HW29848R01, HW29863R01, HW30694, HW30715, HW30762, hw31210, hw31274, HW31280, HW31313, HW32283, HW34300, HW34306, HW34353, HW34976, HW30033R02, HW30310, HW30313, HW30679, HW30684, HW30900, HW30905, hw30938, hw30992, HW31320, HW29941R01, HW30618, HW31190, HW31502, HW32318, HW34304, HW34410, HW34441, HW34562, HW34677, HW34698, HW35089, HW35226, HW29715, HW29817, HW29878, HW29892, HW29934, HW29935R01, HW29973, HW30000, HW30016, HW30017R01, HW30019, HW30048, HW30068, HW30188, HW30189, HW30266R01, HW30561, HW30585, HW30697, HW31331, HW31461, HW31391, HW32020, HW32178, HW32533, HW34702, HW34781, HW35167, HW30228, HW30247, HW30287, HW30326, HW30686, HW30689, hw30927, HW31227, HW31470, HW29831, HW31705, HW34823, HW35442, hw29785R01, HW30433R01, HW30612, HW31006, HW31315, HW31536, HW32105, HW34083, HW34476, HW29599, HW30025R02, HW30993, hw31022, HW32417, HW34199, HW34650, HW35010, HW35011, HW35066, HW35082, HW35241, HW35273, HW31535, HW31870, HW32184, HW32362, HW32512, HW34219, HW34221, HW34477, HW35461, HW35462, HW29824R01, HW30242, HW30436R01, HW30623, HW30888, HW31185, HW31311, HW31400, HW35106, HW29856R02, HW29861R01, HW29884, HW29912, HW30296, HW30346, HW30454, HW30729, HW35195, HW31990, HW32300, HW32334, HW32351, HW32487, HW34735, HW35242, HW29527, HW29728, HW30119, HW30163, HW30584, HW32140, HW34390, HW35169, HW29294, HW29621, HW29879, HW29885, HW29979R02, HW30021R01, HW30507, HW30544, HW30580, HW30615, HW30682, HW30691, HW30692, HW30725, HW31051, HW31329, HW31334, HW31335, HW34216, HW30041, HW34168, HW34589, HW34593, HW34684, HW35118, HW29497R01, HW29598, HW29957, HW30072, HW30073, HW30074, HW30075, HW30299, HW30300, HW30306, HW30761, HW30774, hw30940, hw30987, hw31197, HW31464, HW29531, HW29634, HW29951, HW30002, HW35072, HW30309, HW31369, HW32251, HW32376, HW34856, hw31040, HW29322, HW32082, HW32083, HW34171, HW35437.hw30999, HW31225, HW35059, HW35135, HW34081, HW35232, HW31568, HW34580, HW34787, HW34860, HW34896, HW35048, HW29625, HW29662, hw29852R01, HW29908R01, HW29924, HW30049, HW30052, HW30057, HW30097, HW30124, HW30133, HW30179, HW30251, HW30264R01, HW30429, HW30432, HW30489, HW30566, HW30678, HW30690, HW30753, hw30926, HW30969, hw31027, HW31231, hw31236, HW31278, HW32107, HW34105, HW29868, HW29950, HW30176, HW30182, HW30260R01, HW30428, HW30530, HW30688, HW31054, HW31528, HW31611, HW31709, HW32382, HW34158, HW34322, HW29944, HW30012, HW30271R01, HW30305, HW30447, HW30577, HW30657, hw30931, hw31204, HW31221, HW31469, HW29809, HW31281, HW31634, HW31689, HW34218, HW34770, HW35436, HW35456, HW29624, HW30291, HW30464, HW30487, HW30637, HW30899, HW31007, hw31012, HW31397, HW31576, HW31764, HW32087, HW32151, HW32428, HW34413, HW34504, HW34814, HW35126, HW35243, HW35471, HW35473, HW29812, HW29854, HW30005, HW30067, HW30425, HW30465, HW30718, HW30719, HW30744, HW30902, HW30914, HW31116, hw31200, HW31201, HW31474, HW31646, HW31695, HW31785, HW32312, HW32381, HW34160, HW34258, HW34479, HW34811, HW34999, HW35116, HW29773, HW29822, HW30276R01, HW30307, HW30351, HW30420, HW30437, HW30452, HW30470, HW30481, HW30488, HW30495, HW30496, HW30534, HW30545, HW30550, HW30709, HW30777, HW31042, hw31056, HW31066, HW31067, HW31068, HW31316, HW31946, HW35249, HW35255, hw29611R01, HW30304, HW30321, HW30343, HW30466, HW30479, HW30533, HW30540, HW30570, HW30572, HW31206, HW31612, HW31624, HW31652, HW31653, HW32463, HW34233, HW34389, HW35469, HW29806, HW30190, HW30192, HW30316, HW30331R01, HW30341, HW30438, HW30459, HW30536, HW30895, hw31019, hw31222, HW31542, HW31923, HW32449, HW34153, HW34534, HW34599, HW29716, HW29858R02, HW30444, HW30476, HW30647, HW30683, HW30695, HW30752, HW30892, hw31028, hw31044, HW31396, HW31567, HW34485, HW29838R01, HW29888, HW30006, HW30184, hw31023, HW31617, HW31829, HW31880, HW32074, HW32399, HW35280, HW29637, hw29647R01, HW29913R01, HW29942, HW29952, HW30290, HW30419, HW30535, HW30579, HW30648, HW30676, HW30717, HW31477, HW31347, HW30234, HW30680, HW30890, hw30991, hw30998, HW31055, HW31866, HW32262, HW32431, HW34453, HW34577, HW34578, HW35288, HW35433, HW31606, HW31716, HW32044, HW32063, HW34183, HW34240, HW34473, HW31337, HW31345, HW32323, HW32325, HW29628, HW31139, HW32422, HW30245, HW31762, HW31971, HW29745, HW30055, HW30221R01, HW30286R01, HW30288R01, HW30519, HW30568, HW30569, HW30588, HW30910, HW30920, HW30924, hw31025R01, HW31172, HW31270, HW31395, HW31405, HW32491, HW34164, HW34658, HW30130, HW30256, HW30434, HW31188, HW31312, HW31467, HW31468, HW31498, HW30379, HW30574, HW29920R01, HW31153, HW31807, HW32146, HW32370, HW34174, HW34499, HW34683, HW34825, HW29703R02, HW29881, HW30272R01, HW30348, HW30546, HW30710, HW30785, HW31041, HW31240, HW31245, HW34690, HW30698, HW30699, HW31160, HW31169, HW31782, HW31798, HW31864, HW32402, HW34496, HW34921, HW35159, HW35263, HW35287, HW29772, HW30001, HW30162, HW30720, HW30732, HW30912, HW31230, HW31302, HW31897, HW32089, HW30669, HW30708, HW31001, HW31062, HW31318, HW31481, HW32486, HW34858, HW29740, HW29808, HW29871, HW30236, HW30674, HW30901, HW31279, HW34080, HW32500, HW34391, HW34466, HW34467, HW34601, HW34775, HW35219, HW35238, HW35403, HW35413, HW29713R03, HW30731, HW31308R01, HW34463, HW30078, HW32379, HW34312, HW34406, HW34883, HW35427, HW35448, HW29221, HW29336R01, HW29340R01, HW29450, HW29452, HW29501R01, HW29619, HW29966, HW30134, HW30329, HW30532, HW30575, HW30621, HW30624, HW30707, HW30748, hw31052R01, HW31118, HW31246, HW30240R01, HW30279R01, HW30539, HW31300, HW31871, HW32110, HW35425, HW29811R01, HW31562, HW31775, HW31800, HW32188, HW32276, HW32535, HW32550, HW34163, HW34195, HW34465, HW34815, HW35260, HW29945, HW30161, HW30187, HW30232, HW30338, hw30930, HW31269, HW32332, HW34407, HW34537, HW34826, HW30904, HW31626, HW32136, HW32390, HW34116, HW34143, HW35222, HW31199, HW35181, HW30642, HW31356, HW29622, HW29726, hw29851R01, HW30262R01, HW30284R01, HW30314, HW30335, HW30778, HW30792, HW30913, HW30965, hw30989, HW31239, HW31692, HW31898, HW32126, HW32249, HW32363, HW32442, HW34646, HW34788, HW34795, HW35029, HW35045, HW35047, HW35235, HW35236, HW35438, HW35447, HW29163, HW30388, HW30422, HW30431, HW30446, HW30478, HW30482, HW30490, HW30543, HW30548, HW30644, HW