EMBLEM MRI S-ICD, Subcutaneous Implantable Defibrillator… recallZ-0152-2023 · Boston Scientific Corporation · initiated Sep 30, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is an incorrect manufacturing date/timestamp within the software which causes an inaccurate display of battery capacity.
FDA root cause category: Under Investigation by firm
What the company is doing
The recalling firm emailed Dear Health Care Professional or Hospital Administrator retrieval letters dated 9/30/2022 to their sales representatives on 9/30/2022 who began hand delivering them immediately in the U.S. for delivery to those consignees who were believed to have unimplanted devices in their possession. The letter described the issue and said their Boston Scientific sales representative is available to support this retrieval and listed the following steps to be taken: (1) Immediately segregate all affected product listed in Appendix A of the letter to ensure it will not be used; (2) Complete the attached Reply Verification Tracking Form and return it via the email address provided or via FAX; and (3) Ship product back to Boston Scientific, and the address was provided. OUS notification began 10/3/2022. For the U.S. patients who have already been implanted with the device, correction letters are to be hand-delivered to the doctors the week of 10/17/2022 and the sales representatives will send confirmation of delivery back to the recalling firm. No reply form was enclosed with the correction letter. These letters will be personalized/customized with the name of the following or implanting physicians or clinics. The letters will be dated October 2022 and contain the patient's name, date of birth, and the model/serial number and GTIN implanted. The letter will describe the issue, clinical impact, and recommendations regarding follow-up interval, patient notes entry, and medial records. Prophylactic replacement is not recommended as the issue does not affect the S-ICDs performance or the actual battery longevity.
Recall details
- Recall number
- Z-0152-2023
- Status
- Open, Classified
- Initiated by the company
- Sep 30, 2022
- Posted by the FDA
- Oct 31, 2022
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 142 devices
- Product code
- NVY
- FDA submissions
- P110042
- Distribution
- Worldwide distribution - US Nationwide distribution in the countries of AL, AR, CA, CO, DE, GA, IL, IA, MA, MI, MN, NV, NJ, NY, OH, PA, TN, TX, and VA. The countries of Canada, France, Germany, Latvia, Netherlands, Poland, Sweden, Switzerland, and United Kingdom.
Product description
EMBLEM MRI S-ICD, Subcutaneous Implantable Defibrillator, Model A219, sterile.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
UPN: 00802526584404, Serial numbers: 169280, 169298, 169494, 169753, 169776, 169934, 169941, 170003, 169547, 169151, 169315, 169514, 169851, 169879, 169292, 169696, 169786, 169788, 169852, 170002, 169400, 169405, 169898, 169973, 169885, 169335, 169233, 169608, 169304, 169410, 166452, 169278, 169319, 169321, 169651, 169688, 169864, 169920, 169235, 169277, 169834, 169472, 169417, 169202, 169329, 169426, 169511, 169529, 169630, 169757, 169963, 167092, 169236, 169301, 169310, 169316, 169393, 169436, 169485, 169537, 169543, 169565, 169567, 169571, 169579, 169582, 169599, 169601, 169606, 169663, 169771, 169238, 169323, 169680, 169270, 169265, 169372. UPN: 00802526581519, Serial numbers 169312, 169409, 169421, 169428, 169521, 169561, 169493, 169654, 169653, 169635, 169731, 169477, 169495, 169496, 169474, 169491, 169468, 169455, 169442, 169446, 169435, 169447, 168979, 169271, 169384, 169353, 169371, 169351, 169203, 169469, 169471, 169314, 169342, 169475, 169448, 169404, 169407, 169450, 169396, 169470, 169356, 169373, 169402, 169413, 169441, 169539, 169545, 169551, 169674, 169267, 169350, 169589, 169307, 169595, 169596, 169600, 169286, 169603, 169605, 169206, 169340, 169502, 169505, 169510, 169348.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EMBLEM™ MRI S-ICD A219Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 recalls36 deaths15,404 reports
Other devices under the same FDA submission
9 devices cleared under P110042 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- EMBLEM™ S-ICD A209Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)1 recall7 deaths6,788 reports
- EMBLEM™ S-ICD 3501Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)7 deaths4,936 reports
- EMBLEM™ S-ICD 3401Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)5 deaths1,039 reports
- SQ-RX® PULSE GENERATOR 1010Cameron Health, Inc. · Implantable cardioverter defibrillator (non-CRT)3 deaths596 reports
- EMBLEM™ S-ICD 3401Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)1 death532 reports
- The Q-TRAK™ Subcutaneous Electrode 3400Cameron Health, Inc. · Implantable cardioverter defibrillator (non-CRT)2 deaths394 reports
- SQ-RX® PULSE GENERATOR 1010Cameron Health, Inc. · Implantable cardioverter defibrillator (non-CRT)3 deaths262 reports
- EMBLEM™ S-ICD Electrode Delivery System 4712Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)1 death7 reports
- EMBLEM™ S-ICD Programmer 3200Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 90985).
- Z-0151-2023 · Open, ClassifiedEMBLEM S-ICD, Subcutaneous Implantable Defibrillators, Model A209, sterile.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.