Monofilament Nylon Suture (McKesson Branded) and Coated… recallZ-1173-2023 · Covidien, LP · initiated Dec 2, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
FDA root cause category: Packaging process control
What the company is doing
Medtronic issued an "URGENT: MEDICAL DEVICE RECALL" letter to US consignees on 01-Dec-2022 via UPS 2-day delivery. OUS countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and actions to take: Actions: " Identify and quarantine all unused and non-expired affected non-absorbable sutures. " Please note the affected product may be located within Custom Suture Packs. Please reference Custom Suture Packs to identify affected product. " Please complete the enclosed Customer Confirmation Form (even if your account no longer has affected product on hand) and email to [email protected].> " Return all unused and non-expired affected product in your inventory to Medtronic as indicated in the Shipping and Return Instructions below. " Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product has been transferred or distributed. Shipping and Return Instructions: " Product purchased directly from Medtronic please contact [email protected] for Return Good Authorization (RGA). Your local Medtronic Representative can assist you as necessary in initiating the return of this product. " Credit or Replacement for the returned affected product will be issued based on the RGA number. " Return all unused and non-expired affected product(s) in your inventory to: Medtronic, Attn: Field Returns Dept. 195 McDermott Road, North Haven, CT 06473 USA. " If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. If you have any questions regarding this communication, please contact your Medtronic Representative or Customer Service at 800-962-9888.
Recall details
- Recall number
- Z-1173-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 2, 2022
- Posted by the FDA
- Feb 25, 2023
- Recalling firm
- Covidien, LP
- Quantity
- 12732 units
- Product code
- GAR
- FDA submissions
- K900531
- Distribution
- US Nationwide Distribution and OUS Foreign distribution to countries of: Albania Argentina Armenia Australia Austria Azerbaijan Belarus Belgium Bosnia And Herzegovina Brazil Bulgaria Cambodia Cameroon Canada Canary Islands Chile China Colombia Croatia Czech Republic Denmark Egypt Finland France Georgia Germany Ghana Greece Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy JAPAN Jordan Kazakhstan Kenya Korea, Republic Of Kosovo Kuwait Latvia Lebanon Lithuania Macao Malaysia Malta Mauritius Mexico Moldova, Republic Of Morocco Myanmar Namibia Netherlands New Zealand North Macedonia Norway Pakistan Panama Paraguay Philippines Poland Portugal Reunion Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Syrian Arab Republic Taiwan Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Viet Nam Zimbabwe
Product description
Monofilament Nylon Suture (McKesson Branded) and Coated Braided Silk Suture (McKesson Branded) (1) S661GX 5-0 NYLON 18 DS18 (2)S683GX 4-0 SILK 18 DS18 (3) S683GX 4-0 SILK 18 DS18 (4)S697GX 6-0 NYLON 18 DSM11 (5)S697GX 6-0 NYLON 18 DSM11 (6)S913BX 5-0 NYLON 10 DS12 (7)S931BX 4-0 NYLON 10 DS12
Code information (models, lots, serial numbers)
GTIN: 20612479214235, 20612479214419, 20612479214495, 20612479213283, 20612479214235, 20612479213269 Lot # (1) D1L0912FY (2) D1G1840FY (3) D2B1262FY (4) D1M2377FY (5) D1M2847FY (6) D2D1386FY (7) D2B0756FY
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- McKesson S931BXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Synthetic, Polyamide2 recalls2 reports
- McKesson S661GXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Synthetic, Polyamide2 recalls2 reports
- McKesson S697GXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Synthetic, Polyamide2 recalls1 report
- McKesson S913BXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Synthetic, Polyamide2 recalls0 reports
- McKesson S683GXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Silk2 recalls0 reports
Other devices under the same FDA submission
95 devices cleared under K900531 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Monosof SN-3695Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls3 reports
- Monosof SN-3697Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall32 reports
- Monosof SN-5697Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall10 reports
- Monosof CN-830-LCovidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall9 reports
- Monosof GNJ-283Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall5 reports
- Monosof SN-872Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall4 reports
- Monosof CN-724Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall2 reports
- Monosof SN-667GCovidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall2 reports
- Monosof SN-247Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall1 report
- Monosof SN-5691Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall1 report
- Monosof SN-5694Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall1 report
- Monosof SN-5696Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 91324).
- Z-1172-2023 · Open, ClassifiedDermalon Monofilament Nylon Suture: (1) 88861741-11 DERMALON* 6-0 BLU 45CM C1 (2) 88861747-11 DERMALON* 6-0 BLU 45CM C12 (3) 88861749-24 DERMALON* 5-0 BLU 45CM P13 (4) 88861756-11…
- Z-1174-2023 · Open, ClassifiedMonosof Monofilament Nylon sutures: Product Description CN-724 MONOSOF* 1 BLK 100CM GS25 X36 CN-791 MONOSOF* 2-0 BLK 100CM GCC90X36 CN-791-12 MONOSOF* 2-0 BLK 10…
- Z-1175-2023 · Open, ClassifiedNovafil Monofilament Polybutester Suture Product Description: 8886 4002-41 NOVAFIL 3-0 BLU 90CM C14 8886 4400-13 NOVAFIL 6-0 BLU 45CM SBE2 8886 4400-23 NOVAFIL 5-0 BLU 45CM SBE2 8…
- Z-1176-2023 · Open, ClassifiedSofsilk Coated Braided Silk Suture Product Description CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36 CS211 CS-490 SOFSILK* 2-0 BLK 75CM GS11 X36 CS490 CS-562 SOFSILK* 2-0 BLK 75C…
- Z-1177-2023 · Open, ClassifiedSurgidac Uncoated Braided Polyester suture Product Description D-1764K SURGIDAC* 5-0 WHI 45CM SS24DA
- Z-1178-2023 · Open, ClassifiedSurgilon Braided Nylon suture Product Description: 88861883-51 SURGILON* 2-0 BLK 75CM C16 88861915-71 SURGILON* 1 BLK 75CM GS11 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 …
- Z-1179-2023 · Open, ClassifiedSurgipro Monofilament Polypropylene Sutures Product Description CP-535 SURGIPRO* 1 BLU 100CM GS24 X36 CP-824 SURGIPRO* 1 BLU 150CM GS26 X24 CP-825 SURGIPRO* 2 BLU 150CM GS26 X24 VP…
- Z-1180-2023 · Open, ClassifiedSurgipro II Monofilament Polypropylene Sutures Product Description Model Number VP-421X SURGIPRO*II 4-0 90CM V20 DA7P VP421X VP-435-X SURGIPRO II 4-0 90CM KV5 DA7P VP435X VP-541X S…
- Z-1181-2023 · Open, ClassifiedTi-Cron Coated Braided Polyester Suture Product Description 88862775-31 TICRON* 4-0 BLU 75CM CV327DA 88862880-41 TICRON* 3-0 WHI 75CM Y31DA 88863012-61 TICRON* 0 BLU 90CM C16 X36 8…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.