30714, HW30716, hw30944, HW31205, hw31229, HW31238, HW31243, HW32346, HW32527, HW32532, HW35020, HW35130, HW35294, HW35295, HW31151, HW34212, HW34956, HW29445, HW29529, HW30164, HW31218, HW31306, HW35051, HW35440, HW28725R01, HW30443, HW30702, HW30923, hw31033, HW31039, HW31900, HW34175, HW34228, HW34311, HW34399, HW34660, HW29194, HW29849, HW30020R01, HW30123, HW30186, HW30292, HW30293, HW30564, HW31392, HW34574, HW34575, HW34576, HW34967, HW35293, HW29631, HW29753, HW30237, HW30249, HW30537, HW30562, hw30937, hw30946, HW31501, HW29914R01, HW30125, HW30435, HW30542, HW31615, HW32540, HW34289, HW34548, HW34549, HW35129, hw29613R01, HW29981R02, HW30277R01, HW30301, HW30538, HW30559, HW30746, HW30917, hw31195, HW35454, HW29630, HW31045, hw31211, HW34728, HW34729, HW29719, HW30054, HW30126, HW30308, HW30560, HW30885, HW30896, HW31304, HW31482, HW30169, HW31367, HW29743, HW31740, HW31741, HW31918, HW34014, HW34333, HW34472, HW34545, HW34608, HW34721, HW34828, HW34987, HW35098, hw29706R01, HW29735, HW30172, HW30333, HW30451, HW30668, HW30722, HW30784, HW30915, hw30928, HW31049, HW32345, HW35424, HW31674, HW31677, HW32480, HW34321, HW34611, HW34613, HW35077, HW35190, HW35192, HW35221, HW29799R01, HW29810R01, HW30053, HW30069, HW30217, HW30230, HW30278R01, HW30347, HW30456, HW30675, HW30894, HW31017, HW31323, HW31475, HW32119, HW32313, HW34471, HW34648, HW35076, HW30160, HW30165, HW30218, HW30219, HW30323, HW30328, HW30552, HW30592, HW30659, HW30671, HW31207, HW31333, HW29946, HW31804, HW30480, HW30547, HW30653, HW34245, HW29610R01, HW30216, HW30289R01, hw31109, HW29627, HW29872R02, HW29937, HW30058, HW30060, HW30120, HW30168, HW30263, HW30325, HW30423, HW30556, HW30558, HW30609, HW30670, HW31148, HW31522, HW31647, HW34119, HW34166, HW29335R01, HW29803, hw30933, HW31097, HW31403; GTIN 00888707003261: Serial Numbers: HW40229, HW40405, HW40436, HW40567, HW40568, HW41753, HW40261, HW40601, HW40717, HW40764, HW40880, HW41248, HW41446, HW41692, HW41296, HW40977, HW41061, HW41478, HW41488, HW40145, HW40209, HW40264, HW40299, HW40301, HW40381, HW40389, HW40455, HW40524, HW40527, HW40620, HW40640, HW40643, HW40734, HW40737, HW40738, HW40762, HW40834, HW40288, HW40326, HW41752, HW40222, HW40223, HW40443, HW40922, HW40504, HW40575, HW40576, HW40822, HW41100, HW41118, HW40569, HW40694, HW41823, HW40147, HW40635, HW41476, HW41584, HW41586, HW41588, HW41797, HW41761, HW42169, HW40154, HW40086, HW40139, HW40140, HW40169, HW40176, HW40206, HW40208, HW40226, HW40272, HW40306, HW40323, HW40324, HW40331, HW40340, HW40361, HW40364, HW40466, HW40490, HW40491, HW40492, HW40497, HW40518, HW40590, HW40591, HW40592, HW40593, HW40617, HW40622, HW40718, HW40719, HW40744, HW40813, HW40821, HW40826, HW40827, HW40829, HW40830, HW40831, HW40868, HW41157, HW41158, HW41159, HW41160, HW41162, HW41165, HW41166, HW41167, HW41410, HW41412, HW41414, HW41417, HW41419, HW41420, HW41423, HW41735, HW41748, HW41749, HW41819, HW41820, HW41836, HW41837, HW41867, HW42129, HW42134, HW42254, HW42277, HW42319, HW42320, HW42323, HW42389, HW42768, HW42809, HW42810, HW42819, HW40778, HW42164, HW40488, HW41830, HW42201, HW40074, HW40075, HW40037, HW40199, HW40453, HW41141, HW41490, HW41856, HW40245, HW40440, HW40627, HW40772, HW40798, HW40884, HW40903, HW40907, HW40985, HW41038, HW41067, HW41444, HW41445, HW41884, HW42222, HW42802, HW42804, HW40503, HW41733, HW40051, HW40068, HW40230, HW40517, HW40572, HW40703, HW40749, HW40781, HW40815, HW40891, HW40908, HW40996, HW41032, HW41288, HW41318, HW41457, HW41503, HW41504, HW41636, HW41678, HW41707, HW41708, HW41729, HW41799, HW41841, HW41861, HW42824, HW41629, HW41774, HW40433, HW40595, HW40986, HW40991, HW41006, HW41054, HW41093, HW41154, HW41393, HW41398, HW42315, HW42807, HW40515, HW40585, HW41069, HW41529, HW41598, HW41642, HW41649, HW40238, HW40293, HW40337, HW40384, HW41091, HW41889, HW42173, HW42384, HW40510, HW40918, HW41883, HW40995, HW41107, HW41844, HW42628, HW40197, HW40382, HW40397, HW40894, HW40988, HW41137, HW41451, HW41659, HW41711, HW41805, HW41838, HW42317, HW42343, HW42631, HW42730, HW40460, HW41077, HW41530, HW41689, HW41690, HW40067, HW40079, HW40916, HW41361, HW40451, HW40962, HW40710, HW40805, HW41404, HW41713, HW42172, HW42745, HW40715, HW40776, HW40939, HW41005, HW41456, HW41877, HW42336, HW42820, HW40948, HW41510, HW41524, HW41807, HW42805, HW40009, HW40232, HW40271, HW41858, HW40292, HW40976, HW40987, HW41515, HW40090, HW40270, HW40862, HW41605, HW41354, HW40065, HW40073, HW40528, HW40800, HW40801, HW41148, HW41238, HW41378, HW41387, HW41388, HW41389, HW41541, HW41691, HW41702, HW41821, HW40149, HW40501, HW40633, HW40723, HW40913, HW41242, HW41480, HW41617, HW41618, HW42208, HW40204, HW40520, HW41732, HW42627, HW40201, HW40399, HW40404, HW40708, HW40924, HW40965, HW41146, HW41492, HW41664, HW40033, HW40281, HW40359, HW40606, HW40838, HW41008, HW41185, HW41194, HW41247, HW41280, HW41283, HW41310, HW41411, HW41550, HW40120, HW40143, HW40191, HW40305, HW41865, HW42257, HW40085, HW40168, HW40317, HW40794, HW40883, HW41073, HW41176, HW41179, HW41180, HW42121, HW40807, HW41057, HW41213, HW41507, HW41626, HW40769, HW41234, HW41499, HW42239, HW40265, HW40283, HW41516, HW42346, HW41888, HW42170, HW40036, HW40045, HW40048, HW40177, HW40259, HW40263, HW40644, HW40646, HW40647, HW40750, HW40819, HW41012, HW41138, HW41139, HW41225, HW41298, HW41441, HW41443, HW41517, HW41526, HW41606, HW41669, HW41789, HW41873, HW42125, HW40682, HW40481, HW41634, HW41668, HW41817, HW40109, HW40179, HW40190, HW40214, HW40240, HW40448, HW40902, HW41084, HW41496, HW41549, HW41573, HW40049, HW40097, HW40705, HW40739, HW40925, HW40944, HW41022, HW41072, HW41076, HW41098, HW41131, HW41132, HW41232, HW41237, HW41274, HW41369, HW41394, HW41396, HW41418, HW41424, HW41425, HW41427, HW41428, HW41431, HW41432, HW41435, HW41438, HW42242, HW42281, HW42284, HW42318, HW42734, HW40099, HW41001, HW41483, HW41509, HW41514, HW41521, HW41525, HW41591, HW41635, HW41747, HW41834, HW40142, HW40166, HW40260, HW40279, HW40391, HW40529, HW40803, HW41007, HW41037, HW41102, HW41195, HW41336, HW41650, HW41651, HW41703, HW41730, HW42732, HW40258, HW40314, HW40523, HW40698, HW40861