Why it was recalled
Inaccurate Distance and Area measurements with use of Global Stack viewport.
FDA root cause category: Process design
What the company is doing
The recalling firm issued letters dated 7/15/2022 via traceable means on 7/15/2022 to the following titles within their affected accounts: Director/Manager of Radiology, Director/Manager of Cardiology, Risk Manager/Hospital Administrator, Head of Radiology Department, Head of Cardiology Department, PACS Administrator, Director of IT Department, Head of Biomedical Engineering, Head of Imaging Informatics. The letter explained the issue, the affected product details, and the actions to be taken by the customer/user, which included the following: The customer/user can continue to use the system in accordance with User Manuals and the following actions: It is recommended the customer/user does not rely on measurements displayed in the Global Stack viewport in the Viewer. Users can perform the measurement activity by (1) Using Overview or Series Viewport to perform the measurement; or (2) If using Global Stack viewport, manually calibrate the image to create a measurement calibration reference and then perform necessary measurements. The customer/user is to complete and return the attachment acknowledgement form to the link in the letter or through the QR code. The customer/user was requested to ensure all potential users in their facility are made aware of this safety notification and the recommended actions. GE Healthcare will correct all affected products and a representative will be in contact with the customer/user to arrange for the correction. After the software has been corrected, the customer/user is to destroy all previous versions of the locally stored application installation package(s) immediately.
Recall details
- Recall number
- Z-1566-2022
- Status
- Open, Classified
- Initiated by the company
- Jul 15, 2022
- Posted by the FDA
- Aug 24, 2022
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 1,690 devices
- Product code
- LLZ
- FDA submissions
- K150420, K182419, K211312
- Distribution
- Distribution was made to AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VT, WA, WI, and WV. There was government distribution and no military distribution. Foreign distribution was made to Argentina, Aruba, Australia, Bahrain, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Egypt, France, Germany, Ghana, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, South Korea, Kuwait, Lebanon, Malaysia, Malta, Mexico, Netherlands, Panama, Peru, Poland, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Product description
(1) GE Centricity Universal Viewer 6.0, Model 2088026-9XX; (2) GE Centricity Universal Viewer 7.0, Model 5826659-0XX; (3) GE Universal Viewer 8.0, Model 5865740-00x.
Code information (models, lots, serial numbers)
(1) Centricity Universal Viewer 6.0 - UDI 00840682103800; Software versions 6.0 SP9, 6.0 SP9.0.1 through 6.0 SP9.0.1.11, 6.0 SP9.0.2, and 6.0 SP10 through 6.0 SP10.4. (2) Centricity Universal Viewer 7.0 - UDI 00840682145794, 00840682145558; Software versions 7.0 through 7.0 SP0.0.4.9, 7.0 SP0.0.5, 7.0 SP0.1.0, and 7.0 SP1. (3) Universal Viewer 8.0 - UDI 00195278379610; Software versions 8.0, 8.0 SP0.1.0, and 8.0 SP0.1.1.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- Centricity CPACS 7.0Ge Healthcare · Laparoscope, general & plastic surgery2 recalls1 report
- Universal Viewer 8Ge Healthcare · System, image processing, radiological2 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