Skip to content

HeartWare Ventricular Assist Device (HVAD) System… recallZ-1103-2022 · Heartware, Inc. · initiated Mar 30, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Heartware, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Finnish and Turkish translation errors in HVAD controller and monitor displays as will as Instructions for Use, Patient Manual, and Emergency Responder Guide for the HeartWare Ventricular Assist Device (HVAD) system.

FDA root cause category: Labeling design

What the company is doing

Starting 30-Mar-2022, Medtronic initiated an Urgent Medical Device Notice to VAD Coordinators at customer sites by regional mail and visit in order to inform them of the errors and the corrections impacting the incorrect translation in Finland and Turkey for the HVAD" System. In addition to the Urgent Medical Device Notice a patient template was provided to facilitate the VAD Coordinators in notifying their patients of the issues to the Patient Manual.

Recall details

Recall number
Z-1103-2022
Status
Open, Classified
Initiated by the company
Mar 30, 2022
Recalling firm
Heartware, Inc.
Quantity
667 units
Product code
DSQ
FDA submissions
P100047
Distribution
Finland and Turkey

Product description

HeartWare Ventricular Assist Device (HVAD) System Controller, Model Number 1407DE Kit

Code information (models, lots, serial numbers)

Show the full code information (8k characters)

GTIN 00888707001700, Serial Numbers: CON300826, CON302649, CON304346, CON305438, CON305669, CON309302, CON309306, CON309974, CON312894, CON313111, CON313129, CON313154, CON314788, CON314789, CON314903, CON314929, CON315066, CON315120, CON315384, CON300826, CON304248, CON304346, CON304363, CON304943, CON300612, CON300622, CON302530, CON302655, CON302814, CON303005, CON303057, CON303093, CON303132, CON303368, CON303549, CON303571, CON303936, CON304027, CON304028, CON304282, CON304302, CON304318, CON304322, CON304327, CON304328, CON304358, CON304360, CON304362, CON304381, CON304451, CON304453, CON304587, CON304596, CON304600, CON304610, CON304614, CON304620, CON304628, CON304645, CON304649, CON304793, CON304797, CON304889, CON304965, CON305110, CON305123, CON305124, CON305142, CON305146, CON305148, CON305150, CON305277, CON305282, CON305325, CON315244, CON314763, CON315247, CON314577, CON313809, CON314057, CON315260, CON315259, CON315246, CON314770, CON314120, CON314193, CON314339, CON314021, CON314211, CON314068, CON313626, CON310360, CON314188, CON314084, CON314016, CON314085, CON314027, CON314346, CON314087, CON313967, CON314024, CON314187, CON313993, CON311103, CON314148, CON313513, CON313068, CON313452, CON312745, CON313489, CON312959, CON312899, CON313164, CON312896, CON313107, CON313161, CON313166, CON312898, CON313162, CON313232, CON312948, CON312897, CON313109, CON313183, CON313187, CON313168, CON312900, CON313095, CON313165, CON310155, CON309602, CON309853, CON309850, CON310096, CON310090, CON310102, CON306178, CON310793, CON310499, CON310103, CON310101, CON310500, CON306822, CON310192, CON310100, CON310094, CON306705, CON307902, CON307900, CON304639, CON307903, CON307818, CON304646, CON304658, CON304638, CON307997, CON307367, CON307999, CON307959, CON307960, CON307975, CON308006, CON307998, CON307770, CON307868, CON308144, CON308001, CON307172, CON307166, CON307315, CON306727, CON307536, CON307404, CON307123, CON307571, CON307436, CON307167, CON307435, CON307401, CON306728, CON307132, CON306949, CON305644, CON305822, CON305774, CON305770, CON305651, CON305773, CON305419, CON305554, CON305555, CON305424   GTIN 00888707002813, Serial Numbers: CON315106, CON315108, CON315743, CON315791, CON315792, CON315831, CON316010, CON316450, CON316462, CON316826, CON317363, CON317371, CON317718, CON318002, CON318004, CON318364, CON318751, CON318856, CON318864, CON318994, CON319035, CON319523, CON319524, CON319648, CON400331, CON400332, CON400634, CON400636, CON400637, CON400651, CON401103, CON401111, CON401188, CON401251, CON401264, CON401265, CON401825, CON402006, CON402065, CON402077, CON402302, CON402304, CON402658, CON402659, CON402973, CON402974, CON403346, CON403351, CON403543, CON403546, CON403803, CON403804, CON403883, CON403892, CON403980, CON403986, CON404058, CON405273, CON405274, CON405687, CON405706, CON405707, CON406419, CON406420, CON406490, CON406586, CON406610, CON406634, CON406635, CON406666, CON406717, CON406718, CON406719, CON406720, CON406740, CON406867, CON406869, CON406872, CON406874, CON406875, CON406880, CON407000, CON407001, CON407003, CON407004, CON407008, CON407009, CON407012, CON407015, CON407016, CON407017, CON407018, CON407019, CON407020, CON407021, CON407097, CON407113, CON407115, CON407177, CON407178, CON407179, CON407180, CON407207, CON407208, CON407235, CON407270, CON407487, CON407504, CON407519, CON407537, CON407572, CON407587, CON407685, CON407745, CON407780, CON407790, CON407845, CON407898, CON407911, CON407915, CON407917, CON407918, CON407945, CON407946, CON407947, CON407961, CON407998, CON407999, CON408000, CON408005, CON408031, CON408033, CON408038, CON408039, CON408043, CON408047, CON408048, CON408053, CON408055, CON408056, CON408066, CON408067, CON408068, CON408069, CON408070, CON408072, CON408074, CON408075, CON408076, CON408079, CON408090, CON408107, CON408112, CON408118, CON408122, CON408126, CON408127, CON408133, CON408144, CON408166, CON408168, CON408169, CON408186, CON408194, CON408199, CON408219, CON408230, CON408290, CON408300, CON408301, CON408305, CON408306, CON408307, CON408350, CON408351, CON408363, CON408370, CON408371, CON408372, CON408373, CON408381, CON408382, CON408384, CON408387, CON408388, CON408391, CON408397, CON408398, CON408399, CON408406, CON408407, CON408408, CON408411, CON408431, CON408442, CON408496, CON408497, CON408498, CON408502, CON408503, CON408517, CON408518, CON408519, CON408520, CON408521, CON408522, CON408523, CON408531, CON408533, CON408540, CON409013, CON409100, CON409102, CON409106, CON409109, CON409111, CON409112, CON409115, CON409120, CON409135, CON409138, CON409139, CON409140, CON409145, CON409146, CON409152, CON409153, CON409157, CON409174, CON409175, CON409179, CON409180, CON409183, CON409191, CON409195, CON409196, CON409199, CON409206, CON409212, CON409243, CON409244, CON409266, CON409267, CON409268, CON409328, CON409330, CON409334, CON409336, CON409378, CON409386, CON409387, CON409396, CON409403, CON409406, CON409409, CON409422, CON409430, CON409433, CON409439, CON409452, CON409454, CON409470, CON409476, CON409482, CON409486, CON409487, CON409488, CON409593, CON409672, CON409774, CON409873, CON409877, CON409878, CON409879, CON409880, CON409883, CON409974, CON409975, CON409980, CON409981, CON410954, CON410955, CON410960, CON410962, CON410965, CON410966, CON410967, CON410969, CON410970, CON410991, CON410992, CON410993, CON410996, CON411015, CON411046, CON411047, CON411063, CON411097, CON411098, CON411102, CON411110, CON411112, CON411140, CON411143, CON411144, CON411145, CON411212, CON411213, CON411215, CON411218, CON411220, CON411221   GTIN 00888707006941, Serial Numbers: CON410823, CON410843, CON410844, CON410941, CON410980, CON410982, CON411051, CON411053, CON411058, CON411061, CON411085, CON411088, CON411099, CON411103, CON411104, CON411225, CON411235, CON411236, CON411254, CON411283, CON411334, CON411336, CON411337, CON411610, CON411644   GTIN 00888707008242, Serial Numbers: CON412460, CON412713, CON412766, CON412769, CON412770, CON412771, CON412772, CON412773, CON412798, CON412802, CON412803, CON412806, CON412807, CON412808, CON412815, CON412852, CON412853, CON412854, CON412855, CON412887, CON412888, CON412890, CON412891, CON412892, CON412893, CON412898, CON412902, CON412908, CON412920, CON412925, CON412927, CON412928, CON412932, CON412938, CON412940, CON412944, CON412951, CON412954, CON412956, CON412958, CON412961, CON412962, CON412970, CON413026, CON413131, CON413135, CON413228, CON413229, CON413235, CON413239, CON413240, CON413242, CON413285, CON413291, CON413292, CON413293, CON413301, CON413316, CON413335, CON413341, CON413342, CON413344, CON413391, CON413392, CON413410, CON413428, CON413431, CON413434, CON413451, CON413454, CON413480, CON413541, CON413546, CON413550, CON413552, CON413556, CON413557, CON413569, CON413579, CON413580, CON413632, CON413651, CON413652, CON413653, CON413654, CON413658, CON413667, CON413680, CON413681, CON413682, CON413686, CON413696, CON413697, CON413700, CON413701, CON413702, CON413703, CON413704, CON413705, CON413706, CON413707, CON413708, CON413720, CON413721, CON413737, CON413738, CON413739, CON413747, CON413748, CON413749, CON413751, CON413752, CON413753, CON413758, CON413776, CON413792, CON413793, CON413807, CON413869, CON413870, CON414339, CON414340, CON415019, CON415020, CON415021, CON415040, CON415041, CON415042, CON415046, CON415047, CON415048, CON415050, CON415051, CON415052, CON415053, CON415067, CON415068, CON415087, CON415106, CON415135, CON415137, CON415141, CON415142, CON415143, CON415146, CON415148, CON415835, CON415781

Other devices under the same FDA submission

33 devices cleared under P100047 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 90079).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page