, HW40967, HW40971, HW41017, HW41060, HW41403, HW41421, HW41585, HW41680, HW40992, HW41042, HW41104, HW41105, HW41846, HW42383, HW41391, HW40244, HW40892, HW40897, HW41555, HW41666, HW41715, HW41788, HW41809, HW41485, HW41815, HW40273, HW40757, HW42762, HW41647, HW40334, HW40322, HW40392, HW41558, HW41746, HW41409, HW41613, HW41766, HW40402, HW40700, HW41547, HW41688, HW42993, HW40138, HW40714, HW40157, HW40483, HW40605, HW40683, HW40759, HW41097, HW41564, HW41768, HW41779, HW41868, HW40905, HW41364, HW41543, HW40121, HW40178, HW41212, HW41439, HW40444, HW41551, HW41633, HW41644, HW40171, HW40200, HW40210, HW41223, HW42207, HW40753, HW40930, HW41182, HW41497, HW41563, HW41622, HW41842, HW42136, HW42220, HW42338, HW41018, HW41172, HW41224, HW41538, HW41556, HW42626, HW40152, HW40153, HW40212, HW40256, HW40262, HW41348, HW41578, HW41662, HW41724, HW40291, HW40403, HW40512, HW40530, HW40546, HW40695, HW41351, HW40006, HW41440, HW41453, HW41575, HW40094, HW40159, HW40684, HW41092, HW41278, HW41285, HW41326, HW41337, HW41773, HW41810, HW41811, HW41813, HW41814, HW41862, HW41882, HW40013, HW40329, HW40687, HW40768, HW40808, HW40935, HW41035, HW41066, HW41227, HW41228, HW41229, HW41250, HW41275, HW41289, HW41461, HW41603, HW41812, HW42772, HW40017, HW40672, HW40863, HW40865, HW41381, HW41382, HW41385, HW41630, HW41631, HW41833, HW41876, HW42382, HW41287, HW41552, HW41787, HW41866, HW41124, HW41757, HW41839, HW41569, HW41772, HW42976, HW40563, HW40882, HW41081, HW41331, HW41481, HW41489, HW40874, HW41590, HW41880, HW42124, HW41533.HW40298, HW40942, HW41408, HW41685, HW40061, HW40696, HW40722, HW41321, HW41611, HW41681, HW41786, HW41887, HW41236, HW41339, HW41341, HW41648, HW40046, HW40089, HW40100, HW40122, HW40186, HW40307, HW40390, HW40509, HW40732, HW40832, HW40968, HW41300, HW41493, HW41571, HW41574, HW40499, HW40690, HW41656, HW40777, HW41682, HW41782, HW40181, HW40247, HW40371, HW40745, HW40783, HW40998, HW41025, HW41520, HW41795, HW40841, HW40054, HW41845, HW42127, HW40689, HW41188, HW40482, HW40636, HW40678, HW40870, HW40360, HW40534, HW40537, HW40766, HW40767, HW40941, HW41059, HW41064, HW41594, HW41596, HW41719, HW42260, HW42765, HW42815, HW42994, HW42995, HW40112, HW40192, HW40294, HW40631, HW40755, HW40373, HW40583, HW40857, HW40997, HW41173, HW41174, HW41207, HW41208, HW41240, HW41295, HW41299, HW41345, HW41400, HW41459, HW41460, HW41462, HW41463, HW41464, HW41465, HW41467, HW41468, HW41470, HW41519, HW41522, HW41531, HW41548, HW41577, HW41608, HW41612, HW41614, HW41616, HW41627, HW41679, HW41712, HW42139, HW42142, HW42144, HW42146, HW42147, HW42210, HW42378, HW42736, HW42739, HW40004, HW41607, HW41646, HW41770, HW41800, HW40202, HW40746, HW41058, HW41479, HW41486, HW41818, HW40489, HW41330, HW42727, HW42731, HW41731, HW41885, HW42728, HW40876, HW40025, HW40211, HW40437, HW41024, HW41322, HW41327, HW41355, HW41494, HW41034, HW41560, HW41386, HW42218, HW41475, HW40002, HW40108, HW40130, HW40189, HW40198, HW40441, HW40558, HW40780, HW40809, HW40810, HW40887, HW40917, HW41111, HW41112, HW41115, HW41484, HW41657, HW41822, HW41860, HW42283, HW42324, HW41186, HW40445, HW40533, HW40597, HW40616, HW40621, HW40927, HW40928, HW41090, HW40376, HW40875, HW40994, HW41002, HW41144, HW41360, HW41508, HW41511, HW41589, HW41686, HW40302, HW40303, HW40087, HW41319, HW41334, HW41537, HW41721, HW41756, HW42279, HW42494, HW40295, HW42385, HW42387, HW40728, HW41276, HW41338, HW40255, HW41734, HW41291, HW40395, HW41687, HW41751, HW41528, HW41604; 