Field replacement back up batteries distributed on or after… recallZ-1286-2022 · GE Healthcare, LLC · initiated Apr 25, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Backup batteries can fail earlier than their estimated life.
FDA root cause category: Under Investigation by firm
What the company is doing
An Urgent Medical Device Correction letter dated 04/25/2022 was sent to customers with affected devices. The letter mentions the users can continue to use the ventilator while connected to an AC mains power source that is supported by backup emergency power. The letter provides a series of recommendations for the consignee to follow. The letter also requests that the consignee complete the medical device notification acknowledgement response form and send it back to the firm.
Recall details
- Recall number
- Z-1286-2022
- Status
- Open, Classified
- Initiated by the company
- Apr 25, 2022
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 88,347 batteries
- Product code
- BSZ
- FDA submissions
- K172045
- Distribution
- Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, China, Columbia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Italy, Japan, Kazakhstan, Kenya, Korea, Kuwait, Lao PDR, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Nepal, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.
Product description
Field replacement back up batteries distributed on or after April 1, 2019 for Avance CS2, Avance CS2 Pro, Avance, Amingo, Aespire View anesthesia devices
Code information (models, lots, serial numbers)
Field Replacement Units (FRU) Battery kits - PN: 1009-5682-000-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V) or PN: 5856787-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V PAIR), that may have been installed by the customer into any of the following anesthesia systems: o Avance 1009-9002-000 o Aespire View 1009-9212-000 o Avance CS2 and Avance CS2 Pro 1009-9050-000
Other devices under the same FDA submission
6 devices cleared under K172045 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Aisys CS2Datex-Ohmeda Inc · Gas-machine, anesthesia5 recalls2,048 reports
- Avance CS2Datex-Ohmeda Inc · Gas-machine, anesthesia6 recalls1,540 reports
- Aespire ViewDatex-Ohmeda Inc · Gas-machine, anesthesia1 recall158 reports
- Aespire 7900Datex-Ohmeda Inc · Gas-machine, anesthesia3 recalls48 reports
- Aestiva MRIDatex-Ohmeda Inc · Gas-machine, anesthesia157 reports
- Aestiva 7900Datex-Ohmeda Inc · Gas-machine, anesthesia1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 90137).
- Z-1285-2022 · Open, ClassifiedBack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page