1104JP: GTIN 00763000170967, Serial Numbers: HW40657, HW40658, HW40659, HW40660, HW40662, HW40663, HW40664, HW40665, HW40666, HW40667, HW40668, HW40789, HW40816, HW40881, HW40950, HW40951, HW40952, HW40953, HW40954, HW40955, HW40956, HW40957, HW40958, HW40960, HW40961, HW41045, HW41047, HW41048, HW41050, HW41214, HW41251, HW41252, HW41254, HW41255, HW41256, HW41258, HW41259, HW41260, HW41262, HW41263, HW41264, HW41265, HW41266, HW41269, HW41270, HW41271, HW41308, HW41309, HW41314, HW41316, HW41317, HW41670, HW41672, HW41673, HW41674, HW41675, HW41694, HW41698, HW41699, HW41700, HW41701, HW41783, HW41791, HW41792, HW41803, HW41804, HW41824, HW41826, HW41828, HW41847, HW41848, HW42285, HW42312, HW42352, HW42771, HW42884; GTIN 00888707005838, Serial Numbers: HW42818, HW41784, HW41825, HW41827, HW41829, HW41849, HW41850, HW41852, HW41853, HW41863, HW42882; MCS1705PU: GTIN 00888707005364, Serial Numbers: HW42371, HW42374, HW43227, HW43329, HW43330, HW43334, HW43335, HW42071, HW42296, HW42380, HW42270, HW42272, HW43217, HW43219, HW43220, HW43324, HW43325, HW42267, HW43107, HW43109, HW43115, HW43116, HW43117, HW43188, HW43319, HW43363, HW43364, HW43185, HW43197, HW43275, HW43320, HW43047, HW43048, HW43306, HW43318, HW42418, HW42610, HW42611, HW43040, HW43041, HW43042, HW43258, HW43270, HW43322, HW43352, HW43357, HW43223, HW43226, HW42579, HW42580, HW43123, HW42400, HW42574, HW42576, HW43313, HW43358, HW42448, HW42571, HW43049, HW43050, HW42423, HW42424, HW42205, HW42301, HW42402, HW43051, HW43236, HW42068, HW42069, HW42086, HW42087, HW42174, HW42180, HW42181, HW42182, HW42184, HW42191, HW42192, HW43261, HW42202, HW42375, HW42379, HW43112, HW43316, HW42264, HW42266, HW42362, HW42363, HW42195, HW42420, HW42422, HW41765, HW42070, HW42364, HW42499, HW42578, HW42593, HW43230, HW43037, HW43038, HW43039, HW43052, HW43247, HW43257, HW43279, HW43356, HW42268, HW42577, HW42582, HW43079, HW43125, HW43244, HW43276, HW43307, HW43308, HW43312, HW43098, HW43120, HW43190, HW43191, HW43192, HW42176, HW42179, HW42297, HW42298, HW43129, HW43157, HW43175, HW43336, HW43338, HW42599, HW42600, HW42601, HW42083, HW42084, HW43044, HW43278, HW43354, HW42425, HW42431, HW43090, HW43101, HW43103, HW43187, HW43222, HW42078, HW42081, HW42082, HW42302, HW42366, HW42367, HW42368, HW43351, HW43373, HW43053, HW43145, HW43183, HW42248, HW42261, HW42584, HW42591, HW43176, HW43221, HW43240, HW42359, HW42360, HW42361, HW43194, HW43195, HW43239, HW43262, HW43268, HW42401, HW42403, HW42404, HW42405, HW42438, HW42439, HW42585, HW42586, HW42592, HW43036, HW43043, HW43054, HW43114, HW43232, HW43248, HW43249, HW43251, HW43253, HW43255, HW43361, HW43362, HW43365, HW43366, HW43371, HW43372, HW42397, HW42572, HW42573, HW42581, HW43062, HW43064, HW43282, HW43315, HW43057, HW43059, HW43060, HW43061, HW42595, HW42596, HW43045, HW43046, HW42588, HW42590, HW43178; GTIN 00888707007139, Serial Numbers: HW42372, HW42691, HW42692, HW42693, HW42694, HW42695, HW42696, HW42697, HW42699, HW42701, HW42704, HW42705, HW42706, HW42713, HW42826, HW42832, HW42839, HW42845, HW42846, HW42847, HW42853, HW42924, HW42925, HW42929, HW42930, HW42934, HW42935, HW42936, HW42945, HW42946, HW42950, HW42951, HW42954, HW42956, HW42957, HW42958, HW42959, HW42962, HW42963, HW42972, HW42975, HW42977, HW42978, HW42979, HW42980, HW42982, HW42984, HW42985, HW42986, HW43015, HW43019, HW43027, HW42955, HW43020; GTIN 00888707007160, Serial Number: HW43453; IFU, ERG, and PM catalogue numbers affected: GR00526: GTIN none, Serial Numbers: R021023, R021429, R022801, R023432, R024437, R021282, r023861, R022197, R022370, R024915, R025043, R025705, R023396, R024765, R027316; IFU00358: GTIN none, Serial Numbers: R020473, R020606, R020606, R020606, R021125, R021136, R021369, R024757; IFU00375: GTIN none, Serial Numbers: R021019, R025232, R027762, R028481, R021165, R024178, R024545, R024582, R022376, R024865, R023318, R022436, R027594, R026582, R026753, R027045, R028937, R021663, R026907, R028908; IFU00376: GTIN none, Serial Numbers: R021020, R021167, R022413, R022970, R023478, R024524, R024976, R025507, R029513, R024548, R026621, R027923, R022093, R022284, R021121, R024142, R024200, R021567, R028415, R023279, R023435, R026473, R027670, R028603, R027058, R025102, R024867, R027403, R026271, R026904, R028939, R029066; IFU00520: GTIN none, Serial Numbers: R025351, R026900; IFU00526: GTIN none, Serial Number: R025761R01; IFU00593: GTIN none, Serial Number: R027857; IFU00614: GTIN none, Serial Number: R028505; M017475C001: GTIN none, Serial Number: R029651;
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
- HVAD® Pump Implant Kit MCS1705PUHeartware Inc. · Ventricular (Assist) Bypass7 recalls36 deaths203 reports
Product lines
Other devices under the same FDA submission
30 devices cleared under P100047 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HeartWare® Controller 1403USHeartware Inc. · Ventricular (Assisst) Bypass3 recalls10 deaths1,495 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assisst) Bypass1 recall1,126 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass2 recalls606 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall615 reports
- HVAD® Controller 1403USHeartware Inc. · Ventricular (Assist) Bypass548 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall487 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assist) Bypass1 recall281 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall159 reports
- HVAD® Pump Implant Kit MCS1705PUHeartware Inc. · Ventricular (Assist) Bypass4 recalls1 report
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall150 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assisst) Bypass1 recall130 